510(k) Data Aggregation

    K Number
    K220634

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-06-30

    (118 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    P810046/S226, P810046/S232, K103153, K141236

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NC TREK NEO™ Coronary Dilatation Catheters are indicated for:
    a) Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion
    b) Balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with STsegment elevation myocardial infarction
    c) Balloon dilatation of a stent after implantation (balloon models 2.00 mm - 5.00 mm only)

    Device Description

    NC TREK NEO is a Non-Compliant (NC), Rapid Exchange (RX), Coronary Dilatation Catheter (CDC). The catheter is a rapid exchange co-axial design with a balloon at the distal tip. The proximal lumen provides for inflation of the balloon with contrast medium. The central distal lumen permits the guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outside diameter (OD) of the distal end of the device, including the distal shaft, tip, and the balloon are coated with Hydrophilic Coating.
    Radiopaque markers are positioned on the inner member underneath the balloon to provide accurate positioning of the balloon in the artery. The balloon is designed to provide an expandable segment of known diameter and length at specified pressures. Proximal shaft markers located on the outer shaft aid in gauging the dilatation catheter position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. The Single arm on proximal end of the device provides access to the inflation lumen. It is designed with a Luer-lock fitting to facilitate connection to an inflation device.
    The NC TREK NEO™ Coronary Dilatation Catheter will be available in select combinations of 1.50-6.00mm diameters and 6-25mm lengths.

    AI/ML Overview

    This document concerns the 510(k) premarket notification for the NC TREK NEO™ Coronary Dilatation Catheter (K220634). The information provided does not contain details about acceptance criteria, device performance, and study design typically found in reports for AI/ML-driven medical devices. Instead, it describes a conventional medical device (a PTCA catheter) and establishes substantial equivalence to a predicate device through a comparison of design, materials, manufacturing, sterilization, and non-clinical performance data.

    Therefore, many of the requested fields cannot be filled as they are explicitly related to studies and acceptance criteria for AI/ML devices, which are not applicable here.

    Here's a breakdown of the available information and what cannot be provided:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various performance tests conducted on the device. However, it does not provide a table with specific quantitative acceptance criteria or the reported device performance values for these tests. It only states that the tests supported a determination of substantial equivalence and raised no new safety or performance issues.

    Performance Tests Completed (no specific acceptance criteria or reported values provided in this document):

    • Tip dimensions
    • Crossing Profile
    • Guidewire Lumen
    • Shaft Dimensions
    • Proximal Shaft Marker Location
    • Catheter Length
    • Catheter Preparation
    • Balloon Inflation/Deflation Time
    • Balloon Rated Burst Pressure (RBP)
    • Maximum Compliance Label
    • Balloon Compliance
    • Balloon RBP (In-Stent)
    • Proximal Seal Tensile Strength
    • Outer Member to Hypotube Seal Tensile Strength (Notch-Seal Tensile Strength)
    • Proximal Adaption Tensile Strength
    • Catheter Tip Tensile Strength
    • Inner Member Lumen Collapse
    • Balloon Fatigue Resistance
    • Balloon Fatigue Resistance (In-Stent)
    • Hydrophilic Coating- Dry Adhesion of Coating
    • Hydrophilic Coating- Coefficient of Friction
    • Kink/Flex
    • Torque
    • Particulates
    • Finished Good Sheath Removal
    • Balloon Preparation, Deployment and Retraction

    Biocompatibility and Chemical Characterization Tests Completed:

    • Cytotoxicity
    • Sensitization
    • Irritation
    • Acute Systemic & Pyrogenicity
    • Hemocompatibility

    2. Sample size used for the test set and the data provenance:
    Not applicable. This is a conventional medical device, not an AI/ML device relying on a test set of data. The "test set" here refers to the physical devices manufactured for bench and biocompatibility testing. The exact number of devices or components used for each test is not specified in this summary, nor is data provenance in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. Ground truth and expert consensus are concepts for AI/ML device validation, which is not relevant for this conventional catheter.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable for this type of device and study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    Not applicable. The "truth" for this device's performance is determined by established engineering standards, material properties, and biological compatibility testing, not by expert consensus on clinical data or pathology in the context of an AI/ML algorithm.

    8. The sample size for the training set:
    Not applicable. This device does not involve a training set for an AI/ML algorithm.

    9. How the ground truth for the training set was established:
    Not applicable. This device does not involve a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160961

    Validate with FDA (Live)

    Date Cleared
    2016-07-30

    (116 days)

    Product Code
    Regulation Number
    870.5100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103153, K110133, K131169

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mozec(TM) NC - Rx PTCA Balloon Dilatation Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Mozec(TM) NC - Rx PTCA Balloon Dilatation Catheter is also indicated for post-delivery expansion of balloon expandable stents.

    Device Description

    Mozec™ NC - Rx PTCA Balloon Dilatation Catheter is a sterile, single use rapid exchange catheter consisting of a non-compliant balloon, a soft tip, a coaxial dual lumen distal shaft and a single lumen proximal shaft. Radiopaque platinum/iridium markers contained in the balloon segment facilitate fluoroscopic visualization of the proximal and distal ends of the Mozec™ NC balloon's working length. The Mozec™ NC balloons are offered in diameters ranging from 2.00mm to 4.50mm and lengths varying from 8mm to 38mm.

    AI/ML Overview

    This document describes the Mozec™ NC – Rx PTCA Balloon Dilatation Catheter. The information needed to address the user's request, specifically regarding acceptance criteria and the study that proves the device meets the acceptance criteria, is limited to the non-clinical performance data provided. There is no mention of an AI-based device or a comparative effectiveness study involving human readers. The document pertains to a medical device approval through a 510(k) submission, which focuses on demonstrating substantial equivalence to a predicate device based on non-clinical testing, rather than extensive clinical efficacy trials.

    Here's a summary of the available information based on the provided text, with many fields indicated as "Not Applicable" or "Not Provided" due to the nature of the device and the submission document:

    1. A table of acceptance criteria and the reported device performance

    The document states: "The Mozec™ NC – Rx PTCA Balloon Dilatation Catheter met all predetermined acceptance criteria as specified by applicable standards, FDA guidance documents and test protocols." However, the specific quantitative acceptance criteria and the detailed reported device performance values are not explicitly provided in this summary. The summary lists the types of tests performed, which implies these tests had corresponding acceptance criteria.

    Acceptance Criteria Category (Implied)Reported Device Performance
    Dimensional verificationMet criteria
    Balloon preparation, deployment and retractionMet criteria
    Balloon rated burst pressure and Balloon complianceMet criteria
    Balloon fatigueMet criteria
    Balloon inflation and deflation timeMet criteria
    Bond strengthMet criteria
    Flexibility and kink testMet criteria
    Balloon rated burst pressure (In stent)Met criteria
    Balloon fatigue (In Stent)Met criteria
    RadiopacityMet criteria
    Torque testMet criteria
    Performance testing on the Looper Clip accessoryMet criteria
    Biocompatibility (ISO 10993-1)Met criteria (no safety/efficacy issues)
    - In Vitro Cytotoxicity studyMet criteria
    - Skin Sensitization StudyMet criteria
    - Intracutaneous reactivity testMet criteria
    - Acute Systemic Toxicity StudyMet criteria
    - In vitro Hemolysis test StudyMet criteria
    - In vitro C3a and SC5b-9 complement activation AssayMet criteria
    - Material Mediated Pyrogen TestMet criteria
    - In vivo thromboresistanceMet criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size used for the test set: Not provided. The document mentions "performance testing," but does not specify the number of devices or components tested for each category.
    • Data provenance: Not explicitly stated, but assumed to be internal testing conducted by Meril Life Sciences Private Limited, which is based in India. The testing is non-clinical/bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This request is relevant for AI/diagnostic devices where expert ground truth is established. This document describes a PTCA Balloon Dilatation Catheter, for which ground truth is typically established through engineering and biological testing against established standards, not expert consensus on interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI-assisted diagnostic device. No MRMC study was conducted or is relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or AI device. Performance is assessed through engineering and biocompatibility testing of the physical catheter.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of ground truth: For the non-clinical performance testing, the "ground truth" refers to established engineering standards, FDA guidance documents (e.g., "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters"), and ISO standards (e.g., ISO 10993-1 for biocompatibility). The device's performance is compared against these predetermined specifications and accepted norms for safety and function.

    8. The sample size for the training set

    • Not Applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device