Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN150033

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-26

    (423 days)

    Product Code
    Regulation Number
    876.5981
    Type
    Direct
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K082560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMART Device is intended to aid in weight management in overweight to obese individuals. The device is indicated for individuals with a body mass index (BMI) in the range 27-35 kg/m2 in conjunction with behavioral modification instruction.

    Device Description

    The SMART Device is an oral removable palatal space occupying device that includes an embedded temperature recording sensor. The device is intended to be worn only during meals. to limit bite size and slow the intake of food, thereby reducing the amount of food that is consumed. The SMART Device is available using wire clasps and using a wraparound method (the wraparound technique is an added alternative to accommodate different oral anatomies) where the wires are coated with acrylic and wrapped around teeth in a symmetrical manner to secure the fit of the device (see Figure 1, below). The device is then removed after each meal or snack and the patient or healthcare professional has the option to use the temperature recording sensor embedded in the device to monitor device usage. This component of the SMART Device automatically records adherence, eliminating manual record-keeping by the user. The sensor is operated by a 3 volt lithium coin battery and measures ambient temperature at 5 minute intervals, storing the time and temperature data. The battery and the sensor are both double-encased and hermetically sealed in the acrylic. The SMART Device is provided with the Palatal Mold Kit and SMART Reader components. The Palatal Mold Kit is used to fabricate the SMART Device to match the patient's oral anatomy. The device is made from a mold of the upper oral cavity of the patient taken by a trained healthcare provider using the Palatal Mold Kit. The SMART Reader (See Figure 2, below) is an optional component that connects the sensor to a computer. Usage data are uploaded using infrared data transfer and a record of use can be produced. The user is provided instructions to download software for communicating between a personal computer and the SMART Reader and for displaying the time of day that the SMART Device was used. This software allows information from the embedded sensor to be uploaded to a secure, password protected Patient Portal website. Use of the Patient Portal allows the patient or healthcare provider to assess frequency of device usage.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Sensor Monitored Alimentary Restriction Therapy (SMART) Device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Test/EndpointAcceptance CriteriaReported Device PerformanceMet Criteria?
    Non-Clinical Performance
    Force MeasurementExerted force b(4)PassedYes
    Durability MeasurementNo abrasions, tears, separation, or indications of wear for simulated use of 1.2 years (438 days)PassedYes
    Battery LifeBattery use life ≥ 1.2 years (438 days)PassedYes
    Clinical Performance (Pivotal Study - Primary Endpoint)
    %TBWL (ITT Population)40% of the device group would achieve 5% total body weight loss (%TBWL) as compared to 10% or less of the control group20.6% (21 of 102) lost more than 5% TBWL (p=0.0059) compared to 5.6% (4 of 71) in control.No
    Clinical Performance (Confirmatory Study - Co-Primary Endpoints)
    %TBWL (PP Population) - Co-primary 1At least 40% of the subjects would be observed to have at least 5% weight loss (%TBWL)30% (12 of 40) had at least 5% TBWL. (95% CI: [17%, 47%])No
    Mean %TBWL (PP Population) - Co-primary 2Mean %TBWL would be at least 4%Mean %TBWL was 2.9% [95% CI: 1.8%, 4.1%]No
    Clinical Performance (Pivotal Study - Safety)
    Device-related Adverse EventsIncidence of adverse events (goal of less than 10%)4.9% (5 of 102) device-related AEs. (95% CI: [1.6%, 11.1%])Ambiguous*
    Clinical Performance (Confirmatory Study - Safety)
    Device-related Adverse EventsNot explicitly stated as a numerical acceptance criterion, but low incidence of non-serious events is implied.2 (out of 24 total AEs in 12 subjects) were possibly device-related.Yes
    Clinical Performance (Confirmatory Study - Secondary Endpoint)
    Clinical fit of the deviceImplicit: Good clinical fit100% (20 of 20) subjects assessed for fit by dental professionals were determined to have good clinical fit.Yes
    Clinical Performance (Consumer Preference Study)
    Subject Acceptance/SatisfactionHigh acceptance and satisfactionSubjects reported comfortable use within 1-3 days; acceptance and satisfaction high.Yes

    *Note on Safety Criteria for Pivotal Study: While the observed rate (4.9%) was less than 10%, the 95% confidence interval [1.6%, 11.1%] extended above 10%, meaning it couldn't be definitively claimed that the true rate was less than 10%. However, the events were non-serious and resolved, supporting a favorable safety profile overall.

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    Pivotal Study:

    • Test Set Sample Size (ITT Population): 173 subjects (102 device group, 71 control group)
    • Test Set Sample Size (PP Population): 108 subjects (41 device group, 67 control group)
    • Data Provenance: Prospective, randomized, controlled, open-label study conducted at five US centers.

    Confirmatory Study:

    • Test Set Sample Size (ITT Population): 76 subjects
    • Test Set Sample Size (PP Population): 40 subjects
    • Data Provenance: Prospective, single-arm, multicenter study conducted at 4 US investigational centers.

    Consumer Preference Study:

    • Test Set Sample Size: 24 subjects
    • Data Provenance: A separate consumer and provider acceptance study; presumed prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Pivotal & Confirmatory Studies: The ground truth for weight loss was established through direct measurement of body weight by study personnel at study site assessments. No expert consensus on image interpretation or similar cognitive tasks was required for the primary and secondary endpoints related to weight.
    • Confirmatory Study (SMART Fit Acceptability Sub Study): "Dental professionals assessed clinical fit of the device in 20 subjects." Their specific qualifications (e.g., years of experience) are not provided, but they are identified as "dental professionals."

    4. Adjudication Method for the Test Set

    • Pivotal & Confirmatory Studies: Not applicable for the primary endpoints, which involved objective measurements (weight, device usage via sensor data). Adverse events would typically be adjudicated by the study investigators and reported to the sponsor, but a specific multi-expert adjudication method (e.g., 2+1) is not explicitly described.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This device does not involve human readers interpreting images or data to make a diagnosis or assessment where AI assistance would directly improve their performance on a cognitive task. This is a physical medical device for weight management, and clinical outcomes were measured directly.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Yes, in part. The device's sensor automatically records adherence, providing data independently of human input for that specific function. However, the overall device effectiveness for weight management is and was evaluated in human subjects in conjunction with behavioural modification, making it a human-in-the-loop system for its intended use. The clinical studies measured the device's standalone effect on patients (when used as intended by patients), rather than an algorithm's standalone performance on a dataset.

    7. Type of Ground Truth Used

    • Pivotal & Confirmatory Studies: Objective Clinical Measurements (e.g., measured body weight at baseline and follow-up for %TBWL, sensor data for device usage/adherence) and Adverse Event Reporting by investigators.
    • Confirmatory Study (Fit Sub Study): Expert Assessment by dental professionals for clinical fit.
    • Consumer Preference Study: Patient Self-Report for acceptance and satisfaction, and Objective Clinical Measurements (measured body weight) for weight loss.

    8. Sample Size for the Training Set

    • Not Applicable. The SMART Device is a physical medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The "training" that occurred was the device's development and iterative testing.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As above, there is no "training set" in the context of an AI/ML algorithm. The "ground truth" for the device's design and functionality would have been established through engineering specifications, previous research on palatal devices, and pre-clinical bench testing.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1