Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K163627

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-09-14

    (266 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K100925, K000080, K053394

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Shotel Ankle Arthrodesis Nail System is intended for the following:

    • Charcot Foot
    • Avascular necrosis of the talus
    • Failed total ankle arthroplasty
    • Trauma (malunited tibial pilon fracture)
    • Severe deformity or instability as a result of talipes equinoivarus, cerebral vascular accident, paralysis or other neuromuscular disease
    • Revision ankle arthrodesis
    • Neuroarthropathy
    • Rheumatoid arthritis
    • Osteoarthritis
    • Pseudoarthrosis
    • Post-Traumatic arthrosis
    • Previously infected arthrosis
    • Severe end stage degenerative arthritis
    • Severe defects after tumor resection
    • Pantalar arthrodesis
    Device Description

    The Shotel Ankle Arthrodesis Nail System is an implantable intramedullary system of fusion nails, bone screws, a nail cap and a set of instruments for primary ankle fusion. The curved fusion nail has 6 screw holes that can accommodate up to 6 bone screws and is available in seven (7) diameters in two configurations, left and right. The four (4) distal cross locking holes provide the surgeon with either a static or dynamic cross locking hole in alternative 90 degree orientation configurations. The 5mm fixation bone screws are offered in 27 lengths and the nail caps are available in two configurations; standard nail cap length and a plus 5 mm nail cap length affording the surgeon the ability to slightly adjust the final nail assembly length. All three implant components are manufactured from titanium Ti-6Al-4V and are single use only. Components are offered sterile and non-sterile.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Shotel Ankle Arthrodesis Nail System." It describes the device's indications for use, technological characteristics, and performance data to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's a breakdown of the requested information based on the provided text, with a strong caveat: This document is a regulatory submission for a medical implant (a nail system), not an AI/software as a medical device (SaMD) or diagnostic imaging device. Therefore, many of the requested elements (like acceptance criteria for AI performance, sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance, etc.) are not applicable or not provided in the context of this type of device.

    The "performance data" referred to in this document relates to mechanical and material performance of the implant, not diagnostic accuracy or AI performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for an implantable device)Reported Device Performance (Summary from submission)
    Material: Biocompatibility, appropriate for implant.Manufactured from titanium Ti-6Al-4V. Pyrogen testing confirmed implants meet 20 EU/device limit.
    Mechanical Strength/Durability: Withstand forces during implantation and long-term use for ankle arthrodesis.All necessary testing (primary stiffness, fusion-site compression, primary bending stiffness, mechanical and fatigue testing) performed in accordance with ASTM F1264.
    Design/Functionality: Provide stable fixation for ankle arthrodesis.System of fusion nails, bone screws, nail cap, and instruments for primary ankle fusion. Curved nail with 6 screw holes, 4 distal cross-locking holes (static/dynamic), available in various diameters and configurations (left/right). 5mm fixation bone screws, nail caps in two configurations for length adjustment.
    Sterility: If provided sterile, meets sterility standards.Components offered sterile and non-sterile. (Implied sterility validation if sterile components are offered).

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable. This is not a study assessing a diagnostic or AI device with a test set of data. The "testing" refers to physical and mechanical evaluation of the implant itself, not evaluation on patient data.
    • Data Provenance: Not applicable in the context of diagnostic/AI data. The materials and mechanical tests were performed on the device components.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. Ground truth as typically understood for diagnostic/AI studies (e.g., expert consensus on images, pathology results) is not relevant for this mechanical device submission.

    4. Adjudication method for the test set

    • Not applicable. There is no "test set" in the diagnostic/AI sense to adjudicate.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not a study involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used

    • Not applicable in the diagnostic/AI sense. The "ground truth" for this device's performance would be established through engineering standards (e.g., ASTM F1264) and measurements comparing the device's physical properties to those standards and predicate devices.

    8. The sample size for the training set

    • Not applicable. There is no training set as understood for AI/machine learning.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1