Found 2 results

510(k) Data Aggregation

    K Number
    K151015

    Validate with FDA (Live)

    Date Cleared
    2015-07-29

    (104 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113145

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECHELON Oval System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

    Device Description

    The ECHELON OVAL is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design. The design was based on the ECHELON MRI system. The ECHELON OVAL has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.

    AI/ML Overview

    This request is asking for a detailed breakdown of the acceptance criteria and the study used to validate the ECHELON Oval V5.0 MRI system, based on the provided FDA 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in a traditional table format for performance metrics. Instead, the acceptance is based on qualitative assessments and comparisons to the predicate device.

    Testing TypeAcceptance Criteria (Implied)Reported Device Performance
    Performance Testing - ClinicalAcceptable image quality for clinical use."A radiologist validated that the clinical images have acceptable image quality for clinical use."
    Performance Testing - BenchNew features perform as intended for diagnostic use."We confirmed that each new feature performs as intended for diagnostic use." Specifically for each new feature: - ASL-Perfusion: "Test results confirm ASL-Perfusion acquires perfusion images using labeled blood flowing into the brain tissue without Contrast-Enhanced both in phantom simulations and clinical results." - Beam Sat VASC-ASL: "Test results confirm Beam Sat improves the visibility of the portal vein in making MIP images both in phantom simulations and clinical results." - Breast MRS: "Test results from phantom simulations and clinical results confirm MRS (Magnetic Resonance Spectroscopy) acquires the magnetic resonance signal of in vivo metabolites through chemical shift phenomenon and can detect Choline as metabolite in the breast area." - Enhanced PC: "Test results from phantom simulations and clinical results indicate a reduction in scan time of phase contrast (PC) sequence in 2D and 3D by shorting the TR by optimizing velocity encode gradient and applied parallel imaging (RAPID). As a result of this improvement, we can shorten scan time of '4D flow' which is time-resolved (CINE) three-dimensional (3D) spatial encoding combined with three-directional velocityencoded phase contrast MRI." - Fat and water separation scan (FSE, RSSG, GE): "Test results from phantom simulations and clinical results indicate reliable and uniform fat suppression by utilizing the difference between resonant frequencies due to chemical shift of water protons and fat protons to obtain a water image and a fat image. The chemical shift of a water signal and a fat signal receives two echo signals at the timing which becomes an in-phase and an out-of-phase. By adding and subtracting it, a water image and a fat image are simultaneously acquirable." - k-RAPID: "Test results from phantom simulations and clinical results indicate that k-space parallel imaging technique accelerates the scan by acquiring k-space data with skipped phase encoding and skipped position which is filled with estimated data by the interpolation of neighboring data." - Multi b and DKI: "Test results from phantom simulations and clinical results confirm Multi b DKI images can be acquired in one scan utilizing Tensor 15 and Tensor 30 being added to the number of MPG Axes. Diffusion Kurtosis Imaging (DKI) is the diffusion-weighted imaging technique in restriction." - opFSE / opFIR: "Test results from phantom simulations and clinical results confirm by deriving opFSE and opFIR sequence from the primeFSE and primeFIR image quality is improved." - PBSG: "Test results from phantom simulations and clinical results confirm PBSG which is a sequence based on BASG sequence improves to mitigate the dark band artifact which is unique to BASG sequence. The PBSG sequence makes it possible to acquire BASG images under the condition of inhomogeneous magnetic field with less band artifact." - RADAR-GE/TOF: "Test results from phantom simulations and clinical results confirm the RADAR measurement feature is functioning with the GE and TOF sequence." - T2 RelaxMap:* "Test results from phantom simulations and clinical results confirm that T2* relaxation time can be mapped on morphological image in color by using T2* RelaxMap function. The T2* RelaxMap function consists of (a) acquisition of multi-echo images (up to 32) and (b) analysis of T2* relaxation time."
    Substantial EquivalenceHardware, coils, and functionality are substantially equivalent to the predicate device in design features, fundamental scientific technology, indications for use, and safety and effectiveness."based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed ECHELON Oval V5.0 MRI is considered substantially equivalent to the currently marketed predicate device (ECHELON Oval MRI System (K113145)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states:

    • "Clinical images were collected and analyzed, to ensure that images from the new features meet user needs."
    • "We provide clinical image examples for each new feature..."

    However, the specific sample size (number of images or patients) used for the clinical test set is not provided in the document.

    The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It only mentions "clinical images were collected," implying real patient data, but without further detail.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: "A radiologist" (singular) validated the clinical images.
    • Qualifications of Experts: The document specifies "a radiologist." No further details on their experience (e.g., "10 years of experience") are provided.

    4. Adjudication Method for the Test Set

    The document mentions "a radiologist validated" the images. This suggests a single expert review. There is no mention of an adjudication method such as 2+1, 3+1, or any multi-reader consensus process for the clinical images.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done or reported in this submission. This submission is for a software update to an MRI system itself, not an AI-assisted diagnostic tool for readers. The "AI" features (like k-RAPID for image acceleration or certain processing tasks) are integrated into image acquisition and reconstruction, not designed to directly assist human interpretation in a comparative effectiveness study setting mentioned in the question.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The "Performance Evaluation" section indicates that both "Performance Testing - Clinical" and "Performance Testing - Bench" were conducted for the new features.

    • Bench tests involved "phantom simulations," which represents an algorithm-only evaluation for the technical performance of specific sequences.
    • Clinical tests involved collecting and analyzing clinical images, and "a radiologist validated" them. While a human reviewed the output, the core algorithms for the new features (e.g., ASL-Perfusion generating perfusion images, T2* RelaxMap creating maps) operate in a standalone manner to produce these outputs before human interpretation.

    Therefore, standalone performance tests were conducted for the new features on phantoms, and the resulting clinical images were then validated by a radiologist. The focus was on the technical and image quality output of the new algorithms, not diagnostic accuracy in a standalone AI context compared to a ground truth.

    7. The Type of Ground Truth Used

    • For the bench tests (phantom simulations), the "ground truth" would be the known physical properties and configurations of the phantom, allowing for objective measurement of parameters like signal-to-noise ratio, geometric distortion, etc.
    • For the clinical validation, the ground truth appears to be expert consensus or opinion from "a radiologist" who validated the "acceptable image quality for clinical use." There is no mention of pathology, outcomes data, or a multi-expert consensus forming the ground truth.

    8. The Sample Size for the Training Set

    The document does not describe the development or training of new algorithms in a way that suggests a "training set" in the context of machine learning. The enhanced features are described as software updates and new pulse sequences. Therefore, there is no information provided regarding a training set size. This submission focuses on the performance verification and validation of the software update, not the development process of, for example, a deep learning model.

    9. How the Ground Truth for the Training Set was Established

    Since no training set is mentioned in the context of machine learning algorithm development, this information is not applicable and not provided in the document. The document describes system updates and their performance validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142734

    Validate with FDA (Live)

    Date Cleared
    2015-03-02

    (160 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    ECHELON Oval MRI System (K113145)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRILLIUM Oval MRI System is an imaging device, and is intended to provide the physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

    Anatomical Region: Head, Body, Spine, Extremities
    Nucleus excited: Proton
    Diagnostic uses: T1, T2, proton density weighted imaging, Diffusion weighted imaging, MR Angiography, Image processing, Spectroscopy, Whole Body

    Device Description

    The TRILLIUM OVAL is a Magnetic Resonance Imaging System that utilizes a 2.9 Tesla superconducting magnet in a gantry design. The design was based on the ECHELON Oval MRI system. The TRILLIUM OVAL has been designed to enhance clinical utility as compared to the ECHELON Oval by taking advantage of the stronger magnetic field and stronger gradient field and slew rate.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the "Trillium Oval MR System." It primarily focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed information about acceptance criteria or a specific study proving the device meets those criteria in the context of diagnostic accuracy, particularly for AI/CADe systems.

    This document describes a Magnetic Resonance Imaging (MRI) system, which is a diagnostic imaging device. The "acceptance criteria" discussed are largely related to safety and general performance standards for MRI systems, rather than specific diagnostic performance metrics (e.g., sensitivity, specificity) for a particular clinical application. The "study" refers to non-clinical and some clinical imaging tests to demonstrate that the device meets these engineering and safety standards.

    Here's an analysis based on the provided text, addressing the requested points:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a table of specific acceptance criteria with corresponding performance metrics in the format typically used for evaluating diagnostic accuracy (e.g., sensitivity, specificity, AUC). Instead, it lists non-clinical test standards and mentions that the device was subjected to these tests. The implication is that the device met these standards.

    Acceptance Criteria (Test Standards)Reported Device Performance
    MRI Test Standards (NEMA)Implicitly met, as detailed results are not provided but submission states system was "subjected to" these tests and concludes substantial equivalence.
    NEMA MS 1-2008: Signal-to-noise Ratio (SNR)The device was subjected to testing regarding SNR.
    NEMA MS 2-2008: Two-Dimensional Geometric DistortionThe device was subjected to testing regarding geometric distortion.
    NEMA MS 3-2008: Image UniformityThe device was subjected to testing regarding image uniformity.
    NEMA MS 4-2010: Acoustic Noise Measurement ProcedureThe device was subjected to testing regarding acoustic noise.
    NEMA MS 5-2010: Slice ThicknessThe device was subjected to testing regarding slice thickness.
    NEMA MS 7-1993 (Rev. 1998): Time-Varying Gradient Fields (dB/dt)The device was subjected to testing regarding time-varying gradient fields.
    NEMA MS 8-2008: Specific Absorption Rate (SAR)The device was subjected to testing regarding SAR.
    Additional Test Standards (Safety and Electrical)Implicitly met, as the submission states compliance with these standards.
    AAMI / ANSI ES60601-1:2005/(R) 2012: Basic Safety & Essential PerformanceThe device complies with general requirements for basic safety and essential performance.
    IEC 60601-1-2 Ed 3:2007-03: EMC Requirements and TestsThe device complies with electromagnetic compatibility requirements.
    IEC 60601-2-33 Ed 3.1 2013-04: Particular Requirements for MR EquipmentThe device complies with particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.
    IEC 62304 Ed 1:2006-05: Medical Device Software Life Cycle ProcessesThe device complies with medical device software life cycle processes.
    Additional Non-clinical Testing (SAR Simulations & Other Validations)Results of these simulations are presented or implied to support safety and performance, contributing to the conclusion of substantial equivalence. (e.g., "Validation of Electromagnetic Simulation Summary," "B1 Map Comparison," "Validation for SAR Simulation," "Uncertainty analysis for local SAR," "Comparison of local SAR between TRILLIUM OVAL and ECEHLON C," "Worst case analysis of local SAR," "Simulation Results: SAR Hugo Model," "Simulation Results: SAR Fats Model," "Simulation Results: SAR Hanako Model," "Simulation Results: SAR Roberta Model").
    Clinical Testing (Image Quality)Sample clinical imaging of the head, torso, and extremities using all anatomy coils was performed. This is less about specific quantifiable metrics and more about visual assessment of image quality and suitability for diagnostic interpretation, as specified in FDA guidance for MRI 510(k)s. The implicit acceptance criterion is that the images are of diagnostic quality comparable to the predicate.

    Regarding the Absence of AI/CADe Specifics:

    It's crucial to note that this submission is for a Magnetic Resonance Imaging System (hardware and core software), not an Artificial Intelligence (AI) or Computer-Assisted Detection/Diagnosis (CADe/CADx) software. Therefore, the questions related to AI performance metrics (like sample sizes for test/training sets, expert ground truth adjudication for AI, MRMC studies, standalone algorithm performance, etc.) are not applicable to this specific document. The "device" here is the MRI scanner itself.


    Specific Answers to Your Questions (where applicable for an MRI system submission):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Test Set Sample Size: The document mentions "sample clinical imaging of the head, torso, and extremities using all anatomy coils." It does not specify the number of subjects or images used for this clinical imaging.
      • Data Provenance: Not specified in the document (e.g., country of origin). The document implies these were internally generated as part of the submission process, likely prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • This is not applicable in the context of this 510(k) submission for an MRI system. The "ground truth" for the clinical imaging mentioned would be the visual assessment by "trained physicians" (as mentioned in the indications for use) to confirm diagnostic quality, but the number and qualifications of evaluators are not specified.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. This level of detail on ground truth adjudication is typically for AI/CADe performance studies, not for the submission of a core imaging device.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No. This is not an AI/CADe device. Such a study was not performed or described.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • No. This is not an AI/CADe device. The device is an MRI system, which always requires human operation and interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • For the non-clinical tests (NEMA standards, SAR simulations), the "ground truth" is adherence to established engineering and safety specifications and theoretical models.
      • For the "sample clinical imaging," the implied ground truth is that the images generated are of sufficient diagnostic quality as determined by "trained physicians," comparable to the predicate device. This is a qualitative assessment of imaging capability rather than a definitive diagnosis of a specific condition against a gold standard like pathology.
    7. The sample size for the training set

      • Not applicable. This is not an AI/CADe device that uses a "training set" in the machine learning sense.
    8. How the ground truth for the training set was established

      • Not applicable.