Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K173703

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-15

    (254 days)

    Product Code
    Regulation Number
    884.4160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K121858, K061153, K110020

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sonata® Sonography-Guided Transcervical Fibroid Ablation System is intended for diagnostic intrauterine imaging and transcervical treatment of symptomatic uterine fibroids, including those associated with heavy menstrual bleeding.

    Device Description

    The Sonata System provides radiofrequency (RF) ablation of uterine fibroids using a transcervical approach that is uterine sparing, without incisions or material uterine distension. The system enables a clinician to deliver radiofrequency energy to fibroid tissue resulting in thermal fixation and coagulative necrosis of the tissue. The system combines two technologies - ultrasound for visualization, and radiofrequency energy for ablative therapy - in a single integrated handpiece. The Sonata System is comprised of medical equipment, software, and various single-use and reusable instruments.

    AI/ML Overview

    The provided text describes the acceptance criteria and the clinical study results for the Gynesonics Sonata® Sonography-Guided Transcervical Fibroid Ablation System.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Sonata System)
    Co-primary Efficacy Endpoint: Reduction in Menstrual Blood Loss (MBL) at 12 months Success criterion: Lower Confidence Limit (LCL) ≥ 45% (referring to ≥ 50% Reduction in PBAC and PBAC <250)64.8% of subjects achieved ≥ 50% reduction in PBAC and PBAC <250 at 12-Month Visit. 95% CI (LCL, UCL): 56.3%, 72.6%. Outcome: Met Endpoint Success Criteria. (LCL of 56.3% is ≥ 45%)
    Co-primary Efficacy Endpoint: Surgical Reintervention through 12 Months Success criterion: LCL of the percentage of subject success ≥ 75% (meaning rate of no surgical re-intervention)99.3% Rate of No Surgical Re-intervention through 12 Months. 95% CI (LCL, UCL): 95.1%, 99.9%. Outcome: Met Endpoint Success Criteria. (LCL of 95.1% is ≥ 75%)
    Safety Objective: Absence of serious adverse device effect (SADE), adverse device effect (ADE), or unanticipated adverse device effect (UADE).There was no occurrence of SADE, ADE, or UADE. Two procedure-related serious adverse events (SAEs) in two subjects (one deep venous thrombus, one leukorrhea/pelvic pain managed with antibiotics).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 147 subjects were enrolled, with 142 subjects evaluated for the MBL endpoint and 143 subjects evaluated for the surgical re-intervention endpoint.
    • Data Provenance: The data was obtained from a prospective, longitudinal, multicenter, single-arm cohort study named the SONATA Pivotal IDE Clinical Trial. Subjects were enrolled in clinics in the United States and Mexico.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number or qualifications of experts used to establish ground truth for the clinical trial endpoints. For outcomes such as Pictorial Blood Loss Assessment Chart (PBAC) scores, these are typically self-reported by patients or assessed by study coordinators based on patient reporting. Surgical re-intervention is a factual event, often confirmed by medical records. However, the study involved multiple investigational sites (27) with trained personnel who would have been responsible for data collection and assessment as per the study protocol.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method like 2+1 or 3+1 for the clinical trial endpoints. However, it indicates that the trial was a "single-arm cohort study with each subject serving as her own control," implying comparison of pre- and post-treatment metrics within the same subject. The endpoints were assessed at 12 months post-procedure, and data collection continued for 3 years for select measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned. The clinical study was a single-arm trial evaluating the device's performance against pre-defined success criteria.

    6. Standalone (Algorithm Only) Performance Study

    While the software component (SMART Guide) underwent extensive bench testing including volumetric ablation registration, development and confirmation of ablation dimensions, and effect of parameter variations on ablation dimensions, this was not presented as a standalone clinical performance study but rather as verification and validation of the software's ability to accurately guide the ablation. The study focused on the performance of the integrated Sonata System with human users.

    7. Type of Ground Truth Used

    The ground truth for the clinical study was based on:

    • Patient-reported outcomes: Menstrual Blood Loss (MBL) was quantified using the Pictorial Blood Loss Assessment Chart (PBAC) score. Subject satisfaction, symptom improvement, and likelihood to recommend were also patient-reported outcomes.
    • Observed clinical outcomes: Surgical re-intervention rates were direct clinical events.
    • Objective measurements: Reduction in fibroid volumes (measured at 3 months post-procedure in the FAST-EU trial).
    • Adverse events: Clinically observed and reported events.

    8. Sample Size for the Training Set

    The document specifies the sample size for the pivotal clinical trial (test set) as 147 subjects. It does not explicitly state a separate sample size for a training set for the software or algorithm. The software (SMART Guide) development involved "ex vivo testing with non-perfused bovine tissue model" and "in vivo peri- and pre-Hysterectomy studies data" to set the final SMART Guide dimensions, but a specific "training set sample size" for an AI/ML algorithm in the conventional sense is not provided.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, a specific "training set" for an AI/ML algorithm is not explicitly detailed. However, the "ground truth" for the development of the SMART Guide, which is a key software component providing graphical guidance, was established through:

    • Ex vivo testing: Using bovine tissue models to measure volumetric registration offset and develop the initial dimensions of the Ablation Zone (AZ) and Thermal Safety Border (TSB).
    • In vivo studies: Peri- and pre-Hysterectomy studies provided data used to establish the final dimensions of the AZ and TSB.
    • Computational modeling: Finite Element Method (FEM) was used to determine the sensitivity of ablation dimensions to variations in tissue parameters.

    These methods were used to define and validate the graphical representations (AZ and TSB) within the SMART Guide, ensuring they accurately predict the ablation effect.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140524

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-03-21

    (17 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131209, K112953, K110020

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Teratech Corporation Terason™ uSmart3200T (also known as the BenQ UP200) is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Intra-Operative (abdominal, organs and vascular), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.

    Device Description

    The Terason uSmart3200T ultrasound system is a portable tablet style full feature, general purpose, and diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. The Terason uSmart3200T Ultrasound System is equivalent to the previously cleared version of the uSmart3200T Ultrasound Systems. The modification includes the addition of 3 transducers (15L4, 16HL7, and 8EC4A) with the no change to tablet style computer form factor.

    The Terason™ uSmart3200T ultrasound system was the previously cleared on the date of May 28, 2013 as described in the 510(k) submission (K131209). This system contains a proprietary ultrasound engine for controlling the acoustic output of the transducer and processing the return echoes in real time. This data are then transferred to the tablet computer over a FireWire (aka IEEE 1394) connection for further processing and generation and display of the ultrasound image.

    The Terason™ uSmart3200T ultrasound tablet weighs 4.9 pounds (2.21 Kg) and has an 11.5" backlit touch screen. The tablet dimensions (8.82"(H) x 12.64"(W) x 1.25"(D)) are chosen to allow portability. A Lithium-Polymer battery (integrated into the tablet) provides 2 hours of continuous ultrasound scanning. The tablet includes a docking station (for charging) that uses a medicalgrade power supply. The ultrasound transducer connector is identical to that used in the Terason™ predicate device, the uSmart3200T. Optional accessories include a cart and printer.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Terason uSmart3200T and BenQ UP200 Ultrasound System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets specific acceptance criteria through a clinical study. Therefore, the document does not contain the detailed information requested in terms of acceptance criteria, specific study designs (like MRMC or standalone performance), sample sizes, expert ground truth establishment, or training set details.

    The submission primarily lists the device's technical specifications and how they compare to predicate devices. The "Non Clinical Tests" section refers to compliance with various electrical safety, usability, and acoustic output standards, which serve as the "acceptance criteria" for regulatory clearance in this context. However, these are performance standards, not clinical performance metrics.

    Here's a summary of the available information based on your request, highlighting what is not present in this type of regulatory document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally compliance with international and national standards for medical devices, particularly regarding safety and technological characteristics. The "reported device performance" is demonstrated through this compliance and the assertion of substantial equivalence to predicate devices which have already met such criteria.

    Acceptance Criteria (Standards Met)Reported Device Performance (Compliance)
    IEC 60601-1: General Requirements for SafetyCompliance demonstrated via Intertek Test Record Number 100825075BOX-001
    IEC 62366 & IEC60601-1-6: Usability engineeringCompliance demonstrated via Intertek Projects 100825075BOX-004A & 100825075BOX-003A
    IEC 60601-1-2: Electromagnetic Compatibility (EMC)Compliance demonstrated via Test Record Number 100933162BOX-017 and Radiated Emissions Reports (101224397BOX-001, 101188849BOX-001a)
    CISPR11 Class B: Radiated EmissionsCompliance demonstrated via Radiated Emissions Reports (101224397BOX-001, 101188849BOX-001a)
    IEC 60601-2-37 / EN60601-2-37: Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment (for transducers)Compliance demonstrated via Intertek Reports for each transducer (100404852BOX-002, 100404852BOX-004, 100404852BOX-003)
    NEMA UD 3: Acoustic Output DisplayCompliance mentioned
    Biocompatibility Tests (ISO 10993 Part 5 and Part 10)Compliance demonstrated via biocompatibility reports for new transducers
    AAMI TIR No. 12: Designing, Testing and Labeling Reusable Medical DevicesCompliance mentioned
    GMPs & ISO-13485 quality systemsCompliance asserted
    Track 3 devices, Indications for Use (for predicate devices)Similar indications for use and technological characteristics to predicate devices (K131209, K112953, K110020)

    2. Sample size used for the test set and the data provenance

    This document does not describe a "test set" in the context of clinical performance evaluation with patient data. The tests performed are non-clinical compliance tests against established safety and electrical standards. Therefore, information on sample size and data provenance (country of origin, retrospective/prospective) for a clinical test set is not applicable or present.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document does not pertain to a clinical study requiring expert ground truth for interpretation.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a 510(k) submission for a diagnostic ultrasound system and transducers, which does not involve AI or comparative effectiveness studies with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this device does not involve a standalone algorithm for performance evaluation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no clinical ground truth is established or used in this submission. The "ground truth" here is compliance with engineering and safety standards.

    8. The sample size for the training set

    Not applicable, as this document does not describe a machine learning model or training set.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as above.


    Summary of Approach:

    This 510(k) submission demonstrates substantial equivalence primarily by:

    • Listing the device's technical characteristics and comparing them directly to previously cleared predicate devices (Terason uSmart3200T (K131209), Terason™ t3000 (K112953), Terason™ T3200 (K110020)).
    • Highlighting that the modifications (addition of three transducers and software updates) do not alter the fundamental technological characteristics or intended use.
    • Providing evidence of compliance with a comprehensive set of non-clinical safety, electrical, usability, and acoustic output standards.

    The "study" that proves the device meets the "acceptance criteria" here consists of these non-clinical tests and direct comparisons to predicate devices, rather than clinical performance studies with patient data. The regulatory body (FDA) reviews these submissions to ensure the new device is as safe and effective as existing legally marketed devices, without requiring new clinical efficacy data if substantial equivalence is adequately demonstrated.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131209

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-24

    (25 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K110020, K112953

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Teratech Corporation Terason™ uSmart3200T is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.

    Device Description

    The Terason™ uSmarl3200T ultrasound system is equivalent to the previously cleared version of the t3200 and t3000 Ultrasound Systems described in the following 510(k) submissions (K110020 and K112953). This system contains a proprietary ultrasound engine for controlling the acoustic output of the transducer and processing the return echoes in real time. These data are then transferred to the tablet (previously laptop computer) over a FireWire (aka IEEE 1394) connection for further processing and generation and display of the ultrasound image.

    The Terason™ uSmart3200T ultrasound tablet weighs 4.9 pounds (2.21 Kg) and has an 11,5" backlit touch screen. The tablet dimensions (8,82"(H) x 12,64"(W) x 1,25"(D)) are chosen to allow portability. A Lithium-Polymer battery (integrated into the tablet) provides 2 hours of continuous ultrasound scanning. The tablet includes a docking station (for charging) that uses a medical-grade power supply. The ultrasound transducer connector is identical to that used in the Terason™ predicate device, the t3200. Optional accessories include a cart and printer.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Terason™ uSmart3200T Ultrasound System, detailing its substantial equivalence to previously cleared devices. It focuses on regulatory compliance, safety, and performance characteristics rather than clinical study results establishing diagnostic accuracy against a ground truth.

    Therefore, the following information cannot be extracted from the provided text:

    • A table of acceptance criteria and reported device performance (in terms of diagnostic accuracy metrics). The document refers to compliance with safety and performance standards (e.g., IEC 60601-1) but not clinical performance acceptance criteria like sensitivity, specificity, or AUC against a ground truth.
    • Sample size used for the test set or its provenance.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for the test set.
    • Whether a multi-reader, multi-case (MRMC) comparative effectiveness study was done or its effect size.
    • Whether a standalone (algorithm only) performance study was done.
    • The type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    However, based on the document, I can provide the following relevant information regarding the device's acceptance criteria and the study that proves its compliance with regulatory standards:

    The device's acceptance criteria are primarily defined by compliance with established international and national standards for medical electrical equipment, including safety, usability, electromagnetic compatibility, and acoustic output. The "study" proving the device meets these criteria consists of specific non-clinical tests and reports generated by accredited testing laboratories (Intertek).

    1. A table of acceptance criteria and the reported device performance (focused on technical compliance):

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance/Compliance
    General SafetyIEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for SafetyIntertek Test Record Number 100825075BOX-001 (indicates compliance)
    UsabilityIEC 62366, Medical Devices: Application of usability engineering to medical devicesIntertek Project: 100825075BOX-004 (indicates compliance)
    IEC 60601-1-6, Medical Electrical Equipment Part 1-6: General requirements for safety- Collateral standard: UsabilityIntertek Project: 100825075BOX-003 (indicates compliance)
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2; General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and testsIntertek Test Record Number, 100933162BOX-017 (indicates compliance)
    Specific Safety (Ultrasound)IEC 60601-2-37 / EN60601-2-37 Medical Electrical Equipment Part 2: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipmentTransducer Model 5C2A: Intertek Report 100825075BOX-006 (compliance) Transducer Model 12L5A: Intertek Report 100825075BOX-007 (compliance) Transducer Model 4V2A: Intertek Report 100825075BOX-005 (compliance)
    Acoustic Output DisplayNEMA UD 3 Acoustic Output DisplayTerason uSmart3200T Ultrasound System User Guide (16-3301) (indicates adherence to display requirements)
    Biocompatibility (Transducers)ISO 10993 Part 5 and Part 10Biocompatibility reports for all transducers (indicates compliance)
    Substantial Equivalence (General)Comparison to predicate devices: Terason™ t3000 (K112953) and Terason™ t3200 (K110020)The FDA's 510(k) clearance letter (K131209) states the device is "substantially equivalent" to predicate devices for the stated indications for use.
    Indications for UseGeneral purpose ultrasound imaging and fluid flow analysis for specific clinical applicationsThe device supports Fetal, Abdominal, Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.

    Important Note: This document describes a 510(k) submission, which primarily aims to demonstrate substantial equivalence to a previously cleared device. It does not typically involve new clinical efficacy studies with ground truth establishment in the way that, for example, a novel diagnostic AI algorithm would. The "performance" here refers to meeting established safety and performance standards for ultrasound devices, and demonstrating that changes from predicate devices (e.g., changing from laptop to tablet) do not raise new safety or effectiveness concerns.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1