Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K193339

    Validate with FDA (Live)

    Date Cleared
    2020-01-30

    (59 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K111955

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GTK Veress Needles are intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum prior to the placement of trocars during laparoscopic procedures.

    Device Description

    The structure of the GTK Veress Needles contain inner needle, outer needle, warning block, spring, handle and valve. It contains a spring-loaded blunt stylet mechanism, which is also the major principle mechanism of GTK Veress Needles to achieve the intended use. It is used to establish pneumoperitoneum prior to trocar and cannula insertion in laparoscopic procedures. The veress needle is connected to the insufflators with a luer connector. There are two models of veress needles: VN-120 and VN-150, with different lengths of needle bodies (120mm and 150mm). The GTK Veress Needles are for single use and are supplied sterile. The contact time is no more than 24 hours. The product is sterilized by Ethylene oxide. The manufacture date and expiry date are shown on the labeling. The GTK Veress Needles are only intended for use in a hospital setting/environment.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the GTK Veress Needles, a medical device used in laparoscopic procedures. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than detailing an AI/ML-based device approval. As such, information regarding AI/ML-specific acceptance criteria, ground truth establishment, training/test set sample sizes, and human reader studies (MRMC) is not present.

    However, I can extract the relevant acceptance criteria and the study that proves the device meets the acceptance criteria for the GTK Veress Needles as a hardware medical device, based on the provided document.


    Acceptance Criteria and Device Performance for GTK Veress Needles

    The study demonstrating the device meets acceptance criteria is a series of non-clinical performance tests. The acceptance criteria themselves are implied by the successful completion and meeting of standards for these tests. All listed tests "meet the acceptance criteria," indicating satisfactory performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/StandardReported Device PerformanceComments on Meeting Criteria
    BiocompatibilityISO 10993-5: 2009 (In Vitro Cytotoxicity)ConductedMet acceptance criteria (implied by "The following tests were conducted" and absence of failure)
    ISO 10993-10: 2010 (Skin Sensitization, Intracutaneous Reactivity)ConductedMet acceptance criteria (implied by "The following tests were conducted" and absence of failure)
    ISO 10993-11: 2006 (Acute Systemic Toxicity, Pyrogen test)ConductedMet acceptance criteria (implied by "The following tests were conducted" and absence of failure)
    USP 39 <85> (Bacterial Endotoxin test)ConductedMet acceptance criteria (implied by "The following tests were conducted" and absence of failure)
    Sterilization ValidationISO 11135:2014 (EO Sterilization)ConductedMet acceptance criteria (implied by "The sterilization validation was conducted according to ISO 11135:2014." and absence of failure)
    ISO 10993-7:2008 (EO and ECH Residual)EvaluatedMet acceptance criteria (implied by "The EO and ECH residual...was evaluated by ISO 10993-7:2008." and absence of failure)
    Shelf Life ValidationASTM F 1980-07 (2011) (Accelerated Aging for 3 years)ConductedAll results "meet the acceptance criteria."
    ISO 11737-2:2009 (Sterility test after aging)ConductedAll results "meet the acceptance criteria."
    ASTM D 3078-02:2013 (Vacuum leak test after aging)ConductedAll results "meet the acceptance criteria."
    ASTM F 88 (Seal strength test after aging)ConductedAll results "meet the acceptance criteria."
    ASTM F 1929-12 (Leakage (dye penetration) test after aging)ConductedAll results "meet the acceptance criteria."
    Din 58953-6:2010 (Microbial barrier properties after aging)ConductedAll results "meet the acceptance criteria."
    Physical performance after natural agingConductedAll results "meet the acceptance criteria."
    Performance Test - BenchAppearanceConductedAll results "meet the acceptance criteria."
    SizeConductedAll results "meet the acceptance criteria."
    Tip Pull TestConductedAll results "meet the acceptance criteria."
    Switch Operation TestConductedAll results "meet the acceptance criteria."
    Spring Obturator OperationConductedAll results "meet the acceptance criteria."
    Needle Puncture Force TestConductedAll results "meet the acceptance criteria."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the sample size (number of devices) used for each non-clinical test. It only states that the tests were "conducted" or "evaluated."
    • Data Provenance: The tests are non-clinical, meaning they were likely performed in a laboratory setting by the manufacturer, Guangzhou T.K Medical Instrument Co., Ltd., in China. The data would be internally generated prospective test data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable to this type of device submission. Ground truth in this context refers to established engineering and medical device standards (e.g., ISO, ASTM, USP) for performance, sterility, and biocompatibility. There are no human "experts" establishing a "ground truth" in the way one would for an AI/ML-based diagnostic device (e.g., radiologists annotating images).

    4. Adjudication Method for the Test Set

    Not applicable. The tests are laboratory-based and measured against predefined quantitative or qualitative standards. There is no human subjective assessment requiring adjudication in the context of this submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for diagnostic devices, especially those incorporating AI, to assess the impact of the technology on human reader performance. The GTK Veress Needles are a surgical instrument, not a diagnostic device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm. The device performance is assessed through its physical, material, and functional characteristics.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance tests for this medical device is based on established international and national standards (e.g., ISO, ASTM, USP) and predefined engineering specifications for acceptable performance (e.g., appearance, size, force, leak resistance). The successful completion of these tests against these benchmarks constitutes the "proof" that the device meets its acceptance criteria for substantial equivalence to a predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires training data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1