Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K141294

    Validate with FDA (Live)

    Date Cleared
    2014-08-25

    (98 days)

    Product Code
    Regulation Number
    884.5470
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K112635

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kimberly-Clark* U by Kotex® Security® is an unscented menstrual tampon inserted into the vagina to absorb menstrual fluid.

    Device Description

    This device is a conventional unscented menstrual tampon consisting of an absorbent pledget, a withdrawal cord and an applicator. The absorbent pledget consists of a ribbon of rayon fibers and cotton, overwrapped with a non-woven fabric. A withdrawal cord is placed through the pledget and knotted. The tampon pledget is compressed. The formed pledget is inserted into a two-piece plastic applicator consisting of an inner tube (plunger), and an outer insertion tube (barrel) with flexible petals that form a closed, rounded tip. Each tampon is individually wrapped in a plastic film wrapper and multiple tampons are packaged in sealed cartons for commercial sale. Tampons are available in Regular, Super and Super Plus absorbencies in various counts.

    AI/ML Overview

    This document describes a 510(k) premarket notification for Kimberly-Clark U by Kotex® Security® Unscented Menstrual Tampons*. The purpose of the submission is to demonstrate substantial equivalence to legally marketed predicate devices.

    The information provided focuses on the biocompatibility and absorbency of the device. It does not describe an AI medical device or a clinical study with human readers. Therefore, several sections of your request are not applicable to this document.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / TestStandardReported Device Performance
    Genotoxicity Test (MLA)ISO 10993, Part 3Meets
    Genotoxicity Test (AMES)ISO 10993, Part 3Meets
    Genotoxicity Test (Mouse Micronucleus)ISO 10993, Part 3Meets
    Cytotoxicity TestISO 10993, Part 5Meets
    Mucosal Irritation TestISO 10993, Part 10Meets
    Mucosal Sensitization TestISO 10993, Part 10Meets
    Acute Systemic Toxicity TestISO 10993, Part 11Meets
    Colorant Extraction TestUSP 661Meets
    Syngyna Absorbency Requirements21 CFR §801.430Complies

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes for the specific biocompatibility tests. It mentions that "Testing from the reference device, U by Kotex® Sleek® K112635, was used to support the safety of new applicator colors." This indicates that some of the test data was retrospective, drawn from a previously approved device. The country of origin for the data is not specified, but the submission is to the U.S. FDA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This document is for a medical device (menstrual tampon), not an AI device relying on expert-established ground truth from images or clinical assessments. The "ground truth" here is the physical and chemical properties of the materials and the absorbency performance as measured by standardized tests.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human interpretation or adjudication of results. The results are objective measurements against established standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI medical device and therefore no MRMC study comparing human readers with and without AI assistance was conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is not an AI medical device.

    7. Type of Ground Truth Used

    The ground truth used for this device's performance evaluation includes:

    • Standardized Biocompatibility Test Results: Based on ISO 10993 series and USP 661, these are objective laboratory measurements of material properties and their biological interactions.
    • Syngyna Absorbency Requirements: A regulatory standard (21 CFR §801.430) for measuring absorbency, representing a controlled, objective test.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set. The "training" for this device would be its manufacturing process and quality control.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1