Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K221436

    Validate with FDA (Live)

    Date Cleared
    2023-06-29

    (408 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212031, K103767, K173770

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The breathing set is an accessory to the F&P 950 Respiratory Humidifier. It is intended to provide heated and humidified CPAP to spontaneously breathing premature and full term neonates and infant patients (up to a weight of 10 kg) who require breathing support, due to conditions associated with prematurity (such as Respiratory Distress Syndrome) or other conditions where CPAP is required or desired. It is intended for use within a hospital clinical environment such as a NICU (Neonatal Intensive Care Unit) and PICU (Pediatric Intensive Care Unit), within the limits of its stated technical specifications.

    Device Description

    The subject devices of this 510(k) are the F&P 950 Bubble CPAP Breathing Circuit Kits (950N60J and 950N62J). The F&P 950 Bubble CPAP Breathing Circuit Kits are single patient use and designed for use with the F&P 950 Respiratory Humidifier. The F&P 950 Bubble CPAP (continuous positive airway pressure) Breathing Circuit Kits provide breathing support to spontaneously breathing neonates and infants. The F&P 950 Bubble CPAP breathing circuit kits deliver heated and humidified respiratory gas through an inspiratory limb to the infant via a nasal interface. An expiratory limb connects to a Bubble CPAP generator to generate CPAP. The device is intended to be operated at input gas flows of 4 - 15 L/min with available CPAP levels of 3 - 10 cmH2O.

    AI/ML Overview

    This document outlines the acceptance criteria and supporting studies for the F&P 950 Bubble CPAP Breathing Circuit Kits.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the F&P 950 Bubble CPAP Breathing Circuit Kits are primarily established through compliance with recognized medical device standards and comparison to a predicate device. The performance data section summarizes various tests conducted to demonstrate this compliance.

    Since the document does not present specific quantitative acceptance criteria with corresponding performance metrics in a direct table format, I will reconstruct a table based on the information provided, inferring acceptance as "compliance" or "demonstrated safety/effectiveness" where specific numerical targets are not explicitly stated for the "acceptance criteria" column. The "reported device performance" will refer to the conclusion of the tests.

    Acceptance Criteria CategoryRegulatory/Standard-Based Acceptance CriteriaReported Device Performance
    Intended Use EquivalenceTo deliver heated and humidified respiratory gases to provide CPAP to spontaneously breathing premature and full-term neonates and infants up to 10kg, in a hospital clinical environment (NICU/PICU).Equivalent to Predicate: Subject and predicate devices are both intended to provide CPAP to patients.
    Patient Population EquivalenceNeonatal and infant patients weighing <10 kg.Identical to Predicate: Neonatal and infant patients weighing <10 kg.
    Operating Principle EquivalenceBreathing circuit kit contains a humidification chamber, inspiratory limb, expiratory limb, pressure relief manifold, and Bubble CPAP generator. (For 950N62J, includes Hudson extensions).Identical to Predicate: The subject device's operating principle is described as identical to the predicate (with the addition of Hudson extensions for 950N62J).
    CPAP Range3 – 10 cmH2OIdentical to Predicate: 3 – 10 cmH2O.
    Specified Flow Range4 – 15 L/minIdentical to Predicate: 4 – 15 L/min.
    Over-pressure protection for patient safetyYes – 17 cmH2OIdentical to Predicate: Yes – 17 cmH2O.
    Ambient operating conditions20°C – 26°CIdentical to Predicate: 20°C – 26°C.
    ReusabilitySingle UseIdentical to Predicate: Single Use.
    Duration of Use14 Days (for subject device)Different to Predicate (but accepted): The subject device has been tested to a 14-day duration, which is longer than the predicate's 7 days, indicating demonstrated capability for extended use.
    Shelf Life3 yearsIdentical to Predicate: 3 years.
    Storage conditions-10°C – 50°CIdentical to Predicate: -10°C – 50°C.
    BiocompatibilityCompliance with ISO 10993-1:2018, ISO 18562 (Parts 1, 2, 3, 4), and related standards for external communicating gas pathway (dry and humidified), permanent contact (> 30 days).Compliant: Testing demonstrated an appropriate biocompatibility profile for the device, confirming compliance with specified ISO standards.
    Electrical Safety & Electromagnetic Compatibility (EMC)Compliance with ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2:2014 + AMD1:2020 Ed 4.1, and AIM Standard 7351731 Rev. 2.00 2017-02-23.Compliant: Testing demonstrated appropriate electrical safety and electromagnetic compatibility profile, confirming compliance with specified standards.
    Human Factors & UsabilityConducted in accordance with FDA's "Applying Human Factors and Usability Engineering to Medical Devices" guidance.Safe and Effective: Validation study demonstrated the device to be safe and effective for intended users, uses, and environments according to applicable guidance.
    Humidification Output, Thermal Overshoot, Surface Temp.Compliance with ISO 80601-2-74:2017.Compliant: Performance testing for these parameters was conducted in line with ISO 80601-2-74:2017 to demonstrate substantial equivalence. (Implies compliance, though specific figures are not provided in this summary).
    Resistance to Flow, Compliance, Gas LeakCompliance with ISO 5367:2014.Compliant: Performance testing for these parameters was conducted in line with ISO 5367:2014 to demonstrate substantial equivalence. (Implies compliance).
    Comparative System Level Pneumatic PerformanceDemonstrated equivalence.Demonstrated: Testing conducted to demonstrate substantial equivalence. (Implied successful demonstration).

    2. Sample size used for the test set and the data provenance
    The document does not explicitly state the sample sizes (number of devices or tests performed per device type) for the non-clinical tests (biocompatibility, electrical safety, human factors, bench/performance testing). It generally refers to "testing" or "studies" in the plural.

    The data provenance is from non-clinical tests performed by the manufacturer, Fisher & Paykel Healthcare Limited, to evaluate the device against specified international standards and to demonstrate equivalence to a predicate device. This is prospective testing of the manufactured device components and systems. No country of origin for the data is explicitly mentioned beyond the manufacturer's location in New Zealand, but the tests are conducted against universally recognized international standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not provided in the document. The studies listed are primarily non-clinical engineering and performance tests, not clinical studies requiring expert ground truth for diagnostic or clinical outcomes. For the Human Factors testing, "intended users" would have been involved, but their number and specific qualifications are not detailed.

    4. Adjudication method for the test set
    Not applicable. The tests described are objective, standard-based compliance evaluations or measurements, not clinical studies requiring expert adjudication of subjective outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is a mechanical breathing circuit kit, not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This device is a mechanical breathing circuit kit; there is no algorithm involved.

    7. The type of ground truth used
    For the non-clinical tests, the "ground truth" is defined by the requirements and methodologies outlined in the referenced international standards (e.g., ISO, ASTM, ANSI, IEC). For example, a successful biocompatibility test means the materials met the specified criteria of the ISO 10993 series. A successful electrical safety test means the device operated within the limits defined by IEC 60601-1-2.

    8. The sample size for the training set
    Not applicable. This device is a mechanical medical device, not an AI/machine learning product that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. As above, this device does not utilize AI and therefore has no training set or associated ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163283

    Validate with FDA (Live)

    Date Cleared
    2017-09-20

    (303 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983112, K103767, K151959

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use HAMILTON-H900

    The HAMILTON-H900 humidifier is intended to add moisture and to warm breathing gases during invasive and non-invasive mechanical ventilation. The intended area of use is the intensive care ward or the recovery room.

    The HAMILTON-H900 humidifier is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    Intended Use HAMILTON-BC8010

    The HAMILTON-BC8010 breathing set is intended to be used together with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.

    Intended Use HAMILTON-BC4010

    The HAMILTON-BC4010 breathing set is intended to be used together with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.

    Device Description

    HAMILTON-H900

    The HAMILTON-H900 respiratory gas humidifier is designed to add heat and moisture to respiratory gases, which are then administered to patients who need invasive or noninvasive ventilation. The breathing gas is passed through a humidifier chamber, where the gas is heated and humidified using an external heat source.

    The HAMILTON-H900 humidifier uses breathing circuits that are recognized automatically when connected to the humidifier. The humidifier uses two heating systems as heat sources:

    • -A heating plate that has contact with the metal plate of a special humidifier chamber inserted into the base of the humidifier prior to application, thus heating the water contained inside the chamber.
    • -The controlled heating within the breathing circuits.

    The respiratory gas exiting the humidifier chamber is continuously monitored using temperature sensors that are integrated into the humidifier base. Additionally, a temperature probe inside the breathing tube is used to measure the temperature of the gas delivered to the patient airway.

    The design of the breathing tubes offers heating performance over the entire length by providing constant heating of the entire tube wall, from the chamber exit to the patient connection. A float mechanism, together with detection of the water level inside the humidifier chamber and the autofill mechanism, keeps a constant water level.

    HAMILTON-BC8010

    The HAMILTON-BC8010 is a single-use, dual-limb breathing circuit comprising heated inspiratory and heated expiratory tubes. The humidifier chamber uses a float mechanism to keep a constant water level inside the auto-feed chamber.

    HAMILTON-BC4010

    The HAMILTON-BC4010 is a single-imb breathing circuit comprising a heated inspiratory tube. The humidifier chamber uses a float mechanism to keep a constant water level inside the auto-feed chamber.

    HAMILTON-HC322/HC310

    The HAMILTON-HC322/HC310 are single-use humidifier chambers with a float mechanism to keep a constant water level inside the auto-feed chamber. The HAMILTON-HC322 humidifier chamber is part of the HAMILTON-BC8022/BC4022 breathing sets and the HAMILTON-HC310 humidifier chamber is part of the HAMILTON-BC8010/BC4010 breathing sets, as cleared in K152029.

    AI/ML Overview

    This is a 510(k) summary for a medical device (respiratory gas humidifier and breathing circuits), not an AI/ML medical device. Therefore, the information requested about acceptance criteria, study details, and ground truth establishment, which are typical for AI/ML device submissions, are not primarily available or applicable in the provided text.

    The document discusses substantial equivalence to predicate devices based on technological characteristics and performance, focusing on non-clinical data. It does not describe an AI/ML algorithm that predicts or diagnoses based on medical images or other data, nor does it detail a study that evaluates the performance of such an algorithm against a ground truth established by experts.

    However, I can provide the acceptance criteria and performance as described for this specific device, based on the provided text, recognizing that it's not in the context of an AI/ML model.

    Here's an analysis of the provided text, focusing on the closest equivalents to your requested information, understanding the context is a medical device, not an AI/ML diagnostic:


    The provided text is a 510(k) summary for the Hamilton-H900 Humidifier and Hamilton-BC series breathing circuits. This document outlines the rationale for the device's substantial equivalence to previously cleared predicate devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and performance specifications and non-clinical testing, rather than the type of clinical performance study typically conducted for AI/ML diagnostic or predictive devices.

    Therefore, many of the requested points (e.g., number of experts for ground truth, adjudication methods, MRMC study, standalone performance for an algorithm) are not applicable to this type of device submission.

    Here is the information extractable from the document, focusing on performance characteristics and non-clinical testing rather than AI/ML specific criteria:

    Acceptance Criteria and Reported Device Performance (Non-AI/ML Context)

    The acceptance criteria are implied by the "Characteristic used for comparison" and the "HAMILTON-H900 (Subject device)" column in Table 1 for the humidifier and Table 2 for the breathing circuits. The "reported device performance" is the statement that the subject device meets these established characteristics, which are largely identical to the predicate device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    For HAMILTON-H900 (Humidifier):

    Characteristic Used for ComparisonAcceptance Criteria (Implied by Subject Device Spec)Reported Device Performance
    Indication for UseAdd moisture and warm breathing gases during invasive/non-invasive mechanical ventilation for adult, pediatric, and neonatal patients in ICU/recovery room.Device intended for this use, within limits of stated technical specifications. Patient population expanded from predicate to include pediatric and neonatal.
    EnvironmentHospital use by trained personnel.Confirmed.
    Type of VentilationInvasive and non-invasive.Confirmed.
    Modes of OperationAutomatic Mode (Invasive & Noninvasive), Manual Mode (Invasive & Noninvasive), Standby Mode, Default Mode (Invasive after power on).Confirmed.
    Flow RangeNon-invasive: up to 120 L/min; Invasive: up to 60 L/min.Confirmed.
    Maximum Operating DurationContinuously.Confirmed.
    Accuracy Displayed TemperatureDisplayed Range: 10°C to 60°C; Chamber exit: 10°C to 60°C ± 1°C, 30°C to 41°C ± 0.5°C; Y-piece: 28°C to 43°C ± 0.5°C.Confirmed.
    Resolution of Displayed Temp.0.1 °C.Confirmed.
    Temp Control Setting: AirwayInvasive: 33°C to 42°C; Noninvasive: 28°C to 38°C; in steps of 0.5°C (manually).Confirmed.
    Temp Control Setting: ChamberInvasive: 35°C to 41°C; Noninvasive: 30°C to 35°C; in steps of 0.5°C (manually).Confirmed.
    Disabling Heater WiresYes; expiratory tube disabled with Single Limb Circuits; temp alarm can disable all heating wires.Confirmed.
    Humidity PerformanceInvasive mode: >33 mg H2O/L @ 60 L/min and 26°C ambient; Noninvasive mode: >10 mg H2O/L @ 120 L/min and 26°C ambient.Confirmed.
    Alarm DisplayIndicators for Chamber Temp, Heater Wire, Humidity Alarm, Patient-end Probe (Y-Piece), water level (Low/High), device inclination, tube detection/recognition/connection, chamber insertion.Confirmed.
    Maximum Power283 VA (230V), 293 VA (115V), 268 VA (100V).Confirmed.
    Power Heated Tube22V; 60 W (dual limb), 30 W (single limb).Confirmed.
    Power Heating Plate150 W ± 5 %.Confirmed.
    Heating Plate Overheat Prot.130°C ± 4°C.Confirmed.
    Chamber Insertion DetectionAble to detect if a chamber is inserted.Confirmed.
    Water Level DetectionHigh and Low water level detected via optical sensors.Confirmed.
    Breathing Circuit RecognitionVia electrical connectors inside the breathing tubes.Confirmed.
    Single Use / ReuseCompatible with single use and reusable breathing circuits.Confirmed.
    ClassificationClass I (IEC 60601-1).Confirmed.
    Applied PartType BF.Confirmed.
    Standards / PerformanceCompliance with IEC 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-8:2005+Amd1:2012, ISO 8185:2007, MIL-STD-461F:2007, IEC 62304:2006, IEC 62366:2007+Amd1:2014.Confirmed.

    For HAMILTON-BC8010 / BC4010 (Breathing Circuits):

    Characteristic Used for ComparisonAcceptance Criteria (Implied by Subject Device Spec)Reported Device Performance
    DescriptionBC8010: dual limb, heated with water chamber; BC4010: single limb, heated, with water chamber.Confirmed.
    Classification No. & Product868.5270, BZE.Confirmed.
    Intended UseUse with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.Patient population expanded from predicate to include pediatric and neonatal.
    Patient PopulationPediatric and neonatal patients.Confirmed.
    EnvironmentHospital use by trained personnel.Confirmed.
    Indications for UseInvasive and noninvasive ventilation.Confirmed.
    Compatibility with other DevicesHamilton Medical compatible humidifier (e.g., HAMILTON-H900).Confirmed.
    Where UsedDuring invasive and noninvasive mechanical ventilation, hospital use by trained personnel.Confirmed.
    Target/Patient PopulationPediatric and neonatal patients requiring mechanical ventilation or positive pressure breathing assistance.Confirmed.
    Compressible Volume800 ml (BC8010), 600 ml (BC4010).Confirmed.
    Maximum Chamber Operating Press.20 kPa.Confirmed.
    LengthInspiration: 1.65m total (1.3m heated, 0.35m unheated); Expiration: 1.65m.Confirmed.
    Inner Diameter12 mm.Confirmed.
    Flow Range1 to 30 L/min.Confirmed.
    Flow Resistance@ 2.5 l/min: Inspiration: 0.03 cmH2O/(l/min), Expiration: 0.02 cmH2O/(l/min). @ 15 l/min: Inspiration: 0.06 cmH2O/(l/min), Expiration: 0.04 cmH2O/(l/min).Confirmed.
    Gas Leakage15 mL/min (BC8010), 4.0 mL/min (BC4010).Confirmed.
    Compliance< 1 ml/cmH2O/m.Confirmed.
    Wire ResistanceInspiration: 14.6±0.9Ω; Expiration: 18.2±1.2Ω.Confirmed.
    ConnectorConical according to ISO 5356-1; electrical connector part of breathing tube.Confirmed.
    Breathing Circuit RecognitionTubes recognized by Hamilton Medical Humidifier.Confirmed.
    SterilityNon-sterile, ready for use.Confirmed.
    Single Use / ReuseSingle Use.Confirmed.
    Power Heated Tube33.15W inspiratory tube @100% power (BC8010/BC4010); 26.59W expiratory tube @100% power (BC8010).Confirmed.
    Standards / PerformanceCompliance with various ISO and IEC standards listed (e.g., IEC 60601-1:2012, ISO 8185:2007, ISO 10993 series).Confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    As this is a medical device submission based on substantial equivalence and non-clinical testing rather than an AI/ML study with a "test set" of patient data, specific "sample sizes" in the sense of clinical cohorts are not detailed. The testing involved:

    • Electrical Safety and Electromagnetic Compatibility: Conducted on the HAMILTON-H900 and HAMILTON-BC-series breathing circuits.
    • Mechanical Testing: Conducted on the HAMILTON-H900 and HAMILTON-BC-series breathing circuits.
    • Software Verification and Validation Testing: Conducted for the software within the HAMILTON-H900.
    • Biocompatibility Testing: Performed on "relevant parts of the breathing circuit."
    • Human Factors / Usability Study: Conducted on the HAMILTON-H900.

    The document does not specify the number of units tested for each category, which is typical for non-clinical testing reports summarized in 510(k)s.

    Data Provenance: The document does not specify the country of origin of the data or explicitly state whether the tests were retrospective or prospective. Given these are physical device tests, they would be prospective (performed specifically for this submission). The applicant is based in Switzerland (Hamilton Medical AG, Via Crusch 8, 7402 Bonaduz, Switzerland).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. "Ground truth" in the context of an AI/ML algorithm with expert adjudication is not relevant here. The "ground truth" for this device's performance is objective measurements against engineering specifications and international standards (e.g., temperature accuracy, flow resistance, humidity output).

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication for this type of device testing. Performance is measured against predefined engineering and safety standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is for evaluating human reader performance, often with and without AI assistance, typically in diagnostic imaging. This device is a respiratory gas humidifier and breathing circuit, not a diagnostic AI/ML system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable directly. This device is a physical medical device, not a standalone algorithm. Its performance is evaluated against its physical and electronic specifications.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by engineering specifications, adherence to international standards (e.g., IEC 60601 series, ISO 8185, ISO 5367, ISO 5356-1, ISO 10993 series), and objective measurements during non-clinical testing. For example, temperature accuracy is measured against a calibrated standard, not against a human expert's assessment.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device that requires a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132017

    Validate with FDA (Live)

    Date Cleared
    2014-03-25

    (267 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983112, K020332, K034026, K103767, K122432, K9341401, K913368, K073706, K953711, K131957

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fisher & Paykel HC550 System is designed for use with artificial ventilation systems. Portable volume ventilation systems, pressure support ventilation and Continuous Positive Airway Airway Pressure (CPAP) systems may incorporate an HC550 to provide therapeutic levels of warm humidified air to adult patients with artificial airways or through mask ventilation.

    The operating flow range is 5 to 120L/min depending on the patient interface.

    The HC550 is designed for use in long term care facilities or the home under the prescription of a qualified medical professional.

    Device Description

    The Fisher & Paykel Healthcare HC550 System is designed to condition gases for patients by raising the delivered water vapor content (humidity) and temperature of the gases.

    The HC550 System consists of the following components:

    • . HC550 Respiratory Humidifier
    • . Accessories:
      • a) Breathing circuit (compatible adult breathing circuits as cleared in K983112, K020332, K034026, K103767, K122432)
      • b) Humidification Chamber (as cleared in K9341401 and K913368)
      • c) Heaterwire Adaptor (as cleared in K073706)
      • d) Temperature/Flow Probe (as cleared in K983112)
      • RT008 Air Entrainer (optional oxygen therapy accessory) (as cleared in K953711) e)

    The device consists of an electrically powered heat controller, utilizing a microprocessor with embedded software, to control a heating element that transfers heat to the water in a humidification chamber.

    A dryline tube (part of the breathing circuit) transports respiratory gases from a flow source (e.g. ventilator) to the humidification chamber where the gases are heated and humidified.

    The inspiratory limb of the breathing circuit transports the heated and humidified gases from the humidification chamber to the patient. The inspiratory tube may be electrically heated by means of a heater-wire placed internally to the tube, which is controlled by the HC550 respiratory humidifier.

    The expiratory limb of the breathing circuit transports expired gas from patient. In the case of a dual-heated breathing circuit, this limb may also be heated in the same manner as the inspiratory limb.

    If a heated breathing circuit is used, the heaterwire adaptor provides electrical energy from the respiratory humidifier to the heaterwire in the breathing circuit.

    Temperature probes in the gas path provide feedback on temperature and flow of the gas to regulate temperature and humidity to the patient.

    AI/ML Overview

    HC550 System Acceptance Criteria and Study Details

    This document describes the acceptance criteria and corresponding study results for the HC550 System, a respiratory gas humidifier. The information is extracted from the provided 510(k) Notification K132017.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the HC550 System are based on compliance with the ISO 8185:2007 standard for humidification systems. The device's performance is compared to these requirements.

    Acceptance Criteria (from ISO 8185:2007)Reported Device Performance (HC550 System)
    Humidity performance (for non-invasive mode): ≥ 10 mg/L over recommended flow ranges≥ 10 mg/L over the recommended flow ranges
    Humidity performance (for invasive mode): ≥ 33 mg/L over recommended flow ranges≥ 33 mg/L over the recommended flow ranges
    Enthalpy: < 194 kJ/kg dry gas< 194 kJ/kg dry gas

    2. Sample Size and Data Provenance for Test Set

    The document does not explicitly state a specific sample size for a "test set" in the context of clinical studies. The performance evaluation was conducted through bench testing to demonstrate compliance with ISO 8185:2007.

    • Sample Size for Test Set: Not applicable in the context of a clinical test set. Performance was evaluated through bench testing.
    • Data Provenance: Bench testing results, derived from laboratory experiments, not patient data. No country of origin for specific data is stated, but the submission is from Auckland, New Zealand.
    • Retrospective/Prospective: Not applicable, as no clinical study with patient data was performed.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not applicable. Ground truth was established by international standards (ISO 8185:2007) for device performance, not expert consensus on an annotated dataset.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for Test Set

    • Adjudication Method: Not applicable, as no clinical study requiring adjudication of expert opinions was conducted. Performance was judged against predefined engineering standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed.
    • Effect Size of AI vs. without AI assistance: Not applicable, as this is a medical device and not an AI-assisted diagnostic or treatment system.

    6. Standalone Performance Study

    • Standalone Performance Study: Yes, a standalone performance study was done in the form of bench testing. The algorithm (or device in this case) was tested against established international standards (ISO 8185) to assess its functional performance independently. The results are presented in the table above, demonstrating the device's ability to meet the humidity and enthalpy requirements specified by the standard.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth used was international standard compliance (ISO 8185:2007). This standard sets quantifiable performance requirements for respiratory gas humidifiers.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device does not utilize machine learning or AI models that require a "training set." Its design and performance are based on engineering principles and compliance with established standards.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as no training set was used.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1