    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems.

    The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous graft material.

    Device Description

    The Camber Spine Technologies SPIRA Open Matrix ALIF is an Interbody Fusion Device that has spiral support arches to allow for a hollow chamber to permit packing with autogenous bone to facilitate fusion. The superior and inferior surfaces of the device have a rough surface to help prevent movement of the device while fusion takes place.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the SPIRA™ Open Matrix ALIF intervertebral body fusion device, based on substantial equivalence to predicate devices, rather than a clinical study evaluating its performance with human readers and AI assistance. Therefore, most of the requested information regarding acceptance criteria for AI performance in a clinical setting (e.g., sample size for test sets, experts for ground truth, MRMC studies, standalone performance) is not applicable to this document.

    However, the document does contain information about the mechanical performance testing of the device for regulatory purposes. I can extract and present the relevant information that is available:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with numerical values and reported device performance. Instead, it states the following regarding mechanical testing against predicate devices:

    Acceptance CriteriaReported Device Performance
    Defined by predicate device performanceMet

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for the "test set" in terms of mechanical testing, nor does it detail the provenance of the data (e.g., country of origin, retrospective or prospective). Mechanical testing typically uses a certain number of identical physical devices for each test, but this information is not explicitly provided in the summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the "ground truth" here pertains to mechanical properties of the device, not a diagnosis or interpretation requiring human expert readers.

    4. Adjudication method for the test set

    Not applicable, as this refers to human expert review for establishing ground truth, not mechanical testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC study was not done. This device is an implantable medical device, and the evaluation for its 510(k) clearance focused on demonstrating substantial equivalence in design, materials, function, manufacturing, and mechanical performance to existing predicate devices, not on a comparative effectiveness study involving human readers and AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI diagnostic or assistive device. "Standalone performance" would refer to the mechanical performance of the device itself, which was indeed tested.

    7. The type of ground truth used

    For the mechanical performance testing, the "ground truth" was established by standardized laboratory testing protocols (ASTM F2077-14 and F2267-04) and comparison to the performance of existing, legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this device's regulatory submission, as it is not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set mentioned in the context of this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112801

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-01

    (217 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K080588,K091873

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Qualgenix Ayers Rock Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history of radiographic studies. These patients should have had 6 weeks of non-operative treatment. The Qualgenix Ayers Rock Cervical Cage implants are to be used with autogenous bone graft and implanted via an open, anterior approach. The Qualgenix Ayers Rock Cervical Cage is to be used with supplemental fixation.

    Device Description

    The Ayers Rock Cervical Cage consists of cervical spinal interbody fusion devices as well as instrumentation designed specifically for the implantation of these devices. The Ayers Rock Cervical Cage is manufactured from PEEK OPTIMA LT1 polymer. The Ayers Rock Cervical Cage is for single level anterior spinal use from the C2-C3 to C7-T1 disc levels.

    AI/ML Overview

    The provided text describes a medical device, the Qualgenix Ayers Rock Cervical Cage, and its clearance through the 510(k) pathway. This pathway focuses on demonstrating substantial equivalence to a predicate device rather than conducting clinical trials to establish new safety and effectiveness. Therefore, the information typically requested regarding acceptance criteria based on a study, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies (like MRMC) is not applicable or not provided in this type of submission for this particular device.

    Here's a breakdown of the available information in relation to your request:

    1. A table of acceptance criteria and the reported device performance

    The provided document does not explicitly state acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity) for device performance in a clinical study. Instead, the device's acceptance is based on demonstrating substantial equivalence to predicate devices through preclinical testing and comparison of indications, design, and function.

    Acceptance Criteria (from 510(k) framework)Reported Device Performance
    Substantially equivalent in indications for use, design, and function to predicate devices.The Ayers Rock Cervical Cage was shown to be substantially equivalent to previously cleared devices.
    Compliance with ASTM F2077 (static compression, compression-shear, static torsion, dynamic compression)Preclinical testing performed per ASTM F2077 (static compression, compression-shear, static torsion, dynamic compression) indicates substantial equivalence.
    Compliance with expulsion testingPreclinical expulsion testing indicates substantial equivalence.
    Compliance with ASTM F2267 (subsidence testing)Preclinical subsidence testing (ASTM F2267) indicates substantial equivalence.

    2. Sample size used for the test set and the data provenance

    Not applicable. The clearance is based on preclinical testing and comparison to predicate devices, not a clinical study with a "test set" of patients or data provenance in the context of clinical outcomes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth, in the context of clinical studies, refers to definitive diagnoses or outcomes. For medical devices cleared via 510(k), especially for implants, the "truth" is established through engineering and biomechanical testing demonstrating equivalence to established devices, not expert consensus on patient data.

    4. Adjudication method for the test set

    Not applicable. There was no clinical test set requiring adjudication in the way it's done for diagnostic algorithms.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an implantable cervical cage, not a diagnostic AI device that would involve human readers or AI assistance in reading medical images.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used

    The "ground truth" for the Ayers Rock Cervical Cage is established through preclinical testing results (biomechanical and mechanical testing) demonstrating that its performance characteristics (e.g., static compression, dynamic compression, subsidence, expulsion) are equivalent to those of the predicate devices. The equivalence in design, materials (PEEK OPTIMA LT1 polymer), and intended use also serves as a critical part of the "truth" in the 510(k) context.

    8. The sample size for the training set

    Not applicable. There is no concept of a "training set" for this type of medical device 510(k) submission, as it's not an AI/algorithm-based device.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.