510(k) Data Aggregation

    K Number
    K191201

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-15

    (193 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K083067, K072959, K140934

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EM SA Implant System (DENTURE) is intended to be place in the bone of the upper or lower jaw arches providing support to dental prosthetic devices, specifically for denture stabilization to restore a patient's chewing function. It is intended for single use only.

    The EM SA Implant (NARROW RIDGE) is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.

    Device Description

    The EM SA Implant System is a dental implant made of titanium alloy (Ti-6AI-4V) and is supplied sterile. The surface treatment of EM SA Implant System is SA (Sandblasted and Acid etched) and is intended to be surgically placed in the bone of the upper or lower jaw arches.

    The EM SA Implant System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the OSSTEM Implant Co., Ltd.'s MS SA Implant System (K122171).

    AI/ML Overview

    The provided text is a 510(k) Pre-market Notification for the Hiossen EM SA Implant System. This document asserts substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a standalone study with performance metrics. Therefore, it does not contain the detailed information requested in the prompt regarding acceptance criteria, study design, sample sizes for test and training sets, data provenance, expert ground truth, adjudication methods, or MRMC comparative effectiveness studies.

    Here's an explanation of why the requested information is not available in the provided text:

    • 510(k) Pre-market Notification: This regulatory pathway demonstrates that a new device is "substantially equivalent" to a legally marketed predicate device. This typically involves comparing design, materials, indications for use, and performance characteristics to existing devices, rather than conducting new, comprehensive clinical performance studies to prove superiority or meet novel performance criteria as might be seen for PMA submissions or novel AI devices.
    • Medical Device Type: The EM SA Implant System is an endosseous dental implant, which is a physical device made of titanium alloy. The "device" in this context is a physical implant, not an AI/ML algorithm or software. Therefore, the questions related to AI/ML specific aspects like "multi reader multi case (MRMC) comparative effectiveness study," "standalone (i.e. algorithm only without human-in-the loop performance)," or "sample size for the training set" are not applicable.
    • Focus on Equivalence: The document's primary goal is to establish equivalence. It emphasizes that the new device uses the same materials, manufacturing processes, surface treatments, sterilization methods, and has similar designs to predicate devices. This implies that if the predicate devices meet safety and effectiveness standards, the new device, being substantially equivalent, should also.

    Based on the provided text, here is what can be extracted or deduced regarding the evaluation process for this specific medical device submission, recognizing that it's not an AI/ML device and thus many questions are not directly applicable:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or precise performance metrics in the format typically seen for novel device performance studies. Instead, it aims to demonstrate substantial equivalence to predicate devices, implying that its performance is comparable and therefore acceptable.

    A summary of the comparison to predicate devices, which serves as the basis for "acceptance" (of equivalence):

    FeatureEM SA Implant System (Proposed Device)MS SA Implant System (Primary Predicate, K122171)Hiossen Implant System (Reference Device, K140934)
    Intended UseDenture: Support dental prosthetic devices for denture stabilization to restore chewing function. Single use only. Narrow Ridge: Treat missing mandibular central/lateral incisors to support prosthetic devices (artificial teeth) to restore chewing function in partially edentulous patients. Single use, delayed loading.Denture: Support prosthetic devices to restore chewing function, including denture stabilization. Single use only. Narrow Ridge: Treat missing mandibular central/lateral incisors to support prosthetic devices to restore chewing function in partially edentulous patients. Single use, delayed loading.General: Indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations (cemented retained, screw retained, or overdenture restorations), and final or temporary abutment support for fixed bridgework. Intended for delayed loading. Ultra wide Fixture System is intended for use in the molar region.
    MaterialTitanium alloy Ti-6Al-4V (ASTM F 136)Titanium alloy Ti-6Al-4V (ASTM F 136)Pure Titanium Grade 4 (ASTM F67)
    SurfaceSA (Sandblasted and Acid etched)SA (Sandblasted and Acid etched)SA (Sandblasted and Acid etched - for comparison), RBM (Resorbable Blast Media - also mentioned for MS System)
    SterilizationRadiation SterileRadiation SterileRadiation Sterile
    PackagingSecured in plastic ampule, housed in Tyvek-lidded blister tray, placed in a tamper-evident outer package.Secured in plastic ampule, housed in Tyvek-lidded blister tray, placed in an outer box.Secured in plastic ampule, housed in Tyvek-lidded blister tray, placed in a tamper-evident outer package.
    Design/StructureThreaded Body Design, One Body Implant, specified gingival height neck, Ball-shaped head (for Denture) / Integrated abutment head (for Narrow Ridge). The document states the design is "very similar" to the MS SA.Threaded Body Design, One Body Implant, specified gingival height neck, Ball-shaped head (for Denture) / Integrated abutment head (for Narrow Ridge).Internal Hex-connected, Submerged Fixture, Tapered body shape & Straight body shape (This is a different design philosophy, used as a reference for aspects like sterilization and packaging, and general implant technology).
    Diameters (mm)Denture: Ø2.0, Ø2.5, Ø3.0 Narrow: Ø2.5, Ø3.0Denture: Ø2.5, Ø3.0 Narrow: Ø2.5, Ø2.9Ø3.5-6.8
    Lengths (mm)8.5, 10.0, 11.5, 13.0, 15.0Denture: 10.0, 13.0, 15.0 Narrow: 8.5, 10.0, 13.0, 15.06.2-18.2

    The document concludes that the proposed device is "substantially equivalent with respect to the intended use, structure, material, surface, sterilization and design with the legally marketed primary predicate device" and that "The slight difference in thread diameter and length is a simple size addition. The proposed device does not pose any new or increased risks as compared to the legally marketed primary predicate and reference devices."

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not applicable. No clinical performance testing with a "test set" was submitted for this 510(k) pathway as per the document: "No clinical performance report(s) is being submitted."
    • Data Provenance: Not applicable. The submission relies on non-clinical data (biocompatibility, sterilization, shelf life, surface treatment, fatigue testing) which were either reasoned not necessary to re-test due to equivalence to predicates using identical processes and materials, or demonstrated compliance with international standards (e.g., ISO 11137 for sterilization, ISO 11607-1 for shelf life). These are laboratory/engineering tests, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not applicable. There was no clinical "test set" or requirement for expert-established ground truth mentioned in this 510(k) submission.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. No "test set" was used for clinical performance evaluation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical dental implant, not an AI/ML software or system that would involve human "readers" or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical dental implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable in the clinical trial sense. The "ground truth" for this submission revolves around established engineering and material science standards. For example:
      • Biocompatibility: Established by demonstrating the use of identical manufacturing processes and materials (titanium alloy Ti-6Al-4V, ASTM F 136) as the predicate device, which is a "well known and well established material." Compliance with ISO 10993 (Biological Evaluation of Medical Devices) is implied through equivalence.
      • Sterilization Validation: Established by demonstrating identity to the predicate device in manufacturing process, material, sterile barrier system, and use of gamma radiation under the same conditions, validated following ISO 11137.
      • Shelf Life: Established by demonstrating identical packaging materials & methods as a reference device (Hiossen Implant System, K140934) which had an 8-year shelf life validated per ISO 11607-1, and stating the device is medical grade titanium, non-mechanical, non-active.
      • Surface Treatment Characterization: Established by demonstrating identical manufacturing process, material, and SA (Sandblasted, Acid etched) surface treatment as the predicate device.
      • Fatigue Testing: Established by demonstrating identical manufacturing process, material, and "very similar design" to a predicate device (MS SA) that was fatigue tested in accordance with ISO 14801:2003.

    8. The sample size for the training set

    • Not applicable. This device is a physical implant; there is no "training set" in the context of an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.