Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103760

    Validate with FDA (Live)

    Device Name
    TITAN
    Manufacturer
    Date Cleared
    2011-05-05

    (133 days)

    Product Code
    Regulation Number
    882.1900
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K083861

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Titan with DPOAE440 is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Titan with DPOAE440 includes all ages.

    The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Titan with ABRIS440 is newborns.

    Device Description

    The instrument is audiometric equipment used for testing middle ear, inner ear and brainstem abnormalities.

    The Titan is a combination of:

    • Audiometric device for measurement of aural acoustic impedance (IMP) (Cleared . under K083861)
    • Audiometric device for measurement of otoacoustic emissions (OAE) .
    • Audiometric device for measurement of auditory brainstem responses (ABR) ●

    The Titan can have one of three functions depending on license:

    The Titan IMP test is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear.

    The Titan OAE test generates an acoustic stimulus in the ear canal at a moderate intensity level. The stimulus vibrates a thin membrane in the cochlear, and activates the outer hair cells located on the membrane. This activation is propagated back through the middle ear as sound into the ear canal. Microphone(s) detects OAE-related sounds.

    The Titan ABR test produces a short acoustic stimulus and measures via cutaneous electrodes the auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The test uses an automatic detection algorithm to determine if an ABR is evoked in response to the short acoustic stimulus. The output of the device is either PASS (ABR detected) or REFER (ABR is not detected).

    The Titan with DPOAE440 is intended as clinical audiometer. The Titan with ABRIS440 is intended as a screener audiometer.

    AI/ML Overview

    The Interacoustics Titan with ABRIS440 and Titan with DPOAE440 devices are audiometric equipment used for testing middle ear, inner ear, and brainstem abnormalities. The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document evaluates the Titan with ABRIS440 against the Maico MB 11 (predicate device) and the Titan with DPOAE440 against the DPOAE20 (predicate device). The "Equivalence" column in the tables below serves as the acceptance criteria and indicates whether the new device met the criteria (Same, Similar).

    Titan with ABRIS440 vs. Maico MB 11 (Predicate Device)

    DescriptionMaico, MB 11 (Predicate)Titan with ABRIS 440 (New)Equivalence
    Indications for useThe Maico MB 11 with BERAphone and the MB 11 Standard are intended to generate evoked auditory brainstem response (ABR) -based measurements for newborn hearing screening, providing a PASS/REFER result.The Titan with ABRIS440 is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem.Similar (Wording just a little different)
    Target populationNewbornsNewbornsSame
    Intended userThe User does not require special technical skills or interpretation of the results.The User does not require special technical skills or interpretation of the results.Same
    Anatomical sitesExamination of Ear and hearing nervesExamination of Ear and hearing nervesSame
    Safety standardsIEC 60601-1IEC 60601-1Same
    Performance standardIEC 60645-7 (Type 2)IEC 60645-7 (Type 2)Same
    Device TypeScreening device (PASS/REFER) resultScreening device (PASS/REFER) resultSame
    System Configuration1-channel ABR system operated through a base unit connected via a USB port to any standard computer (Laptop Computer recommended)1-channel ABR system operated through a handheld base unit. The base unit can be operated stand alone or PC controlled through USB or Bluetooth.Similar (Titan can work as stand alone)
    Display informationTesting sequence along with current EEG parameters (number of sweeps, number of artifacts, SNR, and artifact rejection system setting)PASS/REFER status, indicated with value between 0 and 100% where 100% indicates a pass. EEG peak or RMS value, rejection status, residual noise and what transducer(s) are detected.Similar (Minor differences in displayed information)
    Electrode quality checkYESYESSame
    Impedance Test (Quality control)Indicator in three levels. Red: above 10kOhm Yellow: between 10 & 8kOhm Green: below 8kOhmBefore recording: Electrode impedance is measured if they are above 10kOhm, below 10 kOhm or below 3 kOhm.Similar (Difference in mid-range due to measurement frequencies, Titan may require more careful electrode placement)
    Binaural screeningYESYESSame
    Pre-amplifier channels11Same
    StimulusChirpsChirpsSame
    Pre-amplifier Gain86 dB64 dBSimilar (Depends on amplifier characteristics, no influence on measuring results)
    FilteringHW HP 30Hz filters of 1st orderTwo HW HP filters of 1st order: Cut-off frequencies 7Hz and 45HzSimilar (Both look at frequencies above 80Hz; response identical at that frequency. Basic purpose is to eliminate DC components)
    MaskingNoneNoneSame
    TransducersBERAphone TDH39 EAR 3A Insert phonesTitan Probe TDH39 EAR 3A Insert phonesSimilar (Titan probe is dedicated for Titan as BERAphone for MB 11)
    Stimulus rate92.5/s90/sSimilar
    Screening level35dBHL35dBHLSame
    Screening measurement180 seconds180 secondsSame

    Titan with DPOAE440 vs. DPOAE20 (Predicate Device)

    DescriptionDPOAE20 (Predicate)DPOAE440 (New)Equivalence
    Indications for useThe Interacoustics DPOAE20 system is for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emission tone stimuli. The presence of Otoacoustic emissions suggests normal outer hair cell function within cochlea, which in turn suggests normal hearing. OAEs are recorded using an OAE probe which is placed in the ear canal. The OAE response from the ear is recorded and processed by the Eclipse and the DPOAE2O software and then displayed on the computer screen for evaluation.The Titan with DPOAE440 is intended for use in the diagnostic audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions.Similar (Wording just a little different)
    Target populationThe patient group includes all agesThe patient group includes all agesSame
    Intended userEar, Nose, and Throat doctors, Neurology specialists, Audiologist and other health- professionals concerned with measuring auditory functions.The Titan is designed to be used only by skilled personnel such as audiologists, ENT surgeons, and doctors, hearing healthcare professionals or personnel with similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.Same
    Evaluation methodDPOAEDPOAESame
    Safety standardIEC 60601-1IEC 60601-1Same
    EMC standardIEC 60601-1-2IEC 60601-1-2Same
    Performance standardIEC 60645-6 (Type 2)IEC 60645-6 (Type 2)Same
    Test signalsIEC 60645-1/ANSI S3.6, IEC 60645-3IEC 60645-1/ANSI S3.6, IEC 60645-3Same
    Stimulus: Frequency Range500-8000 Hz500 to 10000 HzSimilar (Extended range is a technical improvement due to better technology and probes)
    Stimulus: Frequency Step50 Hz25 HzSimilar (Better resolution possible due to better technology)
    Stimulus: Level30-75 dB SPL (70 dB above 6kHz)30 to 80 dB SPL (75 dB SPL for 6kHz and 65 dB SPL for 8kHz and 10kHz)Similar (Extended range is a technical improvement due to better technique and probes)
    Stimulus: Level Step1 dB SPL1 dB SPLSame
    TransducerDedicated DPOAE20/TEOAE25 probeDedicated IOW ProbeSimilar (New transducer due to hardware design, but same characteristics as predicate)
    Probe detectionNoneAuto detectionDifferent (Probe unit has built-in memory for calibration values)
    Recording Analysis timeMinimum 2 sec to unlimited test time.Minimum 2 seconds to unlimited timeSame
    Recording A/D Resolution16 bit, 7.3 Hz resolution24 bit, 5.38 Hz resolutionSimilar (Better resolution due to better technology)
    Artifact Reject System-30 -> +30 dB SPL or off.-30 to +30 dB SPL or offSame
    Stimulus ToleranceAdjustable between 1 and 10 dBAdjustable between 1 and 10 dBSame
    SNR CriteriaAdjustable between 1 and 20 dBAdjustable between 3 and 25 dBSimilar (Limits adjusted to better fit real-world practice, giving higher evidence for response)
    Probe check window256 points frequency response of the ear canal due to a click stimulus presented with a rate of 100 Hz at 80 dB SPL1024 points frequency response of the ear canal due to a click stimulusSimilar (Better resolution due to better technology used)
    DP-Response window4096 points frequency response4096 points frequency responseSame
    Display General Display gainApplicable during testingApplicable during testingSame
    DisplayStimulus level and typeStimulus level and type, Basic and advanced viewSimilar (A simpler (basic) view has been introduced)
    DatabaseSupported by Interacoustics OtoAccess™ database. Operating under NOAH is possible, using IA NOAHLink.Supported by Interacoustics OtoAccess™ database. Operating under NOAH is possible, using IA NOAHLink.Same

    2. Sample size used for the test set and the data provenance

    The provided document states: "No clinical tests were performed". Instead, the device's substantial equivalence was determined based on its conformity to international standards and comparison to predicate devices, which implies that no specific test set of patient data was used for a clinical performance study as part of this submission. Therefore, data provenance is not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical tests were performed and no ground truth was established by experts for a test set in this submission. The comparison relies on technical specifications and international standards.

    4. Adjudication method for the test set

    Not applicable, as no clinical tests were performed and no ground truth was established by experts for a test set in this submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document does not describe an MRMC study or AI assistance. The device is a diagnostic tool, not an AI-powered diagnostic aide.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    For the Titan with ABRIS440, the device includes an "automatic detection algorithm to determine if an ABR is evoked in response to the short acoustic stimulus. The output of the device is either PASS (ABR detected) or REFER (ABR is not detected)." This indicates a standalone algorithmic performance component. However, no specific standalone performance study results, such as sensitivity or specificity, are presented in this submission. The study relies on comparison to predicate devices and adherence to standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given the statement "No clinical tests were performed," there was no direct patient-level ground truth established for the purpose of this 510(k) submission. The ground truth for the equivalence claims is implicitly based on the established safety and performance characteristics of the predicate devices and adherence to relevant international standards (IEC 60601-1, IEC 60601-1-2, IEC 60645-7, IEC 60645-6).

    8. The sample size for the training set

    Not applicable, as no clinical tests were performed and the device is not described as utilizing a machine learning algorithm that would require a training set in the context of this 510(k) submission.

    9. How the ground truth for the training set was established

    Not applicable, as no training set was used in the context of this 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1