    Intended Use

    Porous Biologic Scaffold is a bone void filler intended for use in osseous defects of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created or the result of traumatic injury to the bone. Porous Biologic Scaffold is indicated to be packed gently into bony voids or gaps of the pelvis and posterolateral spine and is used combined with autograft as a bone graft extender. The device is resorbed and replaced with host bone during the healing process.

    Device Description

    Porous Biologic Scaffold is a resorbable bone void filler that is available in two versions:

    Version 1 is comprised of a porous collagen mesh that is either prefilled or separately packaged with human demineralized allograft bone matrix. In the prefilled format, the demineralized bone matrix is completely enclosed inside the collagen mesh. In the separately packaged format, an open-ended collagen mesh bag is supplied with separately packaged demineralized bone matrix and is filled and sewn closed with absorbable suture by the clinician. Version 1 of the Porous Biologic Scaffold may be hydrated with blood, saline, or bone marrow aspirate, and are combined with autograft bone prior to implantation.

    Version 2 is comprised of a porous resorbable collagen mesh bag either prefilled or separately packaged with activated biphasic ceramic granules. In the prefilled format, the biphasic granules are completely enclosed inside the collagen mesh. In the separately packaged format, an open-ended collagen mesh bag is supplied with separately packaged biphasic granules and is filled and sewn closed with absorbable suture by the clinician. Version 2 of the Porous Biologic Scaffold may be hydrated with blood, saline, or bone marrow aspirate, and are combined with autograft bone prior to implantation.

    Both formats of the Porous Biologic Scaffold are supplied terminally sterile and are available in sizes of 18 x 50 mm, 18 x 100 mm, 24 x 50 mm, and 24 x 100 mm.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K130083

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-17

    (184 days)

    Product Code
    Regulation Number
    878.4493
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K100461, K081002, K960653, K964072

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Demetech Absorbable Poliglecaprone 25 Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

    Device Description

    DemeCAPRONE (Poliglecaprone 25) is a synthetic monofilament absorbable surgical suture composed of Poly (glycolic-co-caprolactone) copolymer (PGCL) and is supplied un-dyed and dyed with D&C Violet #2 below 0.1wt%. DemeCAPRONE (Poliglecaprone 25) synthetic absorbable suture is available in sizes 6-0 through 1 (metric sizes 0.7 - 4). DemeCAPRONE (Poliglecaprone 25) Surgical Suture meets the requirements established by the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures except for diameter.

    AI/ML Overview

    The provided document describes the 510(k) summary for DemeCAPRONE (Poliglecaprone 25) Synthetic Monofilament Absorbable Suture. This document focuses on demonstrating substantial equivalence to predicate devices for a surgical suture, not an AI/ML powered medical device. Therefore, much of the requested information regarding AI/ML device studies (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance) is not applicable to this submission.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text, while acknowledging the inherent differences for a sutures submission:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria for DemeCAPRONE (Poliglecaprone 25) suture are based on demonstrating equivalence to legally marketed predicate devices and conformance to established standards, primarily the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures, and ISO 10993 for biocompatibility.

    Acceptance Criteria (Set by USP & ISO 10993)Reported Device Performance (DemeCAPRONE)
    Material Composition: Composed of Poly (glycolic-co-caprolactone) copolymer (PGCL)Same as predicate devices. Composed of Poly (glycolic-co-caprolactone) copolymer.
    Physical Form: Sterile, flexible, monofilament threadSame as predicate devices. Sterile, flexible, monofilament thread.
    Tensile Strength (< 881>): Meets or exceeds USP performance requirements.Meets or exceeds the performance requirements defined in the United States Pharmacopeia for "Tensile Strength" < 881 >.
    Needle Attachment (< 871>): Meets or exceeds USP performance requirements.Meets or exceeds the performance requirements defined in the United States Pharmacopeia and the current edition USP for "Needle Attachment" < 871 >.
    Suture Length Requirement: Meets 95% of stated label length (per USP).Meets or exceeds the performance requirements defined in the United States Pharmacopeia for "Suture Length Requirement" (95% of stated label length).
    Sterility: Meets the performance requirements defined in the United States Pharmacopeia (current edition U.S.P.).Meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. for sterility.
    Packaging & Labeling: Packed in a sterile, single/double manner with labeling conforming to 21 CFR and current USP.Is packaged in a same or equivalent manner with sterile single or double packaging having labeling conforming to 21 CFR and Current edition of USP.
    Biocompatibility (ISO 10993): Biologically compatible.Is biologically compatible when tested as per ISO-10993. (Specific tests performed: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Genotoxicity - Bacterial Reverse Mutation and Chromosomal Aberration, Bone Marrow Micronucleus, Subchronic Toxicity (4-week, following subcutaneous implantation), Muscle Implantation (12-week)).
    Toxicity (ISO 10993): Non-toxic.Is tested and proved to be non-toxic when tested as per ISO-10993 for toxicity.
    Absorption Profile (In-vivo and In-vitro): Demonstrates substantial equivalence to predicate devices.In-vivo and in-vitro resorption testing was conducted to further demonstrate substantial equivalence to the predicate devices. (Specific data not provided in the summary, but implied to be acceptable for equivalence).
    Extractable Color: Assured conformance to USP methods.Functionality testing assured that the device conformed with extractable color to methods outlined in USP 35. (Specific data not provided in the summary, but implied to be acceptable).
    Suture Diameter: Assured conformance to USP methods (except where noted).Functionality testing assured that the device conformed with suture diameter to methods outlined in USP 35, except for diameter. (This is a noted exception, but still considered acceptable due to overall equivalence).

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size for Test Set: The document does not specify exact sample sizes for each non-clinical test (e.g., number of sutures tested for tensile strength, length, etc.). However, it states that "Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and invivo and in-vitro resorption." This implies that the sample sizes used were in accordance with the specified standards and guidance documents.
    • Data Provenance: The studies were non-clinical (laboratory and animal studies for biocompatibility, in-vivo/in-vitro resorption). The location of the testing facilities is not explicitly stated, but the applicant is Demetech Corporation, Miami Lakes, FL, USA.
    • Retrospective or Prospective: The testing would have been prospective, as new tests were performed on the DemeCAPRONE suture to demonstrate its properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable for this type of device submission. Ground truth, in the context of AI/ML, refers to human expert annotations. For a surgical suture, the "ground truth" is adherence to physicochemical and biological standards (e.g., USP monographs, ISO 10993). The "experts" would be the scientists and technicians conducting the non-clinical tests and interpreting the results against established specifications, not medical domain experts creating ground truth for an algorithm.

    4. Adjudication method for the test set:

    • This is not applicable. Adjudication methods are used in consensus agreement among human annotators, which is not relevant for the objective physical and chemical testing of a surgical suture. The results are quantitative measurements compared against predefined acceptance limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. An MRMC study is for evaluating observer performance (e.g., radiologists interpreting images). This submission is for a physical medical device (suture) and does not involve human "readers" or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This describes the performance of a standalone AI algorithm. The DemeCAPRONE suture is a physical device, not an algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the device's performance is established by objective, established standards and methodologies.
      • Pharmacopeia Standards (USP): These define the chemical, physical, and pharmaceutical quality of drugs and medical devices. For sutures, this includes specifications for tensile strength, diameter, length, sterility, etc.
      • ISO 10993 Biocompatibility Standards: These define methods for evaluating the biological response to medical devices.
      • Predicate Device Performance: Part of the substantial equivalence relies on demonstrating that the new device's properties (including resorption) are similar to those of previously cleared predicate devices.

    8. The sample size for the training set:

    • This is not applicable. There is no "training set" in the context of this 510(k) submission for a surgical suture. Training sets are used for machine learning models.

    9. How the ground truth for the training set was established:

    • This is not applicable. As there is no training set for an AI/ML model, the concept of establishing ground truth for it does not apply.

    In Conclusion:

    The DemeCAPRONE suture's acceptance is based on demonstrating substantial equivalence to predicate devices and conformance to recognized consensus standards (USP) and international standards (ISO 10993). The "study" proving this involves a battery of non-clinical (laboratory and animal) tests rather than clinical trials or AI/ML-specific performance studies. The primary evidence presented is the declaration that DemeCAPRONE "meets or exceeds" the relevant performance requirements and is "Same" in composition, design, and manufacturing process to established predicate devices.