510(k) Data Aggregation

    K Number
    K122639

    Validate with FDA (Live)

    Date Cleared
    2013-05-24

    (268 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SynCage Evolution spacer is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the SynCage Evolution spacer should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

    Device Description

    The Synthes SynCage Evolution spacer is a radiolucent device for use in interbody fusion. It is to be used in conjunction with supplemental fixation to provide structural stability in skeletaly mature individuals. The Synthes SynCage Evolution spacer is fabricated from Invibio® PEEK-OPTIMA® LT-1 (ASTM F2026-10) with four anterior and one posterior tantalum (ASTM F560-08) radiopaque markers. The markers allow intra-operative radiographic assessment of the position of the implant. Implantation is through an anterior or anterolateral approach. The Synthes SynCage Evolution spacer is provided sterile. The Synthes SynCage Evolution spacer is available in three footprints (Small: 32x25mm; Medium, 36x28mm; Large, 40x31mm), four lordotic angles (6°, 10°, 14°, 18°), and a range of heights (9-19mm) to suit individual pathology and anatomical conditions. Pyramidal teeth that assist in stabilization of the construct are located on the inferior and superior surfaces of the spacers. The open architecture of the device allows it to be packed with autogenous bone graft material (i.e., autograft).

    AI/ML Overview

    The provided text describes a 510(k) summary for the Synthes SynCage Evolution Spacer, an intervertebral body fusion device. It details the device's description, intended use, and a comparison to predicate devices, focusing on non-clinical performance data.

    Here's an analysis based on your request:

    1. Acceptance Criteria and Reported Device Performance

    The document does not specify explicit acceptance criteria in terms of numerical thresholds or performance targets for a device meeting clinical or diagnostic accuracy. Instead, it states that the device's performance was evaluated through non-clinical bench testing and compared to predicate devices for substantial equivalence.

    Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):

    Acceptance Criteria CategorySpecific Criteria (Inferred from document)Reported Device Performance
    Mechanical PerformanceStatic Axial Compression (ASTM F2077-11)Substantially equivalent in performance to predicate devices.
    Dynamic Axial Compression (ASTM F2077-11)Substantially equivalent in performance to predicate devices.
    Static Compression Shear (ASTM F2077-11)Substantially equivalent in performance to predicate devices.
    Expulsion (ASTM F2267-04)Substantially equivalent in performance to predicate devices.
    Subsidence (ASTM F2267-04)Substantially equivalent in performance to predicate devices.
    Material BiocompatibilityFabricated from Invibio® PEEK-OPTIMA® LT-1 (ASTM F2026-10) with tantalum (ASTM F560-08) radiopaque markers.Materials conform to specified ASTM standards.
    RadiopacityMarkers allow intra-operative radiographic assessment of implant position.Meets requirement for intra-operative radiographic assessment.

    2. Sample Size for the Test Set and Data Provenance

    The document explicitly states: "Clinical data and conclusions were not needed for this device."
    Therefore, there was no test set of clinical data used to evaluate the device's performance in humans. The evaluation was based solely on non-clinical bench testing.

    • Sample Size for Test Set: Not applicable (no clinical test set).
    • Data Provenance: Not applicable (no clinical data).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    As no clinical data was used for evaluation, there was no ground truth established by experts for a test set. The evaluation relied on engineering principles and bench testing standards.

    • Number of Experts: Not applicable.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    Since no clinical test set was used, there was no adjudication method employed.

    • Adjudication Method: Not applicable.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC study was conducted. The document states: "Clinical data and conclusions were not needed for this device."

    • MRMC Study: No.
    • Effect Size of Human Readers with vs. without AI: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is an intervertebral fusion spacer, not an algorithm or AI system. Therefore, the concept of "standalone performance" for an algorithm is not applicable. The device's performance was evaluated through bench testing.

    • Standalone Study: Not applicable (device is not an algorithm).

    7. The Type of Ground Truth Used

    For the non-clinical performance evaluation, the "ground truth" was established by standardized engineering testing protocols as defined by ASTM F2077-11 and ASTM F2267-04. The performance was then compared to that of legally marketed predicate devices, implying that their established performance under these tests served as a reference for "substantial equivalence."

    • Type of Ground Truth: Bench testing standards (ASTM F2077-11 and ASTM F2267-04) and performance of predicate devices.

    8. The Sample Size for the Training Set

    This document describes a physical medical device, not an AI or machine learning model. Therefore, the concept of a "training set" is not applicable.

    • Sample Size for Training Set: Not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set mentioned or implied for this physical device, this question is not applicable.

    • How Ground Truth for Training Set Was Established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K082406

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-01-14

    (146 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K081636

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Expandable PEEK-OPTIMA® Implant is indicated for vertebral body replacement and intervertebral fusion. When used for vertebral body replacement, The Expandable PEEK-OPTIMA® Implant is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Expandable PEEK-OPTIMA® Implant is also indicated for treating fractures of the thoracic and lumbar spine. The Expandable PEEK-OPTIMA® Implant is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period.

    As an intervertebral body fusion device, the Expandable PEEK-OPTIMA® Implant is indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

    The Expandable PEEK-OPTIMA® Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

    Device Description

    The Expandable PEEK-OPTIMA® Implant is a rectangular, expandable device constructed of medical grade Polyetheretherketone (PEEK), for spinal applications.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving device performance against such criteria. The document is a 510(k) Summary for the "Expandable PEEK-OPTIMA® Implant," outlining its description, indications for use, and a claim of substantial equivalence to predicate devices, primarily based on technological characteristics and mechanical testing for equivalence, not detailed performance criteria or clinical study results.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them, nor can I provide the associated details about sample size, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth information.