Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K130401

    Validate with FDA (Live)

    Device Name
    ALERTWATCH: OR
    Manufacturer
    Date Cleared
    2014-02-07

    (353 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082583

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision.

    Device Description

    AlertWatch:OR is a display and secondary alert system used by the anesthesiology staff - residents, CRNA's, and attending anesthesiologists - to monitor patients in operating rooms. The purpose of the program is to synthesize a wide range of patient data and inform clinicians of potential problems that might lead to immediate or long-term complications. Once alerted, the clinician is instructed to refer to the primary monitoring device before making a clinical decision. AlertWatch:OR should only be connected to AIMS systems and physiologic monitors that have been validated for use with Alert Watch:OR. Alert Watch, Inc. performs the validation for each installation site.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the AlertWatch:OR device, based on the provided document:

    The document does not explicitly state formal acceptance criteria with specific performance metrics (e.g., sensitivity, specificity, accuracy thresholds). Instead, the performance testing section describes verification and validation activities designed to ensure the product works as designed, meets its stated requirements, and is clinically useful.

    1. Table of Acceptance Criteria and Reported Device Performance

    As specific numerical acceptance criteria (e.g., sensitivity > X%, specificity > Y%) are not provided, the table below reflects the described performance testing outcomes.

    Acceptance Criterion (Implicit from Study Design)Reported Device Performance (from "Performance Testing" section)
    Verification: Analysis Output AccuracyProduced desired output for each rule/algorithm using constructed data.
    Verification: Data Display AccuracyProduced desired display for each test case using constructed data.
    Verification: Data Collector FunctionalityLive Collector and Data Collector returned correct data from the EMR.
    Verification: Product Functionality with Historical DataProduct worked as designed using a set of cases from actual patients.
    Validation: Design Review & Software Requirements Specification (SRS) AccuracyProcess and various inputs for creating the product design (SRS) were reviewed. SRS was reviewed for clinical accuracy.
    Validation: Clinical UtilityClinical utility of the product was validated by analyzing case outcomes.
    Validation: Human FactorsSummative Human Factors study conducted to demonstrate the device meets user needs.
    Overall Performance ClaimThe results of the verification and validation activities demonstrate that the AlertWatch:OR complies with its stated requirements and meets user needs and intended uses.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Verification (Step 4: Historical Data): "a set of cases from actual patients" - The exact number of cases is not specified.
    • Data Provenance: "data from the EMR" and "a set of cases from actual patients." The document does not specify the country of origin, nor explicitly whether it was retrospective or prospective, though "historical data" strongly implies retrospective data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth, and it does not explicitly describe a ground truth establishment process involving experts in the traditional sense (e.g., for diagnostic accuracy). The "clinical accuracy" review of the software requirements specification implies expert involvement, but details are missing.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (such as 2+1, 3+1) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or described. The study focused on the device's functionality and utility rather than a direct comparison of human readers with and without AI assistance to quantify improvement.

    6. Standalone (Algorithm Only) Performance Study

    The verification steps, particularly "Verify the analysis output" and "Verify the data display" using "constructed data," and "Verify the product with historical data," indicate that the algorithm's performance was evaluated in a standalone manner (without human-in-the-loop) to ensure it performs "as designed" and produces "desired output/display." However, these are functional verifications rather than a typical standalone diagnostic performance study with metrics like sensitivity, specificity, or PPV/NPV.

    7. Type of Ground Truth Used

    The ground truth for the performance testing appears to be established by:

    • "Desired output" based on the "Software Requirements Specification" for constructed data tests (functional verification).
    • "Works as designed" when tested with "a set of cases from actual patients" (implies comparison to expected system behavior based on its design, rather than a clinical outcome or expert diagnosis acting as a gold standard).
    • "Clinical utility... by analyzing case outcomes" suggests that real-world patient outcomes were used to assess the value of the alerts generated. This hints at an outcome-based ground truth for the validation of clinical utility, but details are scarce.

    8. Sample Size for the Training Set

    The document does not specify the sample size for a training set. The descriptions of verification and validation do not refer to machine learning model training. The device seems to operate based on "rules/algorithms in the Software Requirements Specification" rather than a trained AI model.

    9. How the Ground Truth for the Training Set Was Established

    As there's no mention of a dedicated training set or a machine learning model requiring such, this information is not applicable based on the provided text. The device likely relies on predefined rules and algorithms.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The monitor is intended to be used for monitoring and recording of, and to generate alarms, for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for use during patient transport inside and outside of the hospital environment. The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Predictive Temperature unit is intended for use with adult and pediatric patients in a hospital environment. The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only. The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients. The transcutaneous gas measurement (tcGas) is restricted to neonatal patients only. BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    Device Description

    The Philips IntelliVue Patient Monitors family comprises the multiparameter patient monitor series: MP2, X2, MP5, MP5T, MP5SC, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90 and MX600, MX700, and MX800. Each monitor consists of a display unit including built-in or separate central processing unit (CPU) and physiological measurement modules. All monitors share the same architecture of CPU units and exactly the same software is executed on each monitor. The monitors measure physiological parameters such as: Sp02, pulse, ECG, arrhythmia, ST, QT, respiration, invasive and noninvasive blood pressure, temperature, CO2, spirometry, C.O., CCO, tcp02/ tcpCO2, S02, Sv02, Scv02, EEG, and BIS. They generate alarms, record physiological signals, store derived data, and communicate derived data and alarms to the central station. IntelliVue series MP2, X2, MP5, MP5T, MP5SC, MP20, and MP30 are robust, portable, lightweight, compact in size and modular in design patient monitors with interfaces to dedicated external measurement devices. Models MP2, X2, MP5, MP5T, and MP5SC also incorporate multiple built-in physiological measurements. IntelliVue series MP40, MP50, MP60, MP70, MX600, MX700, and MX800 are patient monitors with built-in central processing unit, flat panel display and interfaces to dedicated external measurement devices. Models MX600, MX700, and MX800 have widescreen displays. IntelliVue series MP80 and MP90 are patient monitors with flat panel display and central processing unit as separate components. They have interfaces to dedicated external measurement devices.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, structured as requested:

    Acceptance Criteria and Device Performance Study for Philips IntelliVue Patient Monitors (Software Revision J.04)

    Overview:
    The submission describes a software modification to existing Philips IntelliVue Patient Monitors (MP2, X2, MP5, MP5T, MP5SC, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90, MX600, MX700, and MX800) to introduce a new SpO2 intelligent alarm delay feature called 'Smart Alarm Delay'. The study aims to demonstrate that this modified device is as safe and effective as the predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a quantitative table of acceptance criteria with corresponding performance metrics for the 'Smart Alarm Delay' feature in the format often seen for diagnostic devices (e.g., sensitivity, specificity, accuracy). However, the "Summary of V&V activities" section outlines the general performance goals and outcomes.

    Acceptance Criteria CategorySpecific Criteria/GoalReported Device Performance
    Clinical EvaluationUsers' understanding of the 'Smart Alarm Delay' feature as described in the Instructions for Use (IFU)."The vast majority of test persons understood the implications of using the new SpO2 Smart Alarm Delay feature."
    Users' perception of the feature's usefulness and clinical meaningfulness."They regarded it as a helpful alternative to the existing SpO2 standard alarm delay."
    Functionality TestingEffectiveness of implemented design risk mitigation measures (from Hazard Analysis)."The test results have confirmed the effectiveness of implemented design risk mitigation measures."
    Safe, effective, and according to specifications and IFU for SpO2 alarm derivation and delays of modified software."All specified criteria have been met. The test results have confirmed that the SpO2 alarm derivation and the SpO2 alarm delays of the modified IntelliVue Monitors have functioned safe, effective and according to the specifications and Instructions for Use."
    Regression TestingFunctionality of related, unmodified software parts."All specified criteria have been met. The test results have confirmed that the SpO2 parameter of the modified IntelliVue Monitors have functioned safe, effective and according to the specifications and Instructions for Use."
    Functionality of alarms of the IntelliVue Patient Monitors."All specified criteria have been met. The test results have confirmed that the alarms of the modified IntelliVue Monitors have functioned safe, effective and according to the specifications and Instructions for Use."
    Performance (Accuracy/Safety)Device performance, accuracy, and compliance with SpO2 standard ISO 9919."The modification does not affect device performance in general and device accuracy in particular... The modification does also not affect any safety and performance aspects covered by the SpO2 standard ISO 9919. Therefore, verification and validation executed on the subject IntelliVue Patient Monitors according to the standard ISO 9919 prior to the minor modification... is still valid and covers the modified devices."

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: The document does not specify the exact number of individuals (test persons) involved in the clinical evaluation. It refers to "two user groups - one consisting of physicians and one consisting of nurses" and later "the vast majority of test persons." This suggests a qualitative assessment rather than a statistically powered performance study.
    • Data Provenance: Not explicitly stated, but clinical evaluation of user understanding implies prospective testing with healthcare professionals. The country of origin for this specific clinical evaluation is not mentioned.

    3. Number of Experts and Qualifications for Ground Truth of the Test Set

    • The "clinical evaluation" appears to focus on user comprehension and acceptance, not on establishing a traditional clinical "ground truth" for diagnostic accuracy.
    • Number of Experts: Two user groups were formed: "one consisting of physicians and one consisting of nurses." The exact number of individuals within each group is not provided.
    • Qualifications of Experts:
      • Physicians
      • Nurses
      • No specific years of experience or subspecialty are mentioned.

    4. Adjudication Method for the Test Set

    Not applicable. The clinical evaluation described is a qualitative assessment of user understanding and perception, not a diagnostic accuracy study requiring adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document describes a software modification to an existing patient monitor to add an intelligent alarm delay feature. The "clinical evaluation" focused on user understanding and acceptance of this feature, not on comparing reader performance with and without AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    No, not in the traditional sense of a standalone diagnostic algorithm performance study. The modification is an alarm delay feature within an existing monitoring system. The document states:

    • "The new 'Smart Alarm Delay' feature is isolated from the SpO2 measurement algorithm, i.e. signal acquisition and numeric processing."
    • "The devices hardware and all accessories including, but not limited to the SpO2 sensors remain completely unchanged."
    • "The modification does not affect device performance in general and device accuracy in particular."
    • Performance aspects covered by ISO 9919 from prior V&V are considered still valid.

    This indicates that the fundamental SpO2 measurement accuracy itself was not re-evaluated as a standalone algorithm performance, as the algorithm for SpO2 measurement remained unchanged. The focus was on the alarm delay logic and its user-facing implications.

    7. Type of Ground Truth Used

    For the "clinical evaluation" regarding the 'Smart Alarm Delay' feature, the "ground truth" appears to be user understanding and subjective opinion as gathered directly from physicians and nurses. For the core SpO2 measurement, the ground truth and performance validation are based on prior verification and validation activities conducted according to ISO 9919 for the predicate device, which are deemed still valid.

    8. Sample Size for the Training Set

    Not applicable. This submission is for a software modification adding an alarm delay feature, not a machine learning or AI algorithm that requires a dedicated training set for model development. The 'Smart Alarm Delay' is described as being "based on the same fundamental principle" as the predicate's 'SatSeconds' alarm management technique, implying a rule-based or empirically derived logic rather than a learned model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for a machine learning model was described or used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1