    K Number
    K113522

    Validate with FDA (Live)

    Date Cleared
    2012-01-27

    (59 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102366

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PathAssist Light Seeker is intended to locate, illuminate within, and transilluminate across nasal and sinus structures, including the frontal, ethmoid and maxillary sinuses, in patients aged 18 and over.

    Device Description

    The PathAssist™ Light Seeker™ is a fiber optic based, manually operated, reusable sinus seeker that can be connected to a light source to emit light from its distal end. It is labeled non-sterile and must be cleaned and sterilized or cleaned and high level disinfected prior to each use. The Light Seeker comes with two standard light post adapters, which allow the device to be compatible with commonly used 2.5mm light guides (cables).

    AI/ML Overview

    The provided 510(k) summary for the PathAssist™ Light Seeker™ (K113522) focuses on demonstrating substantial equivalence to a predicate device (K110158), rather than a detailed clinical study proving specific performance metrics against pre-defined acceptance criteria for a novel device. As such, many of the requested elements (e.g., sample sizes for test/training sets, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance with explicit metrics, and ground truth establishment for novel algorithms) are not applicable or not reported in this type of submission.

    The "acceptance criteria" in this context are primarily related to maintaining functionality and safety after reprocessing, aligning with the predicate device's performance, and meeting design specifications.

    Here's a breakdown based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Inferred/Reported)Reported Device Performance
    Functional PerformanceDevice performs as intended (locate, illuminate, transilluminate nasal/sinus structures).Performance testing showed that the subject device meets design specifications and performs as intended. Functional testing supported the useful life of the device after repetitive reprocessing cycles.
    Reprocessing EffectivenessDevice maintains functionality and safety after manual cleaning and manual high-level disinfection/sterilization.Reprocessing validation (cleaning, HLD, sterilization) was performed, with validation results referenced from predicate devices ([K102366] for cleaning/HLD, [K110158] for sterilization).
    BiocompatibilityDevice materials are biocompatible (per ISO 10993-1).Biocompatibility was established as having the same technological characteristics as the predicate device (same materials). No new biocompatibility data were submitted.
    Substantial EquivalenceDevice has the same indications for use and fundamental scientific technology as the predicate, with equivalent technological characteristics (principle of operation, design, function, materials).The Light Seeker has the same indications for use and fundamental scientific technology as the predicate device. It is substantially equivalent to the predicate device.
    SafetyDevice is safe for its intended use.Performance testing demonstrated that the device is safe and effective.

    Study Information

    1. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not applicable/Not reported. This submission describes design verification testing and reprocessing validation, not a clinical study on patient data with a "test set" in the sense of an algorithm. The testing involved repetitive reprocessing cycling of the device itself.
      • The document implies laboratory/engineering testing rather than patient-based data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable. No ground truth in the context of clinical expert review of diagnostic output is established or mentioned, as this is a surgical instrument, not an interpretive diagnostic device.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. No clinical test set requiring adjudication is described.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No. This is a Class I manual surgical instrument. No MRMC comparative effectiveness study was performed or is relevant to this device type. There is no AI component.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • No. This is a manual surgical instrument. There is no standalone algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • The "ground truth" for the performance testing of this surgical instrument would be its adherence to engineering specifications and demonstrated functionality (e.g., light output, structural integrity, cleanability, sterilizability) after simulated use and reprocessing cycles. This is based on design specifications and engineering validation.
    7. The sample size for the training set

      • Not applicable. No training set for an algorithm is described.
    8. How the ground truth for the training set was established

      • Not applicable. No training set for an algorithm is described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110158

    Validate with FDA (Live)

    Date Cleared
    2011-06-07

    (139 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102366,K071845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PathAssist Light Seeker is intended to locate, illuminate within, and transilluminate across nasal and sinus structures, including the frontal, ethmoid and maxillary sinuses, in adults aged 18 and over.

    Device Description

    The PathAssist Light Seeker is a fiber optic based, manually operated, reusable sinus seeker that can be connected to a light source to emit light from its distal end. It is labeled non-sterile and must be cleaned and sterilized prior to each use. The PathAssist Light Seeker comes with two standard light post adapters, which allow the device to be compatible with commonly used 2.5mm light guides (cables).

    AI/ML Overview

    The Entellus Medical PathAssist Light Seeker is a manual surgical instrument intended to locate, illuminate within, and transilluminate across nasal and sinus structures in adults. The device demonstrated substantial equivalence to predicate devices through various performance tests.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on establishing "substantial equivalence" to predicate devices, rather than defining specific numerical acceptance criteria for performance metrics that would be applicable to a diagnostic or AI-driven device. The "performance data" section details the types of testing performed to confirm the device's design specifications and intended function.

    Acceptance Criteria (Implied)Reported Device Performance
    Technological Equivalence (Principle of Operation, Design, Function, Materials, Biocompatibility, Reusability, Reprocessing) to predicate devices [Sinus Seekers, 510(k) exempt] and/or [Relieva Luma Sinus Illumination System, K071845].Met: The subject device and predicate device [Sinus Seekers, 510(k) exempt] are both shaped like a sinus seeker, advanced under endoscopic visualization, made of stainless steel, non-sterile, reusable, and require cleaning/sterilization. Both are biocompatible per ISO 10993-1. The subject device and predicate device [Relieva Luma Sinus Illumination System, K071845] both transmit light from proximal to distal tip via light fibers, can be connected to a standard light source via a light cable and adapter, and are biocompatible per ISO 10993-1. Performance testing validated that the device meets design specifications and performs as intended.
    BiocompatibilityMet: Biocompatibility testing was performed, and the device is biocompatible per ISO 10993-1.
    Design Verification (Dimensional and Functional Testing)Met: Design verification testing, including dimensional and functional testing, was performed to support the useful life of the device.
    Simulated Use Design ValidationMet: Simulated use design validation testing was performed in a cadaver model.
    Sterilization EfficacyMet: Steam sterilization testing was conducted. The device is labeled non-sterile and must be cleaned and sterilized prior to each use. Cleaning validation testing performed on device [K102366] was referenced.
    Compatibility with Light GuidesMet: Compatibility testing confirmed the device's ability to connect to commonly used 2.5mm light guides (cables) via two standard light post adapters.
    Storage and Transportation StabilityMet: Storage and transportation testing was performed.

    2. Sample size used for the test set and the data provenance

    The submission refers to "simulated use design validation testing in a cadaver model." However, it does not specify the sample size (number of cadavers) used for this test set.

    The data provenance is from simulated use testing in a cadaver model, which would be considered prospective in the context of device validation. The country of origin of the data is not specified, but given the submitter's location (Maple Grove, MN, USA), it's likely from the US.

    It's important to note that no animal or clinical data were submitted. This is typical for Class I devices seeking substantial equivalence where non-clinical testing is deemed sufficient.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not provide details on the number of experts used or their specific qualifications for evaluating the cadaver model testing. Given the nature of this type of testing (design validation for a manual surgical instrument), it would typically involve experienced surgical professionals or engineers assessing the device's function and performance within the cadaveric anatomy. However, this information is not explicitly stated.

    4. Adjudication method for the test set

    The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the cadaver model testing. For this type of design validation, observations and assessments are usually made by the testing personnel, and consensus would be implied if multiple individuals were involved, but a formal adjudication process is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a manual surgical instrument for illumination and transillumination, not an AI-driven or diagnostic imaging tool. Therefore, the concept of human readers improving with or without AI assistance is not applicable to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone (algorithm only) performance study was not done. The PathAssist Light Seeker is a physical, manually operated device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the "simulated use design validation testing in a cadaver model," the "ground truth" would implicitly be the anatomical structures within the cadaver and the observed functionality of the device (its ability to locate, illuminate, and transilluminate as intended) as assessed by the testing personnel. This is a form of direct observation and functional assessment rather than a diagnostic 'ground truth' like pathology or expert consensus on image interpretation.

    8. The sample size for the training set

    Not applicable. This device is a manual surgical instrument and does not involve machine learning or an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As a manual surgical instrument, there is no training set or machine learning algorithm for which ground truth would be established.