    Intended Use
    1. Class III and V restorations
    2. Restoration of primary teeth
    3. Core build-up
    4. Cases where radiopacity is required
    5. Base material for Class I and Class II cavities using a sandwich laminate technique.
    Device Description

    GC Fuji Direct is a light-cured resin-modified restorative glass ionomer cement. The device consists of two pastes filled in a Paste Pak Cartridge. Paste A and B are dispensed with Paste Pak Dispenser II and hand mixed. Paste A and B can also be automixed by attaching the GC Fuji Direct Mixing Tip to the Paste Pak Dispenser II and applied directly to the prepared cavity. The bioactive material sets by acid-base reaction of fluoroalumino-silicate glass and Polyacrylic acid, and polymerization of methacrylate monomers.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, structured to include all your requested points:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    PropertyAcceptance Criteria (ISO 9917-2:2010)Reported Device Performance (GC Fuji Direct)
    AppearanceNo visible sign of extraneous matter, Free of any gelationConforms to requirements
    Working timeThe indenter shall make a complete circular indentation.Conforms to requirements
    Setting timeThe indenter shall fail to make a complete circular indentation.Conforms to requirements
    Flexural strengthMore than 25 MPaConforms to requirements
    RadiopacityMore than equivalent thickness of aluminumConforms to requirements
    Shade and color stabilityShade: Shall match with the manufacturer's nominated shade guide; Color stability: No significant change from the reference specimenConforms to requirements

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes bench testing against ISO standards and biocompatibility assessments. Therefore, there isn't a "test set" in the context of a clinical study with human or biological samples in the way this question typically implies for AI/diagnostic devices. The data provenance is related to laboratory testing, likely conducted by or for GC America Inc., a US-based company, referencing the specified international standards. No country of origin for the specific test data is given beyond the company's location. The testing appears to be prospective (as part of the 510(k) submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a bench test and biocompatibility assessment against technical specifications rather than an evaluation requiring expert-established ground truth on a test set (e.g., medical image interpretation). The "ground truth" for these tests is the quantitative and qualitative requirements of the ISO standard.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. This is a bench test against technical specifications, not a study requiring adjudication of expert interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This device is a dental cement, not an AI software/diagnostic device that would typically involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This device is a dental cement, not an algorithm, so this question is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance testing is adherence to the technical specifications defined in ISO 9917-2:2010 (Dentistry - Water-based cements - Part 2: Resin-modified cements) and the protocols outlined in ISO 10993-1:2009, ISO 10993-5, and ISO 10993-10 for biocompatibility. These are objective, predefined standards for material properties and biological response.

    8. The sample size for the training set

    N/A. This is a dental material, not an AI model, so there is no "training set."

    9. How the ground truth for the training set was established

    N/A. As there is no training set for an AI model, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173500

    Validate with FDA (Live)

    Device Name
    NMF004A
    Manufacturer
    Date Cleared
    2018-03-08

    (115 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K123631, K082434

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Liner or base
    2. Blocking out undercuts
    3. Repair of (in) direct aesthetic restorations, temporary crown & bridge, defect margins are in enamel
    4. Sealing hypersensitive areas
    5. Fissure sealant
    6. Direct restorative for Class I, II, III, IV, V cavities
    7. Core build-up
    Device Description

    NMF004A is a light cured nano-filled radiopaque composite resin filled in a syringe. The device is used for the restorations of both anterior and posterior teeth. The material is available in 30 shades.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental device named NMF004A. This document asserts the device's substantial equivalence to previously cleared predicate devices by outlining its characteristics, indications for use, and performance bench tests against a recognized standard (ISO 4049:2009).

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document states that "The applicant device complies with all the requirements of ISO 4049: 2009 (Dentistry - Polymer-based restorative materials)." While specific measured values for NMF004A are not explicitly provided in the table, the text implies that NMF004A's performance meets these requirements. The table below presents the acceptance criteria from ISO 4049:2009 as listed in the document.

    PropertyAcceptance Criteria (from ISO 4049:2009)Reported Device Performance (NMF004A)
    Sensitivity to ambient lightRemain physically homogeneousComplies
    Depth of cureOpaque shade: > 1.5mmOther shade: > 2.0mmComplies
    Flexural strength> 80 MPaComplies
    Water sorption< 40 µg/mm³Complies
    Solubility< 7.5 µg/mm³Complies
    Color stability after irradiation and water sorptionNo more than slight change in color.Complies
    RadiopacityGreater than the same thickness of aluminumComplies

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample sizes used for the performance bench tests. It also does not provide information regarding the data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The "performance bench tests" described are laboratory tests against ISO standards, not studies involving expert evaluation of clinical data for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable and therefore not provided in the document. The performance tests are objective measurements against ISO standards, not subjective evaluations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not provided and is not applicable. The device, NMF004A, is a tooth shade resin material, not an AI-assisted diagnostic tool or an imaging device. Therefore, an MRMC study comparing human readers with and without AI assistance would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not provided and is not applicable. NMF004A is a physical dental material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance bench tests is the established requirements of the ISO 4049:2009 standard (Dentistry - Polymer-based restorative materials). These are objective, scientifically defined physical and chemical properties of dental materials.

    8. The sample size for the training set:

    This information is not provided and is not applicable. The device is a physical material, not a machine learning model; therefore, it does not have a "training set."

    9. How the ground truth for the training set was established:

    This information is not provided and is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123631

    Validate with FDA (Live)

    Device Name
    MFP-051
    Manufacturer
    Date Cleared
    2013-07-23

    (242 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K082434, K091388, K032921, K042819

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Direct restorative for class I, II, III, IV, V cavities.
    2. Direct restorative for wedge-shaped defects and root surface cavities.
    3. Direct restorative for veneers and diastema closure.
    Device Description

    MFP-051 is a light-cured, nano-filled radiopaque composite resin filled in either a syringe or unitip. The device is a universal type. The material is available in 17 shades. The device is used for the restorations of both anterior and posterior teeth.

    AI/ML Overview

    The provided document is a 510(k) summary for a dental restorative material (MFP-051). It focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study of the device's performance against specific acceptance criteria in a clinical or simulated clinical setting. As such, most of the requested information regarding acceptance criteria and performance studies is not available in this document.

    Here is a breakdown of the information that can be extracted or deduced from the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Feature/TestAcceptance Criteria (Implied by ISO 4049)Reported Device PerformanceAssessment
    Substantial EquivalenceEquivalent to predicate devices in function, principle, technology, safety, and intended use."The curing mechanism of the new and predicate devices is substantially equivalent... in function, and similar in principle, increr (sic), and intended use. This supports that the compatibility and safety of the applicant device are substantially equivalent to the predicate devices."Met
    Material Properties (Implied by ISO 4049 and predicate comparison)Flexural strength, water sorption, and solubility should be equivalent to predicate devices."The applicant device also shows equivalence in flexural strength, water sorption and solubility, which indicate the stability of materials in oral environment."Met
    Mechanical/Physical PropertiesCompliance with ISO 4049:2009 (Dentistry - Polymer-based restorative materials)."The applicant device complies with all the requirements of ISO 4049: 2009..."Met
    ComponentsComponents should be previously used in predicate devices or generally accepted for dental materials."All the components of the applicant device, MFP-051, have already been used in the predicate devices."Met
    Indications for UseSimilar to predicate devices.Proposed indications are direct restorative for Class I, II, III, IV, V cavities, wedge-shaped defects and root surface cavities, veneers, and diastema closure, which are consistent with predicate device indications.Met

    2. Sample size used for the test set and the data provenance:

    • Not explicitly stated in the document. The document refers to meeting ISO standards and equivalence, implying testing was done, but details about the sample size, test methods, or data provenance (e.g., country of origin, retrospective/prospective) are not provided for specific performance tests. This submission relies on demonstrating compositional and functional similarity to already cleared devices and compliance with an international standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This document does not describe a study involving expert assessment or "ground truth" establishment in the context of clinical or diagnostic performance. The evaluation is based on material properties and regulatory standards.

    4. Adjudication method for the test set:

    • Not applicable. See point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a dental restorative material, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to AI assistance is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical dental material, not an algorithm.

    7. The type of ground truth used:

    • For the material properties (flexural strength, water sorption, solubility), the "ground truth" would be established by validated laboratory testing methods against the requirements of ISO 4049:2009.
    • For the claim of "substantial equivalence" of components and indications, the ground truth is the characterization of the predicate devices and their established safety and effectiveness.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that requires a "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    In summary:

    This 510(k) submission for MFP-051 demonstrates substantial equivalence by showing that the device complies with ISO 4049:2009, contains components already used in predicate devices, and shows equivalence in key material properties (flexural strength, water sorption, solubility) to similar legally marketed devices. It does not contain details of specific clinical performance studies with acceptance criteria, sample sizes, or expert adjudication as would be relevant for a diagnostic or AI device. The "acceptance criteria" primarily revolve around meeting an international standard (ISO 4049) and demonstrating technical and functional similarity to predicate devices.