Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K190051

    Validate with FDA (Live)

    Device Name
    TD Clip
    Date Cleared
    2019-06-07

    (147 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN170090, K103704, K111066

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD Clip is intended to be used with MATRx or MATRx plus Titration Trays to form a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for a total of less than 30 nights.

    Device Description

    The TD Clip is a temporary oral appliance. The patient uses the temporary oral appliance to alleviate mild to moderate obstructive sleep apnea and/or snoring while waiting for a custom oral appliance to be manufactured. The dentist fabricates temporary titration trays consisting of upper and lower dental trays. The trays are custom fit to the patient using polyvinyl siloxane impression material. The healthcare provider may use the clip assembly to fasten the titration trays at the target therapeutic position to form the TD Clip temporary oral appliance. Patients may use the temporary oral appliance as treatment while waiting for their custom oral appliance. Adjustment of the position requires the setting of a new TD Clip by the healthcare professional.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are described below for the TD Clip.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Intraoral Material PerformanceNo presence or increased risk of breakage that could lead to unacceptable harm over the maximal study length (30 nights of use).Evaluated under simulated aging, including environmental conditions and representative worst-case loading schemes. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    TD Clip PerformanceMaintained the set protrusive positioning over 30 nights of use.Evaluated under repeated use and exposed to normal sleep forces over 30 nights of use, including static and cyclic loading under worst-case use. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Tray Removal Simulation100% of users were able to remove the Titration Trays from their mouth quickly and without causing injury to themselves or the Titration Trays.Evaluated to ensure users could remove the trays quickly and without injury. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Biocompatibility TestingMaterials in contact with mucosal membrane adhere to ISO 10993-1 and associated FDA guidance.All materials in contact with the mucosal membrane were assessed and tested in accordance with ISO 10993-1 and the associated FDA guidance. The device met the implied acceptance criteria.
    Risk ManagementNo unacceptable risks identified.During risk management and performance testing, no new risks were identified. The risk management concluded that the TD Clip had no unacceptable risks.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified for any of the performance tests. The document only mentions "mechanical and environmental use conditions" and "repeated use and exposed to normal sleep forces over 30 nights of use".
    • Data Provenance: The tests described are "Bench testing" and "simulated mechanical and environmental use conditions," indicating these were laboratory/engineering tests rather than human clinical trials. Thus, there is no country of origin or retrospective/prospective distinction for data provenance in the traditional sense; the data is from an engineering test setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. The ground truth for these performance tests would be based on engineering specifications and direct measurements against those specifications, rather than expert interpretation of medical data. The evaluation was for mechanical and material performance.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical studies with human evaluators to establish ground truth for ambiguous cases. These were bench tests against predetermined engineering criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

    • No MRMC comparative effectiveness study was done. Clinical testing was explicitly stated as "not performed as the device has no technological differences from the predicate."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The TD Clip is a physical medical device (an intraoral appliance) and not an algorithm or AI software. Therefore, the concept of "standalone performance" for an algorithm doesn't apply here. The performance tests focused on the physical integrity and function of the device itself.

    7. The Type of Ground Truth Used:

    • The ground truth for the performance testing was based on engineering specifications and predefined performance limits (e.g., no breakage, maintenance of protrusive positioning, quick and injury-free removal, adherence to biocompatibility standards). This is a form of objective, empirical measurement and adherence to established material and mechanical standards.

    8. The Sample Size for the Training Set:

    • Not applicable. As a physical medical device undergoing bench testing, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170090

    Validate with FDA (Live)

    Device Name
    MATRx plus
    Date Cleared
    2018-08-23

    (245 days)

    Product Code
    Regulation Number
    872.5571
    Type
    Direct
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MATRx plus device may also be used with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    MATRx plus uses these recordings to produce a report for the HCP that can be used to prospectively identify patients with mild to moderate obstructive sleep apnea who may be suitable for therapy with an oral appliance and to recommend a target mandibular position.

    The use of the device does not replace the need for follow-up testing to determine the initial and ongoing effectiveness of the therapy as recommended by clinical practice guidelines.

    Device Description

    This De Novo request is for an expanded indication (new intended use) for use with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    The device is an automated temporary oral appliance that manipulates the mandible through multiple protrusion levels during an overnight sleep study in the home. The device uses the same interface as the MATRx plus Home Sleep Apnea Test, also from Zephyr Sleep Technologies, cleared under product code MNR, regulation 21 CFR 868.2375, cleared in K163665. These components include the recorder, tablet, oximeter, effector belt, and nasal cannula. The MATRx plus device for a titration study includes a mandibular positioner to determine an optimal titration position based on inputs of airflow and an oxygen desaturation index (ODI) of less than 10 events per hour. The device analyses the collected information and generates a report to assist the Healthcare Provider in the clinical management of oral appliance therapy for patients with mild to moderate obstructive sleep apnea. From this report, patients may be informed on the future efficaciousness of oral appliance (OA) therapy.

    Specifically, the clinical management may be assisted by this type of oral appliance assessment study by prospectively identifying patients that are expected to achieve therapeutic success with oral appliance therapy. The oral appliance assessment study also includes the provision of a recommended target protrusive setting for the mandible through use of a legally marketed intraoral appliance (regulated under 21 CFR 872.5570, product code LQZ). The target protrusive setting is the amount of mandibular protrusion where the patient is expected to achieve efficacious therapy with the intraoral appliance and is used by the HCP to more effectively complete the titration process.

    The use of the MATRx plus device for an oral appliance assessment study does not replace the clinical titration process or the need for follow-up testing to determine the initial and ongoing therapy as recommended by clinical practice guidelines.

    The MATRx plus test identifies patients as suitable for intraoral therapy if a mandible protrusion level can be found where the patient has an oxygen desaturation index (ODI) of less than 10 events per hour with a 4% desaturation criterion.

    AI/ML Overview

    Here's a structured breakdown of the acceptance criteria and the study that proves the MATRx plus device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Clinical Endpoints

    Acceptance Criteria CategorySpecific Acceptance Criteria (Effectiveness, Safety, Usability)Reported Device Performance (as stated in the document)
    Effectiveness1. Predictive Accuracy (Primary Endpoint): Sensitivity and specificity statistically significant to reject the null hypothesis that sensitivity ≤0.6.Sensitivity: 94.2% (95% CI: 0.86 – 1.00)
    Specificity: 75.0% (95% CI: 0.40 – 1.00)
    2. Target Protrusive Position (Secondary Endpoint): Statistically significant to reject the null hypothesis that target accuracy ≤0.6.Identification of Efficacious Target Protrusion: 86.3% of participants were correctly predicted as responders and successfully provided with an effective target protrusion.
    SafetyNo significant safety concerns reported by coordinator, dentist, or participant.No significant safety concerns reported. Adverse events were temporary discomfort (in teeth, joint, jaw, minor gum irritations, temporary bite changes, soreness in teeth) that resolved within a few hours.
    Only minor discomfort reported on post-study questionnaires.Temporary discomfort reported (resolved within hours).
    No software failure resulting in injury.No software failure reported that resulted or could result in injury.
    No movement beyond full retrusion and maximum protrusion set values.No movement beyond full retrusion and maximum protrusion set values reported.
    UsabilityAccurate prediction for OA therapy and effective titration level demonstrated by actual use testing (setup, placement, use, removal by participant in home).Human factors assessment conducted as part of the clinical trial showed participants and providers evaluated for comprehension-based tasks. The study supported usability.

    Bench Performance Testing

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance (as stated in the document)
    Mandibular Positioner, Accuracy Data(i) Device will be able to send a position feedback signal to the controller that reports a new position within 0.5mm of the actual physical position of the trays/mandible; and the accuracy will be maintained under expected clinical loads. (ii) The mandibular positioner will be able to physically endure the use life of the device under normal operating load without failure and maintain the accuracy. Simulated use life testing was conducted based on days of single patient use, movements per patient use, and approximated number of insertion and removal cycles per patient use.The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Mandibular Positioner, Force Limit TestMandibular positioner will be incapable of applying a force that could induce serious injury under a worst-case fault scenario of 70 Newtons.The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Device Force Limitation Simulation(i) Demonstrate the ability of users to voluntarily stall the mandibular positioner in a fault position (i.e., full stroke speed). (ii) Demonstrate that without voluntary action to resist the force, the trays dislodge from their seated position on the teeth prior to injury or maximum force (i.e., 70N).The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Impression Material BreakageDemonstrate the ability of the impression material used to form the temporary oral appliance to maintain integrity over the use life of the temporary oral appliance.The data provided showed each test was passed according to its pre-specified acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Clinical Study: The document provides "49" for "Successful Identification of Responders" and "2" for "Failure to Identify Responders" in the "Predictive Success Category," and "3" for "Successful Identification of Responders" and "6" for "Failure to Identify Responders" in the "Predictive Failure Category." This totals 49 + 2 + 3 + 6 = 60 participants in the mild to moderate OSA population for the effectiveness endpoints.
      • Human Factors: 15 patients were recruited for each study.
    • Data Provenance: The MATRx plus is by Zephyr Sleep Technology, located in Calgary, Alberta, Canada. The clinical study was conducted in a clinical setting (dental professional's office for deployment, home environment for use). The study is prospective, as it involved recruiting participants, conducting the MATRx plus study, and then following up with custom oral appliance therapy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts:
      • Effectiveness Ground Truth (Oral Appliance Outcome): While not explicitly stated how many individual experts reviewed each case, the process involved a "Sleep Health professional" reviewing baseline HSAT data and a "licensed dentist" conducting intraoral exams and subjective interviews post-study. The final determination of "successful treatment with an oral appliance" (ODI < 10 with custom appliance) and "unsuccessful treatment" (ODI ≥ 10) was based on "automated analysis of a full night home sleep test," which implies an objective measurement, potentially overseen or interpreted by these professionals. For the "Identification of Efficacious Target Protrusion," if the initial target wasn't successful (ODI < 10), the oral appliance was advanced and retested, indicating involvement of a clinical expert.
      • Human Factors: Not explicitly stated, but the process involved "Participants and providers were evaluated for comprehension-based tasks."
    • Qualifications of Experts:
      • Sleep Health Professional: Reviewed baseline HSAT data. (Specific qualifications like "radiologist with 10 years of experience" are not provided, only the role title).
      • Licensed Dentist: Evaluated for adequate dentition, fitted titration trays, measured protrusion values, conducted intraoral exams, and performed subjective interviews post-study. (Specific qualifications like years of experience are not provided).

    4. Adjudication Method for the Test Set

    The document does not explicitly describe a formal "adjudication method" (like 2+1 or 3+1 consensus) for establishing the ground truth of the test set results. The primary effectiveness endpoint, successful/unsuccessful treatment with an oral appliance, was defined retrospectively by an "automated analysis of a full night home sleep test" using the ODI < 10 criterion. This suggests an objective, rule-based approach rather than a multi-expert consensus for the final outcome. However, "All baseline data were reviewed by a Sleep Health professional," and the follow-up for target protrusion involved clinical discretion ("adjusted accordingly based on clinical expertise of the dentist"), indicating expert involvement in data interpretation and treatment adjustments.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the MATRx plus device in predicting OA therapy success and recommending a target mandibular position. It did not compare human readers' performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone study assessing the algorithm's performance was conducted. The "Predictive Accuracy" effectiveness endpoint explicitly states: "The MATRx plus predictive accuracy endpoint was assigned if the sensitivity and specificity was found to be statistically significant..." and "The predication was done using a trained binary classifier." This refers to the algorithm's ability to identify responders and recommend a target protrusion on its own, based on the collected physiological data. While human factors and clinical oversight were part of the overall process, the core predictive power evaluated was that of the device's algorithm.

    7. Type of Ground Truth Used

    The primary ground truth for the effectiveness endpoint ("Successful treatment with an oral appliance") was outcomes data: achieving an Oxygen Desaturation Index (ODI) of less than 10 events per hour with a custom oral appliance, determined by automated analysis of a full night home sleep test.
    For the target protrusion accuracy, the ground truth was the clinically determined effective protrusion (the position where ODI < 10 was achieved with the custom OA, potentially after clinician-guided titration).

    8. Sample Size for the Training Set

    The document does not provide the sample size for the training set. It only describes the pivotal clinical trial (Phase II) which appears to be the test set for evaluating the device's performance. It mentions "The predication was done using a trained binary classifier," implying a training phase occurred, but details about the data used for training are absent from this text.

    9. How the Ground Truth for the Training Set Was Established

    The document does not describe how the ground truth for the training set was established, as it does not detail the training process or the data used for it. It only mentions that the prediction was made using a "trained binary classifier," implying ground truth data was used for its training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1