Found 4 results

510(k) Data Aggregation

    K Number
    K153412

    Validate with FDA (Live)

    Device Name
    Sleep Profiler
    Date Cleared
    2016-03-14

    (110 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    Apnea Risk Evaluation System (ARES), Model 610 (K112514)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sleep Profiler is intended for use for the diagnostic evaluation by a physician to assess sleep quality and score sleep disordered breathing events in adults only. The Sleep Profiler is a software-only device to be used under the supervision of a clinician to analyze physiological signals and automatically score sleep study results; including the staging of sleep, detection of arousals, snoring and sleep disordered breathing events (obstructive apneas, hypopneas and respiratory event related arousals). Central and mixed apneas can be manually marked within the records.

    Device Description

    The Sleep Profiler is a software application that analyzes previously recorded physiological signals obtained during sleep. The Sleep Profiler software can analyze any EDF files acquired with the Advanced Brain Monitoring X4 System and the X8 System models SP40 and SP29. Automated algorithms are applied to the raw signals in order to derive additional signals and interpret the raw and derived signal information. The software automates recognition of: Sleep stages Rapid Eye Movement (REM) and nREM (N1, N2, N3) and wake, Heart/pulse rate, Snoring loudness, Sleep/wake, Head movement and position, Snoring, sympathetic, behavioral and cortical arousals, ECG,EOG, EMG waveform, SpO2, Airflow, Respiratory Effort, Apneas and Hypopneas, Oxygen desaturations. The software identifies and rejects periods with poor electroencephalography signal quality. The full disclosure recording of derived signals and automated analyses can be visually inspected and edited prior to the results being integrated into a sleep study report. Medical and history information can be input from a questionnaire. Responses are analyzed to provide a pre-test probability of Obstructive Sleep Apnea (OSA). The automated analyses of physiological data are integrated with the questionnaire data, medical and history information to provide a comprehensive report. Several report formats are available depending on whether the user has acquired more than one night of data, wishes to obtain a narrative summary report or provide patient reports. The Sleep Profiler software can be used as a stand-alone application for use on Microsoft Windows 7 & 8 operating system platforms (desktop model). Alternatively, the user interface (i.e., entry or editing of information) can be delivered via a web-portal (portal model). The capability to enter or edit patient information, call the application to generate a study report, and/or download a report is provided using either the desktop PC application or web-portal application. The same analysis and report generation software is used for both the desktop and web-portal applications.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Sleep Profiler device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    EndpointAcceptance Criteria (Equivalent to Predicate Device)Reported Device Performance (Sleep Profiler)
    AHI for OSA DiagnosisMinimum targeted positive likelihood ratios for AHI > 5 and AHI > 15 are 3.5 and 5.0 respectively (equivalent to predicate ARES).Overall AHI: - AHI ≥ 5: Positive Likelihood Ratio = 6.67 - AHI ≥ 15: Positive Likelihood Ratio = 33.0 REM AHI: - AHI ≥ 5: Positive Likelihood Ratio = 8.84 - AHI ≥ 15: Positive Likelihood Ratio = 18.33 (These values exceed the minimum targeted likelihood ratios indicating equivalence)
    Sleep Staging (Agreement with Expert Consensus)Comparison of auto-detected staging to PSG results obtained by expert raters, showing equivalent agreement to the predicate Sleep Profiler (K130007). No specific numeric thresholds are explicitly stated as acceptance criteria, but generally high concordance is expected for substantial equivalence.Overall (n=43 subjects, 3 raters): - Wake: Positive Agreement 0.73, Negative Agreement 0.94 - N1: Positive Agreement 0.25, Negative Agreement 0.93 - N2: Positive Agreement 0.77, Negative Agreement 0.84 - N3: Positive Agreement 0.76, Negative Agreement 0.94 - REM: Positive Agreement 0.74, Negative Agreement 0.97 (The document states "met" for this endpoint by comparison to the predicate.)

    2. Sample Sizes and Data Provenance for Test Set

    • Sample Size (for AHI/OSA detection): 60 subjects for overall AHI, 40 subjects for REM AHI.
    • Sample Size (for Sleep Staging): A subset of 43 subjects from the AHI study, with at least 4 hours of raw X8 diagnostic recording time.
    • Data Provenance: The document states "signals acquired with the X8 System concurrent to polysomnography (PSG)." This implies the data was prospectively collected for this evaluation, likely from a clinical setting, but the country of origin is not specified.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Number of Experts:
      • For AHI/OSA detection: "one rater per study" for PSG results.
      • For Sleep Staging: "weighted majority agreement of three raters" for the 43-subject subset.
    • Qualifications of Experts: Not explicitly stated beyond being "rater(s)" for PSG and "expert scoring" for sleep staging. Standard practice for such studies would imply board-certified sleep technologists or physicians experienced in PSG scoring.

    4. Adjudication Method for the Test Set

    • For AHI/OSA detection: "one rater per study." This implies no formal adjudication/consensus process among multiple independent reviewers, as only a single rater's PSG results were used as ground truth for each case.
    • For Sleep Staging: "weighted majority agreement of three raters." This indicates an adjudication method where the consensus of three experts established the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study is explicitly described where human readers' performance with and without AI assistance is compared. The study primarily focuses on the standalone performance of the Sleep Profiler software against expert-scored PSG.

    6. Standalone Performance Study

    • Yes, a standalone study was performed. The "Sleep Profiler software accuracy was clinically validated with signals acquired with the X8 System concurrent to polysomnography (PSG)." The results presented for AHI and sleep staging are for the algorithm's performance without human intervention, compared to human-scored PSG.

    7. Type of Ground Truth Used

    • Expert Consensus / Human Scoring: The primary ground truth for both AHI detection and sleep staging was derived from Polysomnography (PSG) results scored by human experts/raters. For sleep staging, it was specifically based on the "weighted majority agreement of three raters."

    8. Sample Size for the Training Set

    • The document does not explicitly state the sample size used for the training set of the Sleep Profiler software. It describes the clinical validation study (test set) but not the development data.

    9. How Ground Truth for the Training Set Was Established

    • The document does not provide details on how the ground truth for the training set was established. Typically, for such devices, training data would also be derived from expert-scored PSG, similar to the test set, but this information is not included in the 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152040

    Validate with FDA (Live)

    Date Cleared
    2015-10-30

    (100 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K112514, K072201, K120047, K131383

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X8 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit, and display physiological signals from adult patients. All X8 models (SP40, SP29, and XS29) acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals. Model SP29 additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. The X8 system only acquires and displays physiological signals; no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability.

    Device Description

    The X8 System is indicated for acquiring, recording/storing, transmitting, and displaying physiological data in patients. It can be used by patients in the home, healthcare facility, or clinical research environment. Patients can move within their home or healthcare environment without having to remove the device (e.g. walk to the restroom).

    The X8 System is comprised of the X8 device which is worn on the patient's head and body, accessories, the Device Manager software, and the X-Series Basic-Utility Software. The study records are saved to the PC in EDF format and are available for analysis by third party software applications, such as Persyst Reveal (K011397).

    The X8 System combines hardware, firmware, and software to acquire physiological signals. It acquires physiological data through a battery powered headset worn by the patient and provides a flexible platform for applying sensors and acquiring signals from multiple locations on the head or body, transmitting and recording the signals, and providing visual and auditory indications to ensure high quality data are obtained.

    Model SP40 Sleep Profiler is applied by the patient to acquire and record electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals during sleep. This model utilizes the X8 Sleep Profiler Strip.

    Model SP29 Sleep Profiler PSG2 is applied by the patient to acquire and record electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals during sleep. This model additionally includes a nasal pressure transducer and cannula (for airflow), thoracic and abdomen respiratory effort, and pulse rate and oxyhemoglobin saturation from the finger. This model utilizes the X8 Sleep Profiler Strip.

    Model XS29 X8 Stat is applied by a technician to acquire, record, transmit, and/or display electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG) signals, with optional accelerometer, acoustical, and photoplethysmographic signals acquired during non-sleep conditions. This model utilizes either the X8 Midline or the X8 Referential strip.

    The Device Manager software application provides a means to communicate with the X8 Device, transfer study data, and format a device. The software transfers data saved in the memory of the X8 device using either a PC application or a web-based data entry application operating in a cloud server environment.

    The X-Series Basic-Utility Software acquires, presents, and stores physiological signals from the X8 Device. The software has a modular architecture that allows the users to interact using either the Graphical User Interface (GUI) provided with the installation, or programmatically via a Software Development Kit. Additional functionalities provided by X-Series Basic-Utility Software include impedance measurements, custom markers, and interface with the Bluetooth Receiving Dongle.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: X8 System (Sleep Profiler (SP40), Sleep Profiler PSG2 (SP29), Stat X8 (XS29))
    Study Type: Prospective Comparison and Self-Application Study of X8 System Model SP29

    Acceptance Criterion (Primary Endpoint)Reported Device Performance and Outcome
    A) Comparison of Signals (X8 System airflow and respiratory effort vs. predicate device Compumedics Somte (K072201))
    No more than 10% of the breathing events recorded with the X8-PSG2 airflow signal were inferior to predicate signal.Achieved: No events recorded with the X8-PSG2 airflow signal (0%) were found inferior to the predicate signal.
    No more than 20% of the breathing events recorded with the X8-PSG2 respiratory effort will be inferior to the predicate signal.Achieved: The X8 thorax and abdomen belts were inferior to the predicate in only 4.0% (9/225) and 1.3% (3/232) of the events, respectively (both well below 20%).
    B) Demonstrate that high quality signals can be obtained when the X8 System is self-applied with the user instructions.
    At least 80% of subjects will be able to acquire at least one night of data (i.e., the entire period they were in bed).Achieved: 91% (10 of 11 subjects) were able to acquire at least one night of data for the entire night.
    At least 70% of each night of recording time will be valid across the oximetry, nasal pressure, and effort belt signals. (While not identified as a primary endpoint for cardio-respiratory signal quality, high EEG quality was also assessed.)Achieved: The percentage of good data obtained for oximetry, nasal pressure (airflow), and respiratory effort (thorax and abdomen) exceeded 90% on each night. High EEG quality was also obtained in over 90% of the recording time on each night.
    At least 70% of subjects did not report X8-PSG2 audio alerts (for signal quality) substantially affected (i.e., strongly agreed) their ability to stay asleep.Achieved: 90% (9 of 10 subjects) did not "strongly agree" that the PSG2 made it difficult for them to stay asleep. (Note: One subject was excluded from this calculation, resulting in 10 subjects for this specific endpoint, even though 11 were initially reported for data acquisition ability).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • For comparison of airflow and respiratory effort signals (part A of the study): The specific number of breathing events analyzed is provided (225 for thorax, 232 for abdomen), but the number of subjects contributing to these events is not explicitly stated.
      • For self-application and signal quality (part B of the study): 11 subjects.
      • For audio alerts acceptance criterion: 10 subjects (one subject was excluded from this analysis).
    • Data Provenance: The study was described as "prospective," indicating that the data was collected specifically for this study under controlled conditions defined prior to data collection. The document does not specify the country of origin for the data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts or their qualifications for establishing the ground truth for this comparative study. However, the ground truth for part A (comparison of signals) was implicitly the readings from the predicate device, Compumedics Somte (K072201), an FDA-cleared device. For part B (self-application and signal quality), the ground truth for signal quality likely involved assessment against predefined criteria or by qualified personnel, but this is not detailed.

    4. Adjudication Method for the Test Set

    The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). For the comparative study (part A), the comparison was against an "FDA cleared device," implying the predicate device's output served as the reference standard. For self-application (part B), signal quality was assessed, presumably against accepted norms for such physiological signals, but no multi-expert adjudication process is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted or described in this document. The X8 System "acquires and displays physiological signals; no claims are being made for analysis of the acquired signals with respect to the accuracy, precision, and reliability." Therefore, this device does not involve AI analysis or human reader interpretation for diagnostic claims, but rather accurate signal acquisition.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device's intended use is to "acquire, record, transmit, and display physiological signals." It does not include automated analysis or diagnostic algorithms. Therefore, a standalone algorithm performance study was not applicable and not done. The focus of the performance study was on the quality and equivalence of the acquired physiological signals.

    7. The Type of Ground Truth Used

    • For comparison of airflow and respiratory effort signals (Part A): The ground truth was based on the signals obtained from an FDA-cleared predicate device (Compumedics Somte (K072201)).
    • For self-application and signal quality (Part B): The ground truth for "valid" recording time and "high EEG quality" would have been established by assessing the acquired physiological signals against accepted physiological standards and criteria for signal integrity, likely by trained professionals. The specific method of establishing this ground truth is not detailed beyond the term "good data."

    8. The Sample Size for the Training Set

    The document describes "prospective studies" for verification and validation. Given that the device's function is signal acquisition and display, and no claims for analysis accuracy or reliability are being made, there isn't a "training set" in the context of machine learning. The studies described are performance evaluations of the hardware and software's ability to accurately capture and transmit signals.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, no "training set" in the machine learning sense is described for this device. The ground truth for performance evaluation was established through comparison with a predicate FDA-cleared device and assessment against physiological signal quality standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140190

    Validate with FDA (Live)

    Device Name
    NIGHT SHIFT
    Date Cleared
    2014-05-29

    (125 days)

    Product Code
    Regulation Number
    872.5570
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K100160, K112514

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Night Shift is indicated for prescription use for the treatment of adult patients with positional obstructive sleep apnea with a non-supine apnea-hypopnea index < 20, and to reduce or alleviate snoring. It records position, movement, and sound so that positional changes in sleep quality and snoring can be assessed.

    Device Description

    The Night Shift is worn around the neck to reduce the amount of time the user sleeps in the supine position as a treatment for positional obstructive sleep apnea. Night Shift combines hardware and firmware to detect when the user attempts to sleep in the supine position and can initiate vibro-tactile feedback with increasing intensity until the user shifts to a non-supine position. The initiation of positional feedback from the Night Shift is turned on is programmable to allow the user to fall asleep (if they must) on their back. Each night the Night Shift is worn, it monitors sleep position (% time supine), behavioral sleep efficiency, and snoring levels (% time snoring > 40 and 50 dB) as well as the frequency, duration and intensity of the feedback (when applied). These data can be optionally transferred via the USB port to the Night Shift Web Portal where the user can generate a report to assess how well the positional feedback is working. A "trial" protocol can include one night with no feedback to establish a baseline and two nights with feedback to assess compliance/efficacy. Utilization information is saved on the device that allows reports to be generated that compares daily use by month and monthly averages for one year. The portal also allows the device to be reformatted (to eliminate all previously recorded data) for a new user, adjust the feedback settings to a new user's personal preferences, and/or upgrade the firmware. For large healthcare organizations that limit internet access, desktop software is provided as an alternative to the portal.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Advanced Brain Monitoring, Inc. Night Shift device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    EndpointAcceptance CriteriaReported Device PerformanceConclusion
    Effectiveness of Night Shift therapy (Primary)65% of PT compliant participants with baseline overall AHI > 10 and non-supine AHI < 20 will demonstrate a clinically important reduction in sleep apnea severity based on a minimum 50% reduction.85.2% (23 out of 27) participants with pre-treatment positional obstructive sleep apnea with a non-supine AHI < 20 achieved a >50% decrease in AHI.Met (85.2% vs. 65% target)
    Safety80% of participants will complete the study with no adverse events resulting in voluntary dropping.100% of compliant subjects successfully completed the study. No adverse events were reported.Met (100% vs. 80% target)
    Accuracy of Supine Position MeasurementComputation of percent time supine by Night Shift is within +/- 5% of the percent time supine by video recordings plus chest sensor in 73% of subjects.Night Shift was within 5% of chest/video supine time in 92% of the studies.Met (92% vs. 73% target)
    Treatment ComplianceAt least 80% of participants will be compliant (use Night Shift for a minimum of 5.5 hours/night or length of their time in bed, five nights/week).100% of the participants wore the Night Shift for a minimum of 20 nights across the 28 nights of intended use.Met (100% vs. 80% target)
    Reduction in Supine TimeAt least 70% of participants will average less than 15% time supine across the four weeks of home use.97% of the participants averaged less than 15% of time in bed in the supine position when therapy was delivered.Met (97% vs. 70% target)
    Improved Epworth Sleepiness Score (ESS)50% of PT compliant participants will show an improved ESS of ≥ 2.50% of participants exhibited an improvement of 2 or more, and 50% showed no change. None of the ESS scores worsened by 2 or more.Met (50% vs. 50% target, with no worsening)
    Improved Functional Outcomes of Sleep Questionnaire (FOSQ) totalFOSQ total will improve by ≥ 2 points in at least 50% of subjects.57% exhibited an improvement of 2 or more, 23% showed no change, and 20% showed a worsening of 2 or more.Met (57% vs. 50% target)
    Mean Sensitivity (sleep) and Specificity (wake) for Night ShiftThe mean sensitivity (sleep) and specificity (wake) for Night Shift will be 0.85 and 0.50, respectively.The endpoint was met based on the sensitivity and specificity of 90% and 58% across 65 studies.Met (90% sensitivity vs. 0.85, 58% specificity vs. 0.50)
    Night Shift Total Sleep Time (TST) within predicate range73% of subjects will be within the range of the predicate when subtracting PSG Total Sleep Time (TST) from Night Shift TST (i.e., range 151 and -129 minutes).99% of the studies had TST derived from Night Shift within the maximum error (based on two standard deviations of the TST error for the predicate device) vs. PSG TST.Met (99% vs. 73% target)
    Night Shift Sleep Efficiency (SE) within predicate range73% of subjects will be within the range of the predicate when subtracting PSG Sleep Efficiency (SE) from Night Shift SE (i.e., range 19.1 and -17.2%).92% of studies had SE values derived from Night Shift within the maximum error (based on two standard deviations of the SE error for the predicate device) vs. PSG SE. 80% of subjects had sleep onset values <15-minutes. 82% of subjects had wake after sleep onset (WASO) values <45 minutes.Met (92% vs. 73% target)
    No consistent patterns of increased N1 and cortical arousals or decreased N3 and REM.There are no consistent patterns of increased N1 and cortical arousals or decreased N3 and REM.87% showed a decrease in N1, 80% a decrease in cortical arousals, 17% an increase in N3, and 33% an increase in REM sleep. Only 3% of subjects showed increase in N1, 7% an increase in cortical arousals, 13% a decrease in N3, and 17% a decrease in REM sleep.Met
    Snoring > 50 dB to identify AHI ≥ 10 (sensitivity > 0.80, specificity > 0.65)The percent time snoring > 50 dB can be used to identify patients with an AHI ≥ 10 with a sensitivity > 0.80 and a specificity > 0.65.When the percentage of time snoring above 50 dB exceeds 10% of sleep time, the sensitivity was 0.85 and the specificity exceeded 0.58.Not Met (Specificity of 0.58 did not meet >0.65 target)
    Identification of treatment success/failure based on AHI, ESS, PHQ9, ISI, GAD7, FOSQThose successfully or unsuccessfully treated with Night Shift can be identified via a combination of changes in the AHI, daytime drowsiness (ESS), depression (PHQ9), Insomnia (ISI), anxiety (GAD7) and quality of life (FOSQ).Evaluating trends across these measures, 50% of subjects showed a substantial improvement as a result of Night Shift therapy and an additional 10% showed improvement, and 33% showed no change. None showed a worsening and two cases (7%) showed substantial overall worsening of subjective measures.Met (with caveat) - "numbers of failures were too few to characterize"

    Study Details for Clinical Tests:

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size for Primary Effectiveness Endpoint: 27 patients with pre-treatment positional obstructive sleep apnea with a non-supine AHI < 20 were included in the analysis for the primary effectiveness endpoint.
      • Test Set Sample Size for other Endpoints: 30 subjects (the 27 patients plus an additional 3 subjects who had a pre-study non-supine AHI >20).
      • Data Provenance: Not explicitly stated, but the study was a clinical study conducted to evaluate safety and efficacy, implying prospective data collection. The location (country of origin) is not mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the given text. Ground truth for sleep studies typically involves highly trained sleep technologists and physicians interpreting polysomnography (PSG) data. However, the document does not specify the number or qualifications for this particular study.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided in the given text.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study evaluated the device's performance in treating sleep apnea and recording sleep parameters, not how human readers improve with or without AI assistance in interpreting diagnostic data from the device. The Night Shift is a therapeutic and monitoring device, not an AI diagnostic interpretation tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, the study primarily assessed the standalone performance of the Night Shift device. While it is a prescription device, its effectiveness was measured by its ability to reduce supine sleep and associated AHI, as well as its accuracy in measuring sleep parameters (position, TST, SE) independently. Human interaction is primarily for setup, compliance, and physician review of the generated reports, but the core therapeutic and monitoring function is standalone.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for sleep parameters (AHI, supine time, TST, SE) appears to be Polysomnography (PSG), a gold standard for sleep disorder diagnosis. For the "Accuracy of Supine Position Measurement" endpoint, the ground truth was "video recordings plus chest sensor." For subjective measures (ESS, FOSQ), the ground truth was the participant's self-reported scores.
    7. The sample size for the training set:

      • This information is not provided in the given text. The document describes a clinical validation study, not the development or training phase of an algorithm.
    8. How the ground truth for the training set was established:

      • This information is not provided as the training set details are not mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130007

    Validate with FDA (Live)

    Device Name
    SLEEP PROFILER
    Date Cleared
    2013-04-17

    (105 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K112514

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sleep Profiler is intended for the diagnostic evaluation by a physician to assess sleep quality in adults only. The Sleep Profiler is a software-only device to be used under the supervision of a clinician to analyze physiological signals and automatically score sleep study results, including the staging of sleep, detection of arousals and snoring.

    Device Description

    The Sleep Profiler is a software application that analyzes previously recorded physiological signals obtained during sleep. The Sleep Profiler software can analyze any EDF files meeting defined specifications, including signals acquired with the Advanced Brain Monitoring X4 System. Automated algorithms are applied to the raw signals in order to derive additional signals and interpret the raw and derived signal information. The software automates recognition of: a) sleep stage, b) snoring frequency and severity, c) pulse rate, d) cortical (EEG), sympathetic (pulse) and behavioral (actigraphy and snoring) arousals. A single channel of electrocardiography, electrooculargraphy, electromyography, or electroencephalography can be optionally presented for visual inspection and interpretation. The software identifies and rejects periods with poor electroencephalography signal quality. The full disclosure recording of derived signals and automated analyses can be visually inspected and edited prior to the results being integrated into a sleep study report. Medical and history information can be input from a questionnaire. Responses are analyzed to provide a pre-test probability of Obstructive Sleep Apnea (OSA) (a condition that cannot be diagnosed with Sleep Profiler) so an appropriate referral to a sleep physician is made. The automated analyses of physiological data are integrated with the questionnaire data, medical and history information to provide a comprehensive report. Several report formats are available depending on whether the user has acquired more than one night of data, wishes to obtain a narrative summary report or provide patient reports. The capability to enter or edit patient information, call the application to generate a study report, and/or download a report is provided using either the desktop PC application or in a web-based module which emulates the desktop functionality. The same analysis and report generation software is used for both the desktop and web-portal applications.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Advanced Brain Monitoring, Inc. Sleep Profiler (K130007). This submission is for modifications to an already cleared device (K120450) to introduce a web-based module for patient information entry and report generation/download. The core analysis and report generation software remains unchanged from the predicate device.

    Therefore, the acceptance criteria and study information provided in this document focus exclusively on the non-clinical testing performed to demonstrate that the new web-based module functions equivalently to the desktop application. There is no clinical study described here that would involve human readers, ground truth establishment through expert consensus or pathology, or outcome data for performance metrics like sensitivity, specificity, etc.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Key Metric for Software Verification)Reported Device Performance
    Confirmation of identical performance using either the desktop or portal for key functions:The results of the verification and validation activities demonstrate that the software meets requirements for safety, function, and intended use, including:
    - Enter questionnaire responses- Performance is identical for entering questionnaire responses via desktop or portal.
    - Edit study data- Performance is identical for editing study data via desktop or portal.
    - Initiate generation of a study report- Performance is identical for initiating generation of a study reports via desktop or portal.
    - Download a study report- Performance is identical for downloading study reports via desktop or portal.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The verification and validation activities tested the functionality of the web-based module against the desktop application, indicating a comparative test of functionalities. It does not refer to a "test set" in the context of clinical data.
    • Data Provenance: Not applicable in the context of clinical data, as this was a non-clinical software verification study. The tests would have involved functional verification of the software.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. This submission focuses on software functionality verification, not clinical performance requiring expert ground truth.

    4. Adjudication Method for the Test Set

    • Not applicable. This was a non-clinical software verification, not a clinical study requiring adjudication of expert interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study was not conducted or described in this document. The submission explicitly states that "The modifications ... did not require clinical studies to support substantial equivalence."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, indirectly. The Sleep Profiler is described as a "software-only device" that "automates recognition of: a) sleep stage, b) snoring frequency and severity, c) pulse rate, d) cortical (EEG), sympathetic (pulse) and behavioral (actigraphy and snoring) arousals." The verification discussed here specifically confirms that the web-based module performs these functions identically to the pre-existing desktop version, which does perform these analyses automatically without continuous human intervention during the analysis phase. However, it's intended to be "used under the supervision of a clinician" and the reporting can be "visually inspected and edited prior to the results being integrated into a sleep study report," implying a human-in-the-loop for final interpretation. The performance of the automated algorithm itself was established in the predicate device (K120450) and is not being re-evaluated for K130007.

    7. The Type of Ground Truth Used

    • Not applicable for clinical performance. For the software verification described, the "ground truth" was the expected functional output and behavior of the established desktop application (K120450). The web-based module was verified to produce identical results.

    8. The Sample Size for the Training Set

    • Not provided/not applicable. This submission does not discuss algorithmic development or training sets. It is a modification to an already existing software application. The predicate device (K120450) would have had this information.

    9. How the Ground Truth for the Training Set Was Established

    • Not provided/not applicable. As above, this information relates to the original algorithmic development, not the current submission for a web-based module.