    Intended Use

    The ERIC Retrieval Device is indicated to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The ERIC™ (Embolus Retriever with Interlinked Cage) Retrieval Device is a mechanical thrombectomy device designed to restore blood flow by removing clots from vasculature in patients suffering from acute ischemic stroke. The device consists of retrieval spheres secured on a pusher wire that are designed to capture and remove blood clots from the neurovasculature. The device is inserted into a microcatheter to navigate to the target location and retrieve the thrombus while the device is withdrawn from the vessel.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for the ERIC Retrieval Device:

    I. Acceptance Criteria and Reported Device Performance

    Test Description / OutcomeAcceptance Criteria (Implied)Reported Device Performance
    Bench/Verification Testing
    Dimensional Testing (Expanded Diameter & Device Overall Length)Device attributes (overall device length and expanded outer diameter (OD) of the shaped section) must meet specified criteria and not raise new questions regarding safety and efficacy compared to predicate.Pass. The longer overall length and smaller OD of the subject device offerings do not affect the performance of the device.
    Fluoroscopic Guidance Marker Testing (Radiopacity)Markers must be sufficiently visible under fluoroscopy.Pass. Both subject and predicate devices are sufficiently visible under fluoroscopy.
    Advance/Retraction Force TestingAdvance and retract forces in a tortuous model must be comparable to predicate devices.Pass. The advance and retract forces of the subject device were comparable with the forces measured for the predicate device.
    Re-Sheathing TestingAbility to re-sheath the device must be comparable to predicate devices.Pass. The ability to re-sheath the subject device is comparable to that of the predicate device tested.
    Radial Force TestingRadial force must be comparable to predicate devices.Pass. The radial force of the subject device is comparable to that of the predicate device tested.
    Tensile Strength TestingPeak tensile strength to failure in different sections must be comparable to predicate devices.Pass. The system tensile strength of the subject device is comparable to that of the predicate device tested.
    Kink Resistance TestingKink resistance must be equivalent to predicate devices.Pass. Kink resistance of the subject device is equivalent to that of the predicate device tested.
    Austenite Finish (Af) TestingAf temperature must be less than product use temperature (body temperature) to satisfy clinical application requirements.Pass. The Af temperature of the subject device is less than the product use temperature (body temperature) and, thus, satisfies requirements for clinical applications.
    Simulated Use/Performance TestingAbility to reliably deploy and use the device in a tortuous benchtop model must be comparable to predicate devices.Pass. Simulated use testing was comparable with that of the predicate device.
    Corrosion Resistance TestingMetallic components intended for fluid path contact must show no signs of corrosion.Pass. Corrosion resistance testing of the subject device showed no signs of corrosion.
    Particulate Evaluation TestingParticulate generation in a tortuous benchtop model must be comparable to predicate devices.Pass. Particulate evaluation was comparable with that of the predicate device.
    Torque Response TestingCore wire of the subject device must rotate freely with the proximal sphere and have equivalent torqueability compared to the predicate device.Pass. Torque response testing indicated that the core wire of the subject device rotates freely with the proximal sphere and, thus, has equivalent torqueability compared to the predicate device.
    Biocompatibility EvaluationDevice must be non-cytotoxic, non-irritating, non-sensitizing, systemically non-toxic, non-pyrogenic, non-hemolytic, non-activating (complement activation), non-thrombogenic, and have no effect on coagulation of human plasma and hematological parameters. (Compliance with ISO 10993-1 and FDA Biocompatibility Guidance)Pass. Demonstrated non-cytotoxic, non-sensitizer, non-irritating, systemically non-toxic, non-pyrogenic, non-hemolytic, no effect on coagulation of human plasma, non-activating, no effect on hematological parameters, and non-thrombogenic.
    Sterilization, Shelf-Life, and Packaging IntegrityAchieve a minimum sterility assurance level (SAL) of 10^-6 for electron beam sterilization (specified as 10^-9) and bacterial endotoxin < 2.15 EU/device. Packaging must maintain functionality and sterility for the indicated shelf-life.Pass. Successfully validated electron beam sterilization process achieving a minimum SAL of 10^-9 with bacterial endotoxin < 2.15 EU/device. Shelf-life studies per ASTM F1980 conducted, establishing product and packaging functionality and sterility.
    Animal Testing (Safety & Performance)Tissue response, perforation/dissection, and downstream thromboembolic/ischemic injury must be comparable to the predicate device. Devices must be safe and equivalent to the predicate device.Pass. Histological evaluation showed similar tissue response to test and control devices. No evidence of perforation or dissection. No evidence of downstream thromboembolic or ischemic injury. Chronic study animals remained healthy. Study endpoints met; ERIC devices shown to be safe and equivalent.
    Clinical Study - Primary Effectiveness Outcome (mTICI 2b-3)The primary effectiveness outcome (successful reperfusion, mTICI 2b-3) success rate for the ERIC device cohort must be statistically non-inferior to the mTICI rate for the Trevo/Solitaire cohort.82.2% (76.9-87.5%) for ERIC (166/202) vs. 80.7% (78.2-83.1%) for Trevo/Solitaire (842/1044). When considering missing subjects/rescue as failures: 63.1% for ERIC (130/206) vs. 60.4% for Trevo/Solitaire (639/1058). Conclusion: Substantially equivalent.
    Clinical Study - Primary Safety Outcome (sICH within 24-48h)The rate of occurrence of symptomatic intracerebral hemorrhage (sICH) within 24-48 hours post-procedure for the ERIC device cohort must be statistically non-inferior to the sICH rate for the Trevo/Solitaire cohort.4.1% (1.1-7.2%) for ERIC (7/169) vs. 8.4% (6.6-10.2%) for Trevo/Solitaire (81/965). (Note: A higher proportion of missing follow-up CT/MRI in the ERIC group means the actual sICH rate could be higher than the reported CI). Conclusion: Substantially equivalent.
    Clinical Study - Secondary Outcome (Good Clinical Outcome - 90-day mRS 0-2) (for substantial equivalence support)The 90-day mRS of the ERIC arm must be substantially equivalent to the control arm based on available follow-up data.49.3% (40-57.7%) for ERIC (67/136) vs. 44.9% (41.7-48.2%) for Trevo/Solitaire (413/919). Conclusion: Substantially equivalent.

    II. Sample Size and Data Provenance (Clinical Study)

    • Sample Size for Test Set:
      • ERIC Device Cohort: N = 206 patients
      • Trevo/Solitaire Cohort: N = 1058 patients
    • Data Provenance: Prospective, multi-center, observational study (ETIS - Endovascular Treatment in Ischemic Stroke follow-up Evaluation Observational Cohort Study, NCT03776877). Data collected from all 7 active sites from the beginning of the study to September 2018.
    • Country of Origin: France.
    • Retrospective or Prospective: Prospective.

    III. Number of Experts and Qualifications (Clinical Study Ground Truth / Patient Selection)

    • "All patients were selected for endovascular thrombectomy based on evaluation by a multi-disciplinary team of physicians including neuroradiologists."
    • Number of Experts: Not explicitly stated, but implies a team approach.
    • Qualifications of Experts: Included "neuroradiologists." Specific years of experience are not mentioned.

    IV. Adjudication Method (Clinical Study Test Set)

    • The document does not explicitly describe an adjudication method for the test set outcomes (mTICI, sICH, mRS). This was an observational study based on data collected over time. Outcomes would have been determined by the treating physicians and documented in the study records.

    V. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. The clinical study compared the performance of the ERIC device against predicate devices (Trevo/Solitaire) in terms of patient outcomes (reperfusion rates, sICH, mRS) as measured in standard clinical practice, not specifically assessing the improvement of human readers with AI assistance. This device is a mechanical thrombectomy device, not an AI-assisted diagnostic tool.

    VI. Standalone (Algorithm Only) Performance

    • Not applicable. This device is a medical device (mechanical thrombectomy device), not an algorithm or AI. The performance metrics presented are for the device's clinical efficacy and safety when used by medical professionals, not a standalone algorithm.

    VII. Type of Ground Truth Used (Clinical Study)

    • Clinical Outcomes/Expert Assessment and Imaging:
      • Successful reperfusion (mTICI 2b-3): Based on angiographic imaging assessed by clinical experts (likely neuroradiologists or interventionalists within the clinical setting).
      • Symptomatic Intracerebral Hemorrhage (sICH): Based on clinical assessment and follow-up CT scan/MRI, interpreted by medical professionals.
      • Good Clinical Outcome (90-day mRS 0-2): Modified Rankin Scale (mRS) score assessed clinically by medical professionals at 90 days.

    VIII. Sample Size for Training Set

    • Not applicable. The ERIC Retrieval Device is a hardware medical device, not an AI/ML algorithm that requires a training set. The clinical study investigated its performance against established predicate devices.

    IX. How the Ground Truth for the Training Set Was Established

    • Not applicable. As described above, there is no training set for this type of hardware medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162539

    Validate with FDA (Live)

    Date Cleared
    2016-11-10

    (59 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K113455, K123378, K141491

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Solitaire™ 2 Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
    2. The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
    Device Description

    The Solitaire™ 2 Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ 2 Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA). M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ 2 Revascularization Device facilitates clot retrieval and has Iridium radiopaque markers on the proximal and distal ends. The devices are supplied sterile and are intended for single-use only.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the Solitaire™ 2 Revascularization Device meets those criteria. Here's a structured breakdown of the requested information:

    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents two main categories of performance data: biocompatibility and bench testing. Clinical data is used for comparative effectiveness rather than direct acceptance criteria for the device itself in this specific section.

    Biocompatibility Testing:

    Test CategoryTest DescriptionMethodAcceptance CriteriaConclusion
    CytotoxicityL929 MTT CytotoxicityISO 10993-5Viability is ≥70%.Acceptance criteria met
    SensitizationGuinea Pig Maximization SensitizationISO 10993-10Test article does not elicit a sensitization response.Acceptance criteria met
    IrritationIntracutaneous Irritation TestISO 10993-10Differences in the mean test and control scores of the extract dermal observations are < 1.0.Acceptance criteria met
    Acute Systemic ToxicityAcute Systemic Injection TestISO 10993-11No abnormal clinical signs and weight loss in excess of 10%.Acceptance criteria met
    Materials Mediated Rabbit Pyrogen-Temperature rise ≥0.5°C (Note: This looks like a rejection criterion, implies "no temperature rise ≥0.5°C" would be the acceptance)Acceptance criteria met
    Hemo-compatibilityHemolysisISO 10993-4No significant differences between the test article extract and negative control article results. The test article is considered non-hemolytic.Acceptance criteria met
    Partial Thromboplastin Time-Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway.Acceptance criteria met
    Platelet and Leukocyte Count-Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material.Acceptance criteria met
    Complement Activation C3a and SC5b-9 Assay-Levels of C3a and SC5b-9 are comparable and less than the positive control.Acceptance criteria met
    Thrombosis-Thrombo-resistance properties are acceptable in clinical use.Acceptance criteria met
    GenotoxicityBacterial Mutagenicity TestISO 10993-3Test article is considered non-mutagenic.Acceptance criteria met
    In-vitro Mouse Lymphoma Assay-Test article is considered non-mutagenic.Acceptance criteria met
    In-vivo Mouse Micronucleus Assay-Test article is considered non-mutagenic.Acceptance criteria met

    Performance Bench Testing:

    Test DescriptionMethodAcceptance CriteriaConclusion
    Delivery ForcePeak delivery force was measured through a representative tortuous anatomical model.Stent must be below delivery force specification.Acceptance criteria met
    Re-sheathing ForceRetrieval force was measured through a representative tortuous anatomical model.Stent must be below re-sheathing force specification.Acceptance criteria met
    Total System LengthTotal system length was measured from the proximal tip of the push-wire to the distal-most tip of the finger marker coils.System length must meet product specification.Acceptance criteria met
    Fluorosafe Marker LengthThe total length of the fluorosafe marker was measured.Fluorosafe marker length must meet product specification.Acceptance criteria met
    Multiple Re-sheathing DurabilitySamples were evaluated on their ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use (IFU).Device must reliably deploy and resheath up to four times.Acceptance criteria met
    Body Marker TensileBody Marker tensile strength testing is performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol distal finger of the device.Body marker should be greater than or equal to existing tensile strength specification.Acceptance criteria met
    Body Marker RadiopacityVerification analysis of body markers.The radiopaque body markers must be visible using standard catheter laboratory equipment.Acceptance criteria met
    Proximal Marker to Distal MarkerThe length of the laser cut and electro-polished stents are measured 100% in process.Length of stent must meet all inspection criteria.Acceptance criteria met
    Torque ResponseSamples were evaluated to determine the number of turns required to produce 1 rotation of the distal tip of the device.Device turns must be less than or equal to torque response criteria.Acceptance criteria met
    Torque StrengthSamples were evaluated to determine the number of turns required to break the device in a representative tortuous model.Device turns must be greater than torque strength criteria.Acceptance criteria met
    System TensileSamples were evaluated to determine the tensile strength of the full system.Devices must be greater than or equal to the system tensile criteria.Acceptance criteria met
    Af TemperatureIn-process 100% tracking of heat set parameters used to set final Af.Temperature should be less than or equal to existing Af temperature specification.Acceptance criteria met
    Radial ForceThe radial force was measured 100% in-process.Stent must be within existing radial force specification.Acceptance criteria met
    Kink ResistanceKink resistance testing verified the device flexibility across various levels of tortuosity.Device must be able to maintain vessel wall apposition at a minimum radius bend.Acceptance criteria met
    Clot Removal in a Simulated Neurovascular ModelBoth hard and soft clot retrieval success was evaluated in an in vitro tortuous anatomical model.Overall success rate of the device (usability and effectiveness) must be equal to or better than the predicate* device.Acceptance criteria met
    Particulate Under Simulated Use ConditionsDevice was evaluated for particulate generation under simulated use in a representative tortuous anatomical model per USP<788>.Device was evaluated for particulate generation under simulated use in a representative tortuous anatomical model per USP<788> (Same as method, implying conforming to USP<788> limits).Acceptance criteria met

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Clinical Study (SWIFT PRIME):
      • Total Randomized Subjects: 196 (98 in each group: IV t-PA alone vs. IV t-PA + Solitaire)
      • Analysis Cohort (after exclusions): 161 subjects (84 in the IV t-PA + Solitaire™ group and 77 with IV t-PA only). Further refined to 144 subjects for primary and secondary efficacy endpoints after additional exclusions.
      • Data Provenance: Global, multicenter. The study was a prospective, randomized, open, blinded endpoint (PROBE) clinical study (IDE G120142).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not explicitly state the exact number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical study's endpoints. However, it does mention:

    • Blinded evaluation of modified Rankin Scale (mRS) for neurological disability outcomes. This implies that the mRS scores, which serve as a critical component of the ground truth for effectiveness, were assessed by experts who were blinded to the treatment arm.
    • Clinical Events Committee adjudication for adverse events. This indicates a panel of experts reviewed and categorized adverse events.
    • Core Laboratory assessed data for symptomatic ICH, infarct volume, and reperfusion ratio. This suggests specialized facilities with expert staff were responsible for these assessments.

    While the specific count and detailed qualifications are not provided, the involvement of blinded evaluators, a Clinical Events Committee, and a Core Laboratory indicates that ground truth was established through expert assessment according to established protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document mentions "Clinical Events Committee adjudication" for adverse events. While it doesn't specify a 2+1 or 3+1 method, "adjudication" implies a process where a committee of experts reviews and resolves discrepancies in event classification or assessment. It suggests a structured review by multiple parties, but the exact number of reviewers per case or tie-breaking mechanism is not detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done, or any mention of AI assistance. This study compared a medical device (Solitaire™ 2 Revascularization Device) with or without IV t-PA, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done, as this document is about a mechanical thrombectomy device, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical study was established using a combination of:

    • Clinical Outcomes Data: Primarily, the 90-day global disability assessed via the blinded evaluation of Modified Rankin Scale (mRS). This is a widely accepted functional outcome scale often based on trained interviewer assessment.
    • Imaging-based Assessments: Volume of cerebral infarction, reperfusion ratio, and arterial revascularization (TICI 2b or 3) assessed by a Core Laboratory. This implies expert interpretation of medical images.
    • Adjudicated Adverse Events: Reviewed and categorized by a Clinical Events Committee.

    8. The sample size for the training set

    The document does not describe the development or training of an algorithm or AI. Therefore, there is no training set sample size mentioned. The clinical study (SWIFT PRIME) is a comparative effectiveness study for a medical device in patients.

    9. How the ground truth for the training set was established

    As there is no training set for an algorithm or AI described in the document, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123378

    Validate with FDA (Live)

    Date Cleared
    2012-11-30

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113455

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solitaire™ 2 Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Device Description

    The Solitaire™ 2 device is designed to restore blood flow in subjects experiencing ischemic stroke due to large intracranial vessel occlusion within 8 hours of symptom onset. It is indicated for subjects who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy. The distal nitinol portion of the device facilitates clot retrieval and has Pt-Ir radiopaque markers on the proximal and distal ends.
    The following modifications have been made to the device in support of this application:
    • The attachment zone has been redesigned for greater tensile strength.
    • The marker band has been redesigned to aid the crimping process.
    • The pushwire now contains a fluorosafe marker.
    • The Solitaire™ 2 Device uses one piece of PTFE tubing.

    AI/ML Overview

    This K123378 submission for the Solitaire™ 2 Revascularization Device is a medical device submission for a thrombus retriever, not an AI/ML device. Therefore, the requested information regarding acceptance criteria and studies for an AI/ML device (e.g., sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance) is not applicable.

    The submission focuses on demonstrating substantial equivalence to a predicate device (Solitaire™ FR Revascularization Device) based on modifications to the device design and subsequent non-clinical bench testing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a non-clinical bench testing submission for a medical device modification, the "acceptance criteria" are implied by successful completion of the specified tests, demonstrating that the device modifications do not adversely affect its performance compared to the predicate. No specific numerical thresholds (e.g., sensitivity, specificity) for performance are provided as would be expected for an AI/ML diagnostic or prognostic tool.

    Acceptance Criteria (Implied by Successful Testing)Reported Device Performance (Summary)
    Delivery Force within acceptable limitsTest performed to support changes
    Withdrawal Force within acceptable limitsTest performed to support changes
    Total System Length within specificationsTest performed to support changes
    Fluoro Safe Marker Length within specificationsTest performed to support changes
    Distal Tip to Fluoro Marker Length within specsTest performed to support changes
    Durability maintainedTest performed to support changes
    Radiopacity maintainedTest performed to support changes
    Torque Response maintainedTest performed to support changes
    Torque Strength maintainedTest performed to support changes
    System Tensile Strength maintainedTest performed to support changes
    Performance after 1-Year Accelerated Aging StudyDelivery Force, Withdrawal Force, Durability, Torque Response, Torque Strength, System Tensile Strength maintained

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: Not applicable. These were non-clinical bench tests on the device itself, not clinical data sets. The "sample" would refer to the number of devices or components tested for each parameter. This information is not detailed in the summary provided.
    • Data provenance: Not applicable in the context of clinical data. The tests were performed in a lab setting by the manufacturer (Micro Therapeutics, Inc. d/b/a ev3 Neurovascular).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth in this context typically refers to clinical diagnosis or outcome. For bench testing, the "ground truth" is defined by the engineering specifications and established test methods for device performance parameters.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a clinical study involving reader adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For this device, the "ground truth" for non-clinical testing is defined by the engineering specifications and performance standards for medical devices of this type, which are assessed through a series of defined bench tests. There is no biological or clinical "ground truth" in the AI/ML sense in this submission.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device.

    In summary, the K123378 submission demonstrates substantial equivalence for the Solitaire™ 2 Revascularization Device primarily through extensive non-clinical bench testing, focusing on the mechanical and physical performance characteristics of the modified device components. It does not involve any AI/ML components or clinical data analyses as would be relevant for the questions posed.