Found 3 results

510(k) Data Aggregation

    K Number
    K252285

    Validate with FDA (Live)

    Device Name
    Well-Root PT
    Manufacturer
    Date Cleared
    2025-09-19

    (58 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Pulp Capping
    • Repair of Root Perforation
    • Repair of Root Resorption
    • Root-end Filling
    • Apexification
    Device Description

    Well-Root PT is ready-to use, premixed, bioceramic paste developed for pulp capping, permanent root canal repair and filling. Well-Root PT is delivered by the capsule containing the mixture of calcium silicate cement, liquid and several compound particles. It sets in the root without the mixing process with additional water.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242934

    Validate with FDA (Live)

    Date Cleared
    2024-12-12

    (78 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications.

    • Repair of Root Perforation
    • Repair of Root Resorption
    • Root End Filling
    • Apexification
    • Pulp Capping
    Device Description

    i-MTA BP Bioceramic Root Canal Sealer and Repair Materials is a ready-to-use premixed bioceramic paste developed for permanent root canal sealer and repair and surgical applications. i-MTA BP is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. i-MTA BP does not shrink during setting and demonstrates excellent physical properties. i-MTA BP is packaged in a preloaded syringe.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP), based on the provided FDA 510(k) summary:

    This device is not an AI/ML device, therefore, some of the requested information (like effect size of human readers with vs. without AI, number of experts for ground truth, adjudication methods, details on training set) are not applicable. The information provided relates to a traditional medical device's performance against established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Requirement)Standard/GuidelineReported Device Performance (i-MTA BP)
    Physical/Chemical Properties:
    Setting timeISO 6876:20122.5h (within 2h-6h)
    SolubilityISO 6876:20122.80% (< 3%)
    Radio-opacityISO 6876:2012> 3mm Al (> 3mm Al)
    Composition (visual foreign objects)Technical SpecificationsCompliant with requirements
    Content (of indicated loading capacity)Technical SpecificationsCompliant with requirements (>93%)
    Size change after curingTechnical Specifications (ISO6876:2001)-0.50% (< 1.0%)
    pHTechnical Specifications12.29 (within 11.5-13.5)
    Acid-soluble arsenic contentTechnical SpecificationsCompliant with requirements (< 2mg/kg)
    Acid-soluble lead contentTechnical SpecificationsCompliant with requirements (< 100mg/kg)
    Heavy metal contentTechnical Specifications< 100 µg/g (< 100 µg/g)
    Biocompatibility:
    CytotoxicityISO 7405:2018Compliant
    SensitizationISO 10993-10:2021Compliant
    IrritationISO 10993-10:2021Compliant
    Acute Systemic ToxicityISO 10993-11:2017Compliant
    TK Gene Mutation TestISO 10993-3:2014Compliant
    Bacterial Reverse Mutation TestISO 10993-3:2014Compliant
    Subchronic Systemic ToxicityISO 10993-11:2017Compliant
    ImplantationISO 10993-6:2016Compliant
    Pulp Capping TestingISO7405:2018Compliant
    Endodontic Usage TestISO 7405:2018Compliant

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific performance test (e.g., how many samples were tested for setting time, solubility, etc.). The provenance of the data is not specified beyond being generated from "Performance testing - Bench" and "Biocompatibility Testing" which implies laboratory testing directly on the subject device. There is no indication of retrospective or prospective clinical data from specific countries for these bench tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the type of device and testing presented. The "ground truth" for this device is based on objective measurements against international and technical standards, not expert consensus on medical images or diagnoses.

    4. Adjudication Method for the Test Set

    Not applicable. The tests involve objective measurements against predefined criteria, not subjective interpretation requiring adjudication among experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically for diagnostic imaging devices where human readers interpret medical images. This device is a dental material, and its performance is evaluated by bench testing and biocompatibility.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the device's performance was evaluated in a standalone manner (without human-in-the-loop performance) through bench testing and biocompatibility studies. As it is a physical material, there is no "algorithm" involved, but its intrinsic properties were tested directly.

    7. The Type of Ground Truth Used

    The ground truth used for performance validation is objective measurement against established international standards (e.g., ISO 6876:2012, ISO 10993 series, ISO 7405:2018) and internal Technical Specifications.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232299

    Validate with FDA (Live)

    Device Name
    CeraPutty
    Date Cleared
    2023-09-29

    (59 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092715

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Repair of Root Perforation
    • Repair of Root Resorption
    • Root End Filling
    • Apexification
    • Pulp Capping
    Device Description

    CeraPutty (Bioceramic Root Repair Material) is a ready-to-use, premixed bioceramic root repair material developed for permanent root canal repair and filling applications. It is made of Calcium silicate composite, zirconium dioxide which is a radiopaque agent, and a thickening agent. CeraPutty sets and hardens by the moisture inside the oral cavity. CeraPutty does not shrink during setting and demonstrates excellent physical properties. CeraPutty does not require additional mixing for procedures, and its white color which allows aesthetic treatment. Ceraputty is packaged in a preloaded syringe.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "CeraPutty," which is a root canal filling resin. It describes the device, its intended use, and its equivalence to a predicate device. This document is a regulatory filing and does not contain information about a study proving the device meets acceptance criteria in the way a clinical or diagnostic performance study would. It presents information regarding the device's technical characteristics and non-clinical testing to demonstrate substantial equivalence to a predicate device, rather than a study against a specific performance benchmark for diagnostic accuracy or effectiveness with human subjects.

    Therefore, many of the requested elements regarding acceptance criteria for a diagnostic device performance study, sample sizes, expert ground truth, adjudication methods, multi-reader multi-case studies, and training set details are not applicable or available in this type of regulatory submission. The document focuses on showing that the new device (CeraPutty) is similar enough to an already approved device (iRoot BP Plus) to be considered safe and effective for the same indications.

    Here's the information that can be extracted or inferred from the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in the typical sense of a diagnostic performance study (e.g., minimum sensitivity, specificity). Instead, it compares the performance of the subject device (CeraPutty) to the predicate device (iRoot BP Plus) based on performance standards (ISO 6876) and specific physical properties. The implicit "acceptance criterion" is that the subject device's performance is comparable to or better than the predicate device for these properties, and within the bounds of the standard.

    Performance MetricAcceptance Criteria (implied by predicate, and ISO 6876)Reported Device Performance (CeraPutty)Reported Predicate Performance (iRoot BP Plus)
    Performance StandardConformed to ISO 6876Conformed to ISO 6876Conformed to ISO 6876
    Setting timeComparable to predicate; within standard44 min5 h 10 min
    SolubilityComparable to predicate; within standard0.1 %0.3%
    Radio-opacityComparable to predicate; within standard4.84.2
    BiocompatibilityYesYesYes
    Shelf Life2 years2 years2 years

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document refers to "Performance tests" and "Biocompatibility Tests" but does not provide details on the number of samples or specimens tested for these non-clinical evaluations.
    • Data Provenance: Not specified. These are laboratory/bench tests, not clinical data from patients. The manufacturer is Meta Biomed Co., Ltd. in South Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. This is not a study requiring expert ground truth in the diagnostic sense. The "ground truth" for the performance tests would be the results obtained from calibrated laboratory equipment according to the specified ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study requiring adjudication of expert interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a material for dental procedures, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a material for dental procedures, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the non-clinical performance tests mentioned (setting time, solubility, radio-opacity) would be derived from measurements obtained via standardized laboratory protocols according to ISO 6876:2012. For biocompatibility, it would be results from standardized biological tests according to ISO 10993 series.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.