Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223445

    Validate with FDA (Live)

    Device Name
    ArtiFascia
    Date Cleared
    2023-08-10

    (269 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092388, K991413

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ArtiFascia is indicated as dura substitute for the repair of dura mater. ArtiFascia is indicated for defects of 25cm² (3.87 in²) or less in area. For example, 6 cm X 4 cm (24 cm²) would be an acceptable defect size.

    Device Description

    ArtiFascia is an absorbable dural repair graft for the repair of cranial dural defects. ArtiFascia is a highly flexible, easy to handle, non-friable soft matrix composed of synthetic non-woven fibers and a non-porous film. ArtiFascia is packaged in a single-use peelable package and is provided sterile, nonpyrogenic. ArtiFascia readily conforms to the surface of the wound area and is applied to the dural defect by using sutures.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text.

    The provided document describes the K223445 premarket notification for the ArtiFascia device, a dura substitute. The studies primarily focus on demonstrating the device's substantial equivalence to a legally marketed predicate device (Cerafix Dura Substitute). As such, the "acceptance criteria" discussed are largely in the context of equivalence testing and meeting safety and performance benchmarks comparable to the predicate, rather than an AI/ML specific performance metric. The studies described are primarily bench testing, biocompatibility studies, and animal studies, followed by a clinical study. There is no mention of Artificial Intelligence or Machine Learning in the provided document, so there will be no information on acceptance criteria or studies related to AI/ML device performance (e.g., sensitivity, specificity, MRMC studies, standalone performance).


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the tests performed and the results demonstrating compliance or superiority to the predicate device. For the clinical study, the primary endpoint served as the key acceptance criterion for effectiveness.

    Test TypeAcceptance Criteria (Implied / Explicit)Reported Device Performance and Compliance
    Bench Testing
    Morphological EvaluationCompliance with predefined acceptance criteria set per the morphology test protocol.All evaluated test articles were found to comply.
    Tensile StrengthWithstand specific tensile forces.The ArtiFascia patch can withstand tensile forces [Mpa] that are far greater than the predefined acceptance criteria.
    Burst PressureWithstand specific applied pressure (twice the expected intracranial pressures).The ArtiFascia patch can withstand applied pressure [PSI] that is far greater than the predefined acceptance criteria.
    Suture Retention StrengthWithstand suture retention forces at least comparable to the predicate device.The ArtiFascia patch can withstand suture retention forces [N] that are (on average) greater than the predefined acceptance criteria, i.e., the average results as obtained for the predicate device.
    ShrinkageNo notable shrinkage when properly used.When suspension, such as suturing, is utilized, no shrinkage is noted.
    In-Vitro DegradationEquivalent physical properties for e-beam and gamma sterilization after 126 days.The results show that, after 126 days, the physical properties of the e-beam are equivalent to gamma-sterilized ArtiFascia.
    BiocompatibilityCompliance with ISO 10993-1:2018 requirements for human use.Non-cytotoxic, Non-sensitizing, Non-irritating, Non-toxic, Non-pyrogenic, No treatment-related adverse effects observed in implantation (up to 12 months in rabbits, with ongoing minor inflammatory response consistent with longer resorption; predicate resorbed by 6 months), Non-hemolytic, Non-mutagenic, Non-genotoxic. Chemical analysis showed safe margins of safety.
    Animal Study DataExcellent local tissue response and tolerability, comparable to predicate.ArtiFascia (e-beam/gamma) and DuraGen Plus showed excellent local tissue response and tolerability. ArtiFascia was well-tolerated throughout degradation. Minor inflammation at 12 months consistent with longer resorption, but comparable to DuraGen's response. Mostly resorbed by 12 months.
    Clinical StudyNon-inferiority to commercially available dural grafts regarding effectiveness and safety. Absence of CSF fistula and pseudomeningocele within 6 months post-operative.Primary Endpoint Met: No reported cases of CSF fistula in either group. Only one case of CSF pseudomeningocele in the control group at 6 months; no cases in ArtiFascia group. Secondary Endpoints Met: Wound healing, device handling (ease of use, strength suturability, seal quality) were assessed. MRI at 6 months showed no adverse findings for ArtiFascia (compared to control with one pseudomeningocele). Safety Endpoints Met: No changes in neurological status or other neurological symptoms at ≥12 months post-surgery (except one control case unrelated to device/procedure). No abnormal findings at surgical site.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Bench Testing:

      • In-Vitro Degradation: 96 samples (overall)
      • Other bench tests: Specific sample sizes are not explicitly stated for individual tests, but implied to be sufficient for statistical significance.
      • Data Provenance: Not explicitly stated, implied to be laboratory-generated.
    • Biocompatibility Studies:

      • Implantation (ISO 10993-6:2007) / First Animal Study: 19 animals (rabbits)
      • Implantation (ISO 10993-6:2016) / Second Bridging Study: 9 animals (rabbits)
      • Other biocompatibility tests: Sample sizes are not explicitly stated but are standard for the ISO methods referenced (e.g., cell cultures for cytotoxicity, animal subjects for sensitization/irritation/toxicity).
      • Data Provenance: Laboratory research, animal studies conducted under Good Laboratory Practices (GLP).
    • Clinical Study (NEOART study):

      • Test Set (Initial Study): 78 subjects treated (58 in ArtiFascia group, 20 in control group). Total enrolled randomized and treated subjects was 85.
      • Test Set (Long-term data collection): 32 subjects underwent MRI and/or neurological assessment (25 ArtiFascia, 7 control).
      • Data Provenance: Multi-center, prospective, randomized, controlled, single-blinded, parallel group study. Conducted at 7 clinical sites outside-of-the-US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Bench Testing: Ground truth is established by standardized test methods and measurements. Expert interpretation is often part of the process (e.g., visual examination for morphological irregularities), but the number and qualifications of experts are not specified.
    • Biocompatibility Studies: Ground truth is based on standard test results and observations (e.g., "Non-cytotoxic," "Non-sensitizing"). Pathologists/toxicologists would interpret results, but specific numbers and qualifications are not provided.
    • Animal Studies: Assessments included clinical assessments of animal well-being and histological examination of the treatment area. Pathologists/veterinarians would be involved in interpretation, but specific numbers and qualifications are not provided.
    • Clinical Study:
      • Primary Endpoint (CSF fistula/pseudomeningocele): Assessed by MRI imaging and direct observation by clinicians.
      • Secondary Endpoint (MRI assessment): "confirmed in an assessment of a blinded, independent radiologist." The specific number of radiologists is not mentioned, nor are their detailed qualifications (e.g., years of experience).

    4. Adjudication Method for the Test Set

    • Bench Testing, Biocompatibility, Animal Studies: Adjudication methods are not explicitly described but are inherent to following validated test protocols and reporting objective measurements.
    • Clinical Study:
      • Primary Endpoint: Assessment was based on MRI imaging and clinical observation. No explicit multi-reader adjudication method (e.g., 2+1, 3+1) is mentioned for the primary endpoint's determination.
      • MRI assessment (Secondary/Long-term): Confirmed by a "blinded, independent radiologist." This implies a single independent read, rather than a multi-reader consensus/adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The provided text does not mention any AI or Machine Learning components. Therefore, no MRMC study involving AI-assisted human readers was conducted or reported.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. The provided text does not mention any AI or Machine Learning components. Therefore, no standalone algorithm performance study was conducted or reported.


    7. The Type of Ground Truth Used

    • Bench Testing: Objective physical measurements and observations based on standardized test methods (e.g., force, pressure, visual inspection, length measurements).
    • Biocompatibility: Results derived from established ISO standard test methods, involving cell culture assays, animal models, and chemical analysis, interpreted by qualified personnel (e.g., toxicologists, pathologists).
    • Animal Studies: Histological examination of tissues and clinical observation of animal well-being.
    • Clinical Study:
      • Primary Endpoint: Clinical observations (wound drainage), and MRI imaging interpreted by clinicians and an independent, blinded radiologist. This is effectively expert consensus/clinical outcome based on imaging and direct observation.
      • Secondary Endpoints: Direct observation for wound healing/device handling, and MRI imaging for tissue changes.
      • Long-term follow-up: MRI imaging and neurological assessment (clinical observation).

    8. The Sample Size for the Training Set

    Not applicable/Not mentioned. Since no AI/ML component is mentioned, there is no discussion of a training set for an algorithm.


    9. How the Ground Truth for the Training Set was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153613

    Validate with FDA (Live)

    Date Cleared
    2016-03-16

    (90 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K092388 DuraGen Plus™ Dural Regeneration Matrix

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cerafix® Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 1.9 in² (12.5cm²) or less in area. For example, 1.2 in x 1.6 in (3 cm x 4 cm) would be an acceptable defect size.

    Device Description

    Cerafix® Dura Substitute is a resorbable implant for repair of dural defects and is to be used with tensionless sutures. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.

    AI/ML Overview

    The provided text describes the Cerafix® Dura Substitute, a medical device intended for the repair of dura mater, and its journey through FDA 510(k) clearance. The document details the device's characteristics, indications for use, and the non-clinical testing performed to establish its substantial equivalence to predicate devices.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for many of the mechanical and biological tests were framed as "Equivalent to Predicate or Reference Device" or "Meets Final Device Specification." For some, specific thresholds were mentioned.

    TestAcceptance CriteriaReported Device Performance
    Mechanical Testing
    ThicknessEquivalent to Predicate or Reference DevicePASS
    Mass per AreaEquivalent to Predicate or Reference DevicePASS
    Tensile StrengthEquivalent to Predicate or Reference DevicePASS
    Suture Pull-Out StrengthEquivalent to Predicate or Reference DevicePASS
    Burst StrengthEquivalent to Predicate or Reference Device; and burst strength greater than anticipated intracranial pressuresPASS (burst strength greater than anticipated intracranial pressures)
    Shrink TemperatureShow stability at applicable temperaturesPASS (showed stability)
    Fiber DiameterMeets Final Device SpecificationPASS (meets specification)
    Pore SizeMeets Final Device SpecificationPASS (meets specification)
    Biocompatibility Testing
    ISO Cytotoxicity MEM ElutionNon-cytotoxicCell culture exhibited no reactivity; non-cytotoxic.
    Guinea Pig Maximization - SensitizationNon-irritating, no sensitization responseDid not elicit a sensitization response; non-irritant.
    Intracutaneous Irritation ReactivityNon-irritatingNon-irritating.
    Hemolysis AssayNon-hemolyticFound to be non-hemolytic.
    Genotoxicity (Mouse Lymphoma Assay)Non-genotoxicEquivalent to negative control; non-genotoxic.
    Genotoxicity (Mouse Micronucleus Assay)Non-mutagenicConsidered non-mutagenic.
    Genotoxicity (Bacterial Mutagenicity)Non-mutagenicConsidered non-mutagenic.
    Pyrogenicity (Rabbit Pyrogen Test)Non-pyrogenicExhibited a negative response; non-pyrogenic.
    Acute Systemic ToxicityNon-toxicConsidered non-toxic.
    Endotoxin TestingLess than 2.15 EU/deviceLess than 2.15 EU/device; non-pyrogenic.
    Subchronic Toxicity (90-day animal study)Non-toxicShowed the device to be non-toxic.
    Chronic Toxicity (180-day animal study)Non-toxicShowed the device to be non-toxic.
    Side-by-Side Animal StudyEquivalent safety and performance to predicate deviceShowed equivalent safety and performance.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each mechanical test (e.g., number of samples tested for tensile strength or burst strength). It mentions "side-by-side bench testing versus the predicate or commercially available reference device" for mechanical tests, and for biocompatibility, it refers to standard ISO/ASTM tests using animals (e.g., guinea pigs, rabbits, mice) and cell cultures. The data provenance is pre-clinical testing, likely conducted in a laboratory setting. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of human data, as this is a pre-market clearance based on non-clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for this device's clearance is based on established scientific principles and comparison to legally marketed predicate devices through defined acceptance criteria in mechanical and biocompatibility testing, not on expert consensus of clinical data.

    4. Adjudication method for the test set

    Not applicable. This device clearance relies on objective laboratory and animal testing, not human-based adjudication of clinical outcomes or images.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to the pre-market clearance of a physical medical implant (dura substitute), not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness is discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an algorithm or AI device.

    7. The type of ground truth used

    The "ground truth" for this regulatory submission is a combination of:

    • Predicate Device Equivalence: The primary ground truth is established by demonstrating that the Cerafix® Dura Substitute's technological characteristics, performance, and safety are substantially equivalent to a legally marketed predicate device (Ethisorb™ Dura Patch) and a reference device (DuraGen Plus™ Dural Regeneration Matrix).
    • Established Scientific Standards: Compliance with ISO and ASTM standards for biocompatibility and mechanical properties (e.g., non-cytotoxic, non-pyrogenic, appropriate burst strength).
    • Animal Study Outcomes: Equivalence in safety and performance based on side-by-side animal implantation studies compared to the predicate device.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this device's pre-market clearance, as it's not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1