    K Number
    K103547

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-03

    (336 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052777, K022181, K042785, K091534

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ophthalmology:
    532nm Laser

    • Retinal Photocoagulation
    • Pan Retinal Photocoagulation
    • Endophotocoagulation
    • Macular photocoagulation to treat leaking vessels
    • Laser Trabeculoplasty
      670nm Laser
    • Retinal/Pan Retinal Photocoagulation
      The intended use has not changed from the predicate devices (K052777, K022181, K042785, and K091534)
    Device Description

    The LightLas 532/670 system consists of a Laser Console where the Multi-Wavelength Laser is housed along with the Electronic Control system and Power Supplies and various Laser Delivery Units (LDU's). The LDU's include:

    • Slitlamp Integrated into CSO model SL980.
    • Slitlamp Attachment for CSO model SL990 and other Haag Streit clones.
    • Slitlamp Attachment for Zeiss model SL30 Slitlamp.
    • Laser Indirect Ophthalmoscope (LIO) using a Heine Omega 180 BIO.
    • Endophotocoagulation handpieces (Endoprobes).
    AI/ML Overview

    This submission describes a medical laser system and focuses on its substantial equivalence to predicate devices, rather than presenting a study with specific acceptance criteria and performance metrics typically associated with AI/software devices. Therefore, much of the requested information cannot be directly extracted from the provided text.

    Here is an analysis based on the available information:

    Key Takeaway: The submission for the LightLas Multi-Wavelength Medical Laser System (LightLas 532/670) focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving performance against specific acceptance criteria through a novel study. This means the device's performance is inferred to be similar to legally marketed devices based on shared technological characteristics and intended use, not through a separate quantitative performance study.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance Criteria (Expected/Implied)Reported Device Performance
    Intended UseEquivalent to predicate devices"The intended use has not changed from the predicate devices." (Explicitly stated)
    Technological CharacteristicsSame general design, fundamental scientific technology, functional elements (treatment wavelengths, treatment power, spot size, cooling system), control systems (door interlock, safety systems, displays) as predicate devices. No new hazards introduced."The LightMed Multi-wavelength Medical Laser system... has the same intended use, general design and fundamental scientific technology as the predicate devices... Also the operating controls and functions are equivalent to these products. They have the same functional elements such as treatment wavelengths, treatment power, spot size and cooling system. Control systems such as the door interlock, and the safety systems and displays are constantly monitored... There are no new hazards introduced..." (Explicitly stated)
    Clinical EffectivenessEquivalent to previously marketed products based on specifications affecting treatment modality."Therefore the Clinical effectiveness of the LightLas 532/670 is equivalent to the previously marketed products as these specifications are the key factors that will affect the treatment modality." (Explicitly stated)
    Safety TestingMet appropriate safety standards."The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the LightLas 532/670..." (Implied that tests were passed, but no specific criteria or results are detailed).
    Performance TestingMet appropriate performance standards."The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the LightLas 532/670..." (Implied that tests were passed, but no specific criteria or results are detailed).
    Functional TestingMet appropriate functional standards."The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the LightLas 532/670..." (Implied that tests were passed, but no specific criteria or results are detailed).

    2. Sample size used for the test set and the data provenance:

    • Not applicable / Not provided. This submission is for a physical medical laser device, not a software or AI device that would typically use a test set of data. The "testing" mentioned refers to engineering and safety performance of the hardware.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not provided. As above, this is not a data-driven AI/software device. The "ground truth" here is the established safety and effectiveness of the predicate devices in clinical practice, as recognized by the FDA through their previous clearances.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable / Not provided. No specific test set with human adjudication is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted diagnostic or therapeutic device that involves "human readers" in the sense of interpreting medical images or data. It is a direct medical laser system for ophthalmic procedures.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable / Not provided. This is not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Indirect Ground Truth: Clinical Efficacy and Safety of Predicate Devices. The "ground truth" for this substantial equivalence claim is the well-established safety and efficacy of the predicate devices. The new device is deemed equivalent because its technological characteristics and intended use are the same, implying it will perform in a clinically similar and safe manner.

    8. The sample size for the training set:

    • Not applicable / Not provided. There is no mention of a training set as this is not a machine learning/AI device.

    9. How the ground truth for the training set was established:

    • Not applicable / Not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100558

    Validate with FDA (Live)

    Date Cleared
    2010-07-22

    (143 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K072034, K083613, K091534, K053604, K090164

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quanta System QUANTA Diode Laser Family, including the QUANTA532, QUANTA808, QUANTA940, QUANTA980, QUANTA1064, QUANTA1320, QUANTA1470, and QUANTA1950 (and all their double wavelength combination and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery(PLDD), Gastroenterology, Head/neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures. The QUANTA980 and QUANTA1470 Diode Laser are further indicated for laser assisted lipolysis.

    Device Description

    The QUANTA Diode Laser Family is designed with 4 major subsystems: (1) an external structure; (2) power electronics; (3) display with control electronics, which controls the power electronics, the user interface and the laser source temperature via a thermostat board: and (4) the laser system with an opto-mechanical block composed of the laser source, the Peltier cooling system with dissipater and fans, the fiber launching system, the red diode aiming beam, and the power calibration system. The fiber is connected to the system through an SMA 905 socket on the front panel. In addition to the four subsystems, the QUANTA Diode Laser Family incorporates several safety features, including a remote interlock, an emergency red push button and a key switch. The external accessories include separate optical fibers and hand pieces for dental, dermatological and surgical applications, or for endovascular applications and laser assisted lipolysis.

    AI/ML Overview

    The Quanta System QUANTA Diode Laser Family for surgical applications has been recognized as substantially equivalent to various predicate devices (K072034, K083613, K091534, K053604, K090164) and does not have specific acceptance criteria or an analytical study presented in the provided text to demonstrate its performance. Instead, the FDA's 510(k) clearance is based on its substantial equivalence to already marketed devices.

    However, based on the provided text, we can describe the intended uses for each laser model, which implicitly sets the performance expectations for these devices.

    Acceptance Criteria and Reported Device Performance

    Since no specific quantitative acceptance criteria or reported device performance metrics (e.g., accuracy, sensitivity, specificity) were provided in the document, the "performance" here refers to the extensive list of intended uses and indications for which the device is cleared. The device "meets the acceptance criteria" by being substantially equivalent to existing devices and being cleared for the broad range of surgical applications outlined.

    Laser ModelAcceptance Criteria (Intended Use/Indications)Reported Device Performance (Implicitly Meets Indications)
    QUANTA532Ophthalmology (Retinal Photocoagulation, Pan Retinal Photocoagulation, Endophotocoagulation, Macular Treatments, Laser Trabeculoplasty), Otolaryngology (Stapedectomy, Stapedotomy, Myringotomy, Lysis of adhesions, Control of bleeding, Removal of acoustic neuromas, Soft tissue adhesion in micro/macro otologic procedures), Dermatology (Vascular lesion, Pigmented lesion).Cleared for all listed indications.
    QUANTA808Urology (BPH), Genitourinary, Thoracic Surgery, Plastic Surgery and Dermatology (vascular lesions, hair removal), General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT, Radiology, Endovascular coagulation, Oral Surgery, Dental procedures. Specific indications include lesions of external genitalia, circumcision, condyloma, bladder tumors, bladder neck incisions, vaporization of prostate, rectal/anal hemorrhoidectomy, mastectomy, dermabrasion, appendectomy, bowel resection, colectomy, liver resection, resection of organs, thyroidectomy, adhesiolysis, hepatobiliary tumors, thoracotomy, cholecystectomy, breast biopsy, PLDD, discectomy, hemostasis (meningiomas), cervical conization, myomectomy, endometrial ablation, ovarian cystectomy, appendectomy, dacryocystorhinostomy, open DCR, tumor excision, blepharoplasty, dissect/coagulate, hemostasis (colonoscopy, esophageal varices), excision of polyps, chondromalacia, synovectomy, menisectomy, pulmonary resection, pericardiectomy, coagulation of blebs/bullae, endoscopic pulmonary applications, tracheal bronchial lesions, benign/malignant pulmonary obstruction, removal of benign lesions (ear, nose, throat), excision of carcinoma (larynx), incision/excision of carcinoma in situ, neck dissection, laryngeal papillomectomy, removal of vocal cord/fold nodules/polyps/cyst, frenectomy, frenotomy, biopsy, pulpotomy, cardiac tissue coagulation/hemostasis, photocoagulation of vascular/dermatological lesions, telangiectasia/veinulectasia, reticular veins/branch varicosities, pyrogenic granuloma, lymphangioma, angiofibromas, superficial benign vascular lesions (telangiectasias, rosacea, angioma, venous lakes, couperosis, cherry angioma, hemangioma, port wine stains, angiokeratoma), benign epidermal pigment lesions (lentigines, epidermal nevi, spider nevi), dermatological surgery (condyloma acuminata, warts, small non-malignant skin tumors, small semi-malignant tumors, warty leucoplasty, ulcers debridment), seborrheic keratosis, mixoid cyst, papillary varix, acne treatment, hair removal (skin type I-V), cut/coagulation/vaporization, resurfacing (non), blepharoplasty, endoluminal/endovenous laser saphenous incompetent vein surgery.Cleared for all listed indications.
    QUANTA940Urology, Plastic Surgery, General Surgery, Dermatology, Gynecology, Pulmonary Surgery, Gastroenterology, ENT, Radiology, Endovascular Surgery, Dental Procedures. Intended for cutting, vaporization, ablation and coagulation of soft tissue (with endoscopic equipment, contact/non-contact open surgery), treatment/removal of vascular lesions/tumors, removal of unwanted hair, endovascular coagulation of greater saphenous vein (superficial vein reflux), treatment of BPH symptoms (28-85cc prostates), incision/excision/vaporization/ablation/coagulation of oral soft tissue (biopsies, exposure of unerupted teeth, fibroma removal, frenectomy, frenotomy, gingival troughing, gingivectomy, gingivoplasty, gingival incision/excision, haemostasis/coagulation, implant recovery, incision/drainage of abscess, leukoplakia, operculectomy, oral papillectomies, pulpotomy, pulpotomy adjunct to root canal therapy, reduction of gingival hypertrophy, soft tissue crown lengthening, treatment of canker sores/herpetic/aphthous ulcers, vestibuloplasty), Laser Periodontal procedures (soft tissue curettage, removal of diseased/infected/inflamed/necrosed soft tissue, sulcular debridement).Cleared for all listed indications.
    QUANTA980Incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, endovenous occlusion of greater saphenous vein. Further indicated for laser assisted lipolysis. Specific indications are the same as 808nm for most categories, with additional specific surgical tasks listed for ENT/Oral Surgery, Arthroscopy, Gastroenterology, General Surgery/Dermatology/Plastic Surgery/Podiatry, Urology, Gynecology, Neurosurgery, Cardiac Surgery, Pulmonary Surgery, Dental Applications, and Endovenous Occlusion.Cleared for all listed indications.
    QUANTA1064Dermatology/Plastic Surgery (Photocoagulation of Colored Vascular Lesions of Skin if Argon Laser unsuccessful), Discectomy (Percutaneous Lumbar Discectomy), Gastrointestinal (Tissue Ablation for Benign/Malignant Neoplasm, Polyps, Colitis, Ulcers, Aniodysplasia, Hemorrhoids; Hemostasis for Varices, Esophagitis, Esophageal Ulcer, Mallory-Weiss Tear, Gastric Ulcers, Duodenal Ulcers, Non-bleeding Ulcers, Gastric Erosions), General Surgery (Skin Incision, Tissue Dissection, Excision (external tumors/lesions), Resection of Internal Organs, Tumors/Lesions, Tissue Ablation, Vessel Coagulation), Genitourinary Surgery (Ablation/Hemostasis for Superficial Urinary Bladder Tumors, Invasive Bladder Carcinoma, Urethral Strictures, Lesions of External Genitalia), Gynecological Tissue Ablation (Endometrial Ablation (menorrhagia), Soft Tissue Excisional Conization, Submuccous Fibroids, Polyps, Septa), Neurosurgery (Hemostasis), Orthopedic Surgery (Soft Tissue incision/excision in Knee, Shoulder), Otorhinolaryngology Surgery (Soft Tissue: Skin Incision, Tissue Dissection, Excision (external tumors/lesions), Resection of Internal Organs, Tumors/Lesions, Tissue Ablation, Vessel Coagulation), Prostatectomy (Soft Tissue Coagulation for BPH), Pulmonary Surgery (Palliative Treatment for Benign/Malignant Pulmonary Airway Obstructions).Cleared for all listed indications.
    QUANTA1320Dermatology (incision, excision, ablation, vaporization with hemostasis of soft tissue, treatment of fine lines/wrinkles, back acne, atrophic acne scars), Endovascular surgery (treatment of reflux of great and small saphenous veins associated with varicose veins and varicosities).Cleared for all listed indications.
    QUANTA1470Endovenous occlusion of the greater saphenous vein. Further indicated for laser assisted lipolysis. Specifically for endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.Cleared for all listed indications.
    QUANTA1950Urology (Open/endoscopic surgery for urethral strictures, bladder neck incisions, ablation/resection of bladder/urethral/ureteral tumors, BPH, TUIP, HoLRP, HoLEP, HoLAP, condylomas, lesions of external genitalia), Gastroenterology (Open/endoscopic surgery for appendectomy, polyps, biopsy, gall bladder/biliary/bile duct calculi, ulcers, pancreatitis, hemorrhoids, cholecystectomy, benign/malignant neoplasm, angiodysplasia, colorectal cancer, telangiectasias, vascular malformation, gastritis, esophagitis, esophageal ulcers, varices, colitis, Mallory-Weiss tear, gastric erosions), Thoracic/Pulmonary Surgery (Open/endoscopic surgery for laryngeal lesions, airway obstruction, polyps/granulomas, palliation of obstructing carcinomas), Gynecology (Open/endoscopic surgery for intra-uterine treatment of submucous fibroids, benign endometrial polyps, uterine septum; soft tissue excision procedures like excisional conization of cervix), Ear, Nose & Throat (Endoscopic endonasal surgery for endonasal/sinus surgery, partial turbinectomy, polypectomy, dacryocystorhinostomy, frontal sinusotomy, ethmoidectomy, maxillary antrostomy, functional endoscopic sinus surgery, lesions/tumors of oral/nasal/glossal/pharyngeal/laryngeal, tonsillectomy, adenoidectomy), Dermatology/Plastic Surgery (Incision/excision/resection/ablation/vaporization/coagulation/hemostasis of soft/mucosal/fatty/cartilaginous tissue for basal cell carcinomas, lesions of skin/subcutaneous tissue, skin tags, plantar warts), Arthroscopy (excision/ablation/coagulation of soft/cartilaginous tissue in minimally invasive spinal surgery, PLDD, foraminoplasty, ablation/coagulation of soft vascular/non-vascular tissue), General Surgery (Open/laparoscopic/endoscopic surgery for cholecystectomy, lysis of adhesions, appendectomy, biopsy, skin incision, tissue dissection, excision of external tumors/lesions, complete/partial resection of internal organs/tumors/lesions, mastectomy, hepatectomy, pancreatectomy, splenectomy, thyroidectomy, parathyroidectomy, herniorrhaphy, tonsillectomy, lymphadenectomy, partial nephrectomy, pilonidal cystectomy, resection of lipoma, debridement of decubitus/stasis ulcers, hemorrhoids).Cleared for all listed indications.

    Regarding the study that proves the device meets the acceptance criteria:

    The provided 510(k) Premarket Notification document does not describe a specific clinical study for the Quanta System QUANTA Diode Laser Family to prove its performance against acceptance criteria in the traditional sense of a new device. Instead, the basis for its market clearance is "Substantial Equivalence" to predicate devices already on the market.

    The document explicitly states under "Performance Data": "None".

    This means that a new clinical study was not conducted or presented to demonstrate the device's efficacy or safety. The FDA's determination of substantial equivalence (K100558) implies that the new device is as safe and effective as the predicate devices based on similarities in intended use, indications for use, technological characteristics, and principles of operation, and that minor technological differences raise no new issues of safety or effectiveness.

    Therefore, the following points cannot be answered from the provided text, as the underlying regulatory pathway for this device did not require such studies or data submission:

    1. Sample size used for the test set and the data provenance: Not applicable, as no performance testing using a test set against ground truth was described in this document.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a laser surgical device, not an AI-based diagnostic tool relying on human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device, not an AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    7. The sample size for the training set: Not applicable, as no machine learning or AI component with a training set was mentioned.
    8. How the ground truth for the training set was established: Not applicable.