510(k) Data Aggregation

    K Number
    K141024

    Validate with FDA (Live)

    Date Cleared
    2015-01-13

    (267 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K082487

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The external pulse oximeter (MD50i and MD50p) is designed for spot checking of the pulse oxygen saturation and pulse rate for adult at hospital. This medical device can be reused. Not for continuous monitoring.

    Device Description

    The MD50I includes three main parts as SpO2 probe, SpO2 module and 30 pin port. The SpO2 probe is used to collect the SpO2 and PR signal from the patient, the SpO2 module is used to process the signal from the probe and transform it to digital signal, the 30 pin port is used to connect with iPhone device for digital signal transmission.

    The MD50I includes two parts of software, Upper Computer software (Mobile App) and Low Computer Software.

    The Upper Computer software is installed in the iPhone device, which is used to receive the measured data (SpO2 and PR) from the MD50I, which has the following functions:

    • Measured data (SpO2 and PR) receiving;
    • Display the measured data (SpO2 and PR) in real time;
    • Graph of the measured data (SpO2 and PR) based on the day/week/month or all of observed data trend.
    • Measured data (SpO2 and PR) reviewed, the day/week/month history data could be reviewed;
    • Note function, the user can make comments after specified history record;
    • History data delete, the user can delete the history records;
    • Personal information edit, the user can input their own information and save it;
    • Share function, the user can select the specified history record to share with others by email;

    The Low computer software is embedded in the MD501, which includes the SpO2 sensor and has the following functions:

    • Collect the SpO2 and PR data from patient;
    • Transform the electrical signal to digit signal;
    • Send the digit signal to the Upper Computer software for display and following process.

    The MD50P includes three main parts as SpO2 probe, SpO2 module and USB plug. The SpO2 probe is used to collect the SpO2 and PR signal from the patient, the SpO2 module is used to process the signal from the probe and transform it to digital signal, the USB is used to connect with PC device for digital signal transmission.

    The MD50P includes two parts of software. Upper Computer software (PC App) and Low Computer Software.

    The Upper Computer software is install in the PC device, which is used to receive the measured data (SpO2 and PR) from the MD50P, which has the following functions:

    • Insert device identified automatically;
    • Measured data (SpO2 and PR) receiving;
    • Display the measured data (SpO2 and PR) in real time;
    • Oxygen saturation, Pulse rate, Pulse bar and waveform are clearly displayed on PC when doing real-time measuring;
    • Save the current data into your PC with the format of "txt", "pdf", "xls";
    • Data print;
    • Personal information edit, the user can input their own information and save it, such as title, patient ID, name, gender, age, height, weight, physician comments;
    • Alarm set, set the high/low SpO2 limit and high/low PR limit;
    • Saved data reviewed.
    • Login interface, the user needs input password and user name to access the account.

    The Low computer software is embedded in the MD50P, which includes the SpO2 sensor and has the following functions:

    • Collect the SpO2 and PR data from patient;
    • Transform the electrical signal to digit signal;
    • Send the digit signal to the Upper Computer software for display and following process.
    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the External Pulse Oximeter Models MD50I and MD50P, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test TypeAcceptance CriteriaReported Device Performance
    Main FunctionThe main function of device shall be meet the design specificationPASS
    SpO2 Accuracy (Simulator)Measurement Range: 70%-90%, Display Range: 0%-99%, Resolution: ±1%PASS
    Pulse Rate (PR) Accuracy (Simulator)Measurement Range: 30-235 bpm, Display Range: 0-255 bpm, Resolution: ±1bpmPASS
    SpO2/PR Response TimeMax. response time shall be less than 45 s; Average response time shall be less than 30 s;PASS
    Environmental Test (Normal, High Temp, Hot/Humid)MD50I: SpO2: 70%~99%, ±2%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm) MD50P: SpO2: 70%~99%, ±3%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm)PASS
    SpO2/PR Accuracy under Low PerfusionMD50I: SpO2: 70%~99%, ±2%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm) MD50P: SpO2: 70%~99%, ±3%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm)PASS
    Shelf Life TestNo abnormal or structure loosen on the sample; The performance meet the design specification; (Implicit: samples after accelerated aging should met these)PASS (Implicit)
    Performance Test after Cleaning and DisinfectionMD50I: SpO2: 70%~99%, ±2%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm) MD50P: SpO2: 70%~99%, ±3%; PR: ±2 bpm (30-99bpm) or ±2% (100-235bpm)PASS
    SpO2 Accuracy (Clinical Trial) - MD50ITo be demonstrated by the study results in comparison to CO-Oximetry. (Implicitly, the Arms value is the performance metric).Arms of 1.48 (70-100% SaO2)
    SpO2 Accuracy (Clinical Trial) - MD50PTo be demonstrated by the study results in comparison to CO-Oximetry. (Implicitly, the Arms value is the performance metric).Arms of 2.08 (70-100% SaO2)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Clinical Study Test Set Sample Size:
      • For MD50I: 11 subjects (out of 12 initially included, 1 excluded due to unstable pulse rate).
      • For MD50P: 11 subjects (out of 12 initially included, 1 excluded due to unstable pulse rate).
    • Data Provenance: The studies were conducted in accordance with international standards (EN ISO 14155-1:2009, EN ISO 14155-2:2009, ISO 9919:2005, EN ISO 9919:2009, BS EN ISO 80601-2-61:2011) and FDA guidance for pulse oximeters. The text does not explicitly state the country of origin for the clinical trial data, but the manufacturer is Beijing Choice Electronic Technology Co., Ltd, China. The studies involved "healthy adult volunteer subjects" and data collection occurred during specified dates in July 2013. This indicates a prospective collection of data for the purpose of the clinical validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • The clinical studies for SpO2 accuracy used arterial blood CO-Oximetry for ground truth. This is a direct measurement, and therefore, no human experts were used to establish the ground truth for SpO2 accuracy in the clinical trial.

    4. Adjudication Method for the Test Set:

    • Not applicable, as ground truth for SpO2 accuracy was established via arterial blood CO-Oximetry, not human expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. The document describes non-clinical performance tests and clinical validation for the device's accuracy in measuring SpO2 and PR, not a study evaluating human reader performance with or without the device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

    • Yes, the clinical accuracy studies (for both MD50I and MD50P) were standalone performance evaluations of the device's SpO2 accuracy against a reference method (CO-Oximetry), without human intervention in the loop of the measurement and calculation process. The non-clinical simulator tests also represent standalone performance.

    7. The Type of Ground Truth Used:

    • Clinical Trial Ground Truth: Arterial blood CO-Oximetry was used to provide functional SaO2, serving as the ground truth for SpO2 accuracy comparison.
    • Non-Clinical Test Ground Truth: An SpO2 simulator was used to provide known SpO2 and PR values.

    8. The Sample Size for the Training Set:

    • The document does not provide information regarding a "training set" or its sample size. The studies described are validation or test sets for the device's accuracy. Pulse oximeters generally use established physical principles (Beer-Lambert law) and signal processing, rather than machine learning algorithms that require explicit training datasets in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as no training set or machine learning algorithm requiring such ground truth establishment is described for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122046

    Validate with FDA (Live)

    Date Cleared
    2013-02-04

    (207 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082487

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MD300W4 wrist pulse oximeter is a portable non-invasive device intended for spot checking ,data collection and recording of arterial oxygen saturation (SpO2) and pulse rate of adult and pediatric patient at home and hospital( including clinical use in internist/ surgery, Anesthesia, intensive care and etc).

    Device Description

    The proposed device MD300W4 consists of MCU circuit, power supply circuit, SpO2 module circuit, display circuit, Flash memory circuit, GPRS module circuit, charging circuit, real-time clock circuit, button circuit.
    It can measure, store, review and display the SpO2% and pulse rate value, time, ID number, pulse bar and battery power status, the connection of probe, and transmit data by GPRS or USB cable.
    The power supply is 4.2V Li-battery with capacity 1250mAh. The device can not be used to measure when it charges for the Li- battery.
    The Pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and a photodetector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is infrared light.
    Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in probe collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
    The proposed device has one model detachable sensor as the accessory of which the skin-contacting material is silicon. It is listed below:
    M-50G (510(k) number: K082487)
    The proposed device is not for life-supporting or life-sustaining, not for implant. The device or probe is not sterile and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.
    The device is software-driven and the software validation is provided in Section 14 Software on page 14-1.

    AI/ML Overview

    Here's an analysis of the provided text to fulfill your request:

    Acceptance Criteria and Device Performance Study for Wrist Pulse Oximeter MD300W4 (K122046)

    Based on the provided 510(k) summary, the device is a modified Wrist Pulse Oximeter, MD300W4, and its performance is evaluated against the predicate device, MD300W (K081125), and established standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MD300W4 are largely aligned with its predicate device and industry standards for pulse oximeters. The reported device performance is presented as compliance with these specifications.

    CharacteristicAcceptance Criteria (Target/Standard)Reported Device Performance (MD300W4 Results)
    SpO2 Accuracy70%-100% ±3% (General)70%-100% ±3% (Claimed by manufacturer)
    90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3% (Specific ranges)90%-100% ±2%, 80%-90% ±3%, 70%-80% ±3%, <70% unspecified (Matches predicate and alleged performance)
    PR Accuracy±2% or 2 bpm, whichever is greater (General)±2% or 2 bpm, whichever is greater (Matches predicate and alleged performance)
    SpO2 Display Range0%-100%0%-100%
    SpO2 Measuring Range70%-100%70%-100%
    SpO2 Resolution1%1%
    PR Measuring Range30~235bpm30~235bpm
    PR Resolution1bpm1bpm
    BiocompatibilityCompliance with ISO 10993 "Biological Evaluation of Medical Devices"Meets the requirement of Biocompatibility, in compliance with ISO 10993.
    Software ValidationCompliance with FDA Guidance for "Content of Premarket Submissions for Software Contained in Medical Devices"In compliance with FDA Guidance.
    Electrical SafetyIEC 60601-1:1988+A1:1991+A2:1995Designed, tested, and will be manufactured in accordance with IEC 60601-1.
    EMCIEC 60601-1-2:2007Designed, tested, and will be manufactured in accordance with IEC 60601-1-2.
    Performance (Pulse Oximeters)ISO 9919:2005Designed, tested, and will be manufactured in accordance with ISO 9919:2005. Clinical test conducted according to ISO9919:2005 Annex EE.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: 12 health adults.
    • Data Provenance: Prospective clinical test conducted in Yue Bei People's Hospital. The country of origin can be inferred as China, given the manufacturer's location and the hospital name.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that SpO2 values were compared to "arterial blood sample oxygen saturation (functional SaO2) as measured by CO-Oximetry," referring to this as the "Golden Standard Co-Oximeter." This implies a technical measurement-based ground truth rather than expert consensus on subjective interpretations.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method. The ground truth is established directly by a "Golden Standard" Co-Oximeter for arterial blood sample oxygen saturation, rather than through subjective expert review requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a pulse oximeter, which directly measures physiological parameters (SpO2 and pulse rate) and does not involve "human readers" interpreting images or data in the context of an "AI" system that would enhance human performance. The study described is a clinical accuracy study comparing the device's readings against a reference method.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was done. The clinical test described involved comparing the device's SpO2 measurements directly against arterial blood sample oxygen saturation measured by a CO-Oximeter. This assesses the device's algorithmic performance in isolation (determining SpO2 from its sensor input) without human intervention in the measurement process itself, beyond the initial setup.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The type of ground truth used was objective measurement data from a "Golden Standard" Co-Oximeter measuring functional SaO2 from arterial blood samples. This is considered a highly accurate and direct physiological measurement.

    8. The Sample Size for the Training Set

    The document does not mention a separate training set or its sample size. The clinical test described appears to be for validation/testing, not for training a machine learning model. For a traditional medical device like a pulse oximeter, the core algorithm for converting optical signals to SpO2 and PR is typically based on established physiological models and calibration, rather than iterative machine learning training on a large dataset in the way an AI diagnostic algorithm would be.

    9. How the Ground Truth for the Training Set Was Established

    As no separate training set is mentioned (or implied as relevant for this type of device's core function), the method for establishing ground truth for a training set is not applicable based on the provided information.

    Ask a Question

    Ask a specific question about this device