Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102402

    Validate with FDA (Live)

    Date Cleared
    2011-03-30

    (218 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K093808, K080572, K072253, K090782

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solus Anterior Lumbar Interbody Fusion (ALIF) System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1 with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The Solus implant is intended to be used with autograft. The device is intended for use with supplemental fixation that is in addition to the integrated blades.

    Device Description

    The Solus Anterior Lumbar Interbody Fusion (ALIF) System is an intervertebral body fusion system for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1 with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Solus System consists of implants with various sizes and heights available in 7º or 12º lordosis to accommodate individual patient pathology. The implants are manufactured from polyetheretherkeytone (PEEK Optima LT)) material conforming to ASTM F2026 and tantalum radiographic markers (ASTM F560) with internal anchoring blades of titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136.

    AI/ML Overview

    The provided text describes the Solus Anterior Lumbar Interbody Fusion (ALIF) System, a medical device for spinal fusion. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices, primarily through mechanical performance testing.

    Here's an analysis of the provided information, specifically addressing the requested fields:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance (Solus ALIF System)
    Mechanical PerformanceSubstantially Equivalent to Predicate Devices
    Axial CompressionTesting performed per ASTM 2077-03.
    Dynamic CompressionTesting performed per ASTM 2077-03.
    Static Compression-ShearTesting performed per ASTM 2077-03.
    Static TorsionTesting performed per ASTM 2077-03.
    Dynamic TorsionTesting performed per ASTM 2077-03.
    SubsidenceTesting performed per ASTM 2267.
    Static Push-OutTesting performed per draft standard ASTM F-04.25.02.02.
    Cadaveric Fatigue (Flexion)Testing performed.
    Cadaveric Fatigue (Lateral Bending)Testing performed.
    Cadaveric Fatigue (Axial Rotation)Testing performed.
    Blade/Vertebral Body InterfaceFinite Element Analysis performed.
    Overall Conclusion"The test results demonstrate that the mechanical performance of the Solus Anterior Lumbar Interbody Fusion (ALIF) System is substantially equivalent to the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state numerical sample sizes for each mechanical test performed. For instance, it mentions "cadaveric fatigue testing" but doesn't specify the number of cadavers or specimens used. The provenance of the test data (e.g., country of origin, retrospective or prospective) is also not specified, though it would typically be internal testing by the manufacturer or contracted labs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and is not provided in the document. The "ground truth" for this type of device (spinal implant) is established through standardized mechanical testing according to ASTM standards, not through expert consensus on medical images or patient outcomes in the same way an AI diagnostic device would.

    4. Adjudication Method for the Test Set

    This is not applicable as the "ground truth" is established through engineering and mechanical testing, not through human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers (e.g., radiologists) interpret images with and without AI assistance. The Solus ALIF System is a physical implant, and its performance is assessed through mechanical testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is not applicable. The Solus ALIF System is a physical medical device, not an algorithm. Its "standalone performance" refers to its mechanical integrity and function as tested in a laboratory setting.

    7. The Type of Ground Truth Used

    The "ground truth" for the Solus ALIF System's performance is established through standardized mechanical testing. This includes:

    • Physical testing according to ASTM standards (e.g., ASTM 2077-03 for axial compression, dynamic compression; ASTM 2267 for subsidence; draft standard ASTM F-04.25.02.02 for static push-out).
    • Cadaveric fatigue testing.
    • Finite Element Analysis (FEA) for the blade/vertebral body interface.

    The performance is then compared to predicate devices to demonstrate substantial equivalence, implying that the established "ground truth" for the predicate devices (which have been deemed safe and effective) serves as the benchmark.

    8. The Sample Size for the Training Set

    This is not applicable. The Solus ALIF System is a physical medical device, not an AI or machine learning model that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1