510(k) Data Aggregation

    K Number
    K140392

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-13

    (180 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103277

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use the 3M Attest™ Rapid Readout Biological Indicator 1295 in conjunction with the 3M Attest™ Auto-reader 490H as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in STERRAD® NX and 100NX systems.

    Device Description

    The 3M Attest™ 1295 Rapid Readout Biological Indicator for Vaporized Hydrogen Peroxide Sterilization (pink cap, referred to hereinafter as the 1295 BI) is a self-contained biological indicator specifically designed for rapid and reliable routine monitoring of STERRAD® vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3M Attest™ Auto-reader 490H (hereinafter referred to as the 490H Auto-reader). The 1295 BI is a single-use device.

    The Attest™ 1295 BI is composed of a polycarbonate sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. A chemical process indicator printed with stripes which change from blue to pink upon exposure to vaporized hydrogen peroxide is located on the top of the cap.

    The 1295 BI utilizes the same fundamental technology that exists in current 3M Attest™ Rapid Readout and Super Rapid Readout BIs. All Attest™ Rapid Readout and Super Rapid Readout BIs for steam utilize the a-glucosidase enzyme system, which is generated naturally within growing Geobacillus stearothermophilus organisms. The a-glucosidase enzyme in its active state is detected by measuring the fluorescence produced by the enzymatic hydrolysis of a non-fluorescent substrate. The resultant fluorescent by-product is detected in the 3M Attest™ 490H Auto-reader. The detection of fluorescence upon incubation of the 1295 BI in the 490H Auto-reader indicates a sterilization failure.

    The 1295 BI is similar in design to the 3M Attest™ 1492V Super Rapid Readout Biological Indicator for Steam cleared as K121484 for dynamic-air-removal (prevacuum) steam sterilization cycles. Minor modifications were made to 1492V that resulted in the 1295 BI for STERRAD® vaporized hydrogen peroxide sterilization cycles.

    The Attest™ Auto-reader 490H is similar in design to the Attest™ Auto-reader 490 that has been cleared for use with the Attest™ 1491 Super Rapid Readout Biological Indicator under K103277 and Attest™ 1492V Super Rapid Readout Biological Indicator under K121484.

    The Attest™ Auto-reader 490H is designed to incubate at 60°C and automatically read the Attest™ 1295 BI for a fluorescent result within 4 hours. The fluorescent readout at 4 hours met the FDA's requirement of > 97% alignment with the result after the conventional incubation time of 7 days. The Auto-reader 490H is also designed to allow further incubation of the 1295 BI for an optional visual pH color change result of the growth media at 7 days.

    AI/ML Overview

    The provided document describes the 3M Attest™ Rapid Readout Biological Indicator 1295 (BI) and 3M Attest™ Auto-reader 490H. The primary study presented is a substantial equivalence comparison to predicate devices, rather than a direct study demonstrating absolute performance against highly specific, quantitative acceptance criteria for all aspects of a typical diagnostic device. However, some performance characteristics are outlined, particularly in comparison to the predicate.

    Here's an analysis based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for the new device as a set of a priori thresholds. Instead, it demonstrates substantial equivalence by comparing its characteristics and performance to legally marketed predicate devices, and against the FDA's Guidance for Industry and FDA Staff, Biological Indicator (BI) Premarket Notification [510(k)] Submissions.

    The table below summarizes the key performance characteristics mentioned for the Attest™ 1295 BI compared to its "Intended Use Predicate," the STERRAD® CycleSure® 24 BI (K123017).

    CharacteristicAcceptance Criteria (Predicate Performance / Guidance)Reported Device Performance (Attest™ 1295 BI)
    Intended UseMonitoring vaporized hydrogen peroxide sterilization processes (e.g., STERRAD® NX, 100NX).Same as predicate.
    Organism (Species, Strain)Geobacillus stearothermophilus traceable to ATCC™ 7953.Same as predicate.
    Viable Spore Population≥ 10^6^Same as predicate.
    D-value (Resistance)D_2.5 mg/L_ ≥ 1 second (for predicate at 2.5 mg/L H2O2)D_10 mg/L_ ≥ 1 second (at 10 mg/L H2O2).
    Survival Time≥ 5 seconds≥ 5 seconds
    Kill Time= 60 seconds (for predicate at 2.5 mg/L H2O2)= 7 minutes (at 10 mg/L H2O2) - Note: Much longer due to higher challenge.
    Reduced Incubation Time (Fluorescent Result)>97% correlation to 7-day visual pH color change (for predicate: 24 hours visual pH).Fluorescent result in 4 hours (>97% correlation to 7-day visual pH color change not explicitly stated for 1295, but inferred by meeting FDA guidance).
    Chemical Process IndicatorChanges from red to golden yellow or bronze.Changes from blue to pink.
    Incubation Temperature55-60°C60 +/- 2 °C
    Shelf-life6 months6 months (with plans for extension).

    Note for Auto-reader 490H: The document states the fluorescent readout at 4 hours met the FDA's requirement of > 97% alignment with the result after the conventional incubation time of 7 days. This is a key "acceptance criterion" for the rapid readout functionality, implying the accuracy of the 4-hour fluorescent result vs. the traditional 7-day visual result.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document states, "Multiple lots of Attest™ 1295 BIs were evaluated for performance." However, it does not provide a specific number or breakdown of the sample size (e.g., number of BIs, number of sterilization cycles, replicates) used for the resistance testing or reduced incubation time validation.
    • Data Provenance: Not explicitly stated. Given it's a 510(k) submission, the studies would likely be conducted in a controlled laboratory setting (e.g., 3M's R&D facilities) in the country of origin of the manufacturer (USA, St. Paul, MN). The data would be prospective as it involves controlled testing of the new device's performance against defined parameters.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • This type of device (biological indicator) relies on a biological process (spore inactivation) and a chemical reaction for ground truth, rather than expert review of images or clinical assessments.
    • Ground Truth Establishment: The ground truth for the BI performance studies is based on established microbiological methods for assessing spore viability (e.g., lack of growth after incubation for positive control, confirmed kill for negative control) and comparison to the predicate device's known performance characteristics. No human "experts" in the sense of clinicians or radiologists are involved in establishing the ground truth for this device's function.

    4. Adjudication Method

    • Not applicable in the conventional sense. The "ground truth" for a biological indicator is determined by the growth or non-growth of spores, and the optical detection of fluorescence by the auto-reader. This is a direct, objective measurement, not subject to human interpretation or adjudication in the study itself.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human reader performance is being evaluated, often with or without AI assistance. The 3M Attest system is a standalone biological sterilization indicator and auto-reader, not a diagnostic device involving human interpretation of complex medical cases.

    6. Standalone Performance Study

    • Yes, a standalone performance evaluation of the algorithm (or device's intrinsic function) was performed.
      • The "Reduced Incubation Time" validation demonstrates standalone performance, showing the 4-hour fluorescent readout reliably predicts the 7-day visual pH color change. This is the core "algorithm" of the rapid readout auto-reader.
      • The resistance characteristics (D-value, Survival/Kill Time) are also standalone performance metrics for the BI itself.
      • The comparison to the predicate device is also a form of standalone evaluation against a known standard.

    7. Type of Ground Truth Used

    • The ground truth used for performance evaluation is microbiological growth/non-growth of spores (indicating sterilization efficacy), and chemical pH change in growth media (for the traditional 7-day visual result).
    • For the rapid readout function, the fluorescent signal's correlation to the 7-day visual pH change is the key ground truth.

    8. Sample Size for the Training Set

    • The document does not explicitly mention a training set in the context of an AI/ML algorithm development. This device uses a physiological detection mechanism (enzyme activity leading to fluorescence) rather than a machine learning model that would typically require a training set. The "development of a test cycle" and "evaluating multiple lots" suggests R&D and validation efforts, but not a distinct training set for an AI model.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as a conventional AI/ML training set is not described or implied for this device. The development process would involve standard microbiology and engineering principles to ensure the BI and auto-reader accurately detect spore viability and fluorescence, respectively.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123546

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-18

    (119 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103277, K121484

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 3M Attest™ 390 Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicators for Steam, 1291 and 1292, at 60°C for a final fluorescent result at 1 hour for 1291 and 3 hours for 1292.

    The 3M Attest™ 390G Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicator for Ethylene Oxide, 1294, at 37°C for a final fluorescent result at 4 hours.

    Device Description

    The Attest™ 390 and 390G Auto-readers are used in conjunction with Attest™ Rapid Readout Biological Indicators by trained personnel in central sterilization departments to monitor sterilization processes. The 390 and 390G Auto-readers, like its predecessors the Attest™ 290, 290G and 490 Auto-readers, have two functions. First, the devices serve to incubate the biological indicator (BI) at the temperature appropriate to the organism. Second, the devices determine whether a potential sterilization failure may have occurred by measuring the intensity of fluorescence produced by an enzymatic reaction associated with spore growth inside the selfcontained biological indicator.

    The 390 and 390G Auto-readers include a Crusher Well in which the Attest™ Rapid Readout biological indicators are activated. After activation, the user places the biological indicator into one of the 10 color-coded Incubation Wells. The 390 and 390G Auto-readers have an incubation block controlled to the appropriate temperature for their respective biological indicators. The Auto-readers contain a LCD panel that shows the time remaining for incubation. Fluorescence detection at each incubation well is accomplished by the use of a UV LED and a photo diode sensor. As soon as a positive result is detected, a '+' sign is shown as an indication of sterilization cycle failure, accompanied by an audible alarm. At the end of the specified incubation time, if a negative result is detected, a '-' sign is shown on the LCD panel as indication of an adequate sterilization cycle. When the test is complete, as indicated by the (+) or (-) symbol on the Auto-reader LCD panel under the incubating BI, the customer records the result and disposes of the biological indicator.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Auto-readerTestingAcceptance CriteriaResult
    390Qualification testing with Attest™ 1291 BIMeets FDA's requirements for Reduced Incubation Time with > 97% alignment with the conventional incubation time of 7 days for the following readout time: • Fluorescent result in 1 hour. Survival/Kill: All BIs survive after exposure at the survival time and all BIs are killed after exposure at the kill time.Pass
    390Qualification testing with Attest™ 1292 BIMeets FDA's requirements for Reduced Incubation Time with > 97% alignment with the conventional incubation time of 7 days for the following readout time: • Fluorescent result in 3 hours. Survival/Kill: All BIs survive after exposure at the survival time and all BIs are killed after exposure at the kill time.Pass
    390Auto-reader Maintenance of Incubation TemperatureMaintain 60+/- 2ºC over the specified incubation period.Pass
    390GQualification testing with Attest™ 1294 BIMeets FDA's requirements for Reduced Incubation Time with > 97% alignment with the conventional incubation time of 7 days for the following readout time: • Fluorescent result in 4 hours. Survival/Kill: All BIs survive after exposure at the survival time and all BIs are killed after exposure at the kill time.Pass
    390GAuto-reader Maintenance of Incubation TemperatureMaintain 37+/- 2ºC over the specified incubation period.Pass
    BothElectrical Safety (IEC 61010-1, IEC 61010-2-010)Compliance to specified international safety standards.Pass
    BothElectromagnetic Compatibility (USA Title 47, FCC Part 15, IEC 61326)Compliance to specified electromagnetic compatibility standards for radiated emissions, conducted emissions, and general EMC requirements.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the exact sample size (number of biological indicators or test runs) used for the qualification testing. It only mentions "Multiple lots of 3M Attest™ Rapid Readout biological indicators were evaluated for performance."
    • Data Provenance: The document does not specify the country of origin of the data. It is a premarket notification for submission to the FDA, suggesting the evaluation was conducted for US regulatory purposes. The data is presented as retrospective in the sense that the testing was completed prior to the submission for regulatory approval.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The concept of "experts" and "ground truth" in the typical AI/medical device context (e.g., radiologists interpreting images) does not directly apply here. The device being evaluated is an auto-reader for biological indicators (BIs) used to monitor sterilization processes.

    • Ground Truth Establishment: The ground truth for the biological indicator performance testing is established by comparing the rapid readout results (fluorescent result at 1, 3, or 4 hours) to the "conventional incubation time of 7 days." This conventional incubation time is the established method for determining sterility failure or success, and the biological indicators themselves are a well-defined standard for monitoring sterilization.
    • Experts: No external human experts are mentioned as establishing this ground truth. The "ground truth" for BI performance is inherent in the established standard for biological indicator readout over 7 days.

    4. Adjudication Method for the Test Set:

    Not applicable in the traditional sense. The "adjudication" (comparison) is automated: the device's rapid fluorescent result is compared to the established 7-day conventional incubation result for the biological indicators. The acceptance criteria specify a "> 97% alignment" for this comparison.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices where human interpretation directly impacts the outcome (e.g., diagnostic imaging). This device is an auto-reader for biological indicators, where the primary function is automated detection of fluorescence and temperature maintenance. There's no human "reader" in the loop for the primary function being tested for performance against the acceptance criteria.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Yes, a standalone study was performed. The entire evaluation described is of the automated device (the 3M Attest™ 390 and 390G Auto-readers) in conjunction with its respective biological indicators. The device functions automatically to incubate, read fluorescence, and provide a positive/negative result without human intervention in the interpretation of the BI's sterility status.

    7. Type of Ground Truth Used:

    The ground truth used is based on established standard biological indicator performance (conventional 7-day incubation results) and physical measurement standards (temperature maintenance, electrical safety, EMC). Specifically, for BI qualification, it is a comparison against the long-standing 7-day incubation period result, which serves as the "gold standard" for determining bacterial growth in the BI.

    8. Sample Size for the Training Set:

    The document does not provide any information regarding a "training set" or its sample size. This is expected because the device likely operates based on a fixed algorithm for fluorescence detection and temperature control, rather than a machine learning model that would require a dedicated training phase. The device's performance is validated against established biological indicator principles.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As no training set is mentioned or implied for a machine learning model, there is no ground truth established for a training set. The device follows pre-defined detection parameters and physical controls.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121484

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-19

    (154 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K103277

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use the 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V in conjunction with the 3M™ Attest™ Auto-reader 490 to qualify or monitor dynamic-air-removal (prevacuum) steam sterilization cycles of 4 minutes at 270°F (132°C) and 3 minutes at 275°F (135°C).

    The 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V provides a final fluorescent result in 1 hour. An optional visual pH color change result is observed in 48 hours.

    Device Description

    The 3M Attest™ 1492V Super Rapid Readout Biological Indicator (SRBI) is a selfcontained biological indicator designed to be used with the 3M Attest™ 490 Auto-reader to qualify or routinely challenge dynamic-air-removal (prevacuum) steam sterilization cycles at 270°F (132°C) and at 275°F (135°C).

    The Attest™ 1492V SRBI is composed of a polycarbonate sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. On each 1492V SRBI cap is a chemical process indicator that changes color from pink to light brown when exposed to steam.

    The 1,492V SRBI is a further improvement over the current 3M Attest™ Rapid Readout technology. Both the Attest™ Rapid Readout BIs and the Attest™ Super Rapid Readout BIs utilize the a-glucosidase enzyme system, which is generated naturally within growing G. stearothermophilus organisms. The a-glucosidase enzyme in its active state is detected by measuring the fluorescence produced by the enzymatic hydrolysis of a non-fluorescent substrate. The resultant fluorescent by-product is detected in the 3M Attest™ 490 Autoreader. The detection of fluorescence upon incubation of the 1492V SRBI in the 490 Autoreader indicates a steam sterilization failure.

    The 1492V SRBI is similar in design to the 3M Attest™ 1491 Super Rapid Readout Biological Indicator for Steam cleared as K 103277 for gravity displacement steam sterilization cvcles. Minor modifications were made to 1491 that resulted in the 1492V SRBI for dynamic-air-removal (prevacuum) steam sterilization cycles.

    The Attest™ 490 Auto-reader has been cleared for use with the Attest™ 1491 Super Rapid Readout Biological Indicator for Steam under K103277. The current submission extends the use of the 490 Auto-reader to the 1492V SRBI.

    The Attest™ 490 Auto-reader is designed to incubate at 56°C and automatically read the Attest™ 1492V SRBI for a fluorescent result within 1 hour. The 490 Auto-reader is also designed to allow further incubation of the 1492V SRBI for an optional visual pH color change of the growth media at 48 hours. Both the fluorescent readout at 1 hour and the optional visual readout at 48 hours met the FDA's requirement of > 97% alignment with the result after the conventional incubation time of 7 days.

    AI/ML Overview

    1. Acceptance Criteria and Device Performance

    Biological Indicator TestAcceptance CriteriaReported Device Performance
    Characterization of spores> 90% Genetic similarity to Geobacillus stearothermophilus ATCC™ 7953Pass
    D-ValueGreater than or equal to 10 seconds at 132°CGreater than or equal to 8 seconds at 135°CPass
    Population (Total Viable Spore Count)Greater than or equal to $10^6$ sporesPass
    Survival/Kill TimesSurvival Time = Calculated survival time* or 1 minute at 132°C and 40 seconds at 135°C, whichever is longer;Kill time = Calculated kill time* at 132°C and at 135°C*ANSI/AAMI/ISO 11138-1:2006/(R) 2010, Annex EPass
    Reduced Incubation TimeMeets FDA's requirements for Reduced Incubation Time with > 97% alignment with the conventional incubation time of 7 days for the following readout times:Fluorescent result in 1 hourOptional visual pH color change at 48 hoursPass
    Hold Time AssessmentD-value does not change when activated 7 days post-sterilizationPass
    Component Inhibition StudiesComponents have no impact on the recovery of 10-100 organismsPass
    Chemical Process IndicatorChemical Process Indicator on the BI changes from pink to light brown upon exposure to steamPass
    Auto-reader Maintenance of Incubation TemperatureMaintain 56+/- 2°C over a period of 7 daysPass

    2. Sample Size and Data Provenance

    The document states that "Multiple lots of 3M Attest™ 1492V SRBIs were evaluated for performance" for the nonclinical testing. However, specific sample sizes for each test in the acceptance criteria, or the country of origin of the data, or whether it was retrospective or prospective, are not explicitly provided in the given text.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for the test set. The ground truth for the biological indicator's performance is established through adherence to recognized standards for sterilization biological indicators (e.g., D-value, population count, survival/kill times).

    4. Adjudication Method

    No adjudication method is mentioned, as the studies are focused on direct biological and physical performance testing against established standards, not human interpretation or consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was conducted or mentioned. This type of study is typically relevant for interpretative devices where human variability in reading is a factor, which is not the case for a biological indicator and auto-reader system designed for automated readout.

    6. Standalone Performance Study

    Yes, standalone performance studies were conducted. The entire nonclinical testing section, including the table of acceptance criteria and results, represents the standalone performance of the 3M Attest™ 1492V Super Rapid Readout Biological Indicator (SRBI) when used with the 3M Attest™ 490 Auto-reader. These tests evaluate the device's intrinsic characteristics and functionality against predefined standards and guidance documents.

    7. Type of Ground Truth Used

    The ground truth used for the tests is based on:

    • Scientific and Biological Standards: For tests like "Characterization of spores," "D-Value," "Population (Total Viable Spore Count)," and "Survival/Kill Times," the ground truth is derived from established microbiological principles, FDA guidance, and international standards (e.g., ANSI/AAMI/ISO 11138 series, USP chapters). These standards define the expected biological behavior and performance characteristics of a biological indicator.
    • Engineering and Physical Standards: For tests like "Auto-reader Maintenance of Incubation Temperature," the ground truth is based on engineering specifications and calibration standards.
    • Regulatory Requirements: For "Reduced Incubation Time," the ground truth is adherence to FDA's requirement for >97% alignment with conventional 7-day incubation.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of an algorithm or AI. Biological indicators and auto-readers are typically validated through performance testing against established physical and biological standards, not by training an algorithm on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set for an algorithm in this document describing a physical biological indicator and an auto-reader.