Search Results
Found 1 results
510(k) Data Aggregation
(464 days)
AlloMap Heart Molecular Expression Testing is an In Vitro Diagnostic Multivariate Index assay (IVDMIA) test service, performed in a single laboratory, assessing the gene expression profile of RNA isolated from peripheral blood mononuclear cells (PBMC). AlloMap Heart Testing is intended to aid in the identification of heart transplant recipients with stable allograft function who have a low probability of moderate/severe acute cellular rejection (ACR) at the of testing in conjunction with standard clinical assessment.
Indicated for use in heart transplant recipients:
· 15 years of age or older
· At least 2 months (>55 days) post-transplant
AlloMap Heart Molecular Expression Testing is an In Vitro Diagnostic Multivariate Index Assay (IVDMIA) test service performed in a single laboratory, assessing the gene expression profile of RNA isolated from peripheral blood mononuclear cells (PBMC). AlloMap Heart Testing is a non-invasive blood test that uses genomic technologies to identify the absence of cardiac rejection. When used in conjunction with standard clinical assessments, AlloMap Heart Testing may help identify patients with stable allograft function who have a low probability of moderate to severe acute cellular rejection (ACR) at the time of Testing.
AlloMap is a panel of 20 gene assays, 11 informative and 9 used for normalization and quality control, which produces gene expression data used to calculate a reported AlloMap test score - an integer ranging from 0 to 40. The Score is an algorithmic composite of the expression of 11 differentially weighted informative genes that have a role in rejection and reflect host immune response, representing multiple diverse molecular pathways. The clinician uses the AlloMap Score and other standard clinical assessments to evaluate the patient's probability of rejection and the need for additional diagnostic evaluations. Compared with patients in the same post-transplant period, the lower the Score, the lower the probability of acute cellular rejection at the time of Testing. AlloMap Heart Testing is only performed at the CLIA-certified and CAP-accredited clinical laboratory at CareDx in Brisbane, California.
The provided text describes the acceptance criteria and the study results for the AlloMap Heart Molecular Expression Testing, specifically focusing on the performance of the Roche LightCycler 480 II (LC480) instrument compared to the predicate ABI 7900HT instrument.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
| Test Category | Acceptance Criteria | Reported Device Performance (LC480 vs ABI 7900HT) |
|---|---|---|
| Software V&V | Per Verification Plan against requirement specification. Meet product requirements and user needs/Intended Use. | Modified Software Analyzer v1.2 and its specifications meet product requirements and conform to user needs and Intended Use. |
| Precision – Between Instrument Variability | Variance ratios of LC480 over 7900HT less than (<) 1 for Low, Medium, and High AlloMap score sample pools. | Variance ratios of LC480 over 7900HT are < 1 for Low (0.459), Medium (0.316), and High (0.132) AlloMap score sample pools. Passed the pre-specified acceptance criteria. Between-instrument variability for LC480 is acceptable and comparable to 7900HT. |
| Precision – Reproducibility | Total variance ratio (σ_LC480^2 / σ_7900HT^2) less than (<) 1, or not statistically significant. | Total variance ratio is < 1 for Low (0.459, p<0.05), Medium (0.316, p<0.05), and High (0.132, p<0.05) AlloMap score sample pools, with F-test p-value < 0.05. LC480 met acceptance criteria and demonstrated acceptable precision, with total variances smaller than 7900HT. |
| Linearity | R2 value for linear regression of normalized factors vs. input log10RNA concentration should be acceptable. The variability of raw scores at each RNA input level should meet established acceptance criteria for the predicate device. | R2 value of 0.989 was obtained. LC480 met acceptance criteria and is comparable to the predicate device, 7900HT. Linearity of the AlloMap Heart test across various RNA inputs is maintained, especially around the current RNA input of 300ng. |
| Method Comparison – Accuracy (Passing-Bablok) | 95% CI of the slope includes 1.0 and 95% CI of the intercept includes 0.0. | For both LC480_7296 vs 7900HT and LC480_7371 vs 7900HT: - Intercept: 0 (-0.53, 0.53) and 0 (-0.34, 0.34) respectively (both include 0). - Slope: 1 (0.99, 1.01) and 1 (0.99, 1.01) respectively (both include 1.0). All prespecified passing criteria were met at a two-sided significance level of 0.05. |
| Method Comparison – Accuracy (ATD Zones) | For each LC480, over 95% of samples across the entire clinical range fall within the ATD zones, with a lower limit of 95% one-sided CI > 90%. | For LC480_7296, 99.4% (lower 95% CI 96.6%) of all samples were within ATD zones. For LC480_7371, 98.2% (lower 95% CI 94.7%) of all samples were within ATD zones. The measurements by LC480 are within the analytical variation of the 7900HT and are considered equivalent. |
| Method Comparison – Clinical Equivalence (Projected NPV/PPV) | Absolute differences between the two system platforms (LC480 vs 7900HT) within 1% or 95% CI including 0 for NPV, and within 2% or 95% CI including 0 for PPV, for both 2-6 months and >6 months post-transplant periods. | The absolute differences between the two system platforms were within 1% or 95% CI including 0 for NPV, and within 2% or 95% CI including 0 for PPV, for both 2-6 months (55-182 days) and >6 months (>182 days) post-transplant periods. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Between Instrument Variability: Not explicitly stated, but implies multiple samples across low, medium, and high score ranges.
- Reproducibility:
- Low Sample Pool: 88 samples
- Medium Sample Pool: 90 samples
- High Sample Pool: 90 samples
- Total: 268 samples
- Linearity: Six different amounts of the same patient RNA sample.
- Method Comparison - Accuracy (Passing-Bablok and ATD Zones): 163 patient RNA samples.
- Method Comparison - Clinical Equivalence (Projected NPV/PPV): Based on the "300 samples in the original clinical validation study from the predicate device, K073482." No specific country of origin is mentioned, but the testing was done at CareDx in Brisbane, CA, USA, implying the samples were processed there. The samples were "previously tested heart transplant patient samples with clinical relevance to AlloMap score... after deidentification," suggesting a retrospective origin.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not describe the use of experts to establish a "ground truth" in the traditional sense of diagnostic interpretation. This is an analytical performance study comparing two instruements, where the ground truth is essentially the quantitative measurement produced by the predicate device (ABI 7900HT) and the clinical relevance (NPV/PPV) established in its original clinical validation. The "ground truth" for the patient samples' AlloMap scores were their scores obtained from the 7900HT.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This was an analytical comparison study of test instrumentation, not a diagnostic interpretation study requiring expert adjudication of cases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device or a MRMC study. It is an analytical validation of an in vitro diagnostic (IVD) assay instrument change.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, in a sense. The study primarily evaluated the standalone performance of the new instrument (LC480) and its adapted software (SAX v1.2) against the predicate instrument (ABI 7900HT) to ensure equivalent analytical and clinical equivalency of the AlloMap score calculation. The AlloMap test itself is an algorithm-based test ("Multivariate Index Assay") performed in a lab environment. The "standalone" evaluation here refers to the instrument's ability to produce equivalent scores.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The primary ground truth for the analytical performance studies was the AlloMap scores generated by the predicate ABI 7900HT instrument. For the clinical equivalence projection (NPV/PPV), it refers back to the clinical outcomes data that supported the original clearance of the predicate device (K073482).
8. The sample size for the training set
The document does not explicitly describe a "training set" for the current submission, as it focuses on the analytical and clinical equivalency verification of an instrument change. The "Software Analyzer, SAX v1.2" was modified to support the LC480, and it applied "two gene-level correction factors" to map LC480 response to 7900HT response. The data used to derive these correction factors (which could be considered analogous to a training step for the software adaptation) is not specified in terms of sample size or methodology. However, the original AlloMap algorithm itself was likely developed and "trained" on a much larger dataset during its initial development and clearance (DEN080007 / K073482), but that information is not part of this specific submission's details.
9. How the ground truth for the training set was established
As mentioned above, the current submission focuses on instrument equivalency. For the underlying AlloMap algorithm, the ground truth for its initial development would have involved patient outcomes related to acute cellular rejection (ACR), likely correlated with biopsy results (pathology) which is the gold standard for ACR diagnosis, and potentially other clinical assessments in heart transplant recipients. This background is implied by the "Indications for Use" and the nature of the "Cardiac Allograft Gene Expression Profiling Test System." However, the specifics for the original training set's ground truth establishment are not provided in this document.
Ask a specific question about this device
Page 1 of 1