Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111254

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-22

    (110 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K073041, K102003

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Relieva Solo Elite Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

    For children aged 17 and under, the Relieva Solo Elite Sinus Balloon Catheter is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.

    Device Description

    The Relieva Solo Elite Sinus Balloon Catheter is a flexible catheter that is intended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon and permits the passage of a sinus guidewire or sinus illumination system to facilitate access to the target sinus ostia. A hypotube is incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter. The subject device also has the capability to irrigate the sinus through the distal tip and three side holes on the distal end.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (Relieva Solo Elite Sinus Balloon Catheter), which typically demonstrates substantial equivalence to predicate devices rather than proving clinical efficacy or diagnostic accuracy with extensive human studies. Therefore, many of the requested elements for AI/device performance studies (like MRMC, expert consensus, training set details) are not applicable to this type of submission.

    However, I can extract the relevant performance data and acceptance criteria based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that the Relieva Solo Elite Sinus Balloon Catheter met various performance acceptance criteria. While the specific numerical acceptance criteria values are not given, the categories of criteria and the confirmation that they were met are provided.

    Acceptance Criteria CategoryReported Device Performance
    Dimensional SpecificationsMet all criteria
    Balloon Burst PressureMet all criteria
    Joint Separation ForceMet all criteria
    Deflation TimeMet all criteria
    Irrigation Flow RateMet all criteria
    Balloon Cycle FatigueMet all criteria
    Sterility (SAL 10^-6)Met criteria (validated per AAMI/ANSI/ISO 11135-1: 2007)
    Ethylene Oxide ResidualsMet ISO 10993-7:2008 requirements
    Shelf LifeEstablished per ASTM F1980-07, ASTM F88/F88M-09, ISTA 2A-11, and ASTM F2096-04 requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify sample sizes for each individual test or the data provenance (e.g., country of origin, retrospective/prospective). These are typically detailed in the test reports themselves, which are referenced but not included in this summary document. The tests conducted are primarily engineering performance tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device submission focuses on engineering performance and substantial equivalence, not diagnostic accuracy requiring expert panel review for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. This device submission focuses on engineering performance tests, not clinical performance requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a manual surgical instrument, not an AI-powered diagnostic device, and no MRMC study is mentioned or relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests would be the established engineering standards and material specifications. For example:

    • Dimensional Specifications: Engineering drawings and measurement standards.
    • Balloon Burst Pressure: Material failure limits and design specifications.
    • Sterility: Microbiological validation guidelines (e.g., AAMI/ANSI/ISO 11135-1).
    • Ethylene Oxide Residuals: Toxicology standards (e.g., ISO 10993-7).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that uses a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1