510(k) Data Aggregation

    K Number
    K191424

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-08-28

    (91 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K133636, K101165, K100736

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Interphalangeal Joint Fusion Devices are intended for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.

    Device Description

    The Interphalangeal Joint Fusion Device Range consists of metallic devices intended to stabilize two bone fragments until bone fusion. They are available in several lengths and angulations. All the implants are made of titanium alloy.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Interphalangeal Joint Fusion Device Range." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving improved diagnostic accuracy or effectiveness through clinical trials. Therefore, the information typically requested in your prompt regarding acceptance criteria for AI performance (like sensitivity, specificity, or AUC), and the detailed methodology of studies proving device performance with AI, is not applicable to this submission.

    Here's why and what information is available:

    • Device Type: This is a physical implant (metallic bone fixation fastener), not an AI/software device.
    • Regulatory Pathway: The 510(k) pathway for traditional medical devices relies heavily on demonstrating substantial equivalence to existing devices based on technological characteristics and performance data, often through non-clinical (mechanical) testing.
    • Focus of the Submission: The submission highlights similarities in intended use, materials, design, dimensions, sterilization, and performance specifications compared to predicate devices.

    Given this context, I will address the relevant points from your prompt based on the provided document.


    Acceptance Criteria and Study for Interphalangeal Joint Fusion Device Range

    The acceptance criteria for this device, as demonstrated through the 510(k) submission, revolve around establishing substantial equivalence to existing predicate devices. This is primarily achieved by showing comparable or superior mechanical performance and similar technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Intended UseDevice's intended use is substantially equivalent to predicate devices.The Interphalangeal Joint Fusion Devices and all predicates have similar intended uses.
    MaterialsDevice materials are the same as or comparable to predicate devices and conform to recognized standards (e.g., ISO, ASTM).Fabricated from Titanium alloy per ISO 5832-3 / ASTM F136, the same material as the predicate devices.
    Design Features/FunctionsDevice's basic design features and functions are similar to predicate devices.Shares similar basic design features and functions with cited predicate devices.
    DimensionsDevice dimensions are similar to predicate devices.Dimensionally similar to the cited predicate devices.
    SterilizationDevice sterilization method and state (sterile/non-sterile) are consistent with predicate devices.Provided sterile, as are the cited predicate devices.
    Mechanical Performance (Non-Clinical)Device demonstrates equivalent or superior mechanical performance compared to predicate devices under specified test conditions (e.g., static bending, dynamic bending, axial pull-out).Mechanical testing confirmed the Interphalangeal Joint Fusion Devices demonstrated as good or better performances than the cited predicate devices under the same test conditions. Results indicated they are as strong or stronger than predicate devices.

    2. Sample Size for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of clinical patient data or AI performance. For mechanical testing, the "samples" would refer to the number of devices tested. The document does not specify the exact number of devices tested for each mechanical test (static bending, dynamic bending, axial pull-out).
    • Data Provenance: The mechanical tests were performed by the manufacturer (NEOSTEO). No other provenance information (e.g., country of origin, retrospective/prospective) is relevant as these are laboratory-based engineering performance tests on the device itself.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable. This submission does not involve AI performance evaluation or a "ground truth" established by human experts in the diagnostic sense. The "ground truth" for the mechanical tests would be the established engineering standards and physical measurements.

    4. Adjudication Method for the Test Set

    • Not applicable. There was no human expert adjudication of results, as this was mechanical testing of a physical device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. An MRMC study was not conducted as this is a physical medical device submission, not an AI-assisted diagnostic or prognostic tool. The concept of "human readers improve with AI vs without AI assistance" does not apply.

    6. Standalone (Algorithm Only) Performance

    • No. This is not an AI/algorithm-only device. Therefore, standalone algorithm performance is not applicable.

    7. Type of Ground Truth Used

    • Mechanical Test Standards/Measurements: The "ground truth" for proving the device meets performance criteria relies on objective measurements from mechanical tests (static bending, dynamic bending, axial pull-out) and comparison against the performance of predicate devices under the same test conditions. Compliance with recognized material standards (ISO 5832-3 / ASTM F136) also contributes to the "ground truth" for material properties.

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no AI algorithm or training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K102072

    Validate with FDA (Live)

    Date Cleared
    2010-11-03

    (103 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K100736, K022599

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEMOMETAL implants (K-Snap® & Ti-Fuse®) are indicated for fixation of bone fracture, for bone reconstruction or arthrodesis in presence of appropriate immobilization. The size of the implant should be adapted to the specific indication.

    Device Description

    MEMOMETAL implants are single-use bone fixation appliances intended to be permanently implanted. MEMOMETAL implants are available in 2 different designs: - Snap-off pin, smooth with a stop and a sharp self drilling tip - Notched with a T-shaped head.

    AI/ML Overview

    The provided 510(k) Premarket Notification for MEMOMETAL Technologies Osteosynthesis implants (K102072) describes performance data related to the mechanical properties of the implants. This submission focuses on demonstrating substantial equivalence to predicate devices through engineering rationale and mechanical testing, rather than clinical outcome studies involving human or animal subjects that would typically involve acceptance criteria for a study with a test set, ground truth, and expert evaluation as you've outlined.

    Therefore, many of the requested categories (2-9) are not applicable to the type of information presented in this regulatory submission. The performance data provided is primarily engineering-based.

    Here's a breakdown of the relevant information from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Flexure LoadingNot explicitly defined as numerical acceptance criteria in the summary. Implied to be comparable to predicate devices.Comparison of mechanical properties of Ti-Fuse® and K-Snap® implants to predicate (K100736 Schilling Metalltechnik GMBH Kirschner and Guide wire) under flexure loading. Result: Data performed, indicating comparability.
    K-Snap® Insertion TestingNot explicitly defined as numerical acceptance criteria (e.g., maximum force, depth). Implied to demonstrate no premature breakage.Testing in cortical bone. Result: Premature breakage doesn't occur.

    Study Description (based on available information):

    This submission relies on engineering rationale and mechanical testing to demonstrate the substantial equivalence of the MEMOMETAL implants (K-Snap® & Ti-Fuse®) to their predicate devices. The study is a comparative mechanical performance study rather than a clinical trial.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in the summary (typical for mechanical testing, where the number of test samples for each mechanical property is usually defined by a standard).
    • Data Provenance: The study was conducted by MEMOMETAL Technologies (a French company). The testing is likely prospective in nature (i.e., designed and performed for this submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth as typically understood for clinical or image-based studies (e.g., expert consensus on diagnoses) is not relevant for this type of mechanical testing. The "ground truth" here is physics and engineering principles, with measurements taken from the devices and comparator.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a human-reader-based study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a human reader or AI-assisted study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The "stand-alone" performance is the mechanical performance of the devices themselves, as tested. There is no algorithm involved in the clinical sense.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth is based on objective mechanical measurements (e.g., force, displacement, material properties) as defined by established engineering and materials science standards for such implants. There is no expert consensus, pathology, or outcomes data used as ground truth for this mechanical study.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of mechanical performance testing for medical devices.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1