Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K150632

    Validate with FDA (Live)

    Date Cleared
    2015-07-06

    (117 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use of the Hypervisor IX Monitoring System include:

    • · Real time viewing of patient clinical data and alarms
    • · Storage and Historical review of patient clinical data and alarms
    • · Printing of real time and historical patient data
    • · Configuration of local settings as well as synchronizing settings across the network to a remote device
    • · Transfer of patient clinical data and settings between several CentralStations

    The Hypervisor IX Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in a hospital or clinical environment to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wireless WiFi (WLAN), and Wireless WMTS. The Hypervisor IX Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors; The Hypervisor IX Monitoring System supports bi-directional configuration of the compatible monitors. No data processing is done by the Hypervisor IX Monitoring System for data received from compatible monitors.

    The Hypervisor IX Monitoring System includes several Telemetry monitoring subsystems all of which operate in the 608M WMTS frequency range within a defined coverage area. All of the supported telemetry systems transmit data to the CentralStation for processing, display, and alarm.

    · The TMS-6016 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data. The CentralStation will support ECG, Heart Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS-6016.

    • The Panorama Telepack-608 transmitter is intended for use on Adult patients to monitor ECG and SpO2 physiological data. The Centralstation will support ECG, Heart Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the Panorama Telepack-608.

    · The TMS60 transmitter is intended for use on Adult and Pediatric patients to monitor ECG and SpO2 physiological data. The physiological data can be reviewed locally on the transmitter. The CentralStation will support ECG, Heart Rate, SpO2, Pulse Rate, Arrhythmia analysis, QT monitoring, and ST Segment Analysis for the TMS60.

    The Hypervisor IX Monitoring System is intended for use in a hospital or other clinical environment under the direct supervision of a licensed healthcare practitioner.

    Device Description

    The Hypervisor IX Monitoring System is a networked patient monitoring system intended for use in a fixed location, installed in a hospital or clinical environment to provide clinicians remote patient monitoring. The network connections between the various devices can be any combination of Ethernet (Wired), Wireless WiFi (WLAN), and Wireless WMTS.

    The Hypervisor IX Monitoring System supports one or more Mindray compatible physiological monitors and will display, store, print, and transfer information received from the compatible monitors; The Hypervisor IX Monitoring System supports bi-directional configuration of the compatible monitors. No data processing is done by the Hypervisor IX Monitoring System for data received from compatible monitors.

    The Hypervisor IX Monitoring System includes three telemetry monitoring systems all of which operate in the 608M WMTS frequency range within a defined coverage area. All of the supported telemetry systems transmit data to the CentralStation for processing, display, and alarm.

    The Hypervisor IX Monitoring System consists the following components:

      1. Central Station
      1. View Station
      1. Work Station
      1. Telemetry Systems (TMS 6016, Telepak-608, TMS60)
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Hypervisor IX Monitoring System, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in a quantifiable manner (e.g., target accuracy, sensitivity, specificity). Instead, it describes general claims of meeting specifications and equivalence to a predicate device. The performance data section focuses on types of testing performed rather than specific quantitative results against pre-defined criteria.

    Feature/TestAcceptance Criteria (Not explicitly stated with quantifiable thresholds in the document)Reported Device Performance
    Functional and System Level TestingImplied: Device meets all functional and system-level specifications."The testing provided an evaluation of the performance of the device relevant to each of the differences between the subject device and the predicate device. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate."
    Compliance with Relevant Consensus StandardsImplied: Device meets all relevant consensus standards."Mindray has conducted testing to ensure the subject device meets relevant consensus standards." (References "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm Guidance for Industry and FDA (10/28/2003)", "Cardiac Monitor Guidance (Including Cardiotachometer and Rate Alarm)", and "Class II Special Controls Guidance Document: Electrocardiograph Electrodes")
    Wireless Functionality TestingImplied: Device meets wireless specifications."Mindray conducted Wireless functionality testing to ensure the performance of the Hypervisor IX Monitoring System meets wireless specifications and is equivalent to the predicate device."
    Substantial Equivalence (Overall System Performance)Implied: Differences do not raise new questions of safety and effectiveness."Based on the detailed comparison of specifications for each of the modifications to the previously cleared Hypervisor Central Monitoring System, the functional and performance testing and conformance with applicable standards as well as the comparison of indications, the differences do not raise new questions of safety and effectiveness and the Hypervisor IX Monitoring System can be found substantially equivalent to the predicate device."
    QT MonitoringImplied: Performance comparable to cleared predicate with QT monitoring."Although QT monitoring was not included in the predicate Hypervisor device, it is present in the cleared Philips ST/AR ST and Arrhythmia Software (K101521), and thus does not constitute a new intended use for a patient monitoring system."

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not provide information regarding:

    • The specific sample size (number of patients, records, or tests) used for performance testing (test set).
    • The country of origin of the data.
    • Whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on:

    • The number of experts, if any, used to establish ground truth for the test set.
    • The qualifications of such experts.

    The testing described (functional, system-level, wireless) appears to be primarily engineering verification against specifications rather than clinical validation requiring expert ground truth on patient data. The statement regarding QT monitoring references a predicate device (Philips ST/AR ST and Arrhythmia Software K101521) for establishing the safety/effectiveness of that function, implying reliance on that previous clearance rather than new clinical ground truthing on this specific device.

    4. Adjudication Method for the Test Set

    The document does not provide information on any adjudication method used for the test set. Given the nature of the described testing, formal adjudication by experts is unlikely to have been part of these specific tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    The document does not indicate that an MRMC comparative effectiveness study was done. This type of study is more relevant for diagnostic AI devices where human reader performance is a key metric. The Hypervisor IX Monitoring System is described as a networked patient monitoring system that displays, stores, prints, and transfers data; it explicitly states "No data processing is done by the Hypervisor IX Monitoring System for data received from compatible monitors." This implies the system's role is data management and display, not AI-driven interpretation or diagnosis that would typically warrant an MRMC study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The document does not describe the device as incorporating an AI algorithm that would typically require standalone performance evaluation. The statement "No data processing is done by the Hypervisor IX Monitoring System for data received from compatible monitors" suggests its function is passive.

    7. The Type of Ground Truth Used

    Based on the description of testing (functional, system-level, wireless compliance, and comparison to predicate specifications), the "ground truth" for the tests performed would likely be:

    • Engineering Specifications: For functional and system-level tests.
    • Regulatory Standards: For compliance with consensus standards.
    • Predicate Device Performance/Clearance: For demonstrating equivalence and justification for new features like QT monitoring.

    There is no mention of ground truth derived from expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    The document does not mention a training set. This is consistent with the description of the device as not performing data processing or incorporating AI that would require machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned or implied, there is no information on how ground truth for a training set would have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143195

    Validate with FDA (Live)

    Date Cleared
    2014-12-22

    (46 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132026, K123074, K132037, K092449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passport 17M and Passport 12M patient monitors are intended for monitoring. displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-mvasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sy()//Scy(), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    • . The arrhythmia detection. ST Segment analysis of Mortara algorithm. BIS, RM, CCO. SvOz/Scv02, and PAWP monitoring are intended for adult and pediatric patients only:
    • ST Segment analysis of Mindray algorithm is intended for adult patients only; ●
    • C.O. monitoring is restricted to adult patients only: ●
    • ICG monitoring is only for use on adult patients who meet the following ● requirements: height: 122 to 229cm, weight: 30 to 155kg.

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpOz), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    The arrhythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;

    ST Segment analysis of Mindray ECG algorithm is intended for adult patients only;

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    Device Description

    The subject Passport Series Patient Monitors includes three monitors:

    • . Passport 12M Patient Monitor
    • Passport 17M Patient Monitor ●
    • T1 Patient Monitor

    All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12M and 17M Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor. The T1 patient monitor is one of the available modules that can be plugged into the Passport 17M or 12M monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17M and 12M.

    Each of the three monitors that are the subject of this 510(k) are modified versions of previously cleared Mindray devices. The Passport 17M and Passport 12M were previously cleared in K132075, on April 18, 2014 and the T1 was previously cleared in K123074 on July 3, 2013.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Mindray Passport Series Patient Monitors (K143195). This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a single study proving that the device meets those criteria.

    The core of the submission is that the subject devices are modified versions of previously cleared Mindray devices (K132075 and K123074). The performance data cited refers to functional and system-level testing to evaluate the performance of the modifications and ensure continued adherence to specifications and relevant consensus standards. It does not contain information about a specific single study with acceptance criteria and results.

    Given the nature of the K143195 submission as an amendment/modification to existing devices, the information provided focuses on the changes and how these changes do not alter the substantial equivalence to the predicate devices. Therefore, the requested information about acceptance criteria and a detailed study proving the device meets those criteria for the entire device as if it were a novel submission is not explicitly present in the provided text.

    However, I can extract information related to the performance characteristics, which serve as implicit acceptance criteria based on the predicate device's performance and general recognized standards for patient monitors.

    Here's an attempt to structure the answer based on the available information, noting where specific details are not provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of acceptance criteria with corresponding performance results in a standalone "study summary" format. Instead, it compares the specifications of the subject devices to the predicate device. The "acceptance criteria" are implied to be the specifications of the predicate device and relevant industry standards. The "reported device performance" is generally stated as "Same" or that the new (modified) features "meet specifications" or "are equivalent to the predicate device."

    Here's an illustrative table based on the provided "Device Comparison Table," highlighting areas that could be inferred as "acceptance criteria" and "performance":

    Feature CategoryParameterImplicit Acceptance Criteria (from Predicate Device K132075)Reported Subject Device Performance (K143195)
    GeneralIndications for UseMonitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters for adult, pediatric, and neonatal patients (with specific exceptions).For Passport 12M/17M, "Same" as predicate device. For T1, Indications are a "subset" of the predicate. New feature: 12-lead ECG interpretation, noted as present in cleared Mindray V Series (K132026), implying equivalence.
    DisplayIntegrated display and touchscreen (17M)17", 1280*1024 pixelsSame
    Integrated display and touchscreen (12M)12", 800*600 pixelsSame
    Integrated display and touchscreen (T1)(Predicate K123074: 5", 480*272 pixels)5", 480*272 pixels (same as K123074)
    Secondary display (Passport 12M/17M)Independent control and display (17M), Linked to integrated display (12M)Same
    Secondary display (T1)N/A in predicate K132075Independent display and control via a VGA port in the T1 docking station (new feature for T1)
    ConnectivityWireless module (Passport 17M)ASUS module for network connectionSame
    Wireless module (Passport 12M)ASUS module for network connectionAdded support for Silex wireless module. "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."
    Wireless module (T1)N/A in predicate K132075Added wireless function using the Cyberlink wireless module. "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."
    Physiological ParamsECG (features)3-lead, 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, heart rateSame
    12-lead ECG interpretationNot supported (Predicate K132075)Supported (All subject devices). Justified by equivalence to K132026. Performance not detailed here.
    ECG J-point auto detectionSupportedSame (17M, 12M). Added (T1). Performance not detailed.
    Respiration rate (Resp) Measurement & AccuracyAdult: 0-120 rpm, Pediatric/Neonate: 0-150 rpm. Accuracy: 7-150 rpm: ±2 rpm or ±2%, whichever is greater; 0-6 rpm: Not specified.Same
    Temperature (Temp) Measurement & Accuracy0-50°C (32-122°F). Accuracy: ±0.1°C or ±0.2°F (without probe).Same
    SpO2 Measurement & Accuracy (Mindray SpO2)Range: 0-100%. Accuracy: 70-100%: ±2% (adult/pediatric), ±3% (neonate); 0-69%: Not specified.Same
    SpO2 Measurement & Accuracy (Masimo SpO2)Range: 1-100%. Accuracy: 70-100%: ±2% (no motion, adult/pediatric), ±3% (no motion, neonate), ±3% (with motion); 1-69%: Not specified.Same
    SpO2 Measurement & Accuracy (Nellcor SpO2)Range: 0-100%. Accuracy: 70-100%: ±2% (adult/pediatric), ±3% (neonate); 0-69%: Not specified.Same
    Pulse rate (PR) from SpO2 ModuleMindray SpO2: 20-254 bpm, ±3 bpm. Masimo SpO2: 25-240 bpm, ±3 bpm (no motion), ±5 bpm (with motion). Nellcor SpO2: 20-300 bpm, 20-250 bpm: ±3 bpm.Same
    Pulse rate (PR) from IBP Module25-350 bpm, ±1 bpm or ±1%.Same
    NIBP Measurement & AccuracyAdult Systolic: 40-270 mmHg, Diastolic: 10-210 mmHg, Mean: 20-230 mmHg. Pediatric Systolic: 40-200 mmHg, Diastolic: 10-150 mmHg, Mean: 20-165 mmHg. Neonate Systolic: 40-135 mmHg, Diastolic: 10-100 mmHg, Mean: 20-110 mmHg. Accuracy: Max mean error: ±5 mmHg, Max standard deviation: 8 mmHg.Same
    IBP Measurement & AccuracyRange: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Same
    Pulse Pressure Variation (PPV)Supported feature of IBP (Passport 12M/17M)Same (17M, 12M). Added (T1).
    CO2 Measurement & AccuracySidestream: 0-99 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-76 mmHg: ±5%, 77-99 mmHg: ±10%. Microstream: 0-99 mmHg; Accuracy: 0-38 mmHg: ±2 mmHg, 39-99 mmHg: ±5% +0.08% of (reading-38). Mainstream: 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%.Specifications unchanged. Modules repackaged to occupy a single slot. (All subject devices).
    Other3 Additional SpO2 SensorsNone (Predicate K132075)Added LNCS NeoPt, LNCS Neo, LNCS Inf. Cleared in K132037 (implies their performance against specific standards is already established).
    4 New NIBP cuffsNone (Predicate K132075)Added CM1306, CM1307, CM1506, CM1507. Substantial equivalent to NIBP cuffs cleared in K132075 and K092449 (implies their performance against specific standards is already established).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not detail specific sample sizes for test sets for individual parameters or features. It states:

    • "Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance."
    • "Mindray has conducted testing to ensure the subject devices meet relevant consensus standards."
    • "Mindray has also followed the FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm issues on October 28, 2003."
    • "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."

    There is no mention of data provenance (e.g., country of origin, retrospective/prospective). The studies appear to be internal V&V testing.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    There is no information provided regarding the number or qualifications of experts used to establish ground truth for any test sets. The testing referenced appears to be engineering/technical validation against specifications and standards rather than clinical studies requiring expert adjudication of data. For interpretations like 12-lead ECG, it is stated that the feature is equivalent to a previously cleared device (K132026), implying that its validation methods would have been established there.


    4. Adjudication Method for the Test Set

    No adjudication method is described.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No MRMC comparative effectiveness study is mentioned. This type of study is typically done for diagnostic imaging devices or algorithms where human interpretation is a key component and the AI is intended to assist, altering reader performance. For a patient monitor, the performance validation is typically against direct physiological measurements and established standards.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The entire device (the patient monitor) is a standalone system in the sense that its algorithms process physiological signals and present data and alarms. The performance data described ("functional and system level testing") would inherently be evaluating the algorithms and hardware in isolation from a human interpretation loop. For example, SpO2 accuracy is measured against a reference, not against a human's interpretation of SpO2. The 12-lead ECG interpretation, being an automated algorithm output, would be assessed for its accuracy as a standalone function.


    7. The Type of Ground Truth Used

    Based on the nature of the device (physiological monitor) and the parameters listed, the ground truth would typically be established by:

    • Reference standards/equipment: For parameters like NIBP, IBP, SpO2, Temperature, CO2, etc., the device's measurements are compared against highly accurate reference measurement equipment or simulators.
    • Known input signals: For features like arrhythmia detection, standardized ECG waveforms with known arrhythmias are often used.
    • Clinical data: For capabilities like 12-lead ECG interpretation, ground truth is usually established by highly experienced cardiologists' interpretations, though the document only refers to equivalence to a previously cleared device for this feature.
    • Specifications: The "acceptance criteria" are derived from the performance specifications of the predicate device and relevant international standards.

    8. The Sample Size for the Training Set

    The document does not describe the development or training of AI/ML algorithms. It is a 510(k) for hardware and software modifications to an existing patient monitor, rather than a novel AI/ML device. Therefore, no information on training set sample size is provided. The functions described (e.g., arrhythmia detection, ST segment analysis) typically rely on established signal processing algorithms, not necessarily machine learning requiring explicit "training sets" in the modern sense.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for AI/ML algorithms, this information is not applicable and not provided in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1