Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093115

    Validate with FDA (Live)

    Date Cleared
    2010-03-19

    (168 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K091097

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kimberly-Clark* KC100 Surgical Gowns are sterile, single use devices intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

    The Kimberly-Clark* KC100 Surgical Gowns are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and Ethylene Oxide (EtO) sterilization.

    Device Description

    Disposable gowns are used in the OR as a protective covering, for operating room staff, from the transfer of microorganisms, body fluids and particulates. Kimberly-Clark's KC100 Surgical Gowns are comprised of a single layer of SMS (spunbond/meltblown/spunbond polypropylene) fabric. The gowns consist of 100% polyester cuffs sewn to the end of the sleeves using polyester thread. The gowns also have a manual closure system.

    AI/ML Overview

    The provided text describes a 510(k) summary for Kimberly-Clark KC100 Surgical Gowns. This document is a premarket notification for a medical device and focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than detailing studies about AI algorithms or software performance.

    As such, the requested information regarding AI device acceptance criteria and study details (sample size for test set, data provenance, number of experts, adjudication method, MRMC studies, standalone performance, ground truth types, training set sample size, and ground truth establishment for training set) cannot be extracted from the provided text because it pertains to the approval of physical surgical gowns, not an AI/ML device.

    However, I can provide the acceptance criteria and study information related to the surgical gowns themselves based on the text:

    1. Table of Acceptance Criteria and Reported Device Performance (for KC100 Surgical Gowns):

    Acceptance Criteria CategorySpecific Tests ConductedReported Device Performance
    Material PropertiesBarrier properties (e.g., fluid resistance, microbial barrier)Met acceptance criteria
    Tensile and tear strengthMet acceptance criteria
    Alcohol repellencyMet acceptance criteria
    FlammabilityMet acceptance criteria
    LintingMet acceptance criteria
    BiocompatibilityCytotoxicityMet acceptance criteria (in compliance with ISO 10993 methods)
    IrritationMet acceptance criteria (in compliance with ISO 10993 methods)
    SensitizationMet acceptance criteria (in compliance with ISO 10993 methods)

    2. Study that Proves the Device Meets Acceptance Criteria:

    The study conducted was a bench testing to demonstrate that the KC100 Surgical Gowns are identical to and meet the same acceptance testing criteria as their predicate device (K091097 proMedical Surgical Gowns).

    • Sample size used for the test set and the data provenance: The document does not specify exact sample sizes for each test. It states "Testing included..." implying a standard set of tests for medical devices of this type. The provenance of the data is not explicitly stated beyond being conducted for the Kimberly-Clark Corporation. This would be considered internal company testing.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are physical product performance tests, not AI-driven diagnostic tests requiring expert ground truth for interpretation.
    • Adjudication method for the test set: Not applicable for physical product testing described.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for a physical medical product, not an AI device.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is for a physical medical product, not an AI device.
    • The type of ground truth used: The "ground truth" here is objective measurements against established international standards (like ISO 10993 for biocompatibility) and performance specifications for surgical gowns.
    • The sample size for the training set: Not applicable. This is not an AI/ML device requiring a training set.
    • How the ground truth for the training set was established: Not applicable.

    In summary, the provided text concerns the regulatory approval of physical surgical gowns based on equivalence to a predicate device through standard product performance and biocompatibility testing, not an AI device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1