Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242212

    Validate with FDA (Live)

    Date Cleared
    2024-08-27

    (29 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K133189, K071691, K173450, K092896, K091734

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    100 mL/200 mL Syringe: The Disposable Syringe (Kit) (100 mL or 200 mL Syringe for the injection of contrast media orsaline for CT and MR imaging, for use on Nemoto injectors. 150 mL Syringe: The Disposable Syringe (Kit) (150 mL Syringe for the injection of contrast media or saline forangiography. for use on Nemoto injectors. J-Tube: J-Tube is to aid in transferring a media from its container (typically a vial or bottle) into the barrel of the Disposable Syringe (Kit). Extension Tubes: The Extension Tube is to provide a connection between the syringe barrel and the needle or catheter into the patient.

    Device Description

    The "Disposable Syringe (Kit)" (Subject Device) is specifically designed for use with the Nemoto power injection systems marketed and sold in the USA using 510(k) references K133189, K071691, K173450, K092896, K091734. The Subject Device is a package provided to the end user that contains one or two plastic syringes, plastic tubing used for connecting syringe to needle or catheter (Extension Tubing) and a plastic J-shaped tube for transferring contrast media or physiological saline into the plastic syringe (J-Tube). The Subject Device, in accordance with the Indications for Use, may be used in conjunction with CT or MRI applications or in conjunction with angiographic applications. The key difference in the application as it relates to the Subject Device is the pressure required and the flow rates required for the Subject Device. The Subject Device when used for the CT and MRI applications must support a maximum pressure of 300PSI and a maximum flow rate of 10mL/second. The Subject Device when used for the angiographic applications must support a maximum pressure of 1200PSI and a maximum flow rate of 30ml/second. The maximum values related to the application are controlled by the Nemoto power injection system which are fully described in the aforementioned 510(k)s.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Disposable Syringe (Kit)) and does not contain information about an AI/ML device. Therefore, the questions related to AI/ML device acceptance criteria and study details cannot be answered based on the provided input.

    However, I can provide a summary of the non-clinical performance data presented for the Disposable Syringe (Kit), which demonstrates it meets its acceptance criteria based on established standards.

    Non-Clinical Performance Data for Nemoto Disposable Syringe (Kit)

    The Nemoto Disposable Syringe (Kit) underwent extensive non-clinical testing to demonstrate substantial equivalence to its predicate devices. The device passed numerous tests in accordance with internal requirements, national standards, and international standards.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test / StandardReported Device Performance
    BiocompatibilityCytotoxicity testing (ISO 10993-5)Passed
    Sensitization testing (ISO 10993-10)Passed
    Irritation testing (ISO 10993-23)Passed
    Acute systemic toxicity testing (ISO 10993-11)Passed
    Hemolysis testing (ISO 10993-4, ASTM F 756)Passed
    Pyrogenicity testing (ISO 10993-11, USP-NF 38 <151>)Passed
    Sterility & ContaminationParticulate Contamination testing (USP <788>)Passed
    Endotoxin testing (USP-NF <85> and <161>)Passed
    Sterilization validation (demonstrates SAL of 10-6)Passed
    Ethylene Oxide Sterilization Residuals (ISO 10993-7)Passed
    Syringe Mechanical PerformanceLubricant testing (modified ISO 7886-1)Passed
    Tolerance on Graduated Capacity (ISO 7886-1)Passed
    Scale testing (ISO 7886-1)Passed
    Numbering of Scales testing (ISO 7886-1)Passed
    Overall length of scale to nominal capacity line (ISO 7886-1)Passed
    Position of Scale testing (ISO 7886-1)Passed
    Barrel testing (ISO 7886-1)Passed
    Plunger Stopper/Plunger Assembly testing (ISO 7886-1)Passed
    Nozzle Lumen testing (ISO 7886-1)Passed
    Dead Space testing (ISO 7886-1)Passed
    Freedom from Air past Plunger Stopper (ISO 7886-1)Passed
    Fit of Plunger Stopper/Plunger in Barrel (ISO 7886-1)Passed
    Liquid Leakage past Plunger Stopper (in-house pressure)Passed
    Chemical testing (ISO 7886-1, ISO 8536-9)Passed
    Extension Tubes & ConnectorsTransparency testing (ISO 8536-9)Passed
    Tensile Strength testing (ISO 8536-9)Passed
    Leakage testing (ISO 8536-9)Passed
    Injection Needle testing (ISO 8536-9)Passed
    Connection force (in-house testing)Passed
    Positive Pressure Liquid Leakage (ISO 80369-7)Passed
    Sub-Atmospheric Pressure Air Leakage (ISO 80369-7)Passed
    Stress Cracking testing (ISO 80369-7)Passed
    Resistance to Separation from Axial Load (ISO 80369-7)Passed
    Resistance to Separation from Unscrewing (ISO 80369-7)Passed
    Resistance to Overriding testing (ISO 80369-7)Passed
    Overall Device PerformanceCompatibility (in-house testing)Passed
    Shelf Life Testing (supports 5 years)Passed
    Transportation Testing (ASTM F88/88M, etc.)Passed
    Usability Engineering (IEC 62366-1)Evaluated

    2. Sample Size for Test Set and Data Provenance:
    The document does not specify the sample sizes used for each non-clinical test. The tests are general performance and materials tests, not based on patient data.

    3. Number of Experts and Qualifications for Ground Truth:
    Not applicable. These are non-clinical engineering and laboratory tests, not clinical studies requiring expert ground truth for interpretation.

    4. Adjudication Method:
    Not applicable. This is not a clinical study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    Not applicable. This is not an AI-assisted diagnostic device.

    6. Standalone (Algorithm Only) Performance Study:
    Not applicable. This is a physical medical device.

    7. Type of Ground Truth Used:
    For the non-clinical tests, the "ground truth" is defined by the specific parameters and thresholds set by the relevant ISO standards (e.g., ISO 10993, ISO 7886-1, ISO 8536-9, ISO 80369-7) and internal requirements. These standards specify acceptable limits for various properties like cytotoxicity, leakage, tensile strength, and dimensions.

    8. Sample Size for Training Set:
    Not applicable. This is a physical medical device, not an AI/ML system requiring a training set.

    9. How Ground Truth for Training Set Was Established:
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201395

    Validate with FDA (Live)

    Date Cleared
    2020-10-28

    (154 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is intended for the injection of contrast media or saline. This syringe use with US legally marketed angiographic injectors.

    Device Description

    The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1. The connection tubes, which is used to connect the syringe and the catheter. The tubes are available in two configurations, which are Type Y and Type T tube. The different between these tubes is tube shape, which is available in Y shape for Y tube and T shape for T tube. J shape tube, which is used to draw contrast media/ saline into the syringe barrel. Spikes, which are used to draw contrast media/saline into the syringe barrel.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: "Sterile High-pressure Angiographic Syringes for Single-use". This type of submission focuses on demonstrating "Substantial Equivalence" to a legally marketed predicate device, rather than proving novel effectiveness or safety through extensive clinical trials.

    As such, the document does not describe a study to prove the device meets acceptance criteria in the way you've outlined for an AI-powered diagnostic device. Instead, it details non-clinical tests to show the proposed device performs comparably to the predicate and meets relevant engineering standards.

    Therefore, many of your requested points regarding Acceptance Criteria and Study Proof (especially related to AI, MRMC, expert consensus, and ground truth establishment) are not applicable to this 510(k) submission for a non-AI medical device.

    However, I can extract the relevant information from the document regarding the "acceptance criteria" (in the sense of standards compliance and comparative performance) and the "study" (non-clinical tests) that proves these criteria are met.

    Here's a breakdown based on the provided document:


    Device: Sterile High-pressure Angiographic Syringes for Single-use

    Device Type: Non-AI Medical Device (Angiographic Syringes)

    Study Purpose: To demonstrate Substantial Equivalence (SE) to a predicate device (Shenzhen BaoAn Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use, K151960) through non-clinical testing and comparison.


    1. A table of acceptance criteria and the reported device performance

    For a non-AI device like this syringe, "acceptance criteria" primarily refer to adherence to recognized international standards and demonstration of comparable performance to a predicate device. "Reported device performance" is the outcome of non-clinical tests demonstrating compliance.

    Acceptance Criteria (Standard / Property)Basis / Test MethodReported Device Performance (Result & SE Justification)
    Performance
    Compliance with Syringe StandardsISO 7886-1 (Sterile hypodermic syringes for single use), ISO 594-1 (Conical fittings with a 6% (Luer) taper), ISO 594-2 (Lock Fitting)Complied with ISO 7886-1, ISO 594-1, and ISO 594-2. This demonstrates performance equivalent to the predicate.
    Seal StrengthASTM F88/F88M-15 (Standard test method for seal strength of flexible barrier materials)Test results met the standard.
    Seal Leaks (Porous Medical Package)ASTM F1929-15 (Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration)Test results met the standard.
    Connection Tube PerformanceISO 594-1 and ISO 594-2Performance test results of connection tube meet the criteria. Even with more product specifications (various lengths), the performance is maintained.
    J-shape Tube LeakageNon-standardized test, implied to ensure functionality.Performance test results showed no leaks.
    Biocompatibility
    Biocompatibility AssessmentISO 10993-1: 2018 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)Overall biocompatibility demonstrated; differences in materials (e.g., Piston, Filter membrane) from predicate are deemed not to affect SE due to successful biocompatibility testing (Analysis 2).
    CytotoxicityISO 10993-5:2009 (Tests for In Vitro cytotoxicity)No Cytotoxicity. (Same as predicate)
    IrritationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation. (Same as predicate)
    SensitizationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Sensitization. (Same as predicate)
    Systemic ToxicityISO 10993-11:2017 (Tests for systemic toxicity)No Acute Toxicity. (Same as predicate)
    HemolysisASTM F 756-17 (Standard practice for assessment of hemolytic properties of material)No Hemolysis. (Same as predicate)
    PyrogenicityUSP 41-NF36:2018 <151> (Pyrogen Test (USP Rabbit Test))No Pyrogen. (Same as predicate)
    Sterility & Particulate Matter
    Bacterial Endotoxins LimitUSP 41-NF36 2018 <85> (Bacterial Endotoxins Limit)Test results met the standard.
    Sterility Assurance (Post-sterilization)ISO 11737-2:2009 (Sterilization of medical devices- Microbiological methods- Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process)Test results met the standard. (EO sterilized, same as predicate).
    Ethylene Oxide Sterilization ResidualsISO 10993-7:2008 (Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals)Test results met the standard. (EO sterilized, same as predicate).
    Particulate MatterUSP <788> (Particulate matters)Complied with USP 788. (Same as predicate)
    Material & Physical Compatibility
    Syringe Volume & Configuration Range Compatibility with Predicate's PerformanceComparison of various single and dual shot volumes (50ml to 200ml) and accessory types (connection tubes, J-shape tube, spikes) with the predicate device.While the proposed device has a wider range of sizes/configurations than the predicate, non-clinical tests confirmed that all these configurations meet the performance criteria of relevant ISO standards. This difference is deemed not to affect SE. (Analysis 1)
    Patient-Contact Material CompatibilityComparison of materials used for various components (Barrel, Piston, Connection Tube, Luer Connectors, Spike, Filter Membrane, J-shape tube) with the predicate. Explicitly, changes in Piston (Medical rubber vs. Polyisoprene rubber) and Filter Membrane (Medical rubber vs. Polypropylene) from the predicate.Differences in Piston and Filter Membrane materials exist. However, the contact classification (blood path, indirect limited contact) remains the same as the predicate. Extensive biocompatibility testing (Cytotoxicity, Irritation, Sensitization, Acute Toxicity, Hemolysis, Pyrogen) on the
    proposed device's materials showed no adverse effects. Therefore, these material differences are not considered to affect substantial equivalence. (Analysis 2)
    Labeling21 CFR Part 801 (Medical Device Labeling)Complied with 21 CFR Part 801. (Same as predicate)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the number of units/samples used for each non-clinical test (e.g., how many syringes were tested for seal strength, or how many batches for endotoxin). It only states that "Non clinical tests were conducted to verify that the proposed device met all design specifications." The specific number of test replicates/samples would typically be detailed in the full test reports, which are not part of this summary document.
    • Data Provenance: The tests are non-clinical (laboratory/bench testing) performed to verify compliance with standards. The company "Youwo (Guangzhou) Medical Device Co., Ltd." is based in Guangzhou, P.R. China, implying the testing was conducted in China, likely at their own facilities or certified labs in China. The data is entirely lab-generated, not from human subjects. The nature of the testing is prospective, in the sense that the tests were performed specifically for this submission on new device samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a non-AI, non-diagnostic medical device. There is no concept of "ground truth" established by human experts in the context of diagnostic interpretation. The "ground truth" for the performance of these syringes is defined by the objective pass/fail criteria of the recognized engineering and biocompatibility standards (e.g., is the seal strong enough? Is there pyrogen presence?). Compliance is measured by laboratory equipment and established protocols, not expert human review of an image or signal.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As per point 3, there's no human diagnostic interpretation to adjudicate. Test results are objective measurements against pre-defined thresholds from standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is not an AI diagnostic tool, nor is it intended to assist human readers in interpreting medical cases. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or AI device. Standalone performance refers to the device's ability to meet its specifications independently, which is what the non-clinical tests implicitly demonstrate (e.g., a syringe's seal strength is measured independently of human operation).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standards-Defined Ground Truth: The "ground truth" here is the adherence to internationally recognized performance and safety standards (e.g., ISO, ASTM, USP). The device is "true" if it passes the specified tests according to the criteria defined within those standards. For example, a "true" biocompatible device produces no cytotoxicity according to ISO 10993-5. This is determined by laboratory measurements and chemical/biological assays, not clinical outcomes or expert diagnostician consensus.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device. There is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/machine learning device. No ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143538

    Validate with FDA (Live)

    Date Cleared
    2015-08-14

    (242 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    Nemoto Sonic Shot GX, K091734, E-Z-EM Empower MR Injection System, K062449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MEDRAD® MRXperion MR Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media and saline during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners that have a magnetic field strength between 0.7 and 3.0 Tesla. Only trained healthcare professionals are intended to operate this device.

    Device Description

    The MEDRAD MRXperion MR Injection System is a software-controlled medical device used for the administration of intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) procedures. Commonly referred to as an automated injection system, it is designed to allow a user to fill disposable syringes and then to perform an injection with a user-programmed volume and flow rate.

    The MEDRAD MRXperion MR Injector consists of two basic modules: a Scan Room Unit (SRU) and a Control Room Unit (CRU).

    The SRU is comprised of an integral injector head, base assembly, pedestal, and a power supply.

    • The injector head of the SRU physically performs the injection. .
    • . The base assembly functions as the interface between the CRU, the injector head, and the SRU power supply.
    • . The injector head and base are located on a pedestal is designed with locking casters to allow the SRU to be moved when not connected to a patient.
    • . To power the base assembly, power is received from the AC mains and is converted to DC voltage by the SRU power supply.

    The CRU consists of an All-in-One Computer (AlOC), pod, their dedicated power supplies, mechanical stand, and an optional hand switch.

    • The AIOC provides a platform for the graphical user interface for the injector as well as ● the optional informatics applications. From the AIOC, an operator can use the touchscreen display to manage protocols, arm and disarm the injector, review injection realtime progress/feedback and history, access eGFR and patient Weight-Based Dosing calculators, and set system configuration options. The operator can also use the features of the optional informatics device from the AIOC.
    • The pod provides injection start and stop functionality and contains the safety controls . for the CRU.
    • . To power the AIOC and pod, each component receives AC mains power from its own dedicated off-the-shelf power supply.
    • The AIOC and pod are mounted on a desk top mechanical stand.
    • . An optional hand switch gives the operator an additional means to start, hold, and stop an injection from the Control Room. The hand switch contains a light that identifies the state of the injector.
    AI/ML Overview

    The provided document describes the MEDRAD MRXperion MR Injection System and its associated syringe kit. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All testing passed and the demonstrated product performance met all prior established acceptance criteria." However, specific numerical acceptance criteria for each test are not explicitly detailed in the provided text. Instead, it lists the types of tests performed and confirms they met their objectives.

    CategoryAcceptance Criteria (Not explicitly quantified in text, stated as "met all prior established acceptance criteria")Reported Device Performance (Summary from text)
    Device PerformanceProduct specifications and performance requirements.Verification of system disposable filling and preparation, protocol management, flow rates, pressures, timers, and calculator accuracy. Device unaffected by environmental conditions.
    Disposables PerformanceMechanical functions and pressure capabilities (ISO-594), compatibility with MR contrast agents and other chemical agents (ISO-8536-4), and packaging (ISO-11607).Testing performed on aged and shipping-conditioned samples (ASTM 4169). Verification of syringe and connector tubing mechanical functions and pressure capabilities, compatibility with MR contrast agents and other chemical agents, and packaging.
    BiocompatibilityRequirements of ISO 10993 for an external communicating, indirect contact, less than 24-hour duration device.Syringe and connector tubing kit compliant with ISO 10993 including tests for Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Injection, Materials Mediated Pyrogen, Hemolysis, Partial Thromboplastin Time, Platelet & Leukocyte Counts, and USP physiochemical testing.
    SterilizationSterility Assurance Level (SAL) of 10-6 (ISO 11137-1, 11137-2, 11137-3).Sterilization conditions validated to provide SAL of 10-6 for syringe and connector tubing kit.
    Safety and CompatibilityConfigurations, specifications, circuitry, compliance with IEC 60601-1 and EMC, electrical safety controls, detection of failures, programming keys, sensors, and MRI compatibility (0.7T to 3.0T).Verification of configurations and specifications, circuitry, compliance with IEC 60601-1 and EMC requirements, electrical safety controls, ability to detect failures, programming keys, sensors. MRI compatibility (0.7T to 3.0T) including homogeneity, signal-to-noise, MR artifacts, and susceptibility.
    Shelf-Life & ShippingSystem performance not affected by three-year aging and shelf-life packaged transit and storage (ISTA 1E).System performance not affected by three-year aging and packaged transit/storage (ISTA 1E).
    ReliabilityCapability to sequentially and repeatedly meet system performance requirements with no degradation from simultaneous Informatics processes.Statistical methods used to demonstrate capability. No degradation when Injection System and Informatics processes run simultaneously.
    Simulated Use & Human FactorsClinical user needs met by design (ANSI/AAMI HE75:2009 and EN 62366:2008), raising no new safety/effectiveness questions.Device and disposables used in a simulated clinical environment. No new safety or effectiveness questions raised.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state numerical sample sizes for most of the individual tests. It uses general statements like "testing was performed."

    • Sample Size: Not specified quantitatively for most tests.
    • Data Provenance: The testing was conducted by the manufacturer, Bayer Medical Care, Inc. The data is retrospective in the sense that the studies were completed to support the 510(k) submission, but it's not data from clinical practice. The country of origin for the data is not specified, but the manufacturer is based in the USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The testing described is non-clinical bench testing, not a study requiring expert-established ground truth in the context of diagnostic interpretation.


    4. Adjudication Method for the Test Set

    This information is not applicable as the described testing is non-clinical bench and performance testing, not a study involving human interpretation or adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. The device is an MR injection system, not an AI diagnostic tool.
    • This section is not applicable to the device described.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No, a standalone algorithm performance study was not done. The device is a physical injection system with software control, not a standalone diagnostic algorithm.
    • This section is not applicable to the device described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical testing, the "ground truth" was established by engineering and quality standards, specifications, and regulatory requirements (e.g., ISO, ASTM, IEC standards, and internal product specifications). For example:

    • Physical measurements (flow rates, pressures, volumes) were compared against defined engineering specifications.
    • Biocompatibility was assessed against ISO 10993.
    • Sterility was assessed against ISO 11137 standards.
    • MRI compatibility was tested against defined performance metrics within specified magnetic field strengths.

    8. The sample size for the training set

    • Not applicable. This device is an injection system and does not involve AI or machine learning algorithms that require a training set in the conventional sense. The "training" here would refer to the development and iterative testing of the device's hardware and software, rather than training a model on data.

    9. How the ground truth for the training set was established

    • Not applicable. As stated above, there is no AI training set as defined in this context. The "ground truth" for the device's design and engineering would be based on established engineering principles, international standards, and internal product specifications validated through extensive bench testing and simulated use.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1