Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123120

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-12

    (70 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120732, K100244

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LenSx Laser System is indicated for use:

    • In the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery.
    • In the creation of a lamellar cut/resection for lamellar keratoplasty, and in the creation of a penetrating cut/incision for penetrating keratoplasty.
    • In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
    Device Description

    The LenSx Laser System uses focused femtosecond laser pulses to create incisions and separate tissue in the lens capsule, crystalline lens, and the cornea. Individual photodisruption locations are controlled by repeatedly repositioning the laser focus. The light pulse is focused into a sufficiently small spot in order to achieve photodisruption of the tissue inside the focus. A tiny volume of tissue, a few microns in diameter, is thereby photodisrupted at the laser focus. The surgical effect is produced by scanning thousands of individual pulses per second to produce a continuous incision or tissue separation. The location of the tissue photodisruption is controlled by moving the focus of the laser beam to the desired surgical target location. A computer-controlled scanning system directs the laser beam throughout a three-dimensional pattern to produce an incision. The laser pulses are delivered through a Patient Interface accessory that is placed on the surface of the cornea and fixes the eye with respect to the delivery system. This 510(k) premarket notification is being submitted to seek clearance for an additional patient interface accessory to be offered for use with cataract procedures in addition to the original patient interface accessory to the LenSx Laser System (cleared under K120732 for cataract and lamellar resection procedures). The SoftFit Patient Interface differs from the original Patient Interface (K120732) in that it has an extended suction ring skirt that enables positioning of a soft contact lens against the internal surface of the patient interface glass. The soft contact lens is similar to a standard daily wear contact lens indicated for the correction of ammetropia, with a slightly modified edge to enable fitting into the SoftFit Patient Interface. The materials and manufacturing processes used for the soft contact lens component of the proposed SoftFit Patient Interface accessory are identical to the processes used for the predicate daily wear soft contact lens (K100244). With the soft contact lens insert in place, the SoftFit Patient Interface is then mounted onto the LenSx laser system for docking onto the eye in a manner that is identical to the original LenSx Laser Patient Interface. As a result, use of the soft contact lens substantially reduces intraocular pressure (IOP) during the laser procedure, enhancing comfort and addressing potential concerns in patients with a history of glaucoma. For cataract procedures, either a LenSx Laser Patient Interface or a SoftFit Patient Interface may be used. The original LenSx Laser Patient Interface is used for keratoplasty and corneal flaps.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an additional patient interface accessory (SoftFit Patient Interface) for the LenSx Laser System. The primary purpose of this submission is to demonstrate substantial equivalence to a predicate device and to show that the new accessory does not raise new questions of safety or effectiveness.

    Here's an analysis of the acceptance criteria and study information based only on the provided text:

    1. Table of acceptance criteria and reported device performance:

    The document doesn't explicitly define formal "acceptance criteria" in a quantitative manner (e.g., "accuracy must be >X%"). Instead, it describes performance evaluations conducted to demonstrate substantial equivalence to the predicate device, particularly focusing on the differences introduced by the SoftFit Patient Interface.

    Performance MetricReported Device Performance (SoftFit Patient Interface)
    Holding Force and Intraocular Pressure (IOP) IncreaseEvaluated. The SoftFit Patient Interface results in a significantly lower intraocular pressure elevation during the laser procedure than the original Patient Interface.
    Accuracy and Reproducibility of Cataract Treatment PatternsEvaluated for depths and geometry when using the SoftFit Patient Interface. (No specific values provided, but the context implies it meets requirements for substantial equivalence).

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample sizes used for the performance evaluations (e.g., number of eyes, number of procedures).

    The data provenance is not explicitly mentioned as retrospective or prospective, nor is the country of origin.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The type of performance evaluation described (holding force, IOP, accuracy of laser patterns) suggests objective measurements rather than expert consensus on a diagnostic outcome.

    4. Adjudication method for the test set:

    Not applicable, as the performance tests described involve objective measurements rather than subjective assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a surgical laser system with a patient interface accessory, not an AI-assisted diagnostic or imaging device that would typically involve a multi-reader multi-case study to assess human reader improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The "device" in question is a surgical laser system designed for in-vivo use with human intervention, not a standalone algorithm. The performance tests assessed the physical and mechanical aspects of the new patient interface.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance evaluations would be:

    • Intraocular Pressure (IOP): Direct measurement using appropriate medical devices.
    • Holding Force: Direct measurement using mechanical testing methods.
    • Accuracy and Reproducibility of Cataract Treatment Patterns (depths and geometry): Measurement of the laser's physical output (e.g., incision depth, shape) using precision instruments (e.g., optical coherence tomography, microscopy, or other metrology tools).

    8. The sample size for the training set:

    Not applicable. The device described is a medical device (laser system accessory) and not an AI/ML model that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1