Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142667

    Validate with FDA (Live)

    Date Cleared
    2015-01-07

    (110 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K073628, K033243, K093027, K072071, K111798

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stern AC Dental Implant System can be used in dental implant applications for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The Stern AC Dental Implant System is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading.

    The Stern AC Dental Implants are compatible with the Straumann Rn Synocta, Straumann SynOcta Meso Abutments, Straumann RC Temporary Abutments, Straumann RC Cementable abutments and Stern IC Solid Abutments.

    The Stern AC Dental Implant System is only intended for use with straight abutments.

    Device Description

    The proposed implant is a self-tapping, double thread screw implant, manufactured from pure grade 4 titanium. The implant is acid etched except for the neck, which is machined to a smooth finish. Stern AC Implants are manufactured with a flared neck, used for one-stage, transgingival implantation. The neck and body include an internal taper and octagon. The implants are manufactured with three body diameters, 3.3 mm, 4.0 mm, and 5.0 mm, all with the same prosthetic head.

    The proposed implants are prosthetically compatible with the Straumann ITI implant system and the Stern IC Dental Implant System previously approved for market:

    Straumann Rn Synocta - K073628 Straumann SynOcta Meso Abutments - K033243 Straumann RC Temporary Abutments - K093027 Straumann RC Cementable abutments - K072071 Stern IC Solid Abutments - K111798

    They are packaged sterile in a double blister, which contains implant with an implant mount screw secured in a titanium tube. The Mount Body is screwed into the using the mount screw, which keeps it in place. This configuration is known as "Direct Delivery", which allows the surgeon to transport the implant from the package to the prepared implant site and save chairside time. The proposed Stern AC Dental Implant will be available in a range of lengths and diameters.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Stern AC Dental Implant System. It focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study proving the device meets specific acceptance criteria with quantifiable performance metrics. Therefore, much of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth cannot be extracted from this document.

    However, based on the information provided, here's what can be inferred:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state numerical acceptance criteria or quantifiable performance metrics for the Stern AC Dental Implant System. Instead, the "performance" demonstrated is that the device is "substantially equivalent" to predicate devices and "performs appropriately for the proposed indications for use." The performance testing mentioned is primarily non-clinical, focusing on material and design characteristics.

    Property/CharacteristicAcceptance CriteriaReported Device Performance
    Intended UseSimilar to predicate devicesAligns with predicate devices (Sterngold Acid Etch Dental Implant System (K023580) and Stern IC Dental Implant System (K111798)).
    MaterialSame as predicate devicesCP Titanium grade 4, same as predicate devices.
    DesignSame or similar functions and characteristics as predicate devicesSelf-tapping, double-thread screw implant, flared neck, internal taper and octagon. Prosthetically compatible with Straumann ITI and Stern IC systems.
    Manufacturing ProcessSame as predicate devicesSame manufacturing process as other Sterngold Dental Implants previously approved.
    Surface TreatmentSame as predicate devicesBlasted with aluminum oxide particles and acid etched (identical to Sterngold Acid Etched Implant and widely used surface like Straumann "SLA").
    SterilizationValidatedSterilization validation was conducted.
    Blasting ResidueAll residue removedSEM photographs enclosed showing all residue of blasting material was removed from the surface.
    BiocompatibilityNot explicitly stated, but implied by material similarityCP Titanium grade 4, a commonly used and biocompatible material for dental implants.
    Mechanical Strength/FatigueNot explicitly stated, but implied by "appropriate performance" for indicationsNot detailed, but the document states "performs appropriately for the proposed indications for use" following the "Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not applicable in the context of this document. The submission is based on demonstrating substantial equivalence through comparison of design, materials, and non-clinical performance to existing cleared devices, not on a clinical "test set" of patients or a defined dataset with ground truth for algorithm evaluation.
    • Data Provenance: The data provided primarily stems from the manufacturer's internal design specifications, material characterization, manufacturing process descriptions, and comparisons to predicate devices. It is entirely retrospective in the sense that it relies on existing knowledge and data for predicate devices and internal validation of the manufacturing processes and material properties of the new device. There is no mention of patient data, clinical trials, or external datasets.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Not applicable. There is no "test set" in the sense of a dataset requiring expert ground truth establishment for algorithm evaluation. The "ground truth" for this submission is FDA's existing clearances for predicate devices and established standards for dental implant materials and design.

    4. Adjudication Method for the Test Set:

    • Not applicable. There is no "test set" requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This document describes a traditional medical device (dental implant) submission, not an AI/software as a medical device (SaMD) submission. Therefore, no MRMC study or AI-related effectiveness analysis was conducted or reported.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/algorithm-based device.

    7. The Type of Ground Truth Used:

    • The "ground truth" in this context is established by:
      • Regulatory Precedent: The FDA's previous clearance of the predicate devices (Sterngold Acid Etch Dental Implant System (K023580) and Stern IC Dental Implant System (K111798)).
      • Industry Standards: Adherence to guidance documents like "Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments."
      • Material Science Principles: The properties of CP Titanium grade 4 and the known performance of specific surface treatments (blasted with aluminum oxide particles and acid etched).
      • Engineering Design Principles: The design features like self-tapping threads, flared neck, and internal connections are evaluated against known engineering principles for dental implants.
      • Laboratory Testing: SEM photographs for blasting residue removal and sterilization validation results.

    8. The Sample Size for the Training Set:

    • Not applicable. As this is not an AI/machine learning device, there is no "training set."

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. There is no "training set" or ground truth for such a set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111798

    Validate with FDA (Live)

    Date Cleared
    2012-06-01

    (340 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K073628, K033243, K093027, K072071

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stern IC Dental implant System is intended for long term surgical implantation in the bone of the patient's upper or lower arch to provide immediate load or delayed load of prosthetic systems, such as artificial teeth, in order to restore the patient's chewing function. The Stern IC Dental Implant System is also indicated for immediate loading with good primary stability and appropriate occlusal loading. The Stern IC Dental Implants are compatible with the Straumann Rn Synocta , Straumann SynOcta Meso Abutments, Straumann RC Temporary Abutments, and the Straumann RC Cementable abutments. The Stern IC Dental Implant System is only intended for use with straight abutments.

    Device Description

    The Stern IC Dental device is a root-form dental implant with a screw-type body, acid etched except for the neck of the implant, which is machined to a smooth finish. It is manufactured from pure grade 4 titanium with a flared neck, used for one-stage, transgingival or subgingival implantation. The neck and body include an internal taper and octagon. The implants are composed of grade 4 commercially pure titanium and are available in a range of lengths and diameters. This submission also includes healing abutments, solid abutments, and cover screws which are used as accessories to the dental implants. The implant components are manufactured from pure, implant-grade titanium alloy and are equivalent in terms of materials, performance specifications, manufacturing, manufacturing equipment, packaging materials and design as other Sterngold abutments and cover screws previously approved via K892124 and K924219. The Stern IC Dental Implants and abutments are compatible with the Straumann Rn Synocta (K073628), Straumann SynOcta Meso Abutments (K033243), Straumann RC Temporary Abutments, (K093027) and the Straumann RC Cementable abutments (K072071). The Stern IC Dental Implant System is not intended for correction of implant fixtures placed at an angle. It is only intended for use with straight abutments. The proposed Stern IC Dental Implant will be available in a range of lengths and diameters.

    AI/ML Overview

    "The provided text does not contain information about studies related to AI/ML device performance or acceptance criteria in the typical sense for such devices. The document is a 510(k) summary for a dental implant system (Stern IC Dental Implant System).

    Therefore, I cannot extract the following information:

    1. A table of acceptance criteria and the reported device performance: This document
      focuses on substantial equivalence to predicate devices, not performance criteria
      against a defined benchmark.
    2. Sample size used for the test set and the data provenance: Not applicable as
      there is no discussion of a test set for an AI/ML algorithm.
    3. Number of experts used to establish the ground truth for the test set and the
      qualifications of those experts:
      Not applicable.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if
      so, what was the effect size of how much human readers improve with AI vs without
      AI assistance:
      Not applicable, as this is not an AI/ML device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was
      done:
      Not applicable.
    7. The type of ground truth used (expert consensus, pathology, outcomes data,
      etc.):
      Not applicable.
    8. The sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    Instead, the document details the substantial equivalence of the Stern IC Dental Implant
    System to predicate devices based on material, design, surface treatment, and intended
    use. The study described is a non-clinical testing to demonstrate appropriate
    performance for the proposed indications for use, following "Guidance for Industry and
    FDA Staff - Class 11 Special Controls Guidance Document: Root-form Endosseous Dental
    Implants and Endosseous Dental Abutments."

    Specifically, the non-clinical testing involved:

    • SEM photographs: To show removal of all residue of the blasting material from the surface.
    • Surface chemical analysis: To confirm no contaminants are left on the surface (conducted on the final device).

    It states that fatigue testing was not required because design and technological features are substantially equivalent to predicate devices. The conclusion is based on an analysis of intended use, material, design, and performance being substantially equivalent to the predicate devices (Sterngold Acid Etch Dental Implant System (K023580) and Straumann ITI Dental Implant System (K012757))."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1