510(k) Data Aggregation

    K Number
    K222397

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-29

    (233 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K151983, K081623

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KLS Martin Level One Rib Fixation System is indicated for use in the stabilization and rigid fixation of rib fractures in the chest wall including reconstructive procedures, trauma, or planned osteotomies in patients 18 years of age or older.

    Device Description

    The KLS Martin Level One Rib Fixation System is comprised of PEEK plates and titanium locking screws intended to provide rigid fixation of bone in the thoracic anatomy. The PEEK plates are pre-contoured to accommodate patient anatomy and are offered in plate thicknesses of 2 mm - 3 mm. The PEEK plates are compatible with the Ti-6Al-4V (ASTM F136) 2.3 mm x 7 mm multidirectional locking screws offered in the system. The plates are manufactured from PEEK (ASTM F2026). The system includes the necessary instrumentation

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the KLS Martin Level One Rib Fixation System. It focuses on establishing substantial equivalence to predicate devices based on non-clinical performance data and discussions of intended use and technological characteristics.

    However, the provided document DOES NOT contain information regarding a study involving AI/algorithm performance for a diagnostic or AI-assisted device. Therefore, I cannot extract the specific details requested in your prompt regarding acceptance criteria and performance studies for an AI device.

    The document primarily addresses the substantial equivalence of a physical medical device (rib fixation system) based on mechanical and biological safety testing, rather than a diagnostic or AI-driven system.

    Therefore, many of the requested points, such as "table of acceptance criteria and the reported device performance" (for an AI device), "sample size used for the test set and the data provenance," "number of experts used to establish the ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "type of ground truth," "sample size for the training set," and "how the ground truth for the training set was established," are not applicable to the content provided.

    The non-clinical performance data mentioned refers to:

    • Comparative head-to-head bench testing: To determine if the subject device performs equivalently to or better than the primary predicate device in pull-out system testing.
    • Comparison to a reference device (Synthes MatrixRIB Fixation System): For pull-out system testing.
    • Biological safety risk assessments: In compliance with ISO 10993-1:2018, concluding the devices are biocompatible.

    The document explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence." This further confirms that no human-in-the-loop or standalone diagnostic performance studies were conducted or reported in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161590

    Validate with FDA (Live)

    Date Cleared
    2016-11-03

    (147 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Synthes MatrixRIB Fixation System is indicated for use in skeletally mature patients with normal or osteoporotic bone for chest wall fixation, where:

    DePuy Synthes MatrixRIB pre-contoured plates are indicated for the fixation, stabilization and reconstruction of:

    • Rib fractures, fusions, osteotomies, and/or resections, including spanning gaps and/or defects
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities

    DePuy Synthes MatrixRIB straight plates are indicated for the fixation, stabilization and reconstruction of:

    • Rib and sternum fractures, fusions, osteotomies, and/or resections, including spanning gaps and/or defects
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities

    DePuy Synthes MatrixRIB sternal plates, 2.8mm thickness, are indicated for the fixation, stabilization and reconstruction of:

    • Sternum fractures, fusions, and/or osteotomies
    • Pectus Excavatum, Pectus Carinatum, and other chest wall deformities

    The DePuy Synthes MatrixRIB intramedullary splints and the universal plate are indicated for the fixation and stabilization of ribs.

    Device Description

    The proposed additions to the MatrixRIB Fixation System include sternal plates (straight plates, I-plates and T-plates) for fixation and reconstruction of the chest wall manufactured from commercially pure titanium (CP Ti Grade 4).

    The previously cleared devices in the MatrixRIB Fixation System (K133616 and K081623) include locking plates, locking screws, and intramedullary splints for the fixation and stabilization of the chest wall for use in patients with normal or osteoporotic bone quality. These implants are available in multiple sizes and are manufactured from titanium alloy (Ti-6A-7Nb).

    AI/ML Overview

    The provided information is a 510(k) Summary for the MatrixRIB Fixation System, which is a medical device for chest wall fixation. This type of document is a premarket notification to the FDA to demonstrate that the device is at least as safe and effective as a legally marketed predicate device.

    Based on the provided text, the acceptance criteria and study information are largely related to non-clinical performance data and not to clinical performance data or AI model performance.

    Here's an analysis of the provided text in response to your request, with an emphasis on what is present and what is explicitly stated as not present:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "The non-clinical performance data demonstrates that the mechanical performance of the proposed MatrixRIB Fixation System is comparable to that of the predicates." However, it does not provide a specific table of acceptance criteria with numerical values or detailed reported device performance. It only lists the types of tests conducted.

    Test Method (Non-Clinical)Acceptance Criteria (Not Explicitly Stated)Reported Device Performance (Not Explicitly Stated)
    Intraoperative Contouring Dynamic Test Method(Implicit: Comparable to predicates)(Implicit: Comparable to predicates)
    Comparative Four Point Static Test Method(Implicit: Comparable to predicates)(Implicit: Comparable to predicates)
    Bacterial Endotoxin Testing(Implicit: Met specified endotoxin limits)Met specified endotoxin limits

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: The document does not specify the sample sizes (number of implants tested) for the non-clinical tests.
    • Data Provenance: The document does not specify the country of origin of the test data or whether it was retrospective or prospective. Given these are mechanical laboratory tests, such classifications are not typically applicable in the same way as clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This question is not applicable. The device is a physical fixation system, not an AI or diagnostic imaging device that requires human expert interpretation to establish ground truth for a test set. The "ground truth" for mechanical testing would be defined by engineering standards and test methodology.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This is not applicable for mechanical, non-clinical performance testing. Adjudication methods are typically used in clinical studies or for establishing ground truth in AI model development where human evaluators are involved.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The MatrixRIB Fixation System is a physical medical implant, not an AI-assisted diagnostic or clinical decision support system. Therefore, no MRMC study, human reader improvement with AI, or effect size related to AI assistance would be performed or reported for this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The MatrixRIB Fixation System is a physical medical implant. There is no algorithm or AI component to this device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    For the non-clinical performance data, the "ground truth" is established through engineering standards and validated test methods. For example, the "Intraoperative Contouring Dynamic Test Method" and "Comparative Four Point Static Test Method" would have defined parameters and criteria based on established biomechanical principles and regulatory guidelines (though these specific criteria are not detailed in the summary).

    8. The sample size for the training set:

    This is not applicable. There is no AI component requiring a training set for this physical medical device.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as point 8.