Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K163481

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-03

    (142 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K160154, K131082, K083714, K102067

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Half Dome Posterior Lumbar Intervertebral body fusion body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L1-L2 to L5-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Half Dome implants are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    The Half Dome Posterior Lumbar Interbody devices are implants developed for the substitution of the classical autogenous bone graft blocks They are available in a range of footprints and heights to suit the individual pathology and anatomical conditions of the patient. The implants have a hollow center to allow placement of autogenous bone graft. The Half Dome cages are designed to be used in conjunction with supplemental spinal fixation instrumentation.

    AI/ML Overview

    This document describes the Half Dome Posterior Lumbar Interbody System, a medical device, and does not provide information about an AI/ML powered device. Therefore, the questions related to AI/ML specific criteria (such as AI performance, training/test sets, ground truth establishment by experts, adjudication, or MRMC studies) are not applicable.

    The acceptance criteria and study information provided pertain to the non-clinical testing performed to demonstrate substantial equivalence to predicate devices, as is typical for 510(k) submissions of this nature where clinical studies are not performed.

    Here's the information based on the provided text, focusing on the device's substantial equivalence:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance
    Mechanical Safety and Performance (Non-Clinical Testing)
    Static Compression (per ASTM F2077)Performed and found equivalent to predicate devices.
    Dynamic Compression (per ASTM F2077)Performed and found equivalent to predicate devices.
    Subsidence (per ASTM F2267)Performed and found equivalent to predicate devices.
    Push Out (per ASTM F-04.25.02.02 DRAFT)Performed and found equivalent to predicate devices.
    Intended UseEquivalent to predicate devices.
    DesignEquivalent to predicate devices.
    MaterialsEquivalent to predicate devices (Vestakeep®i4R PEEK per ASTM F2026, Tantalum per ASTM F560, Titanium alloy (Ti-6Al-4V ELI) per ASTM F136).
    Indications for UseEquivalent to predicate devices.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in terms of a "sample size" for a clinical or AI test set. Non-clinical testing typically involves a set number of physical samples for each test (e.g., several implants for compression testing). The document does not specify the number of implants tested for each mechanical test.
    • Data Provenance: The data is derived from non-clinical bench testing (mechanical tests) performed by the manufacturer, Astura Medical. There is no information on country of origin for this testing, nor is it retrospective or prospective in the clinical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not Applicable. The ground truth for this device's acceptance is established through compliance with established ASTM standards for mechanical testing and comparison to legally marketed predicate devices, not through expert consensus on interpretation of data like in an AI/ML context.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are relevant for clinical studies or expert consensus, which were not performed for this 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI-powered device, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI-powered device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the non-clinical tests, the "ground truth" is defined by the performance requirements and methodologies outlined in the specified ASTM standards (e.g., ASTM F2077, ASTM F2267) and the performance characteristics of the identified predicate devices. The substantial equivalence argument posits that if the new device performs within acceptable limits established by these standards and similar to the predicate devices, it is safe and effective for its intended use.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI-powered device.

    9. How the ground truth for the training set was established:

    • Not Applicable. This is not an AI-powered device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140055

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-02

    (173 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K102067, K032812, K071795

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endoskeleton® TL IBD is indicated for use in spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. This device is intended for use supplemental fixation systems cleared for use in the lumbar spine. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients with non-fusion spinal surgery at the involved level(s) may be treated with the device. It is indicated to be used with autograft bone.

    Device Description

    The Endoskeleton® TL Interbody Fusion Device (IBD) implants are available in a various sizes with a variety of lordotic angles (0, 7 or 12 degrees), to accommodate patient anatomy. Lengths range from 40 to 60mm, widths from 18 to 26m and heights range from 8 to 16mm. Endoskeleton® TL IBD implants are intended for treatment in Lateral Lumbar Interbody Fusion used in single placement treatment placed across the disc space, and are designed with a farge hollow region in the center to house bone graft material. The new bone formation through the implant is intended to provide long-term structural support and biologic fusion at the implanted disc space. The design incorporates "windows" through the implant to permit visualization of the graft material and over time formation of new bone. The superior and inferior surfaces are acid etched to improve fixation to the adjacent bone. An implant holding feature has been incorporated into the trailing surface of the implant to mate with the implant holder, and to facilitate placement of the interbody space. All implantable components are manufactured from medical grade titanium alloy (Ti-6Al-AV-ELI).

    AI/ML Overview

    Here's an analysis of the provided text regarding the Endoskeleton® TL Interbody Fusion Device, focusing on acceptance criteria and supporting studies:

    It's important to note that this document is a 510(k) Summary for a medical device seeking clearance, not a detailed research paper. Therefore, the depth of information regarding clinical studies, ground truth establishment, and expert qualifications is much less than what would be found in a peer-reviewed publication. The focus of a 510(k) is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to prove absolute safety and effectiveness through extensive clinical trials for a novel device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
    Mechanical PerformanceStatic Compression-Shear (per ASTM F2077)Shown to be substantially equivalent to predicate interbody devices.
    Static Torsion (per ASTM F2077)Shown to be substantially equivalent to predicate interbody devices.
    Subsidence (per ASTM F2077)Shown to be substantially equivalent to predicate interbody devices.
    Expulsion (per ASTM F2077)Shown to be substantially equivalent to predicate interbody devices.
    Dynamic Compression (per ASTM F2267)Shown to be substantially equivalent to predicate interbody devices.
    Dynamic Compression-Shear (per ASTM F2267)Shown to be substantially equivalent to predicate interbody devices.
    Dynamic Torsion (per ASTM F2267)Shown to be substantially equivalent to predicate interbody devices.
    Biocompatibility(Implied) Manufactured from medical grade titanium alloy (Ti-6Al-AV-ELI)Accepted by the FDA, implying biocompatibility is established for this material.
    Material/Design Equiv.Indications for Use, Design, Dimensions, Materials (to predicate devices)All inclusive of the subject device (indicated as substantially equivalent).
    Safety/EffectivenessDoes not raise new questions about safety and effectiveness.Concluded that the subject device does not raise new questions.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a "test set" in the context of human subjects or clinical data. The performance testing described (mechanical testing) likely used device prototypes or manufactured samples. The number of samples for each mechanical test is not provided.
    • Data Provenance: Not applicable in the context of a clinical test set. The mechanical testing was performed in vitro, presumably by Titan Spine, LLC or a contracted lab. There is no mention of country of origin for clinical data, as no clinical study is detailed.
    • Retrospective/Prospective: Not applicable, as no clinical study is described. The mechanical tests are inherently retrospective in the sense that they are performed on manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not provided in the document. For mechanical testing, the "ground truth" is typically defined by the test standards (ASTM F2077, ASTM F2267) and the performance of the predicate device. There were no human experts establishing a clinical ground truth for this type of submission.

    4. Adjudication Method for the Test Set

    • This information is not provided and is not applicable to the type of mechanical performance testing described. Adjudication methods are typically used in clinical studies for interpreting ambiguous findings, which is not the case here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a physical intervertebral fusion device, not an AI software or diagnostic imaging tool that would involve human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No, a standalone performance study was not done. This refers to AI algorithms. The device is a physical implant.

    7. The Type of Ground Truth Used

    • For the mechanical testing, the "ground truth" was established by established ASTM standards (F2077, F2267) and the performance characteristics of the legally marketed predicate devices. The goal was to demonstrate that the subject device met or exceeded the performance of the predicates under these standardized conditions.

    8. The Sample Size for the Training Set

    • This information is not applicable. "Training set" refers to data used to train machine learning models. This device is a physical implant, and the submission does not involve AI or machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable. As there is no AI component, there is no training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K112030

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-19

    (340 days)

    Product Code
    Regulation Number
    888.3170
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes Radial Head Prosthesis System is intended for:

    1. The replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with
      a. Joint destruction and/or subluxation visible on x-ray
      b. Resistance to conservative treatment
    2. Primary replacement after fracture of the radial head.
    3. Symptomatic sequelae after radial head resection
    4. Revision following failed radial head arthroplasty.
    Device Description

    The Synthes Radial Head Prosthesis is a two-piece modular system comprised of titanium alloy stem and cobalt chrome head components with an integral screw and side-loading application to allow for in situ assembly. The system consists of a range of lengths and diameters for the stem in both straight and curved configurations as well as heads in a range of diameters and heights to accommodate the surgical need.

    AI/ML Overview

    This document is a 510(k) Summary for the Synthes Radial Head Prosthesis System. It is focused on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria and conducting a study to prove performance against those criteria in the way one might for a novel diagnostic or AI device.

    Therefore, many of the specific questions regarding acceptance criteria, device performance metrics (e.g., sensitivity, specificity), sample sizes, ground truth establishment, expert qualifications, and MRMC studies are not directly applicable or explicitly detailed in this type of submission for a medical implant device.

    However, I can extract the relevant information regarding the performance testing and the approach to demonstrating substantial equivalence.

    Here's an analysis based on the provided text, addressing the applicable points and explaining why others are not present:

    Context: The Synthes Radial Head Prosthesis System is a medical implant (a modular radial head prosthesis) intended for the replacement of the radial head in specific orthopaedic conditions. The 510(k) submission aims to demonstrate that this new device is "substantially equivalent" to existing, legally marketed predicate devices. This means it has the same intended use, fundamental technological characteristics, and similar materials, and that performance testing shows it is as safe and effective as the predicates.


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria for performance metrics typical of, for example, diagnostic accuracy (like sensitivity, specificity, AUC). Instead, the "acceptance criteria" are implied by the demonstration of substantial equivalence to predicate devices through various tests and analyses.

    Acceptance Criteria (Implied)Reported (Demonstrated) Device Performance
    Withstand clinical loads of the radiocapitellar joint (similar to predicates)In vitro performance testing demonstrates the ability of the proposed device to withstand the same clinical loads of the radiocapitellar joint. (Specific numerical thresholds or comparative data against predicates are not provided in this summary but would have been part of the full 510(k) submission).
    Similar surface characteristics to predicate devices (for stems)Macro Surface Analysis (roughness, pore size, pore depth, etc.), Scanning Electron Microscopy (SEM), and X-Ray Photoelectron Spectroscopy (XPS) have been conducted to demonstrate similar macro- and micro-surface characteristics in comparison to existing devices (Titan Endoskeleton Interbody Fusion Devices, Rhausler Plage Anterior Cervical Fusion System, Straumann Dental Implant System, Synthes Epoca Titanium Humeral Stem, Synthes Epoca Titanium Humeral Stem w/ (HA) Coating). The literature review also relates these device surface characteristics to physiological responses. Specific quantitative comparisons are not provided in this summary.
    Same indications for use as predicatesThe proposed Synthes Radial Head Prosthesis has the same indications for use as the predicate Biomet ExploR™ (K051385) and Ascension® (K032686) Modular Radial Head Devices.
    Same fundamental technological characteristics as predicatesThe proposed Synthes Radial Head Prosthesis has the same fundamental technological characteristics as the predicate Biomet ExploR™ (K051385) and Ascension® (K032686) Modular Radial Head Devices. This refers to the two-piece modular system with a titanium alloy stem and cobalt chrome head components with integral screw and side-loading application.
    Similar materials as predicatesThe proposed Synthes Radial Head Prosthesis has similar materials as the predicate Biomet ExploR™ (K051385) and Ascension® (K032686) Modular Radial Head Devices (titanium alloy and cobalt chrome).
    No new issues of safety and effectivenessBased on the discussed similarities in conjunction with performance testing, physical macro and micro surface analysis, surface chemistry analysis, and literature review the proposed Synthes Radial Head Prosthesis System does not raise any new issues of safety and effectiveness in comparison to the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify a "sample size" in terms of patients or anatomical specimens as one would for a clinical trial or diagnostic study. The "test set" here refers to the actual physical devices (or components thereof) that underwent in vitro performance testing and surface analysis. The number of such devices or samples tested is not disclosed in this summary.
    • Data Provenance: The tests are described as "in vitro performance testing," "Macro Surface Analysis," "Scanning Electron Microscopy (SEM)," and "X-Ray Photoelectron Spectroscopy (XPS)." These are laboratory-based studies, not clinical studies involving patient data from specific countries or populations. The data provenance is controlled laboratory conditions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable to this type of device submission. The "ground truth" for a mechanical implant is its physical and mechanical properties, not expert-derived diagnostic labels. The performance is assessed through engineering and materials science testing, not clinical diagnosis by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This question is not applicable. Adjudication methods are typically used to resolve discrepancies in expert interpretations of clinical data or images. This submission focuses on the objective physical and mechanical properties of an implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. The device is a physical implant, not a diagnostic imaging or AI algorithm that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. The device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance is established by engineering and materials science standards and measurements. This includes:
      • Mechanical properties: The ability to withstand clinical loads.
      • Material properties: Chemical composition and physical and macro/micro surface characteristics.
      • Peer-reviewed literature: Used to demonstrate how these material and surface characteristics relate to physiological responses.

    8. The sample size for the training set

    • This question is not applicable. There is no concept of a "training set" in the context of this 510(k) submission for a physical implant. The design and manufacturing processes are based on engineering principles and prior knowledge from predicate devices, rather than machine learning training.

    9. How the ground truth for the training set was established

    • This question is not applicable, as there is no training set. The "ground truth" for the device's design and material selection would be established through established engineering principles, materials science, and conformity to relevant ASTM/ISO standards (though specific standards are not named in this summary).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1