Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K101147

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-21

    (273 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K070004, K072699, K082622

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified Alma Lasers Family of Accent™ Radiofrequency (RF) Systems [Accent, Accent XL, Accent Elite] is intended for use in dermatologic and general surgical procedures.

    The Alma Lasers Family of Accent™ RF Systems [Accent XL, Accent Elite] is indicated for the non-invasive treatment of wrinkles and rhytids using a combined treatment with unipolar and bipolar handpieces.

    The massager component of the Alma Lasers Accent UniForm Handpiece is intended for use with the Modified Alma Lasers Family of Accent™ RF Systems [Accent, Accent XL, Accent Elite] to provide:
    Temporary reduction in the appearance of cellulite. .
    Simultaneous application of the RF energy and mechanical manipulation of the skin is intended for use with the Modified Alma Lasers Family of Accent™ RF Systems [Accent, Accent XL, Accent Elite] to provide:
    Temporary reduction in the appearance of cellulite. .

    Device Description

    The Alma Lasers Accent Elite system similar to Accent™ RF Systems [Accent, Accent XL } is comprised of the following main components:
    . Console
    a Bipolar RF module
    A Unipolar RF module (UniLarge)
    A UniForm (RF and Massage) module
    . Control panel
    Footswitch. .
    Modules are used to deliver radiofrequency energy to the treatment site. Each RF module consists of the following components:
    . Handle - used for holding the module
    . Trigger - activates the radiofrequency energy emission when pressed in Ready mode
    Applicator tip establishes contact with the patient's skin .
    . Thermoelectric cooler - integrated within the module, provides internal module cooling
    . RF emission indicator - blue LED illuminates prior to- and during RF emission
    Umbilical cable contains coolant tubes. RF-power cable and the communication . cable that controls the operation of the module
    . Module connector - connects the module to its port. It incorporates an integrated impedance matching network (IMN) and a memory chip which stores information about the module and the parameter settings.
    The UniForm module additionally employs a massaging mechanism that works in conjunction with the RF energy application.
    The Accent Elite is a computerized system with embedded software that controls its operation. The software also runs the graphical user interface, which enables user-friendly control of the system operation.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (Modified Alma Lasers Family of Accent™ Radiofrequency (RF) Systems [Accent, Accent XL, Accent Elite]), seeking substantial equivalence to predicate devices. It does not describe or conduct a study involving acceptance criteria and reported device performance in the way typically associated with clinical trials or AI/algorithm evaluation.

    Instead, this document focuses on demonstrating substantial equivalence to previously cleared devices based on technical characteristics and safety performance testing rather than clinical efficacy studies with specific performance metrics against a defined ground truth.

    Therefore, many of the requested elements for describing an acceptance criteria study and the methodology behind it (like sample size, data provenance, expert ground-truthing, MRMC studies, standalone performance, training set details) are not applicable or not provided in this type of regulatory submission.

    However, I can extract the information that is present concerning the rationale for substantial equivalence and the listed safety performance testing.

    Here's a breakdown based on the provided document:


    Acceptance Criteria and Device Performance Study Information

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in terms of performance metrics (e.g., accuracy, sensitivity, specificity) for a clinical outcome. Instead, the "acceptance criteria" for this submission are compliance with recognized safety standards and demonstration of substantial equivalence in technical characteristics.

    Acceptance Criteria Type (Implicit)Reported Device Performance (Compliance/Equivalence)
    Technical & Functional EquivalenceThe Modified Alma Lasers Family of Accent™ RF Systems [Accent, Accent XL, Accent Elite] shares: • The same indications for use. • The same operation principle. • Similar technical and functional capabilities as the predicate Accent™ Family of RF Systems (K072699 and K070004). (See detailed comparison table in Section VI for specific RF frequency, treatment energy, operational modes, etc., all matching predicate devices).
    Safety and Essential PerformanceCompliance with applicable FDA Recognized Consensus Standards: • IEC 60601-1-2: 2004: Medical electrical equipment Part 1: General requirements for safety: Electromagnetic compatibility • IEC 60601-1:1988, Amendment 1:1991, Amendment 2:1995: Medical electrical equipment Part 1: General requirements for safety • IEC 60601-2-2:2006: Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgery equipment and high frequency surgical accessories The results of the performance testing showed compliance with these standards.
    Clinical Equivalence (Implicit)Based on the above, the device is considered "safe and effective, and performs at least as safely and effectively as the predicate Alma Lasers Family of Accent™ Radiofrequency (RF) Systems [Accent, Accent XL] for the intended use and indications." (This is a conclusion drawn from the technical and safety equivalence, not from a comparative clinical efficacy study provided in this summary).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable / Not Provided. This document does not describe a clinical performance study with a test set of patient data. The "performance testing" refers to engineering and electrical safety standards compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable / Not Provided. No clinical ground truth was established for a test set in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable / Not Provided. No clinical test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable / Not Provided. This device is an RF system for dermatologic/surgical procedures, not an AI or imaging diagnostic tool. Therefore, an MRMC study is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable / Not Provided. This device does not feature an algorithm for standalone performance evaluation in the context of AI or diagnostic systems.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable / Not Provided. The "ground truth" for this submission relates to compliance with engineering safety standards and established technical specifications, not clinical outcomes or expert consensus on a diagnostic task.

    8. The sample size for the training set

    • Not Applicable / Not Provided. This device is not an AI/machine learning system, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    • Not Applicable / Not Provided. No training set exists.

    Summary of the Study (as described in the 510(k) Summary):

    The "study" described in this document is a regulatory submission demonstrating substantial equivalence rather than a clinical performance study measuring specific outcomes. The primary "studies" conducted were non-clinical performance testing to ensure safety and compliance:

    • Type of Study: Non-clinical performance testing for electrical safety and electromagnetic compatibility.
    • Purpose: To demonstrate that the Modified Alma Lasers Family of Accent™ RF Systems [Accent, Accent XL, Accent Elite] is as safe and effective as its predicate devices, primarily through compliance with international standards (IEC 60601 series).
    • Methodology: The system was tested against the requirements of IEC 60601-1-2: 2004 (electromagnetic compatibility), IEC 60601-1:1988 with amendments (general safety), and IEC 60601-2-2:2006 (particular requirements for high-frequency surgical equipment).
    • Conclusion: The tests demonstrated compliance with these standards, supporting the conclusion of substantial equivalence to the predicate devices. This equivalence is based on the device sharing the same intended use, indications for use, operation principle, and similar technical and functional capabilities, alongside meeting safety criteria.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092191

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-25

    (127 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K083590, K053365, K040135, K033942, K072699, K070004

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EXILIS device is indicated for the primary treatment of dermatologic and general surgical procedures for non-invasive treatment of wrinkles and rhytids. EXILIS is a state-of-the-art device to apply therapy by a non-invasive method of high-frequency field.

    Device Description

    The EXILIS device is indicated for the primary treatment of dermatologic and general surgical procedures for non-invasive treatment of wrinkles and rhytids. EXILIS is a state-of-the-art device to apply therapy by a non-invasive method of high-frequency field.

    The control unit of the device is fitted with a color touch screen, which significantly facilitates the use of the device is equipped with a stylus (touch pen) for comfortable control of the touch screen. The vertically oriented design of the device enables to see the on-screen information from various positions of the operator. In addition, the brightness of the screen can be set to match the lighting in the surgery. The on-screen information will guide you through the entire therapy by means of easy setting of parameters using touch-screen buttons and knobs/keys on the device. For easier control, the applicator is equipped with buttons, enabling to operate the device during therapy, and a display, which shows the set and indicated parameters.

    Any therapeutic parameter can be set easily by simple use of the touch-screen buttons. During the entire therapy time the device informs you about the therapeutic method, the type of the therapy applied, the set power, and other necessary data.

    The EXILIS device enables to insert the patient's name and additional corresponding information in its internal memory and link these data with the pre-defined or customer protocols. At the patient's next visit simply call his/her name on the device and apply the preset therapy. The EXILIS device is placed in a specially designed cart, the shape of which provides the maximum comfort of the handling and use of the device. Four stable castors ensure easy moving of the device in the office or surgery.

    AI/ML Overview

    The acceptance criteria and study information for the EXILIS device (K092191) are detailed below based on the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceComments
    Hardware PerformanceMeets established Software Design Specifications for each test planAll hardware specifications met acceptance criteria of each module and interaction of processes. Device passed all testing.Based on internal testing and verification.
    Software PerformanceMeets established Software Design Specifications for each test planAll software specifications met acceptance criteria of each module and interaction of processes. Device passed all testing.Based on internal testing and validation.
    SafetyMitigation of identified hazards from Hazard Analysis.Device Hazard analysis completed, risk control implemented. Device passed all electrical and safety testing according to national and international standards.Demonstrated through non-clinical testing.
    Electromagnetic Compatibility (EMC)Compliance with applicable voluntary standards for EMC.Device passed all electrical and safety testing according to national and international standards.Demonstrated through non-clinical testing.
    Intended Use EquivalenceSame intended use as predicate devices.The EXILIS device has the same intended use as the predicate devices.Assessed for substantial equivalence.
    Technological EquivalenceSimilar device operation, overall technical and functional capabilities to predicate devices.The EXILIS device shares the same or similar device operation, overall technical, and functional capabilities with predicate devices.Assessed for substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary explicitly states: "There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate devices." (Section 10, Clinical Testing). Therefore, there is no separately identified "test set" in the context of clinical data. The performance evaluation was based on non-clinical testing, verification, and validation of the device's hardware and software.

    • Sample Size: Not applicable in the context of a clinical test set. The validation was against software design specifications and applicable voluntary standards.
    • Data Provenance: The data provenance is internal, based on the manufacturer's (BTL Industries, Inc.) hardware and software development, verification, and validation processes, as well as third-party testing for safety and EMC standards. There is no mention of country of origin for a clinical dataset, nor whether it was retrospective or prospective, as no clinical study was conducted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical testing was performed, there was no clinical "test set" requiring expert consensus for ground truth establishment. The ground truth for the non-clinical testing was based on engineering specifications and regulatory standards.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set requiring adjudication was used. The performance was assessed against engineering specifications and regulatory standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was conducted. The EXILIS device is an electrosurgical cutting and coagulation device, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The EXILIS is a medical device for dermatologic and general surgical procedures, not an algorithm. Its performance is inherent in its operation by a human user.

    7. The Type of Ground Truth Used

    For the non-clinical testing:

    • Hardware and Software Performance: The ground truth was established by the Software Design Specifications and other internal engineering specifications.
    • Safety and EMC: The ground truth was established by national and international safety and electromagnetic compatibility (EMC) standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. Its development involves traditional engineering design and testing, not machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1