Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K120454

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-07

    (297 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103375, K961616, K081634

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SafeAir Smoke Pencil. is designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.

    Device Description

    The SafeAir Smoke Pencil is a sterile single use integrated electrosurgical pencil and smoke evacuation handpiece. The device is designed for general electrosurgical applications and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The integration of the electrosurgical pencil and smoke evacuation enables the operator to activate an electrosurgical current as well as capture the smoke plume simultaneously. The device is constructed of similar materials and design specifications as the predicate electrosurgical devices. The SafeAir Smoke Pencil combines the functions of electrosurgery and smoke evacuation into a single handpiece. The smoke evacuation orifice is located less than 1 cm from the electrosurgical blade to provide optimal smoke removal during cauterization. The smoke evacuation suction sleeve is adjustable to optimize the distance to the tip of the electrosurgical blade. As with the predicate GoldVac, the SafeAir Smoke Pencil is available in two (2) activation switch configurations, a rocker style and a push button style, which activate monopolar cut or coagulate functions. The device is connected to tubing using a dual connector to allow the user to connect to a variety of smoke evacuation systems including filtration or central vacuum systems, thus minimizing exposure of personnel to the surgical smoke plume. The device will be packaged singly for sterile distribution.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the SafeAir Smoke Pencil, an electrosurgical pencil with integrated smoke evacuation. The primary goal of this submission is to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the SafeAir Smoke Pencil are implied by its claim of substantial equivalence to predicate devices. This means its performance characteristics (electrosurgical function and smoke evacuation capability) should be comparable to or not adversely different from the predicate devices.

    Acceptance Criteria CategoryDetails (Implied from Substantial Equivalence Claims)SafeAir Smoke Pencil Reported Performance
    Electrosurgical Function (Cutting/Coagulation)The device should effectively perform cutting and coagulation, similar to predicate electrosurgical pencils. This implies delivering appropriate electrical current to achieve the desired surgical effect without compromising safety."The LiNA SafeAir Smoke Pencil has been shown to perform as intended with the equivalent results as the predicate devices through simulated tissue cauterization and cutting functions.... The suction sleeve or intake portion of the device has been designed to not impede the electrosurgical function of the pencil...**" (Section 9)
    Smoke Evacuation EffectivenessThe device should effectively remove surgical smoke generated during electrosurgery without impeding the electrosurgical function. The volume of air removed and the size of the smoke evacuation orifice should be similar to predicate devices. The smoke evacuation system should minimize exposure to the surgical team and patient."The SafeAir Smoke Pencil removes a similar volume of air and has a similar sized smoke evacuation orifice as the referenced predicate devices. The SafeAir Smoke Pencil smoke evacuation system is effective in removing smoke generated during the electrosurgery process." (Section 6) "No particular requirements specific to the smoke performance exist in the standards but tests conducted with the device have shown no effect or changes to function of the electrosurgical pencil. The suction sleeve or intake portion of the device has been designed to not impede the electrosurgical function of the pencil and provides intake of the surgical smoke for filtering." (Section 9)
    BiocompatibilityPatient-contacting materials should be biocompatible, meeting ISO 10993 standards."Complies with ISO 10993 Yes." (Table in Section 8)
    Electrical Safety and EMCThe device must meet electrical safety and electromagnetic compatibility (EMC) standards, specifically IEC 60601-1 and IEC 60601-2-2."Complies with IEC 60601-1 Yes." and "Complies with IEC 60601-2-2 Yes." (Table in Section 8) "The verification and validation testing of the device electrical safety and EMC testing of the device was found to acceptable and supports the claims of substantial equivalence." (Section 11)
    Material EquivalenceMaterials used for key components (handpiece housing, electrode, suction sleeve) should be the same or sufficiently similar to those used in predicate devices."The LiNA SafeAir Smoke Pencil and all predicate devices use polystyrene for the handpiece housing. The SafeAir Smoke Pencil and all the predicate devices use stainless steel for the electrode material. The predicate devices use either polycarbonate or polystyrene, as with the SafeAir Smoke Pencil, for the suction sleeve." (Section 10)
    SterilizationThe device must be provided sterile and maintain sterility until use. The sterilization method should be validated."Sterile Processing: Ethylene Oxide" (Table in Section 8)

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document describes non-clinical testing on Simulated Tissue. The specific sample size for this testing is not provided.
      • The testing was performed by "LiNA Medical ApS" (the sponsor), which is a Danish company, suggesting the data provenance is likely European, possibly Danish, although no explicit country of origin is stated for the test data itself.
      • The study is a prospective test in the sense that the testing was conducted specifically for this 510(k) submission to demonstrate performance of the new device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the testing was non-clinical and conducted on simulated tissue. There was no ground truth that needed to be established by human experts in the context of diagnostic interpretation. The ground truth was based on physical measurements of electrosurgical effect and smoke evacuation.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This question is not applicable as the testing was non-clinical and did not involve human interpretation or adjudication of results in the traditional sense. Performance was evaluated based on objective measurements and compliance with engineering standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This device is a surgical tool, not an AI-assisted diagnostic device, so such a study would not be relevant.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, standalone performance testing was done. The document states: "The LiNA SafeAir Smoke Pencil has been shown to perform as intended with the equivalent results as the predicate devices through simulated tissue cauterization and cutting functions... The SafeAir Smoke Pencil passed all testing and results were consistent with the intended use of the device." (Section 9 & 10) This refers to the physical and functional performance of the device itself, without human-in-the-loop performance in the sense of a diagnostic interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the non-clinical performance testing was based on engineering and performance metrics for electrosurgical cutting and coagulation effectiveness, smoke evacuation capability (e.g., airflow, smoke capture), and compliance with international standards (ISO 10993, IEC 60601-1, IEC 60601-2-2). It's essentially "physical proof" that the device performs its intended functions and meets safety standards, rather than a diagnostic ground truth.
    7. The sample size for the training set:

      • Not applicable. This device does not involve AI or machine learning algorithms that would require a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set for an algorithm, there is no ground truth for a training set to be established.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1