Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K132519

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-12-06

    (116 days)

    Product Code
    Regulation Number
    872.4760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K111065

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Customized Mandible Recon Plate Kit is intended to be used for rigid internal fixation of primary and secondary mandibular reconstructions. The Customized Mandible Recon Plate Kit is indicated for use in primary mandibular reconstruction with bone graft, temporary bridging until delaved secondary reconstruction and secondary mandibular reconstruction.

    Device Description

    The Customized Mandible Recon Plate Kit includes customized, patient specific implants, the Customized Mandible Recon Plates. Additionally, the kit includes the Instruction for Use (IFU) and a printed version of the Design Proposal approved by the surgeon prior to plate manufacture. It may include an anatomical model, named Mandible Model. The Customized Mandible Recon Plates (CMRP) and the corresponding Mandible Model are designed and manufactured for one specific patient. The products are ordered by a surgeon on a patient-by-patient basis over an internet platform referred to as the "eRequest Lifecycle online ordering system" which was already cleared with K111065 for ordering patient specific polymer implants to correct trauma and/or defects in mandibular, maxillofacial, or craniofacial bone. Based on patient specific anatomical data (CT-scan) and input from the surgeon, a virtual mandible recon plate with its specific dimensions (profile heights, length and 3D run of the plate, number and position of screw holes and bar strengthening) is created using the Customized Mandible Recon Plate Design Process (CMRP-DP) including the Plate Design Software (PDS). Screenshots of the virtual plate are then presented to the surgeon in the Design Proposal (pdf-file). After the surgeon approves the Design Proposal via the above mentioned eRequest Lifecycle online ordering system, the Customized Mandible Recon Plate, as well as the Mandible Model (if ordered by the surgeon), are manufactured according to the specifically set design requirements. Once the manufacturing process is finalized and the devices are cleaned, they are packed, labeled and shipped to the location specified during the ordering process. The CMRP implant is compatible with the Stryker Universal Mandibular System. Additionally. Stryker offers customized cutting and drill guides for use with the Customized Mandible Recon Plate.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Stryker Customized Mandible Recon Plate Kit:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary explicitly states that: "The Subject Device fulfilled all set acceptance criteria for each category in accordance with either ISO or ASTM specifications, or internally predetermined acceptance criteria if no standards were applicable."

    Unfortunately, the document does not provide specific quantitative acceptance criteria values or detailed quantitative device performance results for each test. It only lists the categories of tests performed and the relevant standards. Therefore, I can only provide a table with the types of tests and the general outcome reported.

    Acceptance Criteria CategoryReported Device PerformanceRelevant Standards (where stated)
    BiocompatibilityMet acceptance criteriaDIN EN ISO 10993 ff as valid 2013 (Part 1)
    Cleaning ValidationMet acceptance criteriaEN ISO 15883-1:2009, DIN EN ISO 17664:2004
    Sterilization (validation)Met acceptance criteriaISO 11138-1:2006, DIN EN ISO 11737-1:2009, DIN EN ISO 11737-2:2009, DIN EN ISO 14161:2011, DIN EN ISO 14937:2009, DIN EN ISO 17665-1:2006, DIN ISO/TS 17665-2:2009
    Multiple ReprocessingMet acceptance criteria(Standards for sterilization and cleaning are implied)
    Residual Moisture after SterilizationMet acceptance criteria(Implied by sterilization standards)
    MRI ConditionalMet acceptance criteria(Standard not explicitly listed, but implied)
    Transportation ValidationMet acceptance criteria(Standard not explicitly listed)
    Clinical TransportationMet acceptance criteria(Standard not explicitly listed)
    Mechanical Strength of PlateMet acceptance criteriaASTM or internally predetermined criteria
    Mechanical Strength of Locking Mechanism between Plate and ScrewMet acceptance criteriaASTM or internally predetermined criteria
    Handling of System (End User & End Product Test)Met acceptance criteria(Standard not explicitly listed, likely internal)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the non-clinical tests. It also does not discuss data provenance in terms of country of origin or whether the data was retrospective or prospective, as these non-clinical tests typically involve laboratory testing of manufactured devices or materials, not patient data in the conventional sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to the non-clinical performance and V&V testing described. "Ground truth" in this context refers to clinical outcomes or diagnoses, which were not part of this submission's evidence for efficacy. For mechanical and biocompatibility testing, the "ground truth" is defined by the standards (ISO, ASTM) themselves or internal specifications.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple human readers assess cases and their disagreements need to be resolved. The tests described are laboratory-based conformance tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No. The document explicitly states: "No clinical testing was performed to support this submission." Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No. This device is a physical medical implant (a bone plate) and associated design/manufacturing process, not an algorithm or AI software for diagnosis or analysis. The "Plate Design Software (PDS)" is part of the design process, but its performance wasn't evaluated as a standalone AI diagnostic tool.

    7. The Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" is the established acceptance limits defined by international standards (ISO, ASTM) or internal predetermined specifications. For example, a biocompatibility test result must fall within the range deemed safe by ISO 10993, and mechanical strength must exceed certain load thresholds.

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical implant and manufacturing process. There is no concept of a "training set" as one would have for a machine learning algorithm. The design process for each plate is customized based on patient-specific CT data and surgeon input, but this is not a learning algorithm that requires a "training set" of cases to develop its core function.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device. The design process (CMRP-DP including PDS) relies on established engineering principles, CAD/CAM software, patient-specific imaging data (CT-scan), and surgeon input, rather than a machine learning model trained on a ground-truthed dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121153

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-14

    (151 days)

    Product Code
    Regulation Number
    882.5320
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K053199, K111065

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PEEK Customized Cranial Implant Kit is intended to be used to replace bony voids in the cranial and the craniofacial skeleton.

    The PEEK Customized Cranial Implant Kit is indicated for filling bony voids in the cranial and craniofacial skeleton in patients 12 years of age and older.

    Device Description

    PEEK Customized Implant Kit consists of the PEEK Customized Cranial Implant or the PEEK Customized Craniofacial Implant, the Host Bone Model, and the Design Proposal.

    PEEK Customized Cranial Implant: The PEEK Customized Cranial Implant is a customized patient-specific implant based on CT-data and input by the surgeon. The implant is fabricated from polyetheretherketone (PEEK) and is intended to be used to fill bony voids in the cranial skeleton. It is delivered non-sterile.

    PEEK Customized Craniofacial Implant: The PEEK Customized Craniofacial Implant is a customized patient-specific implant based on CT-data. The implant is fabricated from polyetheretherketone (PEEK) and is intended to be used to fill bony voids in the craniofacial region (orbital rim, zygoma, and adjacent bone). The implant matches the shape and dimensions of the missing bone fragments. It is delivered non-sterile.

    The host bone model is provided as a preoperative guide to demonstrate orientation and fit of the Peek Customized Cranial Implant. The Design Proposal is a presentation of virtual 3-dimensional models of the implant design. The PEEK Customized Cranial Implant is offered in different sizes based on the size of the cranial defect. Depending on the surgeon's preference, the PEEK Customized Cranial Implant may be constructed in varying thicknesses, wall designs, number of dura suture holes and dura suture hole diameters. The PEEK Customized Cranial Implant is fixated to the native bone with Stryker Neuro, Midface or Upperface self-tapping screws. The PEEK Customized Cranial and Craniofacial Implant Kit is bundled with an online ordering system called "eRequest Lifecycle", whereby the user can initiate a case request, upload the patient specific image data, download the Design Proposal and approve the implant design.

    AI/ML Overview

    This document describes the PEEK Customized Cranial Implant Kit, a custom patient-specific implant intended to replace bony voids in the cranial and craniofacial skeleton. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving independent performance against specific acceptance criteria. Therefore, several requested details about acceptance criteria, study design, and ground truth are not explicitly provided in the document as it's a 510(k) submission for a medical device, not an AI/ML algorithm.

    However, based on the provided text, the following information can be extracted or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) submission for a physical medical implant, the "acceptance criteria" are primarily related to mechanical properties, material compatibility, and fit, rather than performance metrics like sensitivity or specificity typically seen in AI/ML performance studies. The "reported device performance" refers to the results of various physical tests.

    Acceptance Criteria (Inferred from tests performed)Reported Device Performance
    Mechanical Strength: Comparable to native skull.Demonstrated mechanical strength comparable to native skull.
    Integrity during surgical manipulation: Not melt or lose integrity upon burring, drilling, and/or screw insertion with recommended tools.Did not melt or lose integrity upon burring, drilling, and/or screw insertion.
    Fit and User Needs: Implants meet user needs and fit defects.Met user needs and fit the defects (handling test in Cadaver Lab with surgeons).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions a "Cadaver Lab" for handling tests and various "performance tests" (stability, trimming, drilling, screw insertion, screw pull-out). However, the specific sample sizes for these tests are not provided. Similarly, the data provenance (e.g., country of origin, retrospective or prospective) for these physical tests is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    For the "handling test performed in a Cadaver Lab," it states that "surgeons demonstrated that the implants met the user needs and fit the defects."

    • Number of experts: Not explicitly stated, but implies multiple surgeons were involved.
    • Qualifications of experts: Identified as "surgeons." No further specific qualifications (e.g., years of experience, subspecialty) are provided.

    4. Adjudication Method for the Test Set:

    Given the nature of the tests, a formalized adjudication method like 2+1 or 3+1 is not applicable. The "handling test" involved surgeons demonstrating fit and meeting user needs, implying a qualitative assessment of suitability. The mechanical tests would have objective pass/fail criteria based on quantitative measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices or AI algorithms where human reader performance is a key metric. This submission is for a physical implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, in the context of the device's design process. The document describes a "Virtual Implant Design Process (VIDP)" using CT-data to generate a 3D model and design the implant. This design process, which is an algorithm-driven component, is described as "validated." The performance data for the physical implant itself (mechanical strength, integrity) would be considered "standalone" as it's testing the final product independent of the human intervention during design approval.

    7. The Type of Ground Truth Used:

    • For mechanical tests (stability, trimming, drilling, screw insertion, screw pull-out): The ground truth would be based on engineering specifications and established mechanical properties of native bone (implicitly, as the device aims for "comparable to native skull" strength).
    • For the handling test in the Cadaver Lab: The ground truth was based on expert consensus/qualitative assessment from surgeons regarding fit and user needs.

    8. The Sample Size for the Training Set:

    This question primarily applies to AI/ML algorithms that learn from data. While the device design uses patient CT data, the document does not describe it as a "training set" in the machine learning sense. The "Virtual Implant Design Process (VIDP)" is described as "validated," implying that the rules, algorithms, and processes for custom implant design were developed and verified, but it doesn't refer to a statistical "training set" of patient cases for an AI model to learn from.

    9. How the Ground Truth for the Training Set Was Established:

    Again, this question is not directly applicable in the AI/ML sense to this 510(k) submission. The "ground truth" for the "Virtual Implant Design Process" would refer to the accuracy and validity of the 3D modeling and design algorithms in replicating the desired anatomical structures and ensuring proper fit based on CT data. This would be established through engineering validation, potentially against known anatomical models or validated software outputs, during the development of the VIDP itself. The document states a "validated 'Virtual Implant Design Process' (VIDP)" is used, but doesn't elaborate on the validation specifics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1