    K Number
    K241977

    Validate with FDA (Live)

    Date Cleared
    2024-10-03

    (90 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K080917

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Permanent obturation of the root canal following vital pulp-extirpation.
    • Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.

    i-MTA SP Injectable Root Canal Bioceramic Sealer is suitable for use in the single cone and placement of intracanal dressings.

    Device Description

    i-MTA SP Injectable Root Canal Bioceramic Sealer is a convenient premixed ready-to-use, injectable white hydraulic bioceramic paste developed for permanent root canal filling and sealing applications. an injectable mineral trioxide aggregate. i-MTA SP is used for root canal sealing and permanent obturation. It is an insoluble and radiopaque material based on a calcium silicates composition, which requires the presence of water to set and harden. i-MTA SP is packaged in an oral irrigator and is supplied with disposable Intra Canal Tips. i-MTA SP may be delivered into the canal via the disposable tips or it can be delivered via traditional methods.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the "Injectable Root Canal Bioceramic Sealer (i-MTA SP)". This document focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving the device meets specific acceptance criteria through a clinical study with AI or human readers.

    Therefore, the requested information regarding acceptance criteria, AI/human reader performance, sample sizes for test/training sets, expert ground truth establishment, or MRMC studies is not available in the provided text.

    The text primarily details:

    • Device Name: Injectable Root Canal Bioceramic Sealer (i-MTA SP)
    • Regulation Number: 21 CFR 872.3820
    • Regulation Name: Root Canal Filling Resin
    • Product Code: KIF
    • Indications for Use:
      • Permanent obturation of the root canal following vital pulp-extirpation.
      • Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.
      • Suitable for use in the single cone and placement of intracanal dressings.
    • Comparison to Predicate Devices: The document explicitly states its purpose is to show "substantial equivalence" based on "similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices."
    • Performance Data Provided:
      • ISO 6876:2012 Dentistry - Root canal sealing materials.
      • Endodontic usage Test- ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
      • Biocompatibility Assessment per FDA Biocompatibility Guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

    No information about AI performance, human reader studies, or the specific criteria for such studies is present in this 510(k) summary. This device is a material used for root canal filling, not a diagnostic or AI-powered device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221811

    Validate with FDA (Live)

    Device Name
    Vista BC Sealer
    Manufacturer
    Date Cleared
    2022-11-21

    (152 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K080917

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permanent obturation of the root canal following vital pulp-extirpation.
    Permanent obturation of the root canal following removal of infected or necrotic pulp and placement of intracanal dressings.

    Device Description

    Vista BC Sealer is a convenient premixed ready-to-use injectable white hydraulic cement paste developed for permanent root canal filling and sealing applications. Vista BC Sealer is an insoluble, radiopaque material based on a calcium phosphosilicate and calcium aluminate cement composition, which requires the presence of water to set and harden. Vista BC Sealer does not appreciably shrink during setting and demonstrates excellent physical properties. Vista BC Sealer is packaged in a preloaded syringe and supplied with disposable single use application tips.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Vista BC Sealer, a root canal filling material. It details non-clinical performance testing but does not include information about acceptance criteria or a study proving that the device meets those criteria in the context of an AI/ML algorithm or a typical clinical study with a test set, ground truth, or expert involvement. The submission focuses on demonstrating substantial equivalence to predicate devices through physical, chemical, and biocompatibility testing.

    Therefore, many of the requested fields regarding acceptance criteria, study design for AI/ML, sample sizes for test/training sets, ground truth establishment, expert involvement, and MRMC studies cannot be answered from the provided document as they are not relevant to this type of device clearance.

    However, I can extract the information related to the performance demonstrated by the non-clinical tests.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria for many of the tests in a table format, but it does state compliance with ISO standards and favorable comparison to the predicate device. Where specific values or qualitative outcomes are mentioned, they are included below.

    Test CategoryAcceptance CriteriaReported Device Performance
    Shelf-Life TestingNot explicitly stated; implied to meet typical storage conditions.Through accelerated shelf life testing, Vista BC Sealer was found to have a shelf-life of two years. Real-time aging is being performed to support this.
    ISO 6876 TestingCompliance with ISO 6876 (standards for flow, set time, film thickness, solubility/degradation, radiopacity).Performed very similarly to the predicate device, ENDOSEAL MTA, thereby confirming substantial equivalence for: * Flow testing * Set time testing (Vista BC Sealer: < 90 minutes; Predicate ENDOSEAL MTA: 12.31 minutes) * Film thickness testing * Solubility and degradation testing * Radiopacity testing
    Transit TestingPackaging should withstand simulated transit conditions, and products should perform satisfactorily post-transit.Packaging configurations are sufficient and withstand simulated transit conditions. Products performed satisfactorily post-transit, confirming transit did not have a negative effect.
    Microbiological TestingNot explicitly stated; implied to mitigate contamination risks and exhibit bactericidal properties.Exhibits bactericidal properties. Results support shelf stability and multiple use of non-patient contacting materials (syringes) as introduced microbes will not remain viable.
    Cytotoxicity TestingNot explicitly stated; implied to be acceptable or superior to predicate.Yielded equivalent, or better, cytotoxicity results compared to the predicate device, ENDOSEAL MTA.
    Sensitization TestingNo evidence of causing delayed dermal contact sensitization.Showed no evidence of causing delayed dermal contact sensitization in guinea pigs. Not considered a sensitizer.
    Intracutaneous Irritation TestingDoes not cause intracutaneous irritation.Does not cause intracutaneous irritation and is not considered an irritant.
    Acute Systemic Toxicity TestingNo mortality or evidence of systemic toxicity.No mortality or evidence of systemic toxicity from systemic exposure. All animals displayed normal behaviors and positive weight gain. Not systemically toxic.
    Genotoxicity TestingNo signs of mutagenicity.Showed no signs of mutagenicity during tests. Confirmed as non-mutagenic.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The tests performed are non-clinical (laboratory and animal studies), not studies on human subjects with test sets in the context of AI/ML or diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The ground truth for these non-clinical tests is based on standardized laboratory measurements and observations in animal models, not expert human assessment of a medical condition.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Adjudication methods are typically used in studies involving human interpretation or clinical endpoints, not in the non-clinical laboratory and animal studies described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, and this device is not an AI/ML diagnostic or assistive device for human readers. It is a root canal filling material.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical tests involved standardized laboratory measurements, physical property testing (e.g., set time, flow, solubility), and biological responses observed in in vitro cultures and in vivo animal models for biocompatibility.

    8. The sample size for the training set

    This information is not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    This information is not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181922

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-22

    (96 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K080917

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permanent obturation of the root canal following root canal treatment.

    Device Description

    Sonendo Material A is a single component injectable paste material intended for permanent obturation of the root canal following root canal treatment. The device is comprised of alginate and a calcium containing compound which crosslinks in the presence of water to form an insoluble polymer. The placement into the root canal can be accomplished with or without the use a root canal point. The material is radiopaque and has physical properties such that it is suitable for root canal obturation. The material is packaged in either a syringe with single use dispensing tips or in a single dose capsule.

    AI/ML Overview

    The provided text describes Sonendo Material A, a root canal filling resin, and its FDA 510(k) clearance based on substantial equivalence to a predicate device, iRoot SP Root Canal Sealer (K080917).

    However, the document does not contain any information about an AI/ML powered device, nor does it provide details about acceptance criteria for an AI/ML product, performance metrics typically associated with AI/ML (e.g., accuracy, sensitivity, specificity, AUC), or studies involving human readers or ground truth established by experts/pathology for an AI model.

    The "Performance Data" section of the document refers to bench testing to demonstrate physical-chemical properties and biocompatibility, which are standard for medical devices like root canal filling resins, not AI/ML software.

    Therefore, I cannot fulfill your request for information related to AI/ML device acceptance criteria and study details. The input document pertains to a traditional medical device (root canal filling resin) and its regulatory clearance process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170950

    Validate with FDA (Live)

    Device Name
    Well-Root ST
    Manufacturer
    Date Cleared
    2017-12-20

    (264 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080917

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Permanent sealing of root canal

    Device Description

    Well-Root™ ST is a convenient premixed ready-to use composition which requires the presence of water to set and harden. The device is contained in a plastic syringe and the system includes a plunger, disposable tips, and a holder.

    AI/ML Overview

    The provided text describes the regulatory clearance of a root canal filling resin, Well-Root™ ST, and its substantial equivalence to predicate devices. It does not present a study with acceptance criteria and reported device performance in the context of diagnostic accuracy, clinical outcomes, or human-in-the-loop performance, as implied by the request. Instead, the "performance data" section refers to bench testing against ISO standards and biocompatibility tests. These are typically used to demonstrate the physical, chemical, and biological safety of a material, not to evaluate the diagnostic or clinical efficacy of an AI-powered medical device.

    Therefore, many of the requested items, particularly those related to diagnostic accuracy, a test set, ground truth, experts, and comparative effectiveness studies, cannot be answered from the provided text.

    Here's an attempt to answer the questions based only on the information available in the provided text, highlighting where the information is not present or not applicable to the context of AI device studies:


    1. A table of acceptance criteria and the reported device performance

    The document lists performance standards it conformed to (ISO 6876 and ISO 10993 series), and mentions that "the test result of the subject device satisfies the requirement of the ISO 6876" and "the results of the bench and the biocompatibility tests performed demonstrates that any of these differences do not raise a new question as to safety and effectiveness." However, specific numerical acceptance criteria and reported performance values for these tests are not detailed in the provided text.

    The document mentions compliance with the following ISO standards and specific tests within them:

    Acceptance Criteria (Standard / Test)Reported Device Performance
    ISO 6876: FlowSatisfies requirements
    ISO 6876: Setting TimeSatisfies requirements
    ISO 6876: Film ThicknessSatisfies requirements
    ISO 6876: Solubility and disintegrationSatisfies requirements
    ISO 6876: Radio-opacitySatisfies requirements
    ISO 6876: Working timeSatisfies requirements
    ISO 10993-5 (Cytotoxicity)Compliant
    ISO 10993-10 (Sensitivity-LLNA, GPMT)Compliant
    ISO 10993-11 (Acute systemic Toxicity)Compliant
    ISO 10993-3 (Genotoxicity: Bacterial reverse mutation (Ames), In vitro mammalian chromosomal aberration)Compliant
    ISO 10993-6 (Implantation)Compliant
    Shelf Life: AppearanceCompliant
    Shelf Life: ISO 6876 tests (Flow, Setting Time, Film Thickness)Compliant

    Note: The text explicitly states, "the test result of the subject device satisfies the requirement of the ISO 6876 and does not raise an issue of safety and/or performance." It also mentions for shelf life "Appearance, ISO 6876 tests (Flow, Setting Time, Film Thickness)" were conducted, implying compliance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable/Not provided. This document describes physical, chemical, and biocompatibility testing of a root canal filling resin, not an AI-powered diagnostic device that would typically have a "test set" of data (e.g., images, patient records). The testing methods for ISO standards usually involve a defined number of samples for each test (e.g., N=3 or N=5 for flow, setting time, etc.), but these specific numbers are not given in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable/Not provided. Ground truth, in the context of diagnostic AI, refers to the confirmed correct diagnosis or outcome. For a physical device like a root canal sealer, "ground truth" would relate to the validated methods of the ISO standards themselves, not expert judgment on cases.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable/Not provided. Adjudication methods are relevant for resolving discrepancies in expert interpretations (e.g., in reading medical images) for establishing ground truth, which is not the type of study described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not a study involving human readers or AI assistance. It is a submission for a physical medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the ISO and biocompatibility tests: The "ground truth" is established by the specified methodologies and criteria outlined within the respective ISO standards (e.g., ISO 6876 for physical properties of root canal filling materials, ISO 10993 series for biocompatibility). These standards define the acceptable range or characteristics for the device's performance.

    8. The sample size for the training set

    Not applicable/Not provided. As this is not an AI device, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable/Not provided. As this is not an AI device, there is no "training set" or corresponding ground truth establishment in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120048

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-20

    (166 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080917

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DRFP propoint obturation points are designed for permanent sealing of root canals when used with DRFP smartpaste Dental Sealer or Brasseler USA EndoSequence® BC Sealer following established endodontic procedures by qualified healthcare professionals.

    Device Description

    The propoints are identical to the points cleared for marketing with K100248. They are made of a radiopaque core and an opaque hydrophilic coating of cross-linked hydrophilic polymers that have been used in the contact lens industry for more than 20 years.

    propoints are offered in the traditional 4% and 6% sizes and for use with variable taper files (such as ProTaper™ and Sendoline S5™). They are biocompatible and fully opaque. The hydrophilic coating expands radially to push the sealer into irregularities and voids present in the root canal system for a tight mechanical seal with the dentine. The coating only expands into available space and does not exert pressure on tooth structures. Dimensional length stability is maintained.

    The points are packaged in clean room conditions and come individually wrapped to prevent cross-contamination. They can be cut to exact tip size and length.

    No heating or compression equipment is required. Post insertion can commence immediately after root canal has been filled. The hydrophilic properties of the device allow extraction without softeners in revision procedures.

    propoints are

    • coated with DRFP smartpaste prior to insertion into the root canal . (K100248) or
    • inserted after the root canal has been filled with EndoSequence® BC. .
    AI/ML Overview

    This report describes the acceptance criteria and study that proves the device meets the acceptance criteria for DRFP Ltd.'s Propoints with iRoot SP Sealer.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Effective bonding with dentin and dentin tubules in root canalYes
    Meets Bench Test Criteria in Dye Ingress TestYes (No sign of dye ingress after sectioning and assessment under a microscope in in-vitro dye penetration tests using heat-sterilized extracted teeth filled with EndoSequence® and propoints and immersed in Brilliant Blue Dye solution for at least one week.)
    Effect of Interaction with Sodium Hypochlorite / EDTA residual levels after rinsing 3 timesNone (Remaining 0.01% sodium hypochlorite after rinsing 3 times had no effect; undiluted 17% EDTA had no effect on the system.)
    BiocompatibleYes (Materials shown to be non-toxic, non-carcinogenic, and biocompatible with tissue fluids according to ISO 10993 and ISO 7405.)
    Reported Failure Rate in Clinical Use0.8% (Out of approximately 2000 treatments, predominantly attributed to tooth rather than material problems.)
    Radiopacity in Clinical X-raysGood (Clinical X-rays validate excellent radiopacity of both devices and confirm effective bonding to tooth structures and provide evidence of good healing.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Approximately 2000 treatments were reviewed for clinical performance.
    • Data Provenance: The clinical data is retrospective, derived from the "DRFP User Panel of practicing dentists". The country of origin for the data is not explicitly stated, but the submission is from DRFP Ltd. in the United Kingdom and their "User Panel" is mentioned, implying data collected from dental practices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: A "DRFP User Panel of practicing dentists" was used. The exact number of dentists on this panel is not specified.
    • Qualifications of Experts: They are described as "practicing dentists." Their specific qualifications (e.g., years of experience, specialization) are not detailed in the provided text.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The failure rate and clinical observations appear to be reported directly by the "DRFP User Panel of practicing dentists."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC comparative effectiveness study, nor is AI involved in this device. This submission is for a conventional medical device (root canal obturation points and sealer).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical medical device (root canal obturation system), not an algorithm or AI.

    7. The Type of Ground Truth Used

    • For Clinical Performance: The ground truth for clinical performance (failure rate, healing, bonding effectiveness) was established through observations and reporting by a "DRFP User Panel of practicing dentists" and confirmed through "clinical x-rays."
    • For Bench Tests: The ground truth for bench tests (dye penetration, sodium hypochlorite/EDTA interaction, radiopacity) was based on laboratory measurements and visual assessment (e.g., microscopic assessment for dye ingress).
    • For Biocompatibility: Ground truth was established by testing according to ISO 10993 and ISO 7405 standards.

    8. The Sample Size for the Training Set

    There is no mention of a "training set" in the context of an algorithm or AI. This submission is for a physical medical device, not an AI/ML product.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as there is no "training set" for an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082943

    Validate with FDA (Live)

    Date Cleared
    2009-01-06

    (96 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K080917, K952614, K061906

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • Repair of Root Perforation
    • Repair of Root Resorption
    • Root End Filling
    • Apexification
    • Pulp Capping
    Device Description

    iRoot BP Injectable Root Canal Repair Filling Material (iRoot BP) is a convenient ready-to-use white hydraulic premixed injectable BioAggregate paste developed for permanent root canal repair and filling applications. iRoot BP is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. iRoot BP does not shrink during setting and demonstrates excellent physical properties. iRoot BP is packaged in a preloaded syringe and is supplied with disposable tips.

    AI/ML Overview

    The device in question, iRoot BP Injectable Root Canal Repair Filling Material, is a dental product and theprovided text describes its 510(k) submission to the FDA. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than establishing performance against specific clinical acceptance criteria in the way an AI/ML device would.

    Therefore, the following information is not applicable or not provided in the given text:

    • Acceptance criteria and reported device performance in a table (as this is not an AI/ML device with quantifiable performance metrics like accuracy, sensitivity, etc.).
    • Sample size for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method for the test set.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone performance (without human-in-the-loop).
    • Type of ground truth used (e.g., expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    Instead, the documentation focuses on non-clinical tests to establish substantial equivalence.

    Here's an analysis of the provided information, framed as closely as possible to your request, while acknowledging the limitations for a non-AI/ML device:

    Device: iRoot BP Injectable Root Canal Repair Filling Material

    1. Acceptance Criteria and Reported Device Performance

    For this type of device, the "acceptance criteria" are generally related to demonstrating substantial equivalence to existing predicate devices through comparable chemical, physical, and biocompatible properties, as well as similar performance specifications. The "reported device performance" are the results of these non-clinical tests compared to the established properties of the predicates.

    Acceptance Criteria (Demonstrates Substantial Equivalence to Predicate Devices)Reported Device Performance (Summary of Non-Clinical Test Results)
    Equivalent chemical composition to predicate devices BioAggregate, iRoot SP, Jeltrate® Plus™ Impression Material, and Clearfil™ Ceramic Primer.iRoot BP's main chemical composition is based on BioAggregate. Additional chemical components are found in iRoot SP, Jeltrate® Plus™ Impression Material, and Clearfil™ Ceramic Primer, which are deemed safe and effective.
    Comparable physical properties to BioAggregate (flow, working time, setting time, dimensional change following setting, solubility, radiopacity).Bench tests for iRoot BP included flow, working time, setting time, dimensional change following setting, solubility, and radiopacity, and demonstrated comparability to BioAggregate and iRoot SP.
    Comparable biocompatibility properties (non-cytotoxic, non-mutagenic, no allergenic potential, good tolerance by subcutaneous tissue) to BioAggregate and iRoot SP.iRoot BP was determined to be non-cytotoxic. Biocompatibility data from BioAggregate and iRoot SP (chemical components are shared) provided evidence that iRoot BP is non-mutagenic, does not cause an allergenic potential after multiple uses, and has good subcutaneous tissue tolerance.
    Similar delivery system to iRoot SP.iRoot BP has a comparable delivery system to iRoot SP (preloaded syringe with disposable tips).
    Same indications for use as predicate devices.iRoot BP has the same indications for use as specified for predicate devices (Repair of Root Perforation, Repair of Root Resorption, Root End Filling, Apexification, Pulp Capping).

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The testing described is primarily laboratory bench testing and biocompatibility assessment, not a clinical trial with a "test set" in the context of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This is not applicable as the evaluation is based on non-clinical laboratory tests and established properties of predicate devices, not expert interpretation of clinical cases.

    4. Adjudication Method for the Test Set

    This is not applicable, as there isn't a test set requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI/ML system and does not involve human readers for interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    For physio-chemical properties and performance specifications, the "ground truth" is derived from established standards for dental materials and the known, FDA-accepted properties of the predicate devices (BioAggregate, iRoot SP, Jeltrate® Plus™ Impression Material, Clearfil™ Ceramic Primer).
    For biocompatibility, the "ground truth" comes from standard biocompatibility testing protocols and the established safety profiles of the chemical components present in the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. This device does not use a training set for machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set mentioned or implied for this device.

    Ask a Question

    Ask a specific question about this device