Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222292

    Validate with FDA (Live)

    Device Name
    F&P myAirvo 3
    Date Cleared
    2024-05-08

    (649 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K102465, K162553

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The myAirvo 3 is for the treatment of spontaneously breathing patients, infant to adult, who would benefit from receiving high flow warmed and humidified respiratory gases. This includes patients who have had upper airways bypassed. The flow may be from 2 - 60 L/min depending on the patient interface. The myAirvo 3 is for patients in homes and long-term care facilities.

    Device Description

    The F&P myAirvo 3 device is a heated humidifier flow source. The subject device is intended to treat spontaneously breathing patients, pediatrics and adults, who would benefit from receiving high flow, warmed and humidified entrained air and oxygen (if required).

    The F&P myAirvo 3 is comprised of two main connected functional units: the blower and the humidifier.

    The blower is a motorized fan assembly that provides air flow. The fan speed is directly related to the delivered flow and it is controlled by the software. The blower assembly output connects directly to a humidification chamber at the front of the device.

    The second functional unit of the F&P myAirvo 3 device is a heated Passover humidifier. The water is contained in a humidification chamber positioned on a heater plate at the front of the unit. The gas is warmed and humidified in the chamber to the dew point set temperature, transported through the heated breathing tube and delivered to the patient through the selected interface.

    AI/ML Overview

    This document is a 510(k) Summary for the F&P myAirvo 3, a respiratory humidifier. It is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, it does not contain acceptance criteria or study data related to the performance of an AI/ML powered device. The provided text primarily focuses on comparative claims against a predicate device and adherence to various safety and performance standards relevant to a medical device's physical and functional specifications.

    Therefore, I cannot extract information about acceptance criteria or a study proving an AI/ML device meets them from this document. The questions posed in the prompt (regarding AI/ML performance, sample sizes, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable to the content provided, as it describes a traditional medical device (a respiratory humidifier) and not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1