510(k) Data Aggregation

    K Number
    K102726

    Validate with FDA (Live)

    Date Cleared
    2011-04-04

    (195 days)

    Product Code
    Regulation Number
    870.4220
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).

    The CARDIOHELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport.

    The CARDIOHELP System in configuration with the QUADROX-iR is intended to be used in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery (for periods for up to six hours).

    Device Description

    The CARDIOHELP System is a compact perfusion system consisting of the following components:

    • the CARDIOHELP-i drives suitable disposables using an integrated pump, controls and monitors the extracorporeal circulation and can communicate with other devices
    • the CARDIOHELP Emergency Drive is used in emergencies to manually drive the disposable if the CARDIOHELP-i fails
    • different accessories:
      • o flow/bubble sensor
      • o level sensor including level sensor pad
      • o venous probe for blood gas monitoring and measurement of venous temperature
      • external sensors for temperature and pressure
      • o different holders
      • o connection cables
    • various disposables, that can be driven by CARDIOHELP-i
      • o HLS/ HIT Set Advanced are tubing sets, that contain the HLS Module Advanced oxygenator with integrated centrifugal pump, which fits into the CARDIOHELP-i. These sets are available in different configurations and will be compiled as customized tubing sets
      • o QUADROX-iR is an oxygenator with integrated centrifugal pump and will be provided with or without integrated arterial filter
    AI/ML Overview

    The provided text details a 510(k) submission for the MAQUET CARDIOHELP System. This submission focuses on establishing substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria in a clinical study with a set number of cases and experts. Therefore, much of the requested information about clinical study specifics (sample sizes, ground truth establishment, expert qualifications, etc.) is not present in this document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of explicit, quantifiable acceptance criteria with corresponding performance metrics like sensitivity, specificity, or accuracy for a specific disease or condition. Instead, the acceptance criteria are generally implied as demonstrating "substantial equivalence" to predicate devices through various technical and performance tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety: Electrical, Mechanical Safety, Biocompatibility, Sterility"Performed within its specifications and within the acceptable limits of the applied performance standards."
    Effectiveness: Performance, Measurement accuracy, Functionality (drive disposables, control/monitor ECC, communicate with other devices)"Performed within its specifications and within the acceptable limits of the applied performance standards." Substantially equivalent to predicate devices in technical characteristics, performance specifications, and materials.
    Usability: Handling, User Interface, Alarms"Tests for Usability" and "Verification of Alarms" were conducted. Implied as acceptable.
    Integrity/Tightness/Shelf Life (Disposables):"Tests for Integrity," "Tightness," and "Shelf life" were conducted. Implied as acceptable.
    Inter-hospital Transportation:"Tests for Inter-hospital transportation" were conducted. Implied as acceptable.
    Software Validation:"Software Validation" was conducted. Implied as acceptable.
    Electromagnetic Compatibility:"Electromagnetic Compatibility" was conducted. Implied as acceptable.
    Packaging:"Packaging" was conducted. Implied as acceptable.

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe a clinical test set with a specific sample size for evaluating a diagnostic AI algorithm. The testing described focuses on the device's engineering performance and safety characteristics. The data provenance is primarily from non-clinical testing conducted by the manufacturer, Maquet Cardiopulmonary AG, in Germany.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. The document describes non-clinical engineering and performance testing, not a study involving expert-established ground truth for a diagnostic task.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not performed as described in this document. This submission is for a medical device (a cardiopulmonary support system), not an AI algorithm for diagnostic image interpretation. Therefore, there's no mention of human readers improving with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This document is for a medical device system, not a standalone AI algorithm. The CARDIOHELP System is a human-operated device, albeit with advanced control and monitoring features.

    7. Type of Ground Truth Used

    The "ground truth" in this context refers to established engineering specifications, industry standards, and the performance characteristics of predicate devices. For example, for "measurement accuracy," the ground truth would be precise reference measurements obtained from calibrated instruments. For "biocompatibility," the ground truth is defined by relevant biocompatibility standards.

    8. Sample Size for the Training Set

    Not applicable. The CARDIOHELP System is a physical medical device, not a machine learning model that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090515

    Validate with FDA (Live)

    Date Cleared
    2009-12-18

    (295 days)

    Product Code
    Regulation Number
    870.4360
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K080470,K082117

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RotaFlow Centrifugal Pump is indicated as a component of the extracorporeal circuit for pumping liquid matter e.g. blood and can be used in conjunction with the RotaFlow Console. The utilization period of this device is restricted to six hours. The device is not designed or intended for use except as indicated.

    Device Description

    The RotaFlow Centrifugal Pump is a non-occlusive pump. The pump has a spinning rotor with flow channels which imparts rotary motion to incoming liquid, directing it through a spiral housing to the outflow port.

    AI/ML Overview

    This submission (K090515) is for a medical device called the MAQUET RotaFlow Centrifugal Pump with Softline Coating. It's important to note that this 510(k) summary does not contain acceptance criteria or a study proving the device meets those criteria in the way you've outlined for AI/software-driven devices.

    This is a premarket notification for a Class III physical medical device (a centrifugal pump for cardiopulmonary bypass). The submission focuses on demonstrating substantial equivalence to existing legally marketed predicate devices, not on showing performance against specific, quantifiable acceptance criteria typically associated with software or AI performance.

    Therefore, many of your specific questions are not applicable to the information provided in this 510(k) summary. I will address the relevant sections and explain why others cannot be answered from the provided text.


    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the 510(k) summary for K090515. For physical devices like this, the "performance" demonstrated for substantial equivalence typically involves engineering tests (e.g., flow rates, pressure handling, durability, biocompatibility) rather than quantifiable metrics like sensitivity or specificity seen in diagnostic AI. The summary states that "Performance" was evaluated, but it does not present specific criteria or results in a table.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. As this is a physical device, "test sets" in the context of data points for algorithm evaluation are not relevant here. The evaluation would involve physical prototypes or manufactured units of the pump.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided and is not applicable to a physical medical device submission like this. Ground truth, in the context of expert consensus, is typically for diagnostic interpretation, not for evaluating the mechanical function of a pump.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided and is not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided and is not applicable. This type of study relates to the performance of diagnostic AI with human readers, not a physical pump.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided and is not applicable. There is no algorithm discussed for diagnostic or interpretative purposes in this submission.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided and is not applicable in the way you've described. For a physical pump, "ground truth" might refer to engineering specifications, validated test methods for flow, pressure, material integrity, and biocompatibility standards. The submission highlights that "Integrity", "Performance", "Biocompatibility", and "Sterility" were evaluated, implying various testing methodologies were used, but details are not given.

    8. The sample size for the training set

    This information is not provided and is not applicable. There is no training set for an AI/ML algorithm in this submission.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable.


    Summary of Relevant Information from the K090515 Submission:

    • Device Name: RotaFlow Centrifugal Pump with Softline Coating
    • Submission Type: 510(k) Premarket Notification
    • Purpose of Study/Evaluation: To demonstrate substantial equivalence to legally marketed predicate devices.
    • Predicate Devices:
      • RotaFlow Centrifugal Pump with BIOLINE Coating (K080470) - for the pump design and function.
      • QUADROX-i Adult microporous membrane Oxygenator with and without integrated Arterial Filter with Softline Coating (K082117) - for the Softline Coating material.
    • Key Differences from Predicate: The only stated difference from the primary predicate is the coating (Softline instead of BIOLINE). The Softline coating itself is identical to a coating already approved in another predicate device.
    • Evaluated Areas (for demonstrating substantial equivalence):
      • Integrity
      • Performance
      • Biocompatibility
      • Sterility
    • Conclusion: The data demonstrated that the RotaFlow Centrifugal Pump with Softline Coating is substantially equivalent to the named predicate devices.

    In essence, this 510(k) is for a manufacturing change (a new coating) on an already cleared device, leveraging the prior approval of both the pump design and the specific coating material on other devices. It does not involve AI or software development that would require the kind of performance metrics and study designs you've detailed.