Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K241855

    Validate with FDA (Live)

    Date Cleared
    2025-03-26

    (272 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K073207, K103032, K083452, K080715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SB Knife Jr2 is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47702, MD-47702L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue including Zenker's diverticulum within the digestive tract using high-frequency current.

    Device Description

    SB Knife Jr2 under application (Hereinafter "the Subject Device" is a single-patient use EtOsterilized hand-held device that connects to an external high frequency (HF) generator and is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue including Zenker's diverticulum within the digestive tract using high-frequency current. Varying generator power, cycle and duration results in a control of these surgical capabilities. The Subject Device is designed to connect with Olympus and ERBE connectors and uses the disposable or reusable Olympus or ERBE grounding plates and connecting cables intended for the generators.

    The Subject Device is not life supporting or life sustaining. It is designed as a short-term limited contact device (less than 24 hours), has no software, and has been certified to comply with IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-1-6 and IEC 60601 2-18 by ISO 17025 accredited laboratory.

    The Subject Device consists of two models as follows:

    MD-47702 Working Length: 1970 mm
    MD-47702L Working Length: 2320 mm

    AI/ML Overview

    The provided text is a 510(k) summary for the SB Knife Jr2, a medical device. It does not contain the information typically found in a study proving a device meets specific acceptance criteria based on performance.

    Specifically, the text describes:

    • The device and its intended use.
    • A comparison to a predicate device (SB Knife®).
    • Differences between the subject device and the predicate device, particularly regarding type, shape, structure, presence of a rotating operation portion, and materials.
    • Non-clinical testing performed (transportation, shelf-life, packaging, sterilization, electromagnetic compatibility, electrical safety, biocompatibility).
    • A statement that clinical testing was not performed.

    Therefore, I cannot extract the requested information about acceptance criteria and a study proving the device meets them from the provided text. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device through technological comparison and non-clinical performance, rather than providing a detailed study of the device's performance against pre-defined acceptance criteria for a new clinical claim or feature that requires such a study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153055

    Validate with FDA (Live)

    Date Cleared
    2016-07-11

    (265 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K030194, K100275, K103032

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Electrosurgical Resection and Vaporization Electrodes Series is a single use bipolar electrode series designed and intended for use in endoscopic urological procedures involving the resection, ablation or removal of soft tissue and coagulation where hemostasis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedures for which the devices can be used are transurethral resection in saline (TURis). transurethral prostatectomy (TURP), transurethral resection of the prostate (TURP) for benign prostatic hyperplasia, transurethral incision of the prostate (TUIP) or bladder neck, transurethral resection of bladder turnors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment (0.9% Sodium Chloride). These devices are not intended to be used to treat cancer of the prostate.

    The ACE Electrosurgical Resection and Vaporization Electrode Series are single use bipolar electrodes designed and intended for use in urological surgical procedures involving the vaporization, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue include use in the prostate, bladder and bladder neck. The specific treatment indications include benign prostate hyperplasia (BPH) bladder cancer, tumors, lesions, and neoplasms. The specific urological indications include transurethral electro vaporization (TUVP, TVP, TUEVP), also known as transurethral vapor resection of the prostate (TUVRP) or transurethral vaporization in saline (TUVis). These devices are intended to be used in an irrigated environment (0.9% Sodium Chloride). These devices are not intended to be used in treating cancer of the prostate.

    Device Description

    The ACE Electrosurgical Resection and Vaporization Electrodes Series consist of an active tip, PTFE color code identification, and an insulator between the electrode tube, a guiding tube, telescope clip and arm (shaft). The design and dimensions of the electrode tips vary to accommodate various procedural conditions. The active tips of the various electrodes may consist of loops or buttons. The system includes single-use electrodes that can be connected to a working element (only WA22366A – K100275), or the use of the single-use electrodes together other compatible electrosurgical cables as applicable. Compatible Generators to any of the electrodes include generators with output specifications that meet the following criteria: 100-120/220-240V, ~50-60Hz/1000VA, 310-380 kHz, 320W/200 Ω, Int 10s/30s.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called "ACE Electrosurgical Resection and Vaporization Electrodes Series." This document primarily discusses the substantial equivalence of the new device to a predicate device, rather than presenting a study that establishes acceptance criteria and then proves the device meets those criteria through performance data.

    Here's an analysis of the information, addressing your points where possible, and noting when the information is not present in this type of FDA submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of acceptance criteria or specific device performance metrics in the way you might expect from a clinical study report. Instead, it relies on demonstrating equivalence to an existing predicate device and conformance to recognized consensus standards.

    Acceptance Criteria (Implied / Stated):

    • Substantial Equivalence: The device must be substantially equivalent to a legally marketed predicate device (K120418, Olympus Winter & Ibe GmbH Electrosurgical Resection and Vaporization Electrodes Series) in terms of:
      • Indications for Use
      • Principles of Operation
      • Technological Characteristics
      • Materials
      • Performance (demonstrated through bench testing and conformance to standards)
    • Safety and Effectiveness: Does not raise new issues of safety or effectiveness.
    • Electrical Safety: Conforms to IEC 60601-1.
    • Thermal Safety: Conforms to IEC 60601-2-2.
    • Biocompatibility: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity.
    • Sterilization: Ethylene Oxide (EO) with a Sterility Assurance Level (SAL) of 10-6.
    • Risk Analysis: Performed in accordance with ISO 14971:2007.

    Reported Device Performance:

    The document states that "Design verification was performed to ensure the device functions according to its intended use, and the results met their acceptance criteria." However, it does not provide the specific numerical results of this design verification. It only confirms that the testing was performed and met the criteria.

    The "Comparison of Characteristics" table (Table 5A) highlights the similarities between the applicant's device and the predicate device across various attributes. This comparison serves as the performance data in terms of demonstrating equivalence.

    CharacteristicACE Electrosurgical Resection and Vaporization Electrodes Series (Applicant Device)Electrosurgical Resection and Vaporization Electrodes Series (Predicate Device, K120418)
    Indications for UseEndoscopic urological surgical procedures involving resection, ablation, removal of soft tissue, and coagulation for hemostasis; specific urological indications (prostate, bladder, bladder neck) including TURis, TURP (for BPH), TUIP, TURBT, cystodiathermy; uses vaporization, ablation, cutting, removal of soft tissue and coagulation for hemostasis; specific soft tissue indications (prostate, bladder, bladder neck) including BPH, bladder cancer, tumors, lesions, neoplasms; TUVP/TVP/TUEVP/TUVRP/TUVis; irrigated environment (0.9% Sodium Chloride); not for treating prostate cancer.HF-Resection Electrodes: Similar indications including resection, ablation, removal for hemostasis; prostate, bladder, bladder neck; TURis, TURP (for BPH), TUIP, TURBT, cystodiathermy; irrigated environment; not for prostate cancer. HF-Resection Electrode for Plasma Vaporization: Similar indications including vaporization, ablation, coagulation, cutting, removal for hemostasis; prostate, bladder, bladder neck; BPH, bladder cancer, tumors, lesions, neoplasms; TUVP/TVP/TUEVP/TUVRP/TUVis; irrigated environment; not for prostate cancer.
    Principles of OperationUses energy for cutting and removal of tissue during urologic procedures; uses bipolar energy to perform electrocautery.Uses energy for cutting and removal of tissue during urologic procedures; uses bipolar energy to perform electrocautery.
    Energy TypeBipolar electrocauteryBipolar electrocautery
    MaterialsTungsten, Tungsten alloy, Aluminum Oxide, Stainless Steel 304, Polytetrafluoroethylene (PTFE)Tungsten, Tungsten alloy, Aluminum Oxide, Stainless Steel 304, Polytetrafluoroethylene (PTFE)
    Biocompatibility TestingCytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, ImplantationCytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity
    SterilitySterile packed, Ethylene Oxide (EO), SAL 10-6Sterile packed, Ethylene Oxide (EO), SAL 10-6
    Overall DesignActive tip, PTFE color code identification, insulator between electrode tube, guiding tube, telescope clip, and arm (shaft).Active tip, PTFE color code identification, insulator between electrode and electrode tube, guiding tube, telescope clip, and arm (shaft).
    Electrode TipsLoops, bands, needles, or buttonsLoops, bands, rollers, needles, or buttons
    Single-UseElectrodes are intended for single use only.Electrodes are intended for single use only. (HF bipolar Cable is reusable).
    Shelf Life3 years3 years
    Electrical SafetyConforms to IEC 60601-1Conforms to IEC 60601-1
    Thermal SafetyConforms to IEC 60601-2-2Conforms to IEC 60601-2-2

    2. Sample size used for the test set and the data provenance

    This document describes a 510(k) submission, which relies heavily on non-clinical (bench) testing and comparison to a predicate device, rather than a clinical trial with a "test set" of patients.

    • Sample Size for Test Set: Not applicable in the context of human subjects or "test sets" as you might find in an AI/imaging study. The "test set" in this context refers to the samples used in bench testing (e.g., electrical safety tests, biocompatibility tests), the number of which is not specified in this summary.
    • Data Provenance: Not applicable in terms of country of origin of patient data or retrospective/prospective studies, as no human clinical data is presented here to support the device's safety and effectiveness. The data provenance discussed relates to the manufacturer (Ace Medical Devices Pvt. Ltd., India) and the predicate device manufacturer (Olympus Winter & Ibe GmbH).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not a study involving expert review for establishing ground truth from patient data. The "ground truth" for this device's safety and effectiveness is established by its similarity to a device already on the market and its conformance to recognized engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no "test set" in the context of human data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrosurgical tool, not an AI-assisted diagnostic or therapeutic device that would involve human readers or AI assistance in that way.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable for the same reasons as above. This is an electrosurgical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the substantial equivalence claim is multi-faceted:

    • Predicate Device Performance: The primary ground truth is that the predicate device (Olympus K120418) has a history of safe and effective use in the market.
    • Consensus Standards: Adherence to recognized consensus standards like IEC 60601-1 (electrical safety) and IEC 60601-2-2 (thermal safety), and ISO 14971:2007 (risk management).
    • Bench Testing: Internal design verification testing results (not detailed here) that confirmed the device functions as intended and met acceptance criteria for engineering specifications.
    • Biocompatibility Testing: Results indicating the materials are biocompatible.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning study.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120418

    Validate with FDA (Live)

    Date Cleared
    2012-06-20

    (131 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K103032, K030194

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HF-Resection Electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation or removal of soft tissue and where hemostasis is required. The specific urological indications include use in the prostate, bladder and bladder neck. The procedures for which the devices can be used for are transurethral resection in saline (TURis), transurethral prostatectorny, transurethral resection of the prostate (TURP) for benign prostatic hypertplasia, transurethral incision of the prostate (TUIP) or bladder neck, transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used for treating cancer of the prostate.

    The HF-Resection Electrode for Plasma Vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving the vaporization, ablation, coagulation, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include use in the prostate, bladder neck. The specific treatment indications include benign prostate hyperplasia BPH, bladder cancer, tumors, The specific urological indications include transurethral and neoplasms. lesions electrovaporization (TUVP, TVP, TUEVP), also known as transurethral vapor resection of the prostate (TUVRP) or transurethral vaporization in saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.

    Device Description

    The HF-Resection Electrodes and HF Resection Button Electrode for Plasma Vaporization consist of an active tip, PTFE color code identification, an insulator between the electrode and electrode tube, a guiding tube, telescope clip and arm (shaft).

    The design and dimensions of the electrodes vary to accommodate various procedural conditions. The active tips of the various electrodes may consist of loops, bands, rollers, needles or buttons.

    The system includes Electrodes, working elements, and an HF bipolar cable that can be connected to an electrical surgical unit.

    HF Resection Electrodes (K100275) can now be used with the newly compatible electrosurgical generator, the ESG-400 (K103032). For the HF Resection Electrodes to be used with the ESG-400, an additional HF cable is needed (WA00014A). The WA00014A bipolar reusable cable is an electrosurgical accessory designed to transfer electrosurgical power to electrosurgical working elements from the electrosurgical generator ESG-400 (K103032). The cable is designed to connect the working elements WA22366A and WA22367A (K100275).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a novel clinical study with acceptance criteria in the typical sense of evaluating a new therapeutic or diagnostic claim. The "study" here is a set of performance tests to ensure the new device (specifically, a new cable) does not raise new safety concerns and maintains the same functionality as the predicate device.

    Here's a breakdown based on the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Functional EquivalenceDevice functions according to its intended use (as per predicate device).Design verification performed; results met acceptance criteria.
    Electrical SafetyNo new concerns regarding electrical safety when used with the new cable and electrosurgical generator.Bench testing according to FDA-recognized consensus standards IEC 60601-1 and IEC 60601-2-2 demonstrated no new electrical safety concerns.
    Risk ManagementRisks are identified, assessed, and controlled to an acceptable level.Risk analysis carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size" in the context of a clinical trial or a dataset for an AI model. This is a submission for a physical medical device (electrosurgical electrodes and a new cable) demonstrating equivalence, not a data-driven AI device.

    • Sample Size (Test Set): Not applicable in the context of a dataset; refers to the testing of the physical device components. The document implies sufficient units were tested to perform design verification, electrical safety tests, and risk analysis.
    • Data Provenance: Not applicable in the context of clinical data. The tests performed are bench tests and risk analyses conducted by the manufacturer, Olympus Winter & Ibe GmbH, in Germany.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. "Ground truth" in this context would refer to the established engineering and safety standards (e.g., IEC standards, ISO 14971) against which the device's performance was measured. These standards are developed by expert committees, but the specific individuals involved in this particular submission's verification of adherence to those standards are not listed.

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of expert adjudication or consensus methods for evaluating the results of the bench tests or risk analysis. The results are compared directly against established engineering and safety standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices (e.g., imaging AI) to compare human reader performance with and without AI assistance. This submission is for surgical instruments, not a diagnostic or AI-assisted device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This is not an AI-driven device. The "performance" refers to the physical and electrical characteristics of the surgical electrodes and cable.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission consists of:

    • Established engineering and performance specifications: The device is expected to function according to its intended use, which is defined by its design.
    • International Consensus Standards:
      • IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
      • IEC 60601-2-2: Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
      • ISO 14971:2007: Medical devices – Application of risk management to medical devices.

    These standards serve as the objective benchmarks against which the device's safety and performance were evaluated.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1