510(k) Data Aggregation

    K Number
    K091600

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-08-25

    (83 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K090601, K090262

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DirectVision" Guide System is indicated for use in the visual examination of the urethra and bladder when placing the PercuCath Urologic Catheter.

    Device Description

    The DirectVision™ Guide System is comprised of two (2) distinct components: a flexible semirigid microendoscope, and a handheld camera with integral LED illumination and camera control unit (CCU). The microendoscope is designed to work with the Schoelly Imaging Inc. FSC2 light source and CCU (K090601). The DirectVision TM Guide System is indicated for use in the endoscopic examination of the urethra and bladder, when placing the PercuCath™ Urologic Catheter (K090262). The catheter provides a sheath around the microendoscope to facilitate its use. The catheter also incorporates a channel to be used for supplying irrigation to the microendoscope tip for clearing away debris from the field of view during use. The microendoscope coupler is connected to the cleared camera and integral light source such that the light source is collinear with the light fiber optic bundle and the camera is collinear with the image fiber optic bundle. The image transmitted through the image fiber optic bundle to the camera is then displayed on a monitor accessory for the clinician to view.

    AI/ML Overview

    The provided 510k summary for the DirectVision™ Guide System does not contain information related to specific acceptance criteria or a study demonstrating the device meets such criteria in terms of performance metrics like diagnostic accuracy, sensitivity, or specificity. This document primarily focuses on establishing substantial equivalence to a predicate device based on design, intended use, and general characteristics.

    Therefore, many of the requested sections cannot be filled from the provided text.

    Here's a breakdown of what can be extracted and what information is missing:


    Acceptance Criteria and Device Performance

    No specific quantitative or qualitative performance acceptance criteria (e.g., minimum visual acuity, field of view, illumination levels) and no reported device performance against such criteria are detailed in this 510k summary. The summary relies on comparison to a predicate device and adherence to general standards.

    Study Information (Missing for Performance)

    The document does not describe a performance study in the analytical or clinical sense that would involve a test set, ground truth, or expert review for the DirectVision™ Guide System.


    Information Not Applicable / Not Provided in the Document:

    1. A table of acceptance criteria and the reported device performance: Not provided. The summary focuses on design and safety compliance rather than performance metrics.
    2. Sample size used for the test set and the data provenance: Not applicable as no performance study is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an endoscope, not an AI-powered diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device does not use an algorithm for standalone performance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    Information Available & Relevant to the Submission:

    While not a "performance study" in the requested sense, the submission details compliance with general performance standards relative to endoscopes.

    Predicate Device Comparison (Indirect 'Performance' Justification):

    The submission establishes substantial equivalence to the Davlite Microendoscope (K020310) based on:

    • Intended Use: Both are for visual examination within the body. The DirectVision™ Guide System specifies use in the urethra and bladder when placing the PercuCath™ Urologic Catheter, while the Davlite Microendoscope has a broader indication for viewing "an interior cavity of the human body."
    • Design: Both are flexible microendoscopes with glass fiber optic bundles for light and image, an objective lens at the tip, and a camera adapter.
    • Body Contact Material: Both use biocompatible materials (Polyimide and glass for DirectVision™, Polymer and glass for Davlite).
    • Performance Standard: Both adhere to ISO 8600.
    • Biocompatibility: Both are declared biocompatible.
    • Sterility: Both are supplied non-sterile.
    • Re-Use: Both are multi-use devices.

    Recognized Consensus Standards Met (Acceptance Criteria for Safety & Design):

    The DirectVision™ Guide System meets requirements of the following recognized consensus standards, which can be seen as "acceptance criteria" for the safety, design, and processing of the device:

    • ISO 8600-1:2005 (Optics and Photonics -- Medical Endoscopes Information to be Provided by the Manufacturer for the Processing of Resterilizable Medical Devices)
    • AAMI TIR12: 2004 (Designing, Testing, and Labeling Reusable Medical Devices for Preprocessing in Health Care Facilities)
    • AAMI / ANSI ST81:2004 (Sterilization Of Medical Devices - Information To Be Provided By The Manufacturer For The Processing Of Resterilizable Devices)
    • AAMI / ANSI ST35:2003 (Safe Handling and Biological Decontamination of Reusable Medical Devices in Health Care Facilities and in Nonclinical Settings)
    • IEC 60601-2-18 (Particular Requirements for the Safety of Endoscopic Equipment)
    • ISO10993-1:2003 (Biological Evaluation of Medical Devices)

    The submission implies that by meeting these standards and demonstrating substantial equivalence to a legally marketed predicate, the device is considered safe and effective for its intended use, without requiring a specific performance study in the document provided.