    K Number
    K113788

    Validate with FDA (Live)

    Device Name
    DEFLEX ACRILATO
    Manufacturer
    Date Cleared
    2012-03-16

    (85 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K053060, K102874

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is indicated for the manufacturing of bases of partial or full removable dentures, r nis products thatoated for the marketing and pneumatic injection of the material, following the Indications of Use that come with the product.

    Device Description

    Deflex Acrilato is a pigmented polimethyl-methacrilate (PMMA) resin that belongs to the group of thermoplastic polymers which is used for the manufacturing of partial, complete and combined dentures. The product is presented in pellets, packed in an aluminum canister (as a primary packaging), and then in a vacuum sealed aluminized plastic bag (as a secondary packaging). the PMMA granules are pigmented so as to achieve aesthetic effects on the manufactured denture.

    AI/ML Overview

    The provided text describes a 510(k) summary for a dental device (DEFLEX ACRILATO, a PMMA resin for denture bases). This summary focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing of material properties, rather than an AI-powered diagnostic device. Therefore, many of the requested categories related to AI performance, expert ground truth, and clinical study designs are not applicable.

    Here's the analysis based on the provided text, indicating "Not Applicable" where the information is not present or relevant to this type of device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission compares the new device to predicate devices, stating that its properties are "similar" or "more resistant" or "much lower." It does not provide specific quantitative acceptance criteria or detailed numerical performance data for these comparisons within the provided text. Instead, it relies on a qualitative comparison to establish substantial equivalence.

    TestStandard or Test Method AppliedReported Device Performance (Relative to Predicate)
    Material CharacterizationInternal laboratory methodComparable to predicate (VERACRYL)
    Mechanical PropertiesEN 20795-1:2009 (ANSI+ADA+12-2002+(R2008))More resistant than predicate (TCS unbreakable)
    Residual MMA Monomer ContentEN 20795-1:2009 (ANSI+ADA+12-2002+(R2008))Similar to predicate (VERACRYL)
    Water AbsorptionEN 20795-1:2009 (ANSI+ADA+12-2002+(R2008))Similar to predicate (VERACRYL)
    Water SolubilityEN 20795-1:2009 (ANSI+ADA+12-2002+(R2008))Much lower than predicate (VERACRYL)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify general "test sets" in the way a diagnostic AI study would. The testing was laboratory-based on material properties.

    • Sample Size: Not specified for each test.
    • Data Provenance: The device submitter is NUXEN S.R.L. based in Buenos Aires, Argentina. The testing was conducted by an "Internal laboratory" or against recognized standards (EN 20795-1:2009, ANSI+ADA+12-2002+(R2008)). The data is prospective in the sense that the manufacturer conducted these tests specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a material science study, not a diagnostic study requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. This is a material science study, not a diagnostic study requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. This is a material science study for a dental resin, not an AI-powered diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

    Not applicable. This is not an algorithm or AI device. The device itself is the product being tested (the PMMA resin). The "performance" is its inherent physical and chemical properties.

    7. The Type of Ground Truth Used

    The "ground truth" here is established by the specified standardized testing methods (EN 20795-1:2009, ANSI+ADA+12-2002+(R2008)) for material properties relevant to dental resins. The data generated from these calibrated tests represents the factual properties of the material.

    8. The Sample Size for the Training Set

    Not applicable. This is a material science product, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.