510(k) Data Aggregation

    K Number
    K211729

    Validate with FDA (Live)

    Date Cleared
    2021-11-04

    (153 days)

    Product Code
    Regulation Number
    888.3660
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K992119, K030710, K093803, K191625

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Anatomic Applications

    1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
    2. Rheumatoid arthritis.
    3. Revision where other devices or treatments have failed.
    4. Correction of functional deformity.
    5. Fractures of the proximal humerus, where other methods of treatment are deemed inadequate.
    6. Difficult clinical management problems, including cuff arthropathy, where other methods of treatment may not be suitable or may be inadequate.

    Reverse Applications
    The Comprehensive Reverse Shoulder is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is neeessary.

    Comprehensive Convertible Glenoid Baseplate components are intended for cementless applications with the addition of screw fixation.

    Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented application. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.

    Device Description

    The proposed device is an orthopaedic total joint intended to replace the damaged or diseased natural shoulder joint in shoulder arthroplasty to provide pain relief and restore function. It is modular in design, consisting of a baseplate held to the bone with bone screws and a modular liner. The device is designed to be implanted as the glenoid component of an anatomic total shoulder with the option to convert to a reverse shoulder configuration without removal of the metal components.

    The current submission is to expand the product offering by the addition of a liner manufactured from Vivacit-E Vitamin E Highly Crosslinked Polyethylene (VEHXPE).

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (shoulder prosthesis) and as such, it focuses on demonstrating substantial equivalence to a previously cleared device, rather than providing the kind of extensive performance study data that would be typical for a novel device or a clinical trial.

    Therefore, the document does not contain the detailed information required to answer many of your specific questions about acceptance criteria and a study proving the device meets those criteria.

    Here's why and what information can be extracted:

    • No "Acceptance Criteria Table" or "Reported Device Performance": The 510(k) process for this type of device (a new material liner for an existing, cleared system) primarily relies on demonstrating that the new material (Vivacit-E Liner) performs equivalently to existing materials in relevant non-clinical tests (e.g., mechanical tests). Clinical studies with specific performance metrics and acceptance criteria are generally not required for substantial equivalence for this type of modification.
    • No "Study that proves the device meets the acceptance criteria" in the sense of a clinical trial: The "Summary of Performance Data" explicitly states "Clinical Tests: None provided." This means there was no multi-reader, multi-case study, no human-in-the-loop performance measurement, and no standalone algorithm performance assessment. This device is a passive implant, not an AI or diagnostic tool.

    Based on the provided text, here's what can be answered or inferred:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable in the traditional sense. This submission is for a material change (Vivacit-E Liner) to an existing device. The performance data listed are non-clinical tests designed to show that the new liner material performs equivalently to existing cleared materials/devices.
    • Types of Non-Clinical Tests Conducted:
      • Dissociation Testing
      • Shear Testing
      • Insertion Testing
      • Biocompatibility Assessment
      • Packaging Assessment
    • "Acceptance Criteria" for these tests: While not explicitly stated with numerical values, the acceptance criterion for a 510(k) submission is that the performance of the new device (Vivacit-E Liner) is "as safe and effective as the legally marketed predicate devices" and that "any differences do not raise new questions of safety and effectiveness." This implies that the results of these non-clinical tests must fall within expected ranges or demonstrate equivalence to the predicate device for each test. Specific numerical thresholds or ranges are not provided in this summary.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for the non-clinical tests. For mechanical testing of medical devices, sample sizes are typically determined by engineering standards (e.g., ISO, ASTM) and statistical power requirements to demonstrate equivalence or meet specified performance criteria.
    • Data Provenance: Not specified, but given it's a submission by Biomet Manufacturing Corp. (a US company) to the FDA, the testing would typically be conducted according to internationally recognized standards in controlled laboratory environments. The data would be specific to components manufactured by the applicant. This is not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is not a study requiring expert readers or ground truth establishment in the clinical image interpretation sense. The "ground truth" for mechanical testing is the physical property being measured (e.g., force, displacement, material degradation) against defined engineering specifications.

    4. Adjudication method for the test set

    • Not Applicable. No human adjudication is involved in the non-clinical mechanical testing of an implantable device like this.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The "Summary of Performance Data" explicitly states: "Clinical Tests: None provided." This device is a passive implant, not a diagnostic or AI-powered device that would involve human readers or image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. As above, this is a physical implant, not a software algorithm.

    7. The type of ground truth used

    • For Non-Clinical Tests: The "ground truth" would be engineering specifications and material properties as defined by relevant standards (e.g., ISO, ASTM) and internal design requirements. For example, for "Shear Testing," the ground truth would be the measured shear strength compared to the designed minimum shear strength or the shear strength of the predicate device.

    8. The sample size for the training set

    • Not Applicable. This is not a machine learning or AI-driven device, so there is no concept of a "training set" for the device itself.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for the device, this question is irrelevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131353

    Validate with FDA (Live)

    Date Cleared
    2013-10-08

    (151 days)

    Product Code
    Regulation Number
    888.3660
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K080642, K120121, K030710, K093803

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Comprehensive® Reverse Shoulder is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.

    The Comprehensive® Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.

    Titanium olenospheres are intended for patients with colbalt alloy material sensitivity. The wear of these devices has not been tested but, based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A cobalt alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy.

    Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.

    Interlok® finish humeral stems are intended for cemented use and the Macro Bond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.

    The device is a single-use implant.

    Device Description

    The Titanium Alloy Glenosphere Prosthesis consists of a series of various sized modular glenospheres with variable offset between 0.5mm and 4.5 mm. Titanium alloy material has been selected to provide the surgeon with an alternate material to treat patients with nickel allergies. Each modular glenosphere consists of a "head" and a taper adaptor. The taper adaptor is impacted into the head in a certain position to achieve the desired amount of offset. The system can be used with Biomet's Comprehensive® Reverse Shoulder System or Biomet's BioModular® Reverse Shoulder System.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, specifically the Comprehensive® Reverse Shoulder - Titanium Glenosphere. It focuses on demonstrating substantial equivalence to previously marketed devices rather than presenting a study proving that the device meets specific acceptance criteria based on performance data.

    Therefore, many of the requested sections about acceptance criteria, detailed study design, and performance metrics (like sample size for test sets, expert ground truth, effect size, etc.) are not applicable or not provided in this type of regulatory submission. The document explicitly states "No clinical data submitted" and "No clinical data was necessary for a determination of substantial equivalence."

    However, I can extract the relevant information that is present:

    1. A table of acceptance criteria and the reported device performance

    Since this is a substantial equivalence submission, the "acceptance criteria" are implicitly defined by the characteristics and performance of the predicate devices. The performance reported here is primarily related to non-clinical bench testing.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (Non-Clinical)
    Functional equivalence to predicate devices"The number of components, sizing, and all dimensions are identical to the predicate." "The results of testing indicate the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices."
    Material compatibility/performance"Titanium alloy material has been selected to provide the surgeon with an alternate material to treat patients with nickel allergies." (Implied acceptance for patients with cobalt alloy material sensitivity).
    Mechanical integrity (e.g., taper locking)Test Conducted: Torsional separation of tapers. (The document states this test was performed, implying the device met internal acceptance criteria for taper integrity, though specific values are not provided in this summary.)
    Wear characteristics"The wear of these devices has not been tested but, based on pin on disk testing, the wear rate is inferior to that of cobalt alloy glenospheres. A cobalt alloy glenosphere is the recommended component for reverse shoulder arthroplasty patients without material sensitivity to cobalt alloy." (This is a caveat rather than a direct performance claim, acknowledging inferior wear for the titanium compared to cobalt, but still implying acceptance for its specific indication in material-sensitive patients.)

    Detailed Study Information (Not applicable/Not provided in this document type):

    As noted, this document is a 510(k) summary demonstrating substantial equivalence, not a clinical study report with detailed performance data against specific acceptance criteria. Therefore, the following information is not present:

    1. Sample sizes used for the test set and the data provenance: Not provided for this type of submission which relies on non-clinical testing for substantial equivalence. No clinical test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as there was no clinical test set requiring expert ground truth for this submission.
    3. Adjudication method for the test set: Not applicable, as there was no clinical test set.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an orthopedic implant, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is an orthopedic implant, not a standalone algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. "No clinical data submitted." For non-clinical tests (like torsional separation), the "ground truth" would be engineering specifications and predicate device performance.
    7. The sample size for the training set: Not applicable. This device is an orthopedic implant, not an AI/algorithm that requires a training set.
    8. How the ground truth for the training set was established: Not applicable.