510(k) Data Aggregation

    K Number
    K250351

    Validate with FDA (Live)

    Date Cleared
    2025-10-10

    (245 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092309

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single Use Electrosurgical Knife KD-612 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.

    The Single Use Electrosurgical Knife KD-620UR is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.

    The Single Use Electrosurgical Knife KD-650 Series is intended to be used to cut tissue using high-frequency current within the digestive tract in combination with an endoscope and electrosurgical units.

    Device Description

    The Single Use Electrosurgical Knife was designed to be used with compatible Olympus endoscopes, A-Cord (power cable), and electrosurgical units to cut tissue using high-frequency current within the digestive tract. The subject device is inserted endoscopically into the body cavity and tissue is cut by applying a high frequency current. The subject device is a sterile (EO) single use electrosurgical knife comprised of a handle and an insertion portion. The handle portion consists of the body and the slider. The insertion portion consists of the distal end with one lumen for the stainless-steel cutting knife and the sheath portion. The slider extends the cutting knife from the distal end of the sheath/tube when pushed and retracts the cutting knife into the sheath/tube when the slider is pulled.

    The Single Use Electrosurgical Knife is available in three different models: KD-612, KD-620UR, and KD-650. The primary difference between the models is the tip shape of the distal end.

    The subject device has the same technological characteristics and similar design as the applicable predicate device.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K182408

    Validate with FDA (Live)

    Date Cleared
    2019-05-15

    (253 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K092309

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KD-625
    These instruments have been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the digestive tract.
    These instruments are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.

    KD-645
    This instrument has been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue using highfrequency current within the digestive tract.
    This instrument is indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.

    Device Description

    These instruments have been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the digestive tract. These instruments are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.
    This system consists of the electrosurgical knives (subject devices), flushing pump, and water tubes (accessories). The subject devices are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope. They will be marketed as sterilized (ETO) devices for single use.

    AI/ML Overview

    This document describes the premarket notification for the Olympus Single Use Electrosurgical Knives KD-625 and KD-645. It does not contain information about an AI/ML device. Therefore, I cannot provide details on acceptance criteria and study designs related to an AI/ML device from this document.

    However, I can extract the general performance data and acceptance criteria mentioned for these electrosurgical knives:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-AI/ML Device):

    Performance Test CategoryAcceptance Criteria / StandardReported Device Performance
    Sterilization/Shelf-lifeFDA's Guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" & ASTM F1980-16 for accelerated aging.Sterilization/shelf-life testing was conducted in accordance with the specified FDA guidance and ASTM F1980-16. Accelerated aging tests were completed. Real-time aging tests for three years will be performed to further support stability.
    BiocompatibilityFDA's Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'."Biocompatibility testing included: Cytotoxicity Study (Colony Assay), Intracutaneous Study (Rabbits), Guinea Pig Maximization Sensitization Test, Systemic Toxicity Study (Mice), and USP Rabbit Pyrogen Study (Material Mediated). These tests were conducted in accordance with ISO 10993-1.
    Electrical Safety & EMCAAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Safety) & IEC 60601-2-18:2009 (Safety) & IEC 60601-1-2 Edition 3: 2007 (EMC).The device system complies with the specified AAMI/ANSI and IEC standards for electrical safety and electromagnetic compatibility.
    Performance Testing (Bench)Established in-house acceptance criteria based on design requirements and risk analysis (ISO 14971:2007).Bench testing was conducted to ensure the device performs as intended and meets design specifications. This included: Evaluation Test for Performance to Cut Digestive Tract Mucosa, Evaluation Test for Lifting of Digestive Tract Mucosa, Histopathological Evaluation Test of Thermal Damage on Digestive Tract, and Evaluation Test for safety against perforation by saline feeding function. All tests assessed according to design requirements.
    Risk AnalysisEstablished in-house acceptance criteria based on ISO 14971:2007.Risk analysis was conducted, and the design verification tests and acceptance criteria were identified and performed as a result of this assessment.

    Since this document is for a traditional electrosurgical knife and not an AI/ML device, the following points are not applicable or cannot be extracted from the provided text:

    1. Sample size used for the test set and the data provenance: Not applicable for this type of device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for AI/ML models typically involves expert annotation of data.
    3. Adjudication method: Not applicable.
    4. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable, as this is for an electrosurgical tool, not an interpretive diagnostic aid.
    5. Standalone (algorithm only without human-in-the loop performance) performance: Not applicable.
    6. Type of ground truth used: Not applicable for an AI/ML device; for this device, ground truth established by physical and functional testing.
    7. Sample size for the training set: Not applicable (no AI/ML model for training).
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152771

    Validate with FDA (Live)

    Date Cleared
    2016-06-17

    (266 days)

    Product Code
    Regulation Number
    876.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092309

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.

    Device Description

    The SB Knife™ is a single-patient use EtO-sterilized hand-held device that connects to an external high frequency (HF) generator and is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current. Varying generator power, cycle and duration results in a control of these surgical capabilities. The SB Knives are designed to connect with Olympus and ERBE connectors and use the disposable or reusable Olympus or ERBE grounding plates and connecting cables intended for the generators.

    The device is not life supporting or life sustaining. It is designed as a short-term limited contact device (less than 24 hours), has no software, and has been certified to comply with IEC 60601-1 and IEC 60601-2-2 and IEC 60601-2-18 by UL Japan

    AI/ML Overview

    This document is a 510(k) premarket notification for the SB Knife, an electrosurgical knife. It does not describe an AI medical device or a study involving acceptance criteria for an AI device's performance. Therefore, I cannot extract the requested information regarding acceptance criteria and studies that prove the device meets these criteria in the context of AI.

    The document discusses the following:

    • Device Name: SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type, SB Knife Jr Type - Long
    • Regulation Number: 21 CFR 876.4300
    • Regulation Name: Endoscopic Electrosurgical Unit and Accessories
    • Regulatory Class: Class II
    • Product Code: KNS
    • Predicates: Olympus Medical Systems, Corp. IT Knife 2/ Hook Knife/ Flex Knife/ Triangle Tip Knife Electrosurgical Knife (K092309)
    • Indications for Use: To be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors, and patient grounding plates to cut tissue within the digestive tract using high-frequency current.
    • Non-Clinical Testing: Comparative bench and animal tissue testing was performed for tissue ablation, resection, incision, cauterization, hemostasis, coagulation, vaporization, and dissection of tissues using Olympus and ERBE monopolar HF generators.
    • Clinical Testing: "Clinical testing was not performed for this medical device."

    Since the SB Knife is a physical electrosurgical tool and not an AI device, the provided information does not contain the specific details requested for an AI acceptance criteria study (e.g., sample size for test/training sets, experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for AI).