Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102513

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-05

    (126 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K091947, K092186, K080131, K091947

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S9 VPAP ST is indicated for the treatment of Obstructive Sleep Apnea.(OSA).in.patients.weighing. . more than 66 lb (30 kg). The S9 VPAP ST is intended for use in the hospital and home.

    Device Description

    S9 VPAP ST System (S9 VPAP ST with H5i) is similar to the predicate device(s), using a blower based positive pressure system with an integrated heated humidifier and heater controller. The device platform, both hardware and firmware come from the S8 Aspen with H4i Plus (K091947) and contains a micro-processor controlled blower system that generates controlled positive airway pressure between 3-25 cmH2O as required to maintain an "air splint" for effective treatment of OSA. The system comprises the flow generator, patient tubing, mask (patient interface) and humidifier. The S9 VPAP ST contains treatment modes that come from earlier versions of ResMed Bilevel devices, these modes are known as Spontaneous (S Mode), Timed (T Mode), Spontaneous/Timed (S/T Mode) and VAuto Mode. Therapy modes come from the VPAP Tx system (K092186), VPAP ST (K080131) for S, S/T, T and VAuto and CPAP Mode comes from the S8 Aspen (K091947).

    AI/ML Overview

    This document is a 510(k) summary for the ResMed S9 VPAP ST, a non-continuous ventilator intended for the treatment of Obstructive Sleep Apnea (OSA). It outlines the device's substantial equivalence to predicate devices and the non-clinical testing performed.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state quantitative acceptance criteria in a table format with specific performance metrics and their corresponding values for each test. Instead, it broadly states that "All bench tests confirmed the predetermined acceptance criteria." The performance is generally reported as meeting these criteria through various tests against predicate devices.

    Acceptance Criteria CategoryReported Device Performance
    Bench Tests (General)Confirmed predetermined acceptance criteria.
    Specific Bench TestsPressure, Flattening, Snore, Hypopnea, and Apnea tests passed against predicate devices using common protocols.
    Compliance with StandardsPassed all test protocols for:
        IEC 60601-1-2:2007 (EMC)Met requirements.
        IEC 60601-1:1988 Ed 2 (Basic Safety)Met requirements.
        ISO 17510-1:2007 (Sleep Apnea Therapy)Met requirements.
        ISO 8185:2007 (Respiratory Humidifiers)Met requirements.
    Safety and Effectiveness"Analysis of comparison of design, function and features of the predicate devices... together with the results of testing demonstrates that the S9 VPAP ST to be substantially equivalent to the predicate devices in terms of meeting performance criteria and function as intended."

    2. Sample Size Used for the Test Set and Data Provenance

    The document indicates that clinical data for the S9 VPAP ST is not required. Therefore, there is no human patient test set as the device's safety and effectiveness were established through comparison to predicate devices and bench testing.

    • Sample Size: Not applicable for a human test set. Bench tests were performed on the device itself.
    • Data Provenance: Not applicable for patient data. The tests were non-clinical (bench testing) and likely performed internally by ResMed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    As no clinical data was required and testing was done non-clinically, there was no expert adjudication process to establish ground truth for a patient test set. The "ground truth" for the non-clinical tests would have been the specifications and performance characteristics of the predicate devices and relevant engineering standards.

    4. Adjudication Method for the Test Set

    Not applicable, as there was no patient test set requiring expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical data for the S9 VPAP ST is not required as the predicate devices have been subjected to clinical trial requirements or validated patient simulation models have been used during the bench testing phases."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The S9 VPAP ST is a physical medical device (a ventilator), not a software algorithm providing diagnostic or treatment recommendations. Therefore, the concept of "standalone performance" in the context of an algorithm's without-human-in-the-loop performance is not directly applicable. The device itself performs its intended function (delivering positive airway pressure). Its performance was evaluated through non-clinical bench testing.

    7. The Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" was based on:

    • The performance specifications and established safety/effectiveness of the predicate devices.
    • Engineering parameters and requirements defined by international standards (IEC, ISO) and internal design requirements.
    • Validated patient simulation models used during bench testing.

    8. The Sample Size for the Training Set

    The device is a hardware-based medical device with integrated software, not a machine learning algorithm that undergoes a "training phase" with a distinct "training set" of data in the typical AI/ML sense. Therefore, no specific sample size for a training set is applicable or mentioned. Its design and functionality are based on established engineering principles and prior device iterations, not statistical learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of machine learning, this question is not applicable. The device's functionality and "knowledge" are engineered into its design, operating principles, and firmware, validated against industry standards and predicate device performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1