Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120350

    Validate with FDA (Live)

    Date Cleared
    2012-11-01

    (269 days)

    Product Code
    Regulation Number
    870.5310
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To power AEDs for which the OEM battery pack was intended. Only qualified service personnel should evaluate, test, or install AED battery packs.

    To provide power for the specific Automated External Defibrillators (AEDs) for which the original battery packs were designed and manufactured, in back-up and portable use.

    Device Description

    The Green4Life Batteries are non-rechargeable battery packs utilized as the primary power source or as a standby or backup power source for Automated External Defibrillators (AEDs). These devices provide a means of supplying electrical power through the chemical reaction of lithium cells. The energy provided depends upon the voltage and capacity rating of a particular pack and the amount of current required by the device into which they are installed. The performance and life span of these batteries depends on operating conditions of temperature, current drain, and discharge method. The goal of these battery packs is to provide a sufficient power source for AEDs for as long as possible under a specified range of environmental conditions.

    The Green4Life Batteries are re-celled battery packs of those identified below in Predicate Devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device Model (Green4Life Battery)Acceptance Criteria: Open Circuit Minimum Voltage @ 20°CReported Device Performance (Implied)Acceptance Criteria: Loaded Circuit Minimum Voltage @ 20°CReported Device Performance (Implied)
    Cardiac Science 9141 (Vnom=6V)6.0VMet5.5V (L=8 ohms)Met
    Cardiac Science 9141 (Vnom=12V)12.0VMet11.0V (L=15 ohms)Met
    Cardiac Science 914612.0VMet11.0V (L=15 ohms)Met
    Philips BT118.0VMet16.6V (L=23 ohms)Met
    Philips M5070 (FRX/Onsite)9.0VMet8.2V (L=11 ohms)Met
    Philips M3863 (FR2)12.0VMet11.0V (L=15 ohms)Met
    Philips 989803150161 (FR3)12.0VMet11.0V (L=15 ohms)Met
    Defibtech DCF-200 (Vnom=3V)3.0VMet2.7V (L=4 ohms)Met
    Defibtech DCF-200 (Vnom=15V)15.0VMet13.8V (L=20 ohms)Met
    Physio Control Lifepak 50012.0VMet11.0V (L=15 ohms)Met

    Additional Acceptance Criteria:

    • Ease of Installation: Battery pack must easily install into the AED and firmly lock into place as with the original battery pack.
    • AED Self-Test Result: Upon installation, the AEDs undergo an operational self-test of the installed battery, and the final battery pack acceptance testing requires a reported "pass" by the corresponding AED.

    Reported Device Performance: The document states, "The Green4Life Battery packs are re-celled packs manufactured as cleared devices... Performance testing within the specific AED was conducted to verify that the replacement battery pack tested within limits and that reset of battery pack information when operated within the designated AED, for those packs equipped with memory devices, was acceptable." It also states, "Final battery pack acceptance testing requires a reported 'pass' by the corresponding AED." This implies that all stated acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state a specific numerical sample size for the "Performance Testing" or "Manufacturing Tests," but it implies testing each battery pack. It mentions "typical sampling plans for similar devices" for incoming inspection of lithium cells, and "100% testing" for assembled packs. For the final acceptance test, "the battery packs will be installed into their corresponding AED for a final acceptance test." This suggests that a representative number of each re-celled battery model were tested.
    • Data Provenance: The testing was conducted by AED Battery Exchange, LLC. The data provenance is internal to the manufacturer, intended to demonstrate equivalence to predicate devices. It is a form of prospective testing for the manufactured re-celled batteries. The company is based in Wauconda, IL, USA.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • This information is not applicable to this type of device and study. The "ground truth" for battery performance is objective and measurable (voltage, resistance, successful AED operation), not based on expert interpretation like in image analysis or clinical diagnosis.

    4. Adjudication Method for the Test Set

    • This information is not applicable. Adjudication methods like 2+1 or 3+1 are used for subjective interpretations (e.g., medical image reads) where disagreements between experts need resolution. For objective measurements like battery performance, the results are definitive.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, often with and without AI assistance. This submission pertains to a re-celled battery pack.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, in essence, standalone testing was done for the battery's performance. The core tests (open circuit voltage, loaded circuit voltage) are performed directly on the battery packs without human intervention beyond setting up the test equipment and recording results. The "final acceptance test" involves the AED's internal self-test, which is an automated algorithm checking the battery's functionality. This is the "algorithm only" aspect of performance evaluation for this type of device.

    7. The Type of Ground Truth Used

    • The ground truth is based on objective, measurable electrical parameters (voltage under open and loaded conditions) and functional validation (the AED's internal self-test reporting "pass"). This is a form of empirical validation against defined performance specifications.

    8. The Sample Size for the Training Set

    • This information is not applicable as this is not an AI/machine learning device that requires a "training set" in the conventional sense. The "training" for the manufacturing process is adherence to established specifications and procedures for re-celling and testing.

    9. How the Ground Truth for the Training Set was Established

    • This information is not applicable for the same reason as point 8. The "ground truth" for the manufacturing process is the successful replication of original battery performance, which is verified by the performance tests described in the acceptance criteria.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1