510(k) Data Aggregation

    K Number
    K181963

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-14

    (114 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    P950019, K090707, K173148, K072326

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ostaPek Interbody Fusion Cages system is indicated for use as an intervertebral body fusion device in skeletally mature patients at one or two contiguous levels of the sacro-lumbar spine (L2-S1) to facilitate fusion in case of degenerative disc disease (DDD) defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The ostaPek Interbody Fusion Cages are placed via an anterior, posterior or lateral approach using autogenous bone graft and/or allograft comprised of cancellous and/or corticocancellous bone graft. The ostaPek Interbody Fusion Cages are to be used with supplemental fixation cleared by the FDA for use in the lumbar spine. Patients should have at least six months of non-operative treatment prior to surgery. Patients with previous non-fusion spinal surgery at involved level may be treated with the device.

    Device Description

    The ostaPek Interbody Fusion Cages system consists of four families of cages used with four surgical approaches: the Erriva® ELIF/TLIF Cage (Extraforaminal Lumbar Interbody Fusion/ Transforaminal Lumbar Interbody Fusion), the Gemitra® TLIF Cage, the PLIF Cage (Posterior Lumbar Interbody Fusion) and PLIF Bullet Cage, and the ALIF Cage (Anterior Lumbar Interbody Fusion). All device configurations are available in multiple sizes to adequately fit patient anatomical variations. They all have bone graft windows, ridges or teeth to resist migration, and x-ray markers.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device called "ostaPek Interbody Fusion Cages." It is not a document about an AI/ML medical device. Therefore, it does not contain the information requested about acceptance criteria, study design for AI evaluation, sample sizes, expert ground truth establishment, or multi-reader multi-case studies.

    The document indicates that the device's substantial equivalence was demonstrated through non-clinical testing and comparison to predicate devices, and explicitly states: "No clinical studies were performed."

    The request asks for details specific to the evaluation of an AI/ML medical device, which are not applicable to the traditional medical device clearance process described in this document.

    Therefore, I cannot provide the requested information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173148

    Validate with FDA (Live)

    Device Name
    ACIF
    Manufacturer
    Date Cleared
    2017-12-01

    (63 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K072326

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACIF system is indicated for use as an intervertebral body fusion device in skeletally mature patients at one or two contiguous levels of the cervical spine (C2-T1) to facilitate fusion in case of degenerative disc disease (DDD) defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The ACIF implants are placed via an anterior approach using autogenous bone graft and/or allograft comprised of cancellous and/or corticocancellous bone graft. The ACIF implants are to be used with supplemental fixation. Patients should have at least six weeks of non-operative treatment prior to surgery.

    Device Description

    The coLigne ACIF system are used to maintain disc space distraction in skeletally mature adults requiring anterior cervical interbody fusion (ACIF). This system includes two design configurations: ACIF oscar, and ACIF oscar dome. The implants are available in a range of footprints and heights to suit the individual pathology and anatomical conditions of the patient. The implants have a hollow center to allow placement of bone graft.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device called ACIF (Anterior Cervical Interbody Fusion) system. It outlines the device's characteristics, intended use, and substantial equivalence to previously cleared devices, a requirement for FDA clearance for Class II medical devices.

    Here's an analysis of the provided text to address your specific questions regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    Based on the provided text, the ACIF device is not an AI/Software as a Medical Device (SaMD). It is a physical intervertebral body fusion device. Therefore, the concept of "acceptance criteria" and "device performance" as typically applied to AI-powered diagnostic or predictive tools (e.g., sensitivity, specificity, AUC) does not apply in this context.

    Instead, the acceptance criteria for this device are related to its physical, mechanical, and biological properties and its substantial equivalence to existing predicate devices.

    Here's a table summarizing the relevant "performance" and "acceptance criteria" as described for this physical device:

    Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit in 510(k))Reported Device Performance (as stated in 510(k))
    Mechanical Safety & PerformanceMust demonstrate equivalent mechanical performance to predicate devices under specified test conditions."Mechanical testing confirmed the ACIF system demonstrated equivalent performance to the cited predicate device under the same test conditions."
    Static Compression (per ASTM F2077)Results "indicate that the coLigne ACIF are equivalent to predicate devices."
    Dynamic Compression (per ASTM F2077)Results "indicate that the coLigne ACIF are equivalent to predicate devices."
    Static Torsion (per ASTM F2077)Results "indicate that the coLigne ACIF are equivalent to predicate devices."
    Dynamic Torsion (per ASTM F2077)Results "indicate that the coLigne ACIF are equivalent to predicate devices."
    Subsidence (per ASTM F2267)Results "indicate that the coLigne ACIF are equivalent to predicate devices."
    Material CompatibilityMust use biocompatible materials with established safety profiles for implantable devices."Polyether ketone ether ketone ketone (ASTM F1876-98), Fiber carbon filaments, Gold (ASTM B562-95)" - stated to be similar to reference device.
    Design & DimensionsMust have similar basic design features, functions, and dimensions to predicate devices."The subject ACIF system and cited predicate devices share similar basic design features and functions." "The subject ACIF system is dimensionally similar to cited predicate devices."
    Intended UseMust have the same or very similar intended use as predicate devices."The subject ACIF system and all the predicates have similar intended uses."
    SterilizationMust meet appropriate sterilization requirements."The subject ACIF system is provided non-sterile and cited predicate devices are sterile and non-sterile for single use only." (This implies the user performs sterilization or it is used as is, and it's consistent with some similar devices.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not applicable in the context of an AI/SaMD test set. For this physical device, the "test set" refers to the specific physical prototypes or batches of the device that underwent non-clinical mechanical testing. The document does not specify the number of individual devices tested, which is typical for a 510(k) summary focused on demonstrating equivalence through standard test methods.
    • Data Provenance: The data comes from non-clinical bench testing performed by or for the manufacturer (coLigne, AG, based in Switzerland). It is not patient data, nor is it retrospective or prospective in the sense of a clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. The "ground truth" for a physical device like this is established by recognized engineering standards (e.g., ASTM F2077, ASTM F2267) and the results of established mechanical testing methodologies. There are no "experts" in the sense of clinical interpreters establishing ground truth as there would be for an image-based AI algorithm. The expertise lies in engineering and materials science to design and conduct these tests.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This concept pertains to resolving discrepancies in expert labeling or diagnoses for AI/SaMD test sets. For a mechanical device, the test results are quantitative (e.g., force, displacement), measured by calibrated equipment, and interpreted against predetermined thresholds or comparative data from predicate devices.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC study is relevant for AI/SaMD devices where human readers (e.g., radiologists) use AI assistance to improve their diagnostic accuracy. This device is a physical implant, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This concept is for AI/SaMD devices. The "performance" of this device is its mechanical integrity and ability to facilitate fusion, which is not an "algorithm."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As explained in point 3, the "ground truth" for demonstrating the safety and effectiveness of this physical device is based on accepted engineering standards (e.g., ASTM F2077, ASTM F2267) and direct comparative mechanical testing against a well-characterized predicate device. The goal is to show the new device performs equivalently in a mechanical sense.

    8. The sample size for the training set

    Not applicable. This device is not an AI/SaMD that requires a "training set" for machine learning.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of medical device.

    In summary: The provided document describes a physical implantable device, not an AI/SaMD. Therefore, many of the questions asked (especially those related to "test sets," "ground truth experts," "MRMC studies," and "training sets") are not applicable to this type of device and its regulatory clearance process. The "study" that proves the device meets acceptance criteria consists of non-clinical, benchtop mechanical testing designed to demonstrate substantial equivalence to legally marketed predicate devices, not clinical trials or AI performance evaluations.