Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163127

    Validate with FDA (Live)

    Device Name
    bellavista 1000
    Manufacturer
    Date Cleared
    2017-07-31

    (265 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    HAMILTON-G5 (K131774), COVIDIEN-PB 840 (K151252), DRAEGER-Oxylog 3000 plus (K103625)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The bellavista 1000 ventilator is intended to provide positive pressure ventilator support to adults and pediatrics (IBW greater than 6 kg).

    Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer

    It is intended for use by qualified, trained personnel under the direction of a physician.

    Device Description

    The bellavista 1000 ventilator (hereafter referred to as "bellavista") is designed for continuous ventilation of adult and pediatric patients.

    The bellavista ventilator is a positive pressure ventilator for intubated patients and patients with mask ventilation. Its pneumatic power is provided by an internal turbine. Supplemental oxygen is provided by an auxiliary oxygen source.

    The following modes of ventilation are available

    • Adults: CPAP, PCV, P-A/C, PC-SIMV, PSV, S, S/T, T, beLevel, APRV, VCV, ● V-A/C, VC-SIMV.
    • Pediatrics the following modes of ventilation are available: CPAP, PCV, P-A/C, ● PC-SIMV, S, T, beLevel, APRV, VCV, V-A/C, VC-SIMV
    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a medical device called the "bellavista 1000" ventilator. It is a letter confirming the device's substantial equivalence to legally marketed predicate devices.

    Crucially, this document does NOT describe the acceptance criteria and study that proves the device meets the acceptance criteria in the way you are asking.

    The 510(k) process for substantial equivalence (SE) reviews a new device against a predicate device. It demonstrates that the new device is as safe and effective as a legally marketed device, not that it independently "meets acceptance criteria" through clinical trials for novel devices.

    Therefore, many of the specific points you requested (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set information) are not applicable to this type of regulatory submission because clinical performance studies, as you've outlined, are generally not required for a 510(k) unless significant differences or new questions of safety/effectiveness arise.

    The document states:

    • "No clinical testing was conducted or required in support of this premarket notification." (Page 17)
    • The comparison is primarily against a predicate device (HAMILTON-C2, K121225) and some reference devices (HAMILTON-G5, COVIDIEN-PB 840, DRAEGER-Oxylog 3000 plus).

    Instead of clinical study data, the document relies on:

    • Performance Testing:
      • Tests according to IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2007, ISO 80601-2-12:2011, IEC 60601-1-8:2006+A1:2012, ISO 80601-2-55:2011.
      • Reliability testing as described in ASTM F1100-90.
      • Validation testing with "wide range of System Acceptance Tests (SAT)".
      • Biocompatibility testing according to ISO 10993-1, with certain scientific methodologies within ISO 18562-3.
    • Comparison of Technical Specifications: Detailed tables comparing settings, alarms, monitoring parameters, and accuracy controls with the predicate device.
    • Usability: Summative usability tests performed with a user group of respiratory therapists.

    Given this, I will answer the applicable parts of your request based on the provided document and clarify why other points are not present.


    Acceptance Criteria and Device Performance (Based on Substantial Equivalence to Predicate)

    The "acceptance criteria" in this context are primarily demonstrating equivalence to the predicate device (HAMILTON-C2, K121225) and compliance with relevant performance standards. The "reported device performance" is presented through detailed comparisons.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/ParameterAcceptance Criteria (Demonstrated Equivalence / Standard Compliance)Reported bellavista 1000 Performance (vs. Predicate/Reference)
    Indications for UseSubstantially equivalent to predicate HAMILTON-C2 (K121225)."Intended to provide positive pressure ventilator support to adults and pediatrics (IBW greater than 6 kg)." (Similar to predicate)
    Patient PopulationSubstantially equivalent to predicate."Adults and pediatrics (IBW greater than 6 kg)." (Predicate has broader population including infants and neonates, but this "does not change the intended use.")
    Environment of UseSubstantially equivalent to predicate."Hospitals, sub-acute care facilities and intra-hospital transfer." (Same as predicate)
    Technology / DesignSimilar system architecture, gas source, flow measurement, power supply, exhalation method to predicate. Compliance with relevant standards.System Architecture: 2 microprocessors + alarm controller (Predicate: 1 microprocessor + alarm controller) Gas Source: Blower + high pressure O2 (Similar) User Interface: Touch screen (Predicate: touchscreen + keys + press-and-turn knob) Flow Measurement: Proximal flow sensor (Similar) Power Supply: AC, DC or battery (Similar) Exhalation: Membrane valve towards ambient (Similar)
    Ventilation ModesSimilar to predicate and reference devices.Lists various modes for adults and pediatrics (e.g., CPAP, PCV, VCV). Most are similar to the predicate or specific reference devices. PSV and S/T not available for pediatrics.
    Setting RangesBellavista setting is a maximum of 10% outside the range of the predicate device's setting (for ventilation modes).CPAP: 4-30 mbar (Predicate: 0-35 mbar – narrower range, but acceptable) Exp Trig: 5-90% (Predicate: 5-80% – higher available range) FiO2: 21-100% (Similar) PEEP: 0-40 mbar (Predicate: 0-35 mbar – similar to reference G5) VtInsp/Vt: 40-2500 mL (Predicate: 20-2000 ml – similar to reference PB840)
    Alarm LimitsBellavista setting is a maximum of 10% outside the range of the predicate device's setting. Compliance with ISO 60601-1-8.Apnea Time: 2-60s (Predicate: 15-60s) FiO2: Upper 24-100%, Lower 18-80% (Similar) MVInsp: 0.1-60 L/min (Similar) PPeak: Upper 7-65 mbar, Lower 1-55 mbar (Similar) Rate: 1-130 bpm (Similar) VtExp: 40-3000 mL (Similar) Leak%: 5-95%, Off (Predicate: n/a - new feature, informative only)
    Monitoring ParametersBellavista measuring range maximum 20% outside predicate; compliance with ASTM F 1100-90 (indicators within 10%, freq. within 1 bpm or 10%).%Spont: 0-100% (Similar) AutoPEEP: 0-100 mbar (Similar) CStat: 0-1000 mL/mbar (Predicate: 0-200 mL/mbar – higher range, but minor difference) FiO2: 18-100 Vol% (Similar) MVExp: 0-250 L/min (Predicate: 0-99.9 L/min – higher range, related to higher peak flows) PEEP/PMean/PPeak/PPlateau: Similar ranges and accuracies.
    Accuracy of ControlsBellavista setting within 10% of predicate range.Volume: ± (5 mL + 6%) (Predicate: ±10% or ±10 ml) (Similar)Inspiratory pressure: ± (0.8 mbar + 4%) (Predicate: ± (1.6 mbar + 4%)) (Similar)PEEP: ± (0.5 mbar + 1%) (Predicate: ± (1.6 mbar + 4%)) (Similar)Oxygen: ± (3.5% FiO2 + 5% of actual reading) (Similar to reference Dräger Oxylog 3000)
    BiocompatibilityMaterials in gas pathway evaluated per ISO 10993-1 and ISO 18562-3, acceptable for intended use.VOC (gas emission), CO, CO2, Ozone and PM25 testing performed. Materials found acceptable.
    Electrical Safety/EMCCompliance with IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2007.Performed and demonstrated compliance.
    ReliabilityPerformed as described in ASTM F1100-90.Testing performed.
    Software ValidationSystem Acceptance Tests (SAT).Testing performed.
    UsabilitySummative usability tests performed with representative user group.Performed with a user group of respiratory therapists.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. This is a 510(k) premarket notification relying on technical specifications and performance testing against standards and a predicate device, not a clinical study with patient samples.
    • Data Provenance: The document does not specify the country of origin for the internal testing data, but the submitter is IMTMEDICAL AG, Switzerland. The testing is non-clinical (laboratory/bench testing) rather than retrospective or prospective patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. Ground truth in the context of clinical expert consensus would be relevant for devices involving image interpretation or diagnosis. For a ventilator, "ground truth" for performance is established through internationally recognized performance standards (e.g., ISO, IEC, ASTM) and physical measurements using test lungs and calibrated sensors.
    • The usability testing was conducted with a user group reported as "respiratory therapists," but no number or detailed qualifications are provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Applicable. This is a method used in clinical studies, particularly for diagnostic devices where subjective interpretation (e.g., by radiologists) needs consensus. It does not apply to the technical performance testing of a ventilator.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study is a type of clinical study for diagnostic devices, especially those using AI, to assess the impact of AI on human reader performance. This document concerns a ventilator, which is a therapeutic device, and explicitly states "No clinical testing was conducted or required."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in spirit, for certain aspects. The performance testing (e.g., accuracy of controls, alarm limits, monitoring parameters) represents the "standalone" or "algorithm only" performance of the ventilator's internal systems against engineering specifications and international standards. This is not "AI algorithm only" but "device functionality only." The document does not detail specific "algorithms" for AI, as this is a traditional ventilator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the device's technical performance is the relevant international performance standards (e.g., IEC, ISO, ASTM) and the measured performance of the legally marketed predicate device. These standards define acceptable ranges for parameters like accuracy, flow, pressure, etc.

    8. The sample size for the training set

    • Not Applicable. This is a traditional medical device (ventilator). The document does not describe the use of machine learning or AI models requiring a "training set" of data. Its design and validation rely on engineering principles, known predicate device performance, and adherence to established standards.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1